Page 2 - Quickie S-11 Troubleshooting Guide; INTRODUCTION
Quickie S-11 Troubleshooting Guide INTRODUCTION ...................................................... 0.1 Specifications VSI Controller ........................................................... 0.2Plugs/Connectors ..................................................... 0.3Main Wiring Diagram/ Tool ...
Page 4 - VSI Controller
QUICKIE S-11 SERVICE MANUAL PAGE 0.2 2005 SUNRISE MEDICAL VSI Controller VSI Controller SMASP0X01 SMASP0P01 VSI Controller Buttons Battery Condition Meter- A series of ten LED’s, which indi-cate charge level. Speed/Profile indicator- A series of five LED’s, whichdisplay speed and profile settings On...
Page 5 - +24 Vdc Solenoid Brake Input; charger port
QUICKIE S-11 SERVICE MANUAL PAGE 0.3 2005 SUNRISE MEDICAL Plugs/Connectors 0 Vdc 9 0 Vdc 24 Vdc 24 Vdc LH Motor - LH Motor + 1 = 24 Vdc 2 = 0 Vdc 3 = Inhibit 1/ Programmer 1 = Brake Solenoid 2 = Brake Solenoid 3 = Motor 4 = Motor 1 = 0 Vdc 2 = Inhibit 3 3 = 24 Vdc RH Motor - RH Motor + +24 Vdc Solen...
Page 6 - Main Wiring Diagram/Tool List; Wiring Diagram
QUICKIE S-11 SERVICE MANUAL PAGE 0.4 2005 SUNRISE MEDICAL RIGHT MOTOR CHARGER AC INPUT Indicator Light Circuit Breaker LEFT MOTOR SOLENOID 24v BRAKE SOLENOID BLACK FUSE FUSE BLACK BLACK ORANGE ORANGE RED RED RED Front Battery Rear Battery 24v BRAKE Main Wiring Diagram/Tool List For correct maintenen...
Page 7 - Section 1; Troubleshooting: No Power; Circuit Breaker Reset; f On/off button is pressed and no light or bar is shown, check
QUICKIE S-11 SERVICE MANUAL PAGE 1 2005 SUNRISE MEDICAL Section 1 Troubleshooting: No Power Check that the voltage is going to the controller, set the meter to dc volts and take a voltage reading from pin 1 (using the red lead) and pin 2 (using the black lead) to the charger port of the VSI controll...
Page 8 - Battery Connection Check
QUICKIE S-11 SERVICE MANUAL PAGE 2 2005 SUNRISE MEDICAL 1 3 5 7 2 4 6 8 9 1.5 Battery Connection Check Check that the female Beau plug on the chair has voltage. Set the meter to dc volts and measure pins 5 (using the red lead of the meter) and 7 (using the black lead of the meter) as shown in (figur...
Page 9 - Circuit Breaker Test; If the meter reads more than 1 ohm, then change the circuit breaker.; If polarity is reversed correct battery wiring.
QUICKIE S-11 SERVICE MANUAL PAGE 3 2005 SUNRISE MEDICAL If the above steps did not correct the problem, change the main harness. 1.9 Main Harness 1.8 Circuit Breaker Test To check the circuit breaker set the meter to ohms and measure the resistance across the circuit breaker as shown in (figure A1.8...
Page 10 - Section 2; Understanding Controller Display; This indicates that all is well.
QUICKIE S-11 SERVICE MANUAL PAGE 4 2005 SUNRISE MEDICAL Section 2 Understanding Controller Display The wheelchair is locked. To unlock the wheelchair, deflect the joystick forwards until the control system chirps. Then deflect the joystick in reverse until the control system chirps. Release the joys...
Page 11 - Section 3; Understanding VSI Controller Diagnostics Codes; If the meter reads between 0 to 1.5 ohms, then replace the controller.
QUICKIE S-11 SERVICE MANUAL PAGE 5 2005 SUNRISE MEDICAL Section 3 Understanding VSI Controller Diagnostics Codes Test Left Motor OpenCheck that the batteries are fully charged and in good condition; and check all cables and connections. Check the connections to the left motor, look for a loose or da...
Page 12 - The left hand motor has a short circuit to a battery connection.
QUICKIE S-11 SERVICE MANUAL PAGE 6 2005 SUNRISE MEDICAL Test Left Motor ShortCheck that the batteries are fully charged and in good con-dition; and check all cables and connections. Check the connections to the left motor, look for a loose or damaged connector. Take a resistance reading from pin 3 t...
Page 13 - Four Bars- Right Motor Disconnected
QUICKIE S-11 SERVICE MANUAL PAGE 7 2005 SUNRISE MEDICAL Understanding VSI Controller Diagnostics Codes (cont.) Test Right Motor Open Check that the batteries are fully charged and in good condition, and check all cables and connections. Check the connections to the right motor, look for a loose or d...
Page 14 - est Right Motor Short; Five Bars - Right Motor Wiring Trip
QUICKIE S-11 SERVICE MANUAL PAGE 8 2005 SUNRISE MEDICAL Understanding VSI Controller Diagnostics Codes (cont.) T est Right Motor Short Check that the batteries are fully charged and in good con-dition, and check all cables and connections. Check the connections to the right motor, look for a loose o...
Page 15 - Seven Bars - Possible Joystick Trip
QUICKIE S-11 SERVICE MANUAL PAGE 9 2005 SUNRISE MEDICAL 1 3 5 7 2 4 6 8 9 Understanding VSI Controller Diagnostics Codes (cont.) A joystick trip is indicated. Make sure that the joystick is in the center position before switching on the control system. Check that the batteries are fully charged and ...
Page 16 - If the meter reads more than 30 volts replace the charger,; Test Motor Connections
QUICKIE S-11 SERVICE MANUAL PAGE 10 2005 SUNRISE MEDICAL Understanding VSI Controller Diagnostics Codes (cont.) Battery FaultCheck that the batteries are fully charged, the correct volt-age and in good condition. Take a voltage reading from pin 1 and pin 2 of the charger port of the VSI controller, ...
Page 17 - Section 4; Step 1 - Controller and Seat
QUICKIE S-11 SERVICE MANUAL PAGE 11 2005 SUNRISE MEDICAL Section 4 Disassembly/Reassembly, and Adjustment Disassemblya. Squeeze down the locking legs of the 3 pin charger plugand unplug it from the rear of the chair. (figure A4.1).b. Rock the controller plug back and forth until it is looseand then ...
Page 24 - Step 9 - Wire Harness
QUICKIE S-11 SERVICE MANUAL PAGE 18 2005 SUNRISE MEDICAL Disassembly/Reassembly, and Adjustment (cont) Disassemblya. Remove all the zip ties on the base frame.b. Unwrap all the wire loom ties securing the harness to theframe.c. For the main harness, remove the two screws that hold-ingthe female beau...