Accuracy Study for the Welch Allyn
®
Model 692/690 SureTemp
®
Plus
Paediatric Axillary Thermometer
"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.
Summary
Clinical accuracy studies were performed using the SureTemp® Plus in the pediatric axillary mode (17 years and younger). One hundred ten temperature measurements were obtained from over 6 clinical sites. Sixteen percent of the data represent newborns ranging in age from just hours to 3 days old. Thi...
Table 1 SureTemp ® Plus Monitor Mode Temperature vs. Predicted Temperature Paediatric Axillary Mode Predicted Temperature Monitor Mode T e mperature Average Error -0,044 °CStandard Deviation 0,185 °C Newborn Data Axillary temperatures ranged from 35.88° C to (37.38° C. The totalnumber of data sets w...