Page 2 - Notices and Trademarks
Notices and Trademarks Copyright 2009 by Honeywell International Inc. Release G July 2009 While this information is presented in good faith and believed to be accurate, Honeywell disclaims the implied warranties of merchantability and fitness for a particular purpose and makes no express warranties ...
Page 4 - Contents Serial; Section Title
Contents Serial number: Section Title Page 1 System Description 5 2 Operation 7 3 Maintenance 9 4 Programming and Advanced Features 11 5 Appendices 15 6 Limited Warranty 17 Introduction This manual has been prepared to help in the use of the Manning EC-P2 Portable Gas Detector. ATTENTION: This manua...
Page 5 - System Description
1 System Description Figure 1: Basic Parts of the Manning EC-P2 Sensor The Manning EC-P2 is a portable, battery operated instrument for the measurement of gas concentrations in ambient air. Outlet Tube Sampling WandAssembly Data Logger Port Standby BatteryRecharge Receptacle Primary Battery The Mann...
Page 6 - continued
Manning EC-P2 Sensor 19546 ECP2 07/09 REVG Copyright © 2009 Honeywell Analytics. All Rights Reserved. 6 1 System Description continued System Specifications Primary Power: Alkaline C cell battery, 48 hours operation. Note: An audible alarm single beep every 60 seconds indicates that either the Alkal...
Page 7 - Operation; Operating the Sensor; Changing the Target Gas; MENU SMPL
2 Operation A Operating the Sensor Figure 2: Normal Operating Screen The Manning EC-P2 is shipped ready for use. To power up the instrument, push the PWR button. The power up sequence displays first the self check, the sensor full scale value, the downscale caution level, warning concentration setpo...
Page 9 - Maintenance; Periodic Maintenance; Calibration; To Enter ZERO / SPAN Mode
3 Maintenance A Periodic Maintenance It is essential that the test and calibration schedule be adhered to. Honeywell Analytics recommends the following maintenance schedule: • Response test prior to each use. Expose sensor to a sample of the target gas to verify proper sensor response and alarm func...
Page 10 - Troubleshooting
Calibration gasflowmeter Vent through flowmeter Bypass tee Calibration gas bottle(use only CertifiedCalibration Gas) 1.0 .8 .6 .4 .2 L P M A I R CAUTION DO NOT BLOCK INLET OR OUTLIET PORTS INLET OUTLET TEST UNIT PRIOR TO USE Fu 3 Maintenance continued Zero Calibration 4 ) Be sure the unit is in clea...
Page 11 - Programming and Advanced Features; Alarm Setpoints, Functions; Warning Setpoint
4 Programming and Advanced Features A Alarm Setpoints, Functions The Manning EC-P2 has three adjustable alarm setpoints. These setpoints are stored on each SMART-CELL. The Caution setpoint is to warn of excessive negative zero drift. The WARNING and ALARM setpoints are to warn of potentially dangero...
Page 12 - Display; To Enter AVG / BLANK Mode
4 Programming and Advanced Features continued B Display Variables The Manning EC-P2 has two adjustable display variables. These variables control signal conditioning of the LCD display. Variables are stored on each SMART-CELL. The AVG (Average) variable determines the number of one second samples to...
Page 13 - SAMP Variable; Data; To Start the Data Logging Session
4 Programming and Advanced Features continued Programming sample mode variables: The sample mode variables are Sample, Measure and Clear as described above. The following procedure will change the Sample variable. To Enter SAMP / MEAS / CLEAR Mode 1) Push power button to turn unit on. Wait for power...
Page 14 - Analog; To Enter RANGE Mode
4 Programming and Advanced Features continued To Stop the Data Logging Session 7) Push the MENU button. 8) Push the LOG button. 9) Push and hold the STOP button until the screen changes to CLEAR / INT / START / MODE (two seconds). 10) Push the DONE button to get back into normal operation mode. Work...
Page 15 - Appendices
Manning EC-P2 Sensor 19546 ECP2 07/09 REVG Copyright © 2009 Honeywell Analytics. All Rights Reserved. 15 5 Appendices APPENDIX A: SMART-CELLs Available GAS RANGE PART NUMBER Alcohol 0-500 ppm EC-P2-SMART-CELL-OH-0/500ppm Ammonia 0-500 ppm EC-P2-SMART-CELL-NH 3 -0/500ppm Ammonia 0-2000 ppm EC-P2-SMAR...