Fluke 16 - Manual

Fluke 16

Fluke 16 – Manual, read for free online in PDF format. We hope this helps you resolve any issues you may have. If you have further questions, please contact us through the contact form.

1 Page 1
2 Page 2
3 Page 3
4 Page 4
5 Page 5
6 Page 6
7 Page 7
8 Page 8
9 Page 9
10 Page 10
11 Page 11
12 Page 12
Page: / 12

Table of Contents:

  • Page 2 – Multimeter with Temperature; Function
  • Page 3 – Specifications
  • Page 4 – Disassembly and Reassembly; Caution; Cleaning; Warning; Recommended Test Equipment
  • Page 5 – Figure 1. Model 16 Disassembled Unit
  • Page 6 – Snap
  • Page 7 – Performance Test; Multimeter; Calibration Adjustments; Initial Steps
  • Page 9 – Ice Bath; Figure 3. Performance Test for Temperature; Clip to Base Contacts
  • Page 10 – Room Temperature Calibration; Replacement Parts; Table 4. Replacement Parts; Parts
Loading the manual

PN 644226 July, 1997 Rev.1, 3/99

© 1997,1999 Fluke Corporation, All rights reserved. Printed in U.S.A. All product names are trademarks of their respective companies.

1

®

16

Multimeter with Temperature

Calibration Information

W

Warning

To avoid electrical shock, remove test leads and any input signals from the Meter before opening the case.

Caution

This Meter contains parts that can be damaged by static discharge. Follow the standard practices for handling
static sensitive devices.

Introduction

The 16 Calibration Information instruction sheet provides the
information necessary to calibrate and verify the Fluke
Model 16 Multimeter with Temperature (hereafter known as
the Meter). This instruction sheet provides the following
information:

Product specifications

Disassembly and reassembly

Performance test procedures

Calibration procedures and adjustments

Replacement parts

See the instruction sheet for operating instructions.

Service

The Meter is warranted to be free from defects in material
and workmanship for one year. The complete warranty
statement is in the Meter’s instruction sheet.

To contact Fluke, call one of the following telephone
numbers:

USA and Canada: 1-888-99-FLUKE (1-888-993-5853)
Europe: +31-402-678-200
Japan: +81-3-3434-0181
Singapore: +65-738-5655
Anywhere in the world: +1-425-356-5500

Or, visit Fluke’s Web stie at www.fluke.com.

Specifications

Specifications are in Table 1. Accuracy is specified for a
period of one year after calibration, at 18ºC to 28ºC (64ºF to
82ºF) with relative humidity to 90%. AC conversions are ac-
coupled, average responding, and calibrated to the rms
value of a sine wave input.

Table 1. Specifications

Maximum Voltage Between any Terminal and Earth Ground (excludes 10% tolerance) 600 V ac or dc

Display

3-3/4-digits, 4000 counts, updates 4/sec

Operating Temperature

-10

°

C to 50

°

C

Storage Temperature

-30

°

C to 60

°

C indefinitely (to -40

°

C for 100 hrs)

Temperature Coefficient

0.1 x (specified accuracy)/

°

C (<18

°

C or >28

°

C)

Relative Humidity

0% to 90% (-10

°

C to 35

°

C)

0% to 70% (35

°

C to 50

°

C)

Altitude Operation/Non-operation

2000 m

Battery Type

9V, NEDA 1604 or IEC 6F22

Battery Life

650 continuous hours with alkaline

450 continuous hours with carbon-zinc

Shock, Vibration

3 meter drops: Sinusoidal vibration up to 3 G @ 55 Hz

Size (HxWxL)

1.35 in x 2.75 in x 5.55 in

(3.46 cm x 7.05 cm x 14.23 cm)

Weight

10 oz (286g)

EMI Regulations

Complies with FCC Part 15B, Class B, EN50081-1, EN50082-1.

Safety

Designed to Protection Class II requirement of UL1244, ANSI/ISA-S82.01 - 1988, CSA C22.2 No 231,
and VDE 0411, and IEC 1010-1 overvoltage category III (CAT III), 600 V

Certification

P

$

"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.

Summary

Page 2 - Multimeter with Temperature; Function

16 Multimeter with Temperature 2 Accuracy specifications are given as follows: ±([% of reading] + [number of least significant digits]). Table 1. Specifications (cont) Function Range Resolution Accuracy f (50 to 400 Hz) 4000 mV 1 4.000 V 40.00 V 300.0 V 400.0 V 600.0 V 1 mV 0.001 V 0.01 V 0.1 V 0.1 ...

Page 3 - Specifications

Specifications 3 Table 1. Specifications (cont) Function Overload Protection 1 Input Impedance (Nominal) Common ModeRejection Ratio (1 k Ω Unbalanced) 2 Normal Mode Rejection 2 f 600 V dc >5 M Ω <100 pF 2 Automatic Selectionand LoZ = >2 k Ω <200 pF(ac coupled) 3 >60 dB at dc 50 Hz or6...

Page 4 - Disassembly and Reassembly; Caution; Cleaning; Warning; Recommended Test Equipment

16 Multimeter with Temperature 4 Disassembly and Reassembly Referring to Figure 1, disassemble the Meter as follows. APhillips-head screwdriver and small flat-blade screwdriverare required. 1. Remove the test leads and set the rotary knob to OFF. 2. Remove the Phillips-head screws from the case bott...