Summary
Accuracy Study for the Welch Allyn ® Model 692/690 SureTemp ® Plus Oral, Adult Axillary, and Rectal Thermometer Clinical accuracy studies were performed using the SureTemp ® Plus in the oral, adult axillary (18 years and older), and rectal modes. Over 220 temperaturemeasurements were obtained from o...
Oral Data Oral temperatures ranged from 97.1° F (36.16° C) to102.6° F (39.22° C). The total number of data sets were91, with 27 of those being febrile. Subjects ranged inage from 11 years to 87 years. The average error was-0.03° F (-0.016° C) with a Standard Deviation of 0.358° F(0.198° C ). Adult A...
0 - 2 years 3 - 10 years 11- 65 years > 65 years 99.5 95.9 99.6 97.6 98.5 96.4 — 100.4 97.9 100.6 98.6 97.1 98.0 96.6 98.4 95.3 97.4 96.0 100.0 97.0 99.7 96.6 99.5 96.4 O Orra all R Re ecctta all A Ax xiilllla arryy E Ea arr 100.4 97.9 99.1 94.5 100.4 97.5 ° F 100.0 97.5 100.2 98.2 98.8 96.6 C Co...