Page 2 - Important Safety Instructions; SAVE THESE INSTRUCTIONS; Power Supply Module Connection Warnings; Equipment Connection Warnings
2 Important Safety Instructions Tripp Lite’s Medical-Grade, Mobile Power Retrofit Kit includes two modules: a Power Supply Module anda Battery Module. When the two modules are connected (and attached to a remote user interface), theyprovide AC power for a variety of equipment on mobile medical carts...
Page 3 - Electric Shock Hazard:
3 Battery Module Warnings • Do not allow objects to contact the Battery Module’s two DC input terminals. Do not short or bridge these terminals together. Serious personal injury or property damage could result. • It is normal for sparks to occur when connecting the Battery Module to the Power Supply...
Page 4 - Mounting
4 Mounting WARNING! Mount the Power Supply Module and Battery Module BEFORE connection. The mounting brackets are only designed to support a Tripp Lite Power Supply Module and asingle Battery Module. Do not mount or place any other devices on the mounting brackets.The mounting brackets will not supp...
Page 5 - Charge the Battery Module
5 Connection / Start-Up Connect the Remote User Interface (RUI) to the Power Supply Module The Power Supply Module will not operate without the RUI connected. Connect the included grey cord* from the RUI jack labeled “COMM1 GREY” to the Power Supply Module jack labeled “COMM1 GREY”. Connect the incl...
Page 7 - Operation
Stationary Operation (Plugged In / Battery Charging) Plug the Power Supply Module into a live AC wall outlet.* The Power Supply Module will deliver ACpower to connected equipment while simultaneously charging the Battery Module. Use connected equipment indefinitely as long as the Power Supply is con...
Page 8 - Features
8 Features Remote User Interface (RUI): “Power” Button: This button turns the Power Supply Module’s outlets (and any connected equipment) on and off. To turn the Power Supply Module’s outlets ON: Press and hold the “Power” button for one second. The alarm will beep once briefly afterone second has p...
Page 9 - Power Supply Module
9 Features continued Power Supply Module continued : USB Port: This port connects the Power Supply Module to any cart- mounted computer (laptop or desktop) with included cable. Use withPowerAlert Data-Saving Software (included on CD) to automaticallysave open patient data files and safely close appl...
Page 10 - Troubleshooting
10 Troubleshooting Problem No AC output power available at outlets. Battery Module not recharging,even with AC utility powerpresent. Low battery alarm sounding. Possible Solutions Turn Unit ON: Turn the Power Supply Module ON using the “Power” Button. See “Connection / Start-Up” section for details....
Page 11 - Safety & Regulatory Statements; Storage
11 Storage & Service Safety & Regulatory Statements Storage For short term storage (when charging the Battery Module between shifts), see “Operation” section. CAUTION! Even after the Power Supply Module is unplugged, its outlets may still deliver current, until it is disconnected from the Ba...
Page 12 - Warranty & Warranty Registration; Limited Warranty
12 Warranty & Warranty Registration Limited Warranty Tripp Lite warrants its Power Supply Module to be free from defects in materials and workmanship for a period of 2 years from the date ofpurchase by end user. Tripp Lite warrants its Battery Module to be free from defects in materials and work...