USN Series
Handle tough applications in any environment...
because tests are required everywhere!
Inspection oriented and easy
to use
• 40 dB dynamic DAC/TCG
enables cor rection for
distance/amplitude var-
iations due to material loss
and beam spread Cur ve
displayed on A-Scan.
(USN 52L/R)
• RF display mode improves
signal evaluation and bond
inspection of dissimilar ma-
terials and assists crack tip
detection and sizing.
(USN 52L/R)
• Selectable rep rate
minimizes wraparound.
• Selectable damping control
provides transducer
matching for broader range
of applications.
• Angle beam trigonometric
functions display sound
path, projection, and depth
measurements to flaw.
• Magnification window
displays measurement
value in lar ge,easy to r ead
digits.
• Gate monitors amplitude
and soundpath distance for
both flaw and thickness
applications.
• Multi-echo or inter face-to-
first thickness measur e-
ment mode, with selectable
flank or peak detection.
• A-Scan freeze, freeze peak,
and zoom functions.
• Lock key prevents ac-
cidental change of settings.
Documentation and recording
• On-board thickness data
logger stores a total of
2,500 readings in up to 99
configurable files.
• RS-232 bidirectional
inter face for transfer ring
data, A-scan displays, and
repor ts to a printer, com-
puter, and for remote control.
• Stores instr ument control
datasets with A-Scans in
memor y for quick and
repeatable calibration.
• Amplitude and time of flight
analog output enables
recording via strip char t or
digitizing device.
(USN 52L/R)
• Direct printing of inspection
repor ts.
Lightweight, advanced and
reliable
• 6 pounds (2.7 kg) total
weight with batteries.
• Sealed case and membrane
keypad are splashproof,
r ugged and easy to clean
• Operates on AC or 6 “D” size
alkaline or NiCad batteries in
a detachable housing
Por table Ultrasonic Flaw Detectors
High Resolution LCD and EL Displays
USNLBRPF8985K
Printed in USA
Krautkramer Branson
50 Industrial Park Road
Lewistown, PA 17044
Tel: 1-717-242-0327
Fax: 1-717-242-2606
E-Mail: [email protected]
Buehler Krautkramer Ltd
University of War wick
Science Park
GB-Coventr y CV4 7HS
Tel: +44-1203-690069
Fax: +44-1203-693032
E-Mail: buehlerkrautkramer
@compuser ve.com
Krautkramer France
ZAC Sans Souci
68, Chemin des Or meaux
69 760 Limonest France
Tel: +33-47-217-9220
Fax: +33-47-847-5698
Krautkrämer GmbH & Co.
Rober t-Bosch-Str.3
D-50354 Hür th (Ef feren)
Postfach 1363
D-50330 Hür th
Tel: +49-2233-601-0
Fax: +49-2233-601-402
E-Mail: [email protected]
www.krautkramer.com
Our facility’s quality
system conforms to
ISO 9001.
ANSI/ASQC Q91.
Krautkramer Italiana,
una divisione della
Krautkramer France SA
68, Chemin des Or meaux
69 760 Limonest France
Tel: +167-790-880
Fax: +167-780-881
USN Series
Flaw Detector
Specifications
Common to All Instruments:
Standard Package
Lightweight Ultrasonic Flaw Detector
Batter y Char ger/AC Power Supply
Six D-Size NiCad Batteries
Detachable Batter y Holder Char ging Frame
Operating Manual
Recommended Accessories
Combination soft protective car r y case with
light shield and shoulder neck strap.
UltraDOC Documentation Software
9 Pin Serial PC Cable
Serial/Parallel Printer Cable
External Char ger Adapter Cable
Additional Batter y Char ging Frame
Additional NiCad Batteries (6)
Attachable Audible Alar m
Batter y Belt Pack
Analog Output Cable
Hard Shell Transpor t Shipping Container
Transducer Kits
Range
0.2" to 200" at steel velocity (5mm to
5000mm), range selectable in fixed steps, or
continuously variable
Velocity
Continuously adjustable from .0394 to .3936
in/µsec (1000-9999 m/second);
2 programmable fixed velocities
Delay
Minus 20 µsec to 999 µsec
Zero Of fset
0 to 1600 µsec
Gain
0 to 110 dB controllable in selectable gain
steps 0.1, 0.5, 1.0, 2.0, 6.0, and 24.0
(24dB step is programmable to other values)
Test Modes
Pulse echo and dual
Pulser
Spike/excitation pulse
Pulse Repetition Fr equency
Selectable, low & high; varies w/range
Rectification
Positive halfwave,negative halfwave,fullwave
Reject (suppression)
Controllable 0 to 80% absolutely linear
Units
Inch or millimeter selectable
Display
“R” Version: 5.75" x 2.65" (146mm x 67mm)
high resolution EL with selectable brightness;
hollow and solid A-Scan
60 Hz update rate
Pixel density 552 x 256
“L” Version: 4.5" x 3.0" (114mm x 76mm)
Transr eflective LCD with backlight
60 Hz update rate
Both displays have Scr een Savers
Pixel density 480 x 320
Gate Monitor
Controllable over entire sweep range in 0.1mm
increments. Alar m indication by LED.
Threshold contr olled 10% to 90% screen
height. Separate second threshold and star t
control in multi-echo measurement mode.
Measur ement Modes
Zero to first echo, 0.001" (0.1mm) resolution
up to 10" (245mm) ; multi-echo in gate with
controllable second threshold; time-of-flight to
peak or flank in gate; amplitude evaluation as
% FSH or above threshold to 127%; dB
reference indicates REF gain and addition or
subtraction; trigonometric flaw location
Memor y
Dataset memor y locations stor e all
instr ument parameters plus A-Scan. Recall
function displays stored dataset.
Data Logger
2,500 r eading sequential thickness measure-
ment storage with clear, review and dir ect
repor t featur e; configurable up to 99 files.
Operating Temperatur e
0°C to 55°C (–20°C to +70°C sur vivable)
Outputs
Gate violation output on open collector, digital
thickness value output on serial inter face
Dedicated Key Functions
COPY -
Activates outputs to printer, transmit
to data logger, output data logger repor t
ZOOM
- Expands A-Scan display area for
increased scr een resolution
MAGNIFY
- Expands area within gate for
increased A-Scan r esolution in selected area
FREEZE
- Captures cur rent scr een image or
allows continuous capture of echo peaks
LOCK
- Locks against inadver tent change of
parameter values except gain control
dB REF
- Recor ds cur rent gain then allows
changes to be separately noted; amplitude
evaluation then related to refer ence echo
amplitude
dB STEP
- Selects gain incr ements
HOME
- Retur ns instr ument to basic menu
Language
Selectable English, Fr ench, Spanish, Italian,
Por tuguese, Ger man, Dutch, Danish, Finnish,
Nor wegian, Swedish, Russian
Connectors
BNC or Lemo selectable at order
Keypad
Inter national symbols
Batter y Power
6 D-size alkaline or NiCad cell
“L” version: up to 12 hrs operation
“R” version: up to 5.5 hrs operation
Size
9.8"Wx5.25"Dx5.75"H(249mmx133mmx146mm)
Weight
6lbs (2.7kg) with NiCad batteries
3.10lbs (1.4kg) without batteries
USN 50 L & R:
Damping
Selectable: High and Low
Amplifier Bandpass
Br oadband: 0.4 MHz to 10.2 MHz
Memor y
30 dataset memor y locations with rapid clear
functions, memor y flag and A-Scan preview
Optional
DAC/TCG
USN 52 L & R:
Damping
Selectable: 50, 75, 150, 1000 Ohms
Rectification
RF display plus all others
Amplifier Bandpass
Selectable: Br oadband 0.4 MHz to 10.2 MHz
Low: .25 MHz to 4.9 MHz
High: 2.3 MHz to 10.8 MHz
DIN: 1.7 MHz to 8.2 MHz
DAC/TCG
40 dB Dynamic Range with 6 dB/µsec slope
Memor y
USN 52L: 100 Dataset memor y locations
USN 52R: 140 Dataset memor y locations
A-Scan pr eview function, alphanumeric
naming of datasets, director y of all datasets
with memor y flag for filled datasets, rapid
clear function, memo functions
Other Functions
Cur vature cor rection, movable gate in freeze
mode, auto-calibration of material velocity
and pr obe zer o
Outputs
Analog TOF & Amplitude, TTL Go/NoGo, Sync
Optional
DGS, JIS evaluation
"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.