I n s t r u c t i o n S h e e t

0 1 2 - 1 1 0 6 0 A

Light Level Sensor

PS-2177

*012-11060*

\

*See the PASCO catalog or web site at www.pasco.com for compatible PASPORT interfaces and Data Acquisition Software..

Introduction

Photometry deals with the measurement of visible light as perceived by human eyes. The PASPORT Light Level
Sensor measures illuminance in lux. The readings of the sensor mimic the wavelength sensitivity of the human
eye, which is most sensitive in the green part of the spectrum (peaking at about 555 nanometers). This is
achieved with a green filter in the sensor.

In photometry, illuminance is the

luminous flux

incident on a surface,

per unit area

. In other words, it is the mea-

sure of the intensity of the incident light, wavelength-weighted to correlate with human brightness perception.
Illuminance is measured in lux, or lumens per square meter. Luminous flux, measured in lumens, is the perceived
power (energy per unit of time) of light. One lumen is defined as the luminous flux of light produced by a light
source that emits one candela of luminous intensity over a solid angle of one steradian. Luminous intensity, mea-
sured in candela, is the wavelength-weighted power emitted in a particular direction per unit solid angle. If a
light source emits one candela, a metric (SI) base unit, it emits monochromatic green light with a frequency of
540 terahertz (THz) that has a radiant energy of 1/683 watts per steradian in a given direction. A one-hundred
watt incandescent light bulb emits about 120 candela. A uniform one candela source emits 12.6 lumens.

Included Equipment

Part Number

Light Level Sensor

PS-2177

Recommended Items

PASPORT Extension Cable

PS-2500

PASCO Interface

Catalog or web site*

Data Acquisition Software

Catalog or web site*

Eight pin mini-

DIN connector

Light Dome

Light Level

Sensor

Range Select

Buttons

"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.

Summary

Page 2 - I n t r o d u c t i o n; About the Sensor; Light Level Ranges; Usage; The sensor can be used for a variety of experiments.

® M o d e l N o . P S - 2 1 7 7 I n t r o d u c t i o n 2 About the Sensor The PS-2177 Light Level Sensor uses a silicon photodiode that produces a voltage that is proportional to light intensity. The photodiode has a spectral response range from 320 to 730 nanometers (nm) and a peak sensitivity at ...

Page 3 - S e t u p; Setup; Hardware Setup; Launch DataStudio; Xplorer and Xplorer GLX Setup; Start

® M o d e l N o . P S - 2 1 7 7 S e t u p 3 Setup Hardware Setup • Connect the Light Level Sensor to a PASPORT interface. If you will be using a computer, connect the PASPORT interface to the computer’s USB port. DataStudio Setup If you will be using the Light Level Sensor with a computer, install t...

Page 4 - Select a Display; Calibration; Accuracy; Technical Support; Product End of Life Disposal Instructions:

® L i g h t L e v e l S e n s o r S e t u p 4 Select a Display To set up a particular display (e.g., digits display), touch BUILD. Touch a quantity from the list, and then touch one of the display icons. Touch OK to open the display, and then touch the Start button to begin collecting data. Calibrat...