Welch Allyn LXi - Manual

Welch Allyn LXi

Welch Allyn LXi – Manual, read for free online in PDF format. We hope this helps you resolve any issues you may have. If you have further questions, please contact us through the contact form.

1 Page 1
2 Page 2
Page: / 2
Loading the manual

Blood Pressure Operation

Press the

BP Start/Stop

button to start a blood pressure

measurement; press it again to stop a current reading.

SureTemp Plus Temperature Operation

1. Remove probe from well and apply a new probe cover. Use the

Navigation

buttons to change the temperature mode then

press the

Select

button.

2. Place probe in contact with patient tissue as directed in the

Directions for Use manual. When done, a beep sounds. The temperature appears
on the display.

3. Eject probe cover. Return probe to well before starting another measurement.

Braun ThermoScan

®

PRO 4000 Temperature Operation

1. Remove probe from holder and apply a new probe cover; when ready, the device

will beep.

2. Fit probe snugly in the ear canal as directed in the Directions for Use manual.
3. Press the Activation button. When done, a long beep sounds. The probe will

display the temperature.

4. Eject probe cover and return probe to the holder. The Spot Vital Signs LXi will

display the temperature.

SpO

2

Operation

Apply sensor to patient. Verify there is a signal on the pulse signal bar graph. The
SpO

2

measurement and pulse rate will update continuously for 10 minutes.

Blood Pressure Operation

Press the

BP Start/Stop

button to start a blood pressure

measurement; press it again to stop a current reading.

SureTemp Plus Temperature Operation

1. Remove probe from well and apply a new probe cover. Use the

Navigation

buttons to change the temperature mode then

press the

Select

button.

2. Place probe in contact with patient tissue as directed in the

Directions for Use manual. When done, a beep sounds. The temperature appears
on the display.

3. Eject probe cover. Return probe to well before starting another measurement.

Braun ThermoScan

®

PRO 4000

Temperature Operation

1. Remove probe from holder and apply a new probe cover; when ready, the device

will beep.

2. Fit probe snugly in the ear canal as directed in the Directions for Use manual.
3. Press the Activation button. When done, a long beep sounds. The probe will

display the temperature.

4. Eject probe cover and return probe to the holder. The Spot Vital Signs LXi will

display the temperature.

SpO

2

Operation

Apply sensor to patient. Verify there is a signal on the pulse signal bar graph. The
SpO

2

measurement and pulse rate will update continuously for 10 minutes.

Blood Pressure Operation

Press the

BP Start/Stop

button to start a blood pressure

measurement; press it again to stop a current reading.

SureTemp Plus Temperature Operation

1. Remove probe from well and apply a new probe cover. Use the

Navigation

buttons to change the temperature mode then

press the

Select

button.

2. Place probe in contact with patient tissue as directed in the

Directions for Use manual. When done, a beep sounds. The temperature appears
on the display.

3. Eject probe cover. Return probe to well before starting another measurement.

Braun ThermoScan

®

PRO 4000 Temperature Operation

1. Remove probe from holder and apply a new probe cover; when ready, the device

will beep.

2. Fit probe snugly in the ear canal as directed in the Directions for Use manual.
3. Press the Activation button. When done, a long beep sounds. The probe will

display the temperature.

4. Eject probe cover and return probe to the holder. The Spot Vital Signs LXi will

display the temperature.

SpO

2

Operation

Apply sensor to patient. Verify there is a signal on the pulse signal bar graph. The
SpO

2

measurement and pulse rate will update continuously for 10 minutes.

Blood Pressure Operation

Press the

BP Start/Stop

button to start a blood pressure

measurement; press it again to stop a current reading.

SureTemp Plus Temperature Operation

1. Remove probe from well and apply a new probe cover. Use the

Navigation

buttons to change the temperature mode then

press the

Select

button.

2. Place probe in contact with patient tissue as directed in the

Directions for Use manual. When done, a beep sounds. The temperature appears
on the display.

3. Eject probe cover. Return probe to well before starting another measurement.

Braun ThermoScan

®

PRO 4000 Temperature Operation

1. Remove probe from holder and apply a new probe cover; when ready, the device

will beep.

2. Fit probe snugly in the ear canal as directed in the Directions for Use manual.
3. Press the Activation button. When done, a long beep sounds. The probe will

display the temperature.

4. Eject probe cover and return probe to the holder. The Spot Vital Signs LXi will

display the temperature.

SpO

2

Operation

Apply sensor to patient. Verify there is a signal on the pulse signal bar graph. The
SpO

2

measurement and pulse rate will update continuously for 10 minutes.

Spot Vital Signs LXi

Operating Instructions

On/Off

Button

Controls power to the device.

Select

Button

Accepts the current option.

Spot Vital Signs LXi

Operating Instructions

On/Off

Button

Controls power to the device.

Select

Button

Accepts the current option.

Spot Vital Signs LXi

Operating Instructions

On/Off

Button

Controls power to the device.

Select

Button

Accepts the current option.

Spot Vital Signs LXi

Operating Instructions

On/Off

Button

Controls power to the device.

Select

Button

Accepts the current option.

Blood Pressure Operation

Press the

BP Start/Stop

button to start a blood pressure

measurement; press it again to stop a current reading.

SureTemp Plus Temperature Operation

1. Remove probe from well and apply a new probe cover. Use the

Navigation

buttons to change the temperature mode then press

the

Select

button.

2. Place probe in contact with patient tissue as directed in the

Directions for Use manual. When done, a beep sounds. The temperature appears
on the display.

3. Eject probe cover. Return probe to well before starting another measurement.

Braun ThermoScan

®

PRO 4000 Temperature Operation

1. Remove probe from holder and apply a new probe cover; when ready, the device

will beep.

2. Fit probe snugly in the ear canal as directed in the Directions for Use manual.
3. Press the Activation button. When done, a long beep sounds. The probe will

display the temperature.

4. Eject probe cover and return probe to the holder. The Spot Vital Signs LXi will

display the temperature.

SpO

2

Operation

Apply sensor to patient. Verify there is a signal on the pulse signal bar graph. The
SpO

2

measurement and pulse rate will update continuously for 10 minutes.

Blood Pressure Operation

Press the

BP Start/Stop

button to start a blood pressure

measurement; press it again to stop a current reading.

SureTemp Plus Temperature Operation

1. Remove probe from well and apply a new probe cover. Use the

Navigation

buttons to change the temperature mode then press

the

Select

button.

2. Place probe in contact with patient tissue as directed in the

Directions for Use manual. When done, a beep sounds. The temperature appears
on the display.

3. Eject probe cover. Return probe to well before starting another measurement.

Braun ThermoScan

®

PRO 4000 Temperature Operation

1. Remove probe from holder and apply a new probe cover; when ready, the device

will beep.

2. Fit probe snugly in the ear canal as directed in the Directions for Use manual.
3. Press the Activation button. When done, a long beep sounds. The probe will

display the temperature.

4. Eject probe cover and return probe to the holder. The Spot Vital Signs LXi will

display the temperature.

SpO

2

Operation

5. Apply sensor to patient. Verify there is a signal on the pulse signal bar graph. The

SpO

2

measurement and pulse rate will update continuously for 10 minutes.

Blood Pressure Operation

Press the

BP Start/Stop

button to start a blood pressure

measurement; press it again to stop a current reading.

SureTemp Plus Temperature Operation

1. Remove probe from well and apply a new probe cover. Use the

Navigation

buttons to change the temperature mode then

press the

Select

button.

2. Place probe in contact with patient tissue as directed in the

Directions for Use manual. When done, a beep sounds. The temperature appears
on the display.

3. Eject probe cover. Return probe to well before starting another measurement.

Braun ThermoScan

®

PRO 4000 Temperature Operation

1. Remove probe from holder and apply a new probe cover; when ready, the device

will beep.

2. Fit probe snugly in the ear canal as directed in the Directions for Use manual.
3. Press the Activation button. When done, a long beep sounds. The probe will

display the temperature.

4. Eject probe cover and return probe to the holder. The Spot Vital Signs LXi will

display the temperature.

SpO

2

Operation

Apply sensor to patient. Verify there is a signal on the pulse signal bar graph. The
SpO

2

measurement and pulse rate will update continuously for 10 minutes.

Blood Pressure Operation

Press the

BP Start/Stop

button to start a blood pressure

measurement; press it again to stop a current reading.

SureTemp Plus Temperature Operation

1. Remove probe from well and apply a new probe cover. Use the

Navigation

buttons to change the temperature mode then

press the

Select

button.

2. Place probe in contact with patient tissue as directed in the

Directions for Use manual. When done, a beep sounds. The temperature appears
on the display.

3. Eject probe cover. Return probe to well before starting another measurement.

Braun ThermoScan

®

PRO 4000 Temperature Operation

1. Remove probe from holder and apply a new probe cover; when ready, the device

will beep.

2. Fit probe snugly in the ear canal as directed in the Directions for Use manual.
3. Press the Activation button. When done, a long beep sounds. The probe will

display the temperature.

4. Eject probe cover and return probe to the holder. The Spot Vital Signs LXi will

display the temperature.

SpO

2

Operation

Apply sensor to patient. Verify there is a signal on the pulse signal bar graph. The
SpO

2

measurement and pulse rate will update continuously for 10 minutes.

Spot Vital Signs LXi

Operating Instructions

On/Off

Button

Controls power to the device.

Select

Button

Accepts the current option.

Spot Vital Signs LXi

Operating Instructions

On/Off

Button

Controls power to the device.

Select

Button

Accepts the current option.

Spot Vital Signs LXi

Operating Instructions

On/Off

Button

Controls power to the device.

Select

Button

Accepts the current option.

Spot Vital Signs LXi

Operating Instructions

On/Off

Button

Controls power to the device.

Select

Button

Accepts the current option.

"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.

Other Welch Allyn Models

All Welch Allyn Other