Page 3 - Table of Contents
Table of Contents Warranty/Disclaimer.........................................................................2Return.Requests/Inquiries.................................................................3 Instructions........................................................................................
Page 4 - • Warranty/Disclaimer; and workmanship for a period of; warranty; receive maximum coverage on each product.
• Warranty/Disclaimer OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA’s Warranty adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shippi...
Page 5 - • Return Requests/Inquiries; proper packaging to prevent breakage in transit.; Warranty Returns; For WARRANTY RETURNS, please have the following information
• Return Requests/Inquiries Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING DELAYS). The...
Page 6 - • Instructions
• Instructions NOTE:.All.units.are.factory.calibrated.prior.to.shipment. 1. Zero Trimming NOTE:.Some.units.are.equipped.with.zero.“knob”. If it becomes necessary to re-adjust the “Zero” on the display, this can be accomplished by turning the trimpot marked “Z” just to the left of the On/Off push but...
Page 7 - Tip
4. Re-Calibration This procedure requires a known pressure source of at least ±0.1% accuracy in order to fully utilize the accuracy potential of the DPG5000 (If not available, the gauge can be returned to OMEGA for re-calibration.) Procedure: A. Ensure the Gauge is at 0 psig (or vacuum if absolute),...
Page 8 - RE-CALIBRATION: The procedure is the same as in step 4, except that
7. 4-20 mA Transmitter Option When equipped with this option the gauge no longer operates from batteries but instead is “Loop Powered”. The loop connection is made to a 6 pin receptacle located at the rear of the unit (a mating plug is supplied). A voltage of between 9 and 32 VDC must be maintained ...
Page 9 - Option; External Power and 0-2 VDC Output Plug
A F B E C D 6 Pin Connector 8. External Power (24 VDC) and 0 to 5 VDC Analog Output Option When equipped with this option the gauge no longer operates from batteries but instead is “externally powered”. Both the 24* VDC input power and the 0 to 5 VDC Analog Output are connected to the 6 pin circular...
Page 10 - Notes
Page 11 - WARNING; : These products are not designed for use in, and should not be; Thank You for Purchasing; Again, thank you for your interest in OMEGA.
OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords our customers the latest in technology and engineering. It is the policy of OMEGA to comply with all worldwide safety and EMC/ EMI regulations that apply. OMEGA is constantly pursuing cert...
Page 12 - OMEGAnet; Servicing North America:; One Omega Drive, River Bend Technology Centre
OMEGAnet ® On-Line Service Internet e-mail www.omega.com [email protected] Servicing North America: USA: One Omega Drive, P.O. Box 4047 ISO.9001.Certified Stamford, CT 06907-0047 TEL: (203) 359-1660 FAX: (203) 359-7700 e-mail: [email protected] Canada: 976 Bergar Laval (Quebec) H7L 5A1 TEL: (514) 856-6928...