Page 3 - Instructions Manual; Conductivity Cells; Index
Instructions Manual Conductivity Cells CDE-440-001/CDE-440-01/CDE-440-1/CDE-440-5/CDE-440-01T/CDE-442 Index Conductivity Cells . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3Mechanical Description . . . . . . . . . . . . . ....
Page 4 - measurement is independent from the total solution volume.; Electrical; frequency supplied by the equipment.
1. Certificado de Garantia 3. Conductivity Cells 2 Cell is composed basically by two boards or two concentric cylinders(electrodes), with well defined geometry. The space between the boards or cylindres creates a solution collumn, being so the measurement is independent from the total solution volum...
Page 5 - or a 1⁄2 “ cable knockout.
Insertion Cell CDE-440-01 Cell Models CDE-440-001, CDE-440-01, CDE-440-1, CDE-440-5 and CDE-440-01T are called Insertion or Immersion Cells. 1. Certificado de Garantia 4. Mechanical Description Cell Model CDE-442 is called Toroydal or Electrodeless Cell, as it does not offer electrodes. The body mat...
Page 6 - Model
1. Certificado de Garantia 5. Technical Specifications Model CDE-440-001 CDE-440-01 CDE-440-1 CDE-440-5 CDE-440-1T CDE-442 Type Insertion Torroid Range (S/cm) 0.01µS to 2mS 0.1µS to 20mS 1µS to 100mS 10µS to 150 mS 0.1µS to 20mS 1mS to 1S Cell Constant (K) cm -1 0.01 0.1 1 5 0.1 NA Temperature(@ Atm...
Page 7 - Certificado de Garantia; ELECTRICAL CONNECTION; . Verify if the cable knockouts are; firmly attached; to its respective interconnection cables. This is
Green Yellow Red Black Cable Knockout Connection Box Connection Box Interconnection Base Interconnection Base 3 4 1 2 1. Certificado de Garantia 6. Installation 5 CN1 CN1 1 1 CN3 CN3 CN2 CN2 CN4 CN4 ++ R S4 8 5 R S4 8 5 S2S2 m A -2 m A -2 2 2 3 3 S3S3 -- ++ -- m A -1 m A -1 A C A C -- ++ TC-401 TC-4...
Page 8 - Interconnection Cable
a) Do not cut on mend the cell cables! Interconnection Cable This cell type is also connected directly to the equipment connection barr, so cable terminal pin type are used for some equipment cases. Colorless from Red Red Black Colorless from Shield Blue Colorless from Blue 1. Certificado de Garanti...
Page 9 - Insertion Type
Torroydal cell CDE-442 generally must be installed in a Transfluency Body made out of lower magnetic permeability material. Special attention must be observed regarding the sample flow being Flanged Threaded Point 1. Certificado de Garantia 6. Installation (cont.) Insertion Type Retractile Insertion...
Page 10 - Cleaning; would be basically cleaning, or if necessary, a new calibration.; PVC TRANSFLUENCY
1. Certificado de Garantia 7. Maintenance Cleaning Calibration Calibration for Low Conductivity Reading(Ultra Pure Water) 8 A ll OMEGA equipments, are shipped from factory tested and calibrated. When under operation, the Conductivity Cell must be maintained periodically (the frequency will depend on...
Page 11 - ) When Calibrating the equipment, the function is not complete
1. Certificado de Garantia 8. Troubleshooting 9 1) When Calibrating the equipment, the function is not complete Verify for humidity presence at probes header or at connection box of the equipment. Another possibility could be the equipments interconnection cable and cell being cut or in short. 2) Re...