Summary
Please read this informationcarefully before use! Before use, confirm the fluid to be used iscompatible with the meter or consult withOMEGA for advice. To prevent damage from dirt or foreignmatter, OMEGA recommends a Y or bas-ket type 60 mesh strainer be installed asclose as possible to the inlet si...
1. OMEGA recommends that when set- ting up pipework for meter installations,a bypass line be included in the de-sign. This provides the facility for ameter to be removed for maintenancewithout interrupting production. (SeeFigure 1) 2 NOTE: Incorrect installation can cause pre- mature wear of meter c...
3 OPERATION SERVICE INSTRUCTIONS When fluid passes through the meter, therotors turn. The gear located on top of oneof the rotors drives the mechanical regis-ters gear train which provides an accuratereadout. Figure 3 Rotor #2 Rotor #1 Rotors must be at 90 ° to each other Disassembly: Ensure that th...