Milwaukee MW600 - Manual
Milwaukee MW600 – Manual, read for free online in PDF format. We hope this helps you resolve any issues you may have. If you have further questions, please contact us through the contact form.
Table of Contents:
- Page 3 – CALIBRATION PROCEDURE:
- Page 6 – ALTITUDE COMPENSATION:
- Page 7 – SALINITY COMPENSATION:
- Page 8 – To replace the membrane; or refill it with electrolyte, see page 1.
- Page 9 – Note; : In order to obtain accurate and stable measurements, it
- Page 12 – Milwaukee
Milwaukee
USER MANUAL
PORTABLE
DISSOLVED OXYGEN METER
MODEL:
MW600
Smart DO Meter
"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.
Other Manuals for Milwaukee MW600
Summary
PROBE POLARIZATION: • A dissolved oxygen probe must be polarized to function properly. • To polarize the probe, the fully assambled probe must be connected to the meter and the meter must be on. • During polarization (and during measurement), approximately 800 mV is applied to the cathode and anode ...
5 °C 02468 10121416182022242628303234363840 0 m 14.613.813.112.411.811.310.810.3 9.99.59.18.78.48.17.87.57.37.16.86.66.4 300 m 14.113.312.712.011.410.910.4 9.99.79.28.88.48.17.87.57.27.16.96.66.46.2 600 m 13.612.912.211.611.010.510.1 9.6 9.28.78.58.17.87.57.37.06.86.66.36.26.0 1200 m 12.712.011.410....
6 SALINITY COMPENSATION: The table below illustrates the change in the solubility of oxygenin air saturated water as a result of chloride concentration orsalinity.The table can also be used to correct the displayed measurement.If you are making measurements in salt water and know thechloride concent...