Page 2 - The HeartStart FR2+ Defibrillator
P h ilip s Med ical Sys tem s The HeartStart FR2+ Defibrillator Clockwise from top right. Battery. Standard long-life or rechargeable battery used to power the FR2+. (Check local regulations for disposal and recycling requirements.) On/Off button. Turns on the FR2+ and starts voice and screen prompt...
Page 4 - Intentionally left blank.
P h ili ps Me dic a l Sy s tem s Intentionally left blank.
Page 5 - H E A R T S T A R T F R 2 +
P h ili ps Me dic a l Sy s tem s H E A R T S T A R T F R 2 + M3860A, M3861A Defibrillator I N S T R U C T I O N S F O R U S E Edition 10
Page 6 - Instructions for Use; Publication date: August 2003; Authorized EU Representative
P h ili ps Me dic a l Sy s tem s M3860A/M3861A HEARTSTART FR2+ Instructions for Use Equipment specifications are subject to alteration without notice. All changes will be in compliance with regulations governing manufacture of medical equipment. Printed in the U.S.A. Publication date: August 2003 Pu...
Page 7 - Contents; Introduction to the HeartStart FR2+
1 P h ili ps Me dic a l Sy s tem s Contents 1 Introduction to the HeartStart FR2+ What is the HeartStart FR2+? ................................................................. 1-1 When Is the HeartStart FR2+ Used? .................................................... 1-2 How Does the HeartStart FR2+...
Page 9 - Appendices; A Accessories for the HeartStart FR2+
Contents P h ili ps Me dic a l Sy s tem s Appendices 3 A Accessories for the HeartStart FR2+ HeartStart Accessories ............................................................................. A-1 Suggested Additional Items ..................................................................... A-2 B...
Page 10 - Notes
M3860A/M3861A HEARTSTART FR2+ P h ili ps Me dic a l Sy s tem s Notes
Page 11 - Introduction to the HeartStart FR2+; The HeartStart FR2+ is an enhanced version of the defibrillator; N O T E; The FR2+ comes with a factory default setup that can be modified.; No manual charge in advanced mode
1-1 P h ili ps Me dic a l Sy s tem s 1 1 Introduction to the HeartStart FR2+ What is the HeartStart FR2+? The HeartStart FR2+ Defibrillator (“FR2+”) is an automated external defibrillator. It is compact, lightweight, portable, and battery powered. It is designed for simple and reliable operation by ...
Page 15 - Overview; Run the battery insertion selftest.; Installing the Battery; C A U T I O N; Follow all instructions supplied with the HeartStart M3863A
2-1 P h ili ps Me dic a l Sy s tem s 2 2 Preparing Your HeartStart FR2+ for Use Overview There are a few basic steps to preparing your HeartStart FR2+ Defibrillator for use: Install data card (optional). Install a battery. Set the clock in the FR2+ (optional). Run the battery insertion selftest. Pla...
Page 16 - not
M3860A/M3861A HEARTSTART FR2+ 2-2 P h ili ps Me dic a l Sy s tem s When the battery is installed, the FR2+ automatically turns on. The Status Indicator displays a flashing black hourglass. The Shock button light and the indicator light for the defibrillator pads connector socket turn on briefly. The...
Page 17 - If you make no selection for 10 seconds, the selftest will; Setting the Clock
P r e p a r i n g Yo u r H e a r t S t a r t F R 2 + f o r U s e 2-3 P h ili ps Me dic a l Sy s tem s 2 N O T E : If you make no selection for 10 seconds, the selftest will automatically run. If you want to select something different from either of these menus, you must do so before the selftest beg...
Page 18 - Running the Battery Insertion Selftest
M3860A/M3861A HEARTSTART FR2+ 2-4 P h ili ps Me dic a l Sy s tem s To manually set the clock: 1. Use the lower Option button to move the highlight bar to the part of the clock setting you want to change. 2. Press the upper Option button repeatedly to scroll through the settings until you reach the o...
Page 19 - Placing and Securing the HeartStart FR2+
P r e p a r i n g Yo u r H e a r t S t a r t F R 2 + f o r U s e 2-5 P h ili ps Me dic a l Sy s tem s 2 3. Press the upper Option button to activate the test. 4. OR make no selection for 10 seconds, and the selftest will start automatically if the FR2+ has been turned off for at least 5 minutes. N O...
Page 21 - Using Your HeartStart FR2+; I M P O R T A N T; Be sure to read the Warnings and Cautions on the last page; These directions apply to both the model M3860A and the model
3-1 P h ili ps Me dic a l Sy s tem s 3 3 Using Your HeartStart FR2+ Overview This chapter describes how to use the HeartStart FR2+ Defibrillator in an emergency incident. Some general things to remember are: Try to relax and stay calm. The FR2+ automatically provides appropriate voice and display pr...
Page 22 - Refer to the drawing on each pad for; For adult patients, one pad goes just below the; ON
M3860A/M3861A HEARTSTART FR2+ 3-2 P h ili ps Me dic a l Sy s tem s Step 1: Preparation Press the On/Off button to turn on the HeartStart FR2+ Defibrillator. Follow the instructions provided by the FR2+ voice and screen prompts in the order indicated. Remove clothing from the patient's upper body. Wi...
Page 23 - Step 2: ECG Analysis and Monitoring; If no shock is advised,; the FR2+ provides voice and screen prompts to tell; CPR may interfere with background monitoring. During CPR,; If a shock is advised,; the FR2+ charges to prepare for shock delivery. It; When the FR2+ is fully charged, you can disarm it at any time by
U s i n g Yo u r H e a r t S t a r t F R 2 + 3-3 P h ili ps Me dic a l Sy s tem s 3 Step 2: ECG Analysis and Monitoring Follow the instructions provided by the HeartStart FR2+ Defibrillator’s voice and screen prompts in the order indicated. As soon as the FR2+ detects that the defibrillator pads are...
Page 24 - Step 3: Shock Delivery; Press the Shock button to deliver the shock.; You must press the button for a shock to be delivered.; If you do not press the Shock button within 30 seconds of being
M3860A/M3861A HEARTSTART FR2+ 3-4 P h ili ps Me dic a l Sy s tem s Step 3: Shock Delivery Press the Shock button to deliver the shock. I M P O R T A N T : You must press the button for a shock to be delivered. The HeartStart FR2+ Defibrillator will not automatically deliver a shock. There are four w...
Page 25 - ECG Display for Ongoing Observation; away from the ECG electrodes.
U s i n g Yo u r H e a r t S t a r t F R 2 + 3-5 P h ili ps Me dic a l Sy s tem s 3 ECG Display for Ongoing Observation At the discretion of emergency care personnel, the M3860A FR2+ with ECG display enabled can also be used with the M3873A/ M3874A FR2+ ECG assessment module. The FR2+ used with the ...
Page 26 - W A R N I N G
M3860A/M3861A HEARTSTART FR2+ 3-6 P h ili ps Me dic a l Sy s tem s cleaning the FR2+ Defibrillator. (See Chapter 4, “Maintaining, Testing, and Troubleshooting Your HeartStart FR2+.”) W A R N I N G : During defibrillation, air pockets between the skin and defibrillator pads can cause patient skin bur...
Page 27 - Improper maintenance may damage the FR2+ or cause it to; Electrical shock hazard. Dangerous high voltages and currents
4-1 P h ili ps Me dic a l Sy s tem s 4 4 Maintaining, Testing, and Troubleshooting Your HeartStart FR2+ Overview This chapter provides information on HeartStart FR2+ Defibrillator maintenance, detailed descriptions of the selftests, and a guide to troubleshooting. Maintenance Maintenance of the FR2+...
Page 28 - After Using the HeartStart FR2+; Check the operation of the FR2+; Check the outside of the FR2+; Check the Status Indicator.
M3860A/M3861A HEARTSTART FR2+ 4-2 P h ili ps Me dic a l Sy s tem s After Using the HeartStart FR2+ After each use of the FR2+, perform the maintenance tasks described in the table above, as well as the following post-use checks before returning the FR2+ to service: Check the operation of the FR2+ by...
Page 29 - Check the connector socket; Cleaning the HeartStart FR2+; Operator’s checklist
Maintaining, Testing, and Troubleshooting Your HeartStart FR2+ 4-3 P h ili ps Me dic a l Sy s tem s 4 Check the connector socket to make sure that defibrillator pads are disconnected from the FR2+ when it is not in use. Check to make sure the data card tray is installed, even if a data card is not b...
Page 30 - O P E R ATO R ' S CH E C K LI ST; date; Supplies Available; • Two sets defibrillator pads,; Status Indicator
M3860A/M3861A HEARTSTART FR2+ 4-4 P h ili ps Me dic a l Sy s tem s O P E R ATO R ' S CH E C K LI ST HeartStart FR2+ Model No.: ______________________Serial No.: __________________________________ HeartStart FR2+ Location or Vehicle I D: _______________________________________________________ date sc...
Page 31 - Testing; Battery insertion selftest; will not
Maintaining, Testing, and Troubleshooting Your HeartStart FR2+ 4-5 P h ili ps Me dic a l Sy s tem s 4 Testing The HeartStart FR2+ Defibrillator has several ways of testing itself and alerting you if it finds a problem. In addition to the selftest performed each time a battery is installed, the FR2+ ...
Page 32 - information; It is recommended that the M3848A FR2+ rechargeable battery not; If the automatic part of the selftest fails:
M3860A/M3861A HEARTSTART FR2+ 4-6 P h ili ps Me dic a l Sy s tem s When you install the battery, the screen tells you whether or not a data card is installed. If so, a screen message displays how much recording time is left until the data card is full. (See Chapter 7, “Data Management and Review,” f...
Page 33 - The Status Indicator shows a flashing or solid red; Do not use the FR2+ until
Maintaining, Testing, and Troubleshooting Your HeartStart FR2+ 4-7 P h ili ps Me dic a l Sy s tem s 4 The screen displays a message that the selftest has failed. After a short time, an error code is displayed. Write down the error code and contact Philips Medical Systems for technical support. The S...
Page 34 - speaker; shock button
M3860A/M3861A HEARTSTART FR2+ 4-8 P h ili ps Me dic a l Sy s tem s The following table describes the parts of the FR2+ tested in the interactive part of the selftest and any action you are asked to take. feature test description speaker Screen prompt: CHECK SPEAKER SOUND (2 beeps) > Listen for th...
Page 35 - Periodic selftests; feature
Maintaining, Testing, and Troubleshooting Your HeartStart FR2+ 4-9 P h ili ps Me dic a l Sy s tem s 4 When the interactive part of the battery insertion selftest is complete, the FR2+ turns off and goes to standby mode to be ready for use. If proper operation of all features has not been verified in...
Page 36 - Battery History
M3860A/M3861A HEARTSTART FR2+ 4-10 P h ili ps Me dic a l Sy s tem s TESTS — how many tests have been run. Four figures are shown: daily (upper left), weekly (upper right), and monthly (lower left) periodic selftests, and battery insertion selftests (lower right). REV — device language, model, and so...
Page 37 - Troubleshooting Guide; Status indicator summary; Perform CPR (if needed) any time there is a delay before the FR2+; Recommended action during an emergency; After completing emergency use; Flashing black hourglass; A self-test error has occurred or the battery is low or depleted.; Solid red X
Maintaining, Testing, and Troubleshooting Your HeartStart FR2+ 4-11 P h ili ps Me dic a l Sy s tem s 4 Troubleshooting Guide Status indicator summary N O T E : Perform CPR (if needed) any time there is a delay before the FR2+ can be used. Recommended action during an emergency If the status indicato...
Page 38 - Troubleshooting during patient use; symptom; The defibrillator pads:
M3860A/M3861A HEARTSTART FR2+ 4-12 P h ili ps Me dic a l Sy s tem s If neither of these actions clears the problem, do not use the FR2+. Attend to the patient, providing CPR if needed, until emergency medical personnel arrive. Troubleshooting during patient use symptom possible cause recommended act...
Page 39 - or; The patient is being moved or jostled.; Voice and screen prompts:
Maintaining, Testing, and Troubleshooting Your HeartStart FR2+ 4-13 P h ili ps Me dic a l Sy s tem s 4 Voice prompts: ANALYZING INTERRUPTED or CANNOT ANALYZE or STOP ALL MOTION • The patient is being moved or jostled. • Radio or electrical sources are interfering with ECG analysis. • The environment...
Page 40 - General troubleshooting; new
M3860A/M3861A HEARTSTART FR2+ 4-14 P h ili ps Me dic a l Sy s tem s General troubleshooting symptom possible cause recommended action Status Indicator: FLASHING RED X Audio signal: CHIRPING • The energy remaining in the battery is low. • The FR2+ has been stored outside the recommended temperature r...
Page 41 - Install a; function
Maintaining, Testing, and Troubleshooting Your HeartStart FR2+ 4-15 P h ili ps Me dic a l Sy s tem s 4 Status Indicator: SOLID RED X Audio signal: NONE • The battery is missing or completely depleted. • The training & administration pack is being used in the administration function (the solid re...
Page 43 - Clinical and Safety Considerations; Clinical Considerations; Indications; Contraindications; Safety Considerations; D A N G E R
5-1 P h ili ps Me dic a l Sy s tem s 5 5 Clinical and Safety Considerations Clinical Considerations Indications The HeartStart FR2+ is indicated for use on victims of sudden cardiac arrest exhibiting the following signs: Unresponsiveness Absence of normal breathing The HeartStart FR2+ is intended fo...
Page 44 - — conditions, hazards, or unsafe practices that can; safety level
M3860A/M3861A HEARTSTART FR2+ 5-2 P h ili ps Me dic a l Sy s tem s C A U T I O N — conditions, hazards, or unsafe practices that can result in minor personal injury, damage to the HeartStart FR2+, or loss of data stored in the device. THESE SAFETY CONSIDERATIONS ARE DIVIDED INTO FOUR GROUPS: SAFETY ...
Page 45 - Do not use the M3849A charger on aircraft.
C l i n i c a l a n d S a f e t y C o n s i d e r a t i o n s 5-3 P h ili ps Me dic a l Sy s tem s 5 safety level possible improper device performance WARN I NG Prolonged or aggressive CPR to a patient with defibrillator pads attached can damage the pads. Replace the defibrillator pads if they are d...
Page 46 - Defibrillation warnings and cautions
M3860A/M3861A HEARTSTART FR2+ 5-4 P h ili ps Me dic a l Sy s tem s Defibrillation warnings and cautions CAUTION Use only Philips-approved data cards. The HeartStart FR2+ may perform improperly if non-approved accessories are used. safety level possible electrical interference with ECG monitoring WAR...
Page 47 - Monitoring cautions; WARN I NG; The HeartStart FR2+ advanced mode’s; feature is; CAUTION
C l i n i c a l a n d S a f e t y C o n s i d e r a t i o n s 5-5 P h ili ps Me dic a l Sy s tem s 5 Monitoring cautions Maintenance cautions safety levels possible burns and ineffective energy WARN I NG Do not allow the defibrillator pads to touch each other or other ECG electrodes, lead wires, dre...
Page 49 - Setup and Advanced Mode Features; Setup Overview; Non-protocol parameters; parameter; speaker volume; (ECG display cannot be changed from default OFF; OFF
6-1 P h ili ps Me dic a l Sy s tem s 6 6 Setup and Advanced Mode Features Setup Overview The “setup” of the HeartStart FR2+ Defibrillator is made up of several programmable aspects, or parameters, of FR2+ operation. Some setup parameters govern specific features that are not related to the patient c...
Page 50 - Automatic protocol parameters; boldface type; the
M3860A/M3861A HEARTSTART FR2+ 6-2 P h ili ps Me dic a l Sy s tem s Automatic protocol parameters The HeartStart FR2+ is designed to follow an automatic protocol that guides you through patient treatment with the defibrillator. The default settings for programmable parameters used in the automatic pr...
Page 51 - NSA action
S e t u p a n d A d v a n c e d M o d e F e a t u r e s 6-3 P h ili ps Me dic a l Sy s tem s 6 NSA action (minutes) MONITOR , 0.5, 1.0, 1.5, 2.0, 2.5, 3.0 MONITOR Sets how the FR2+ behaves following a NO SHOCK ADVISED (NSA) decision: MONITOR — directs the FR2+ to monitor the patient’s ECG following ...
Page 52 - Manual override parameters; advanced
M3860A/M3861A HEARTSTART FR2+ 6-4 P h ili ps Me dic a l Sy s tem s Manual override parameters The HeartStart FR2+ provides several ways of overriding the automatic protocol. The parameters in the following table are used to enable different kinds of manual override. parameter settings default descri...
Page 53 - Using Setup Features; of the CPR pause and return to analyzing. If either the; CPR Timer; or the; NSA Action; setting is programmed to
S e t u p a n d A d v a n c e d M o d e F e a t u r e s 6-5 P h ili ps Me dic a l Sy s tem s 6 Using Setup Features N O T E : To move around the menus displayed, use the Option buttons as follows:• Press the LOWER Option button to move the highlight bar from one item to another on the menu. • Press ...
Page 55 - Receiving setup; Reference Guide
S e t u p a n d A d v a n c e d M o d e F e a t u r e s 6-7 P h ili ps Me dic a l Sy s tem s 6 Use the infrared communications feature of the FR2+ to receive the revised setup from another FR2+. (Instructions are provided later in this chapter.) Use the infrared communications feature of the FR2+ to...
Page 56 - Reading setup; Sending and Receiving Clock Settings
M3860A/M3861A HEARTSTART FR2+ 6-8 P h ili ps Me dic a l Sy s tem s “receiving” FR2+ displays a SETUP COMPLETE message. Your HeartStart FR2+ immediately uses the new setup. Receiving setup from a computer running HeartStart Event Review software is discussed in the directions for use provided with He...
Page 57 - Using Advanced Mode Features
S e t u p a n d A d v a n c e d M o d e F e a t u r e s 6-9 P h ili ps Me dic a l Sy s tem s 6 uninterrupted view of the “eye” in the other. (See the diagram on the back of the first page of this manual.) The two devices should be no more than 1 meter apart. 5. Select SEND TIME from the CLOCK screen...
Page 58 - Using the manual analyze feature
M3860A/M3861A HEARTSTART FR2+ 6-10 P h ili ps Me dic a l Sy s tem s model used. For the M3860A, the user can select ANALYZE , CHARGE , or OFF . For the M3861A the user can select only ANALYZE or OFF . (Detailed directions for use are supplied with the training & administration pack.) C A U T I O...
Page 59 - Using the manual charge feature (M3860A only)
S e t u p a n d A d v a n c e d M o d e F e a t u r e s 6-11 P h ili ps Me dic a l Sy s tem s 6 In the M3861A HeartStart FR2+, the patient’s ECG is not displayed; in the M3860A, the display includes the patient’s ECG and heart rate. Press the lower Option button ( ANALYZE ) to initiate rhythm analys...
Page 61 - Data Management and Review; Recording data in internal memory; To record incident data on a data card, the data card; The FR2+ is designed to be used only with Philips-approved
7-1 P h ili ps Me dic a l Sy s tem s 7 7 Data Management and Review Overview The HeartStart FR2+ is designed to make it easy to manage incident data. Some information is automatically stored in the internal memory of the HeartStart FR2+. More detailed data can be stored on a data card if desired. Th...
Page 62 - To install a data card:; for less
M3860A/M3861A HEARTSTART FR2+ 7-2 P h ili ps Me dic a l Sy s tem s To install a data card: 1. Make sure the data card is clean and dry. 2. Load the data card into its plastic tray, with the tray’s “tongue” fitting over the matching yellow area on the data card. The label on the card should face up. ...
Page 63 - To replace a data card:; before; Reviewing Incident Data; Reviewing data from internal memory
D a t a M a n a g e m e n t a n d R e v i e w 7-3 P h ili ps Me dic a l Sy s tem s 7 To replace a data card: I M P O R T A N T : You must turn the FR2+ off (return it to standby mode) before you remove the data card, to ensure that no incident data are lost. 1. Press the On/Off button to turn off th...
Page 64 - Reviewing data from a data card
M3860A/M3861A HEARTSTART FR2+ 7-4 P h ili ps Me dic a l Sy s tem s Reviewing data from a data card If a data card is installed when the HeartStart FR2+ is turned on for use during an incident, the HeartStart FR2+ automatically records detailed information on the data card. To review this information...
Page 67 - Accessories for the HeartStart FR2+; HeartStart Accessories
A-1 P h ili ps Me dic a l Sy s tem s A A Accessories for the HeartStart FR2+ HeartStart Accessories Accessories for the HeartStart FR2+ available separately from Philips Medical Systems include the following: Spare M3863A FR2 standard battery (recommended) DP2/DP6 adult defibrillator pads M3870A FR2...
Page 68 - Suggested Additional Items
M3860A/M3861A HEARTSTART FR2+ A-2 P h ili ps Me dic a l Sy s tem s Suggested Additional Items It may be useful to keep some additional items with your HeartStart FR2+ for use if needed when an incident occurs. Some suggested supplies include: a pair of paramedic’s shears or scissors * a disposable r...
Page 69 - Technical Reference Manual; for the FR2; HeartStart FR2+ Defibrillator Specifications; Physical; With data card tray and battery installed, meets IEC 529 class IP54.
P h ili ps Me dic a l Sy s tem s B B-1 B Technical Specifications The specifications for the HeartStart FR2+ provided in this chapter apply to both the M3860A and M3861A, unless otherwise noted. Additional information can be found in the Technical Reference Manual for the FR2 Defibrillator, located ...
Page 70 - Waveform Parameters; is the peak
M3860A/M3861A HEARTSTART FR2+ B-2 P h ili ps Me dic a l Sy s tem s Defibrillator EMI (Immunity) Meets EN 60601-1-2 limits (1993), method EN 61000-4-3:1998 Level 2 (3 V/m and 10 V/m at 26 MHz to 1 GHz). Aircraft: Method Meets RTCA/DO-160D:1997 Section 21 (Category M - Charging). category nominal spec...
Page 71 - Energy; newborn; Charge Control; category
Te c h n i c a l S p e c i f i c a t i o n s B-3 P h ili ps Me dic a l Sy s tem s B Energy Using adult defibrillator pads: 150 J nominal into a 50 ohm load.Using infant/child reduced-energy defibrillator pads: 50 J nominal into a 50 ohm load. Sample pediatric energy doses: age energy dose newborn 14...
Page 73 - Display
Te c h n i c a l S p e c i f i c a t i o n s B-5 P h ili ps Me dic a l Sy s tem s B Display Controls and indicators Pacemaker Detection On detection of a pacemaker (in advanced mode or with M3873A/M3874A FR2+ ECG assessment module), provides screen display of PACEMAKER DETECTED alert. M3860A include...
Page 74 - Accessories Specifications
M3860A/M3861A HEARTSTART FR2+ B-6 P h ili ps Me dic a l Sy s tem s Accessories Specifications M3863A FR2 standard battery Controls On/Off buttonShock buttonOption buttons LED Indicators Connector socket LE D, flashes to indicate socket location. LED is covered when defibrillator pad connector is pro...
Page 75 - flashing red X on the front panel of the FR2+ (When low battery
Te c h n i c a l S p e c i f i c a t i o n s B-7 P h ili ps Me dic a l Sy s tem s B (Optional) M3848A FR2+ rechargeable battery (Optional) M3849A charger M3870A and DP2/DP6 defibrillator pads category nominal specifications Battery Type 11.3 VDC, 6.5 Ah, lithium ion. Rechargeable cell using the M384...
Page 77 - Application
Te c h n i c a l S p e c i f i c a t i o n s B-9 P h ili ps Me dic a l Sy s tem s B (Optional) M3873A/M3874A FR2+ ECG assessment module category nominal specifications Application For use with the FR2+ M3860A with ECG display enabled and running version 1.5 software or higher (denoted by FR2+ on the...
Page 79 - Glossary of Symbols and Controls; Control panel and back label; Printed on recycled paper.; symbol; Upper and lower Option buttons.
C-1 P h ili ps Me dic a l Sy s tem s C C Glossary of Symbols and Controls Instructions for Use HeartStart FR2+ M3860A and M3861A Defibrillator Symbols and Controls Control panel and back label symbol description Meets the requirements of the European medical device directives. Printed on recycled pa...
Page 80 - HeartStart FR2+ display screen; Refer to operating instructions.; HR XXX; Number of shocks delivered.; TEMPERATU R E
M3860A/M3861A HEARTSTART FR2+ C-2 P h ili ps Me dic a l Sy s tem s HeartStart FR2+ display screen High voltage. Refer to operating instructions. Meets the requirements of the European medical device directives. This product has passed relevant safety tests by CSA, a nationally recognized test lab. I...
Page 81 - Status indicator; Accessories Symbols; Solid red; Pouch contents: one pair of defibrillator pads.
G l o s s a r y o f S y m b o l s a n d C o n t r o l s C -3 P h ili ps Me dic a l Sy s tem s C Status indicator Accessories Symbols M3870A and DP2/DP6 defibrillator pads symbol description Flashing black hourglass. Ready for use. Solid red X . Not ready for use. (See Chapter 4, “Maintaining, Testin...
Page 83 - Protect from moisture.; Do not mutilate or open.
G l o s s a r y o f S y m b o l s a n d C o n t r o l s C -5 P h ili ps Me dic a l Sy s tem s C M3863A standard battery and M3848A FR2+ rechargeable battery Box contents = 2 pouches. (DP2) Box contents = 6 pouches. (DP6) This side up. Fragile. Protect from moisture. Printed on recycled paper. symbol...
Page 86 - Place the electrodes as shown.
M3860A/M3861A HEARTSTART FR2+ C-8 P h ili ps Me dic a l Sy s tem s M3873A/M3874A FR2+ ECG assessment module M3864A training & administration pack symbol description Use the cable and electrodes before the respective dates shown on this label. Ship and store the product within the temperature ran...
Page 89 - Glossary of Terms; SMART
D-1 P h ili ps Me dic a l Sy s tem s D D Glossary of Terms The terms listed in this Glossary are defined in the context of the HeartStart FR2+ and its use. advanced mode ..................... A programmable treatment mode that permits an authorized user to control when the FR2+ starts rhythm analysi...
Page 90 - as detected through defibrillator pads.; to a shockable rhythm.
M3860A/M3861A HEARTSTART FR2+ D-2 P h ili ps Me dic a l Sy s tem s defibrillation shock. Use only DP2/ DP6, M3870A, M3713A, and M3716A defibrillator pads with the HeartStart FR2+. ECG .......................................... Electrocardiogram, a display or printout of the electrical rhythm of the ...
Page 91 - analysis of the patient’s heart rhythm.
G l o s s a r y o f Te r m s D -3 P h ili ps Me dic a l Sy s tem s D non-shockablerhythm ................................ A heart rhythm that the HeartStart FR2+ determines is not appropriate for shock delivery. NSA .......................................... No Shock Advised decision, made by the He...
Page 92 - shockable heart rhythms.; rhythm and determine whether a shock is advised.
M3860A/M3861A HEARTSTART FR2+ D-4 P h ili ps Me dic a l Sy s tem s rhythm analysis ...................... See “ SMART analysis.” send (data) ............................. A feature of the HeartStart FR2+ that allows use of its infrared (I R) communications port to send data directly to another FR2+ ...
Page 93 - waveform
G l o s s a r y o f Te r m s D -5 P h ili ps Me dic a l Sy s tem s D training &administration pack .......... An optional accessory for the FR2+ that enables training and administrative functions. The integral battery should be charged only using the M3855A charger. waveform .......................
Page 95 - IN; Index
I-1 P h ili ps Me dic a l Sy s tem s IN D E X Index A accessories additional, recommended A-2 adult defibrillator pads A-1 available to order A-1 charger for M3848A rechargeable battery A-1 charger for training & administration pack A-1 data card A-1 data card tray A-1 fabric carrying case A-1 f...
Page 97 - see
I n d e x I-3 P h ili ps Me dic a l Sy s tem s IN D E X G glossary of terms D-1 of symbols and controls C-1 H heart rhythms shockable B-4 HeartStart M3870A FR2 infant/child reduced-energy defibrillator pads A-1 hourglass Status Indicator C-2 how to install the battery 2-1 how to remove the data card...
Page 102 - Philips Medical Systems is part; Philips Medical Systems
P O W E R T O S A V E A L I F E 011120-0010 M 3 8 6 0 - 9 1 9 0 0 Philips Medical Systems is part of Royal Philips Electronics Philips Medical Systems USA Philips Medical Systems2301 Fifth Avenue, Suite 200Seattle, WA 98121, USA(800) 263-3342 Canada Philips Medical Systems281 Hillmount RoadMarkham, ...