Accuracy Study for the Welch Allyn
®
Model 692/690 SureTemp
®
Plus
Oral, Adult Axillary, and Rectal Thermometer
"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.
Summary
Accuracy Study for the Welch Allyn ® Model 692/690 SureTemp ® Plus Oral, Adult Axillary, and Rectal Thermometer Clinical accuracy studies were performed using the SureTemp ® Plus in the oral, adult axillary (18 years and older), and rectal modes. Over 220 temperaturemeasurements were obtained from o...
Oral Data Oral temperatures ranged from 97.1° F (36.16° C) to102.6° F (39.22° C). The total number of data sets were91, with 27 of those being febrile. Subjects ranged inage from 11 years to 87 years. The average error was-0.03° F (-0.016° C) with a Standard Deviation of 0.358° F(0.198° C ). Adult A...
0 - 2 years 3 - 10 years 11- 65 years > 65 years 99.5 95.9 99.6 97.6 98.5 96.4 — 100.4 97.9 100.6 98.6 97.1 98.0 96.6 98.4 95.3 97.4 96.0 100.0 97.0 99.7 96.6 99.5 96.4 O Orra all R Re ecctta all A Ax xiilllla arryy E Ea arr 100.4 97.9 99.1 94.5 100.4 97.5 ° F 100.0 97.5 100.2 98.2 98.8 96.6 C Co...