Page 2 - LCD display symbols; ON Button
2 EN INTRODUCTION Thank you for selecting the Oregon Scientific™ Wrist BloodPressure Monitor (BPW129) as your health product of choice.This product has been designed to provide you with manyyears of reliable service provided it is used correctly, andcan help you measure and track the following metri...
Page 3 - SAFETY AND CARE INSTRUCTIONS; Safety precautions
3 EN SAFETY AND CARE INSTRUCTIONS Safety precautions Please observe the following safety precautions when settingup and using your blood pressure monitor. · This device is intended for adult use only. · This device is intended for non-invasive measuring andmonitoring of arterial blood pressure. It i...
Page 4 - Caring for your blood pressure monitor; ABOUT BLOOD PRESSURE; What is blood pressure?
4 EN Caring for your blood pressure monitor To ensure you receive the maximum benefit from using thisproduct, please observe the following care guidelines. · When not in use, store the main unit in the protectiveplastic container that came with it. · Do not immerse the main unit in water. If it come...
Page 5 - Hypotension value is for reference only.; HOW DOES THE BLOOD PRESSURE ?; This product uses the Oscillometric Measuring method to; Figure 4; diastolic over 95
5 EN hypertensive people do not realize they are at risk until theirhealth is seriously threatened. What is the standard blood pressure classification? The following chart (Figure 3) was taken from the WorldHealth Organization (WHO), and describes the variousclassifications. Please note: · Hypotensi...
Page 6 - GETTING STARTED; Install the batteries by matching the correct polarity, as; Press and hold SET for 2 seconds to enter the setting
6 EN detect your blood pressure. Before every measurement, themain unit establishes a “zero pressure” equivalent to the airpressure. Then it starts inflating the wrist cuff to 180mmHgor higher until it senses that it has blocked your blood in theartery. After that, the deflation process starts, duri...
Page 7 - Roll or push up your sleeve to expose the skin.
7 EN Positioning the wrist cuff It is important to properly position the wrist cuff to ensure that you receive an accurate reading. 1. Remove all accessories (watch, bracelet, etc.) from your left wrist. If your physician has diagnosed you with poor circulation in your left arm, use your right wrist...
Page 8 - To take a measurement:; Choose the position you from which you wish to measure; IF you are... THEN...
8 EN TAKING A BLOOD PRESSURE MEASUREMENT You can choose to take your blood pressure while sitting or lyingdown. Helpful Tips for taking a measurement:· Be sure to set the clock before taking your first measurement, or whenever you replace the batteries, so that the date and time arestored in the mem...
Page 9 - · Press MEM / UP again to see additional records.; · The date and time of measurement are shown with each; TROUBLESHOOTING GUIDE; Lying down
9 EN RECALLING MEASUREMENTS STORED IN THE MEMORY The memory can hold up to 8 records. The records areautomatically stored each time you take a new measurement. Note: · The most recent record (1) is shown first. Each new measurement is assigned to the first (1) record. All otherrecords are pushed bac...
Page 10 - No power
10 EN Display is dimor will not lightup. No power Batteries areexhausted. Replace withnew batteries( 6). Batteries areinsertedincorrectly. Insert thebatteriescorrectly( 6). Lowbatteries s h o w s o nthe display. Batteries arelow. Replace withnew batteries( 6). Errormessage shows onthe display. Ameas...
Page 11 - Problem
11 EN SPECIFICATIONS Application: Measuring method Oscillometric / non-invasive Application For adult use only Measurement location Wrist section of lower arm Memory Maximum 8 records Dimensions: Height x Width x Depth ~ 72mm x 72mm x 24.5mm (2.8in x 2.8in x 1.0in) Weight 130g (4.6 oz) without batte...
Page 12 - Operating environment:; Operation; INFORMATION CONCERNING THE CE
12 EN Operating environment: Operation 10 ° C…40 ° C (50 ° F…104 ° F) Storage / Transport -20 ° C…60 ° C (-4 ° F…140 ° F) Humidity range 10%…83% relative humidity INFORMATION CONCERNING THE CE MARK This device complies with the European regulations basedon the Medical Products Code, and bears the CE...