Page 2 - Servicing North America:; For immediate technical or application assistance:; Czech
Servicing North America: USA: One Omega Drive, Box 4047 ISO 9001 Stamford CT 06907-0047 Certified Tel: (203) 359-1660 FAX: (203) 359-7700 e-mail: [email protected] Canada: 976 BergarLaval (Quebec) H7L 5A1Tel: (514) 856-6928 FAX: (514) 856-6886 e-mail: [email protected] For immediate technical or applicatio...
Page 5 - Table of Contents
Table of Contents Chapter One - Welcome to Infrared Thermometry . . . .1 1.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . .11.2 Product Features . . . . . . . . . . . . . . . . . . . . . . .11.3 Models and Ordering Information . . . . . . . . . . .1 Chapter Two - Operation . . . ...
Page 6 - Product Features; Models and Ordering Information
Chapter One - Welcome to InfraredThermometry 1.1 Introduction Omega Engineering, the world's leading manufacturer ofinfrared temperature systems, presents the low costOS950™ Series of hand held infrared thermometers.These dual-purpose systems function as both NIST trace-able temperature calibrators ...
Page 7 - Chapter Two - Operation; To Change Display Measurement Units
OS956 0 to 1600°F (86 to 871°C) Unit includes aRemote Sensor, 9 V DC Alkaline Battery andOS950 Instruction Manual. Also available withan RS232 Digital Output and an Analog Output. Chapter Two - Operation 2.1 OS950 Controls 2.2 To Change Display Measurement Units The OS950 Models come from the factor...
Page 9 - MAX Mode Temperature Measurements
2.4.2 Temperature Measurement of MovingTargets To measure moving targets that are not possible to touch,use the following procedure: 1. Press and hold the MEASURE button. 2. The display will briefly show [8888]. 3. Bring the nosepiece as close as possible to the target,preferably within 1/32 inch (1...
Page 10 - SCAN Mode Temperature Measurements; Chapter Three - Measurement Tips; Optimizing Target Area Measurements
5. Record the temperature from the display. 6. Release the MEASURE button. The last reading willremain locked on the display for 10 seconds afterreleasing the button. 2.6 SCAN Mode Temperature Measurements Use this measurement mode for thermal scanning of a tar-get or when measuring targets located ...
Page 12 - Chapter Four - Maintenance; Battery Replacement; Chapter Five - Troubleshooting
heat. As long as the ice is not melting, the OS950 willaccurately read its temperature. To make an accurate target measurement, remove any iceor moisture before measuring. Chapter Four - Maintenance The high-performance OS950 is a rugged and durableinstrument, designed for long term performance reli...
Page 14 - Chapter Six - OS950TM Specifications
5.2 Troubleshooting MeasurementDifficulties If you experience measurement problems, review the fol-lowing troubleshooting steps: Is the sensor window clean? Clean the sensor window and reflective cup and repeat themeasurement. Is the target's surface non-metallic? Coat metallic surfaces with a non-m...