Page 2 - Notice; Restricted Rights Legend
ii HP 8517B S-Parameter Test Set Manual Notice The information contained in this document is subject to change withoutnotice. Hewlett-Packard makes no warranty of any kind with regard to this material,including, but not limited to, the implied warranties of merchantability andfitness for a particula...
Page 3 - What You’ll Find in This Manual...; HP 8517B S-Parameter Test Set; Test set description and installation information; Chapters 3 and 4; Principles of operation and specifications
HP 8517B S-Parameter Test Set Manual iii What You’ll Find in This Manual… HP 8517B S-Parameter Test Set Chapters 1 and 2 • Test set description and installation information Chapters 3 and 4 • Principles of operation and specifications Chapters 5, 6, and 7 • Troubleshooting, replacement procedures, a...
Page 4 - Warranty; Certification
iv HP 8517B S-Parameter Test Set Manual Warranty Certification Hewlett-Packard Company certifies that this product met its publishedspecifications at the time of shipment from the factory. Hewlett-Packardfurther certifies that its calibration measurements are traceable to theUnited States National I...
Page 5 - Assistance
HP 8517B S-Parameter Test Set Manual v CONSEQUENTIAL DAMAGES, WHETHER BASED ON CONTRACT,TORT, OR ANY OTHER LEGAL THEORY. Assistance Product maintenance agreements and other customer assistance agreementsare available for Hewlett-Packard products. For assistance, call your local Hewlett-Packard Sales...
Page 6 - Service and Support
vi HP 8517B S-Parameter Test Set Manual Service and Support Any adjustment, maintenance, or repair of this product must be performedby qualified personnel. Contact your customer engineer through your localHP Service Center. You can find a list of HP Service Centers on the web at http://www.hp.com/go...
Page 7 - Safety and Regulatory Information; WARNING; CAUTION; Instrument Markings
HP 8517B S-Parameter Test Set Manual vii Safety and Regulatory Information Review this product and related documentation to familiarize yourself withsafety markings and instructions before you operate the instrument. Thisproduct has been designed and tested in accordance with internationalstandards....
Page 8 - Before Applying Power
viii HP 8517B S-Parameter Test Set Manual Safety EarthGround This is a Safety Class I product (provided with a protective earthingterminal). An uninterruptible safety earth ground must be provided from themain power source to the product input wiring terminals, power cord, orsupplied power cord set....
Page 9 - Typeface Conventions; Italics
HP 8517B S-Parameter Test Set Manual ix Typeface Conventions Not all of the following conventions may appear within this manual,however, refer to this listing whenever you encounter on of the special fontcharacters. Italics • Used to emphasize important information: Use this software only with the H...
Page 11 - Contents; Getting Started
Contents HP 8517B S-Parameter Test Set Manual Contents-1 Notice . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ii Restricted Rights Legend . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ii What You’ll Find in This Manual… . ...
Page 12 - Installation
Contents-2 HP 8517B S-Parameter Test Set Manual 2. Installation Initial Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2 Operating Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2Storing the Test Set . . . . . ....
Page 13 - Specifications; and S
HP 8517B S-Parameter Test Set Manual Contents-3 Making Operational Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13Performance Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14Using Anti-Rotation Clamps . . . . . . . . . . . . . ...
Page 14 - HP 8517B Test Set Replacement Procedures
Contents-4 HP 8517B S-Parameter Test Set Manual Unratioed Power Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14 Figure 5-9. Diagram of IF Signal Path . . . . . . . . . . . . . . . . . . . . . 5-14Isolating Signal Path Problems . . . . . . . . . . . . . . . . . . . . . . . ....
Page 15 - Figure 7-3. Major Assembly Locations in the Standard
HP 8517B S-Parameter Test Set Manual Contents-5 Bias Tees Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11 A7 and A9 Bias Tees . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11 Test Port Coupler Replacement . . . . . . . . . . . . . . . . . ...
Page 17 - How to Use This Manual
HP 8517B S-Parameter Test Set Manual 1-1 1 Getting Started This is the operating and service manual for the HP 8517B S-parameter testset. It is to be used in conjunction with the HP 8510C Network AnalyzerOperating and Service Manual. Together, these manuals provideinformation needed to configure the...
Page 18 - Test Set Description; measurements are required.
1-2 HP 8517B S-Parameter Test Set Manual Getting Started Test Set Description Test Set Description The HP 8517B test set configured with an HP 8510B/C network analyzerand an HP 8360 series source, creates a system capable of makingS-parameter measurements from 45 MHz to 50 GHz. The system isparticul...
Page 20 - Options for the HP 8517B test set are described in Table 1-1, below.
1-4 HP 8517B S-Parameter Test Set Manual Getting Started HP 8517B Test Set Options HP 8517B Test Set Options Options for the HP 8517B test set are described in Table 1-1, below. Table 1-1 HP 8517B Options and Descriptions Options HP 8517B Test Set Option Description Option 001 Adds IF switching capa...
Page 21 - Verifying Test Set Operation; Test Set Operation Verification
HP 8517B S-Parameter Test Set Manual 1-5 Getting Started Verifying Test Set Operation Verifying Test Set Operation The test set is designed to operate with an HP 8510C network analyzer. Youcan verify its operation by: Table 1-2 Test Set Operation Verification Checking System Operation Perform a syst...
Page 22 - Maintaining Measurement Accuracy
1-6 HP 8517B S-Parameter Test Set Manual Getting Started Maintaining Measurement Accuracy Maintaining Measurement Accuracy Precision measurements rely on a precision calibration of the networkanalyzer. As a general rule, the shorter the time lapse between a calibrationand a device measurement, the m...
Page 23 - Instruments Supported by This Manual; Instrument Serial-Number Label Example
HP 8517B S-Parameter Test Set Manual 1-7 Getting Started Instruments Supported by This Manual Instruments Supported by This Manual A serial number label is attached to the rear panel of the test set. There aretwo parts to the serial number: a prefix, and a suffix. See Figure 1-4, andread the followi...
Page 24 - Instrument Firmware Compatibility; Source Firmware Revision Compatibility Matrix
1-8 HP 8517B S-Parameter Test Set Manual Getting Started Instrument Firmware Compatibility Instrument Firmware Compatibility To use the HP 8517B test set at frequencies up to 50 GHz, use anHP 83650B or HP 83651B source. If your network analyzer or source do not meet the required frequencyranges for ...
Page 25 - Service and Support Options Available; NOTE; System installation is not included.
HP 8517B S-Parameter Test Set Manual 1-9 Getting Started Service and Support Options Available Service and Support Options Available A variety of service and support products are available for coverage ofrepair, calibration, and verification issues. Contact your HP sales and serviceengineer for deta...
Page 26 - Accessories; Accessories Available; HP 8517B Test Set Accessories Available
1-10 HP 8517B S-Parameter Test Set Manual Getting Started Accessories Accessories The accessories supplied with the test set, including part numbers, are listedin “Accessories Supplied” in this section and in the “Replaceable Parts List”of this manual. Accessories Available Table 1-5 describes acces...
Page 27 - HP 8517B Test Set Accessories Available (Continued)
HP 8517B S-Parameter Test Set Manual 1-11 Getting Started Accessories NOTE For more information about other 2.4 mm adapters, refer to the “2.4 mmAdapters and Calibration Accessories,” Operating Note, (HP part number11900-90003). HP 85135E 7 mm Flexible Test Port Return Cable Used when measuring one ...
Page 28 - Recommended Test Equipment; Recommended Test Equipment
1-12 HP 8517B S-Parameter Test Set Manual Getting Started Recommended Test Equipment Recommended Test Equipment Additional equipment and accessories required for use with the test set arelisted in Table 1-6. The table lists which items are required to verify theperformance of the test set and which ...
Page 29 - Operating and Safety Precautions; ESD Sensitive Assemblies; Operating Precautions; Do not exceed the input power levels listed below:; Safety Precautions; Maximum Input Power Levels
HP 8517B S-Parameter Test Set Manual 1-13 Getting Started Operating and Safety Precautions Operating and Safety Precautions CAUTION ESD Sensitive Assemblies T est set assemblies are very sensitive to damage from electrostatic discharge. They may or may not continue to function if subjected to electr...
Page 32 - Initial Inspection; C. Keep relative humidity to less than; Storing the Test Set; Acceptable storage temperatures range from –40; Accessories Supplied
2-2 HP 8517B S-Parameter Test Set Manual Installation Initial Inspection Initial Inspection Inspect the shipping container and cushioning materials for damage. If thereis damage, keep the container until you have checked the contents forcompleteness. If shipping materials are damaged, complete the p...
Page 33 - Accessories Supplied with the HP 8517B Test Set; Accessories in Figure 2-1
HP 8517B S-Parameter Test Set Manual 2-3 Installation Initial Inspection Figure 2-1 Accessories Supplied with the HP 8517B Test Set CAUTION Assemblies in the test set are extremely sensitive to damage by staticelectricity. They may or may not continue to function if subjected to anelectrostatic disc...
Page 34 - Installing the Test Set Into a System Rack; Recommended Static-Free Workstation Configuration
2-4 HP 8517B S-Parameter Test Set Manual Installation Installing the Test Set Into a System Rack Installing the Test Set Into a System Rack The HP 85043C system rack is the recommended model to use. Torack-mount the test set into a system configured with the HP 8510C, refer tothe “Installation” chap...
Page 35 - Configuring the Test Set in a System; Configuring an HP 8517B Test Set in a System
HP 8517B S-Parameter Test Set Manual 2-5 Installation Configuring the Test Set in a System Configuring the Test Set in a System Figure 2-3 illustrates the cable connections for installing an HP 8517B in asystem. Refer to the information that follows to install the test set. Figure 2-3 Configuring an...
Page 36 - Connection Instructions and Connector Descriptions
2-6 HP 8517B S-Parameter Test Set Manual Installation Configuring the Test Set in a System Connecting the SystemCables Replacing O-Rings inAnti-Rotation Clamps The anti-rotation clamp’s internal O-ring is a field-replaceable part. There isno need to disassemble the anti-rotation clamp. When the O-ri...
Page 37 - Packaging the Test Set; To repackage the test set, use its original factory packaging.
HP 8517B S-Parameter Test Set Manual 2-7 Installation Packaging the Test Set Packaging the Test Set To repackage the test set, use its original factory packaging. Refer to the test set using its complete model and serial numbers in anycorrespondence to an HP Sales and Service Office. Containers and ...
Page 39 - Operating the HP 8517B Test Set; Front-Panel Features of the HP 8517B Test Set
HP 8517B S-Parameter Test Set Manual 3-1 3 Operating the HP 8517B Test Set This chapter contains illustrations and descriptions of the HP 8517B front-and rear-panel features. Information about operating multiple test sets(Option 001) is also provided. Front-Panel Features Figure 3-1 Front-Panel Feat...
Page 40 - Rear-Panel Features of the HP 8517B Test Set
HP 8517B S-Parameter Test Set Manual 3-2 Rear-Panel Features Figure 3-2 Rear-Panel Features of the HP 8517B Test Set (6) a1 LED This LED indicates that the test set is internally switched to the S 11 or S 21 mode and source power is switched to Port 1. (7) a2 LED This LED indicates that the test set...
Page 41 - Controlling Multiple Test Sets; apply its IF signal to the HP 8510C, or it can
HP 8517B S-Parameter Test Set Manual 3-3 Controlling Multiple Test Sets HP 851X Series, Option 001 test sets allow an HP 8510C to alternatelycontrol up to four test sets connected to it. For example: • While a measurement is underway on test set number 1 (equipped withOption 001), a test device can ...
Page 42 - RF and IF Switching with Two Test Sets
HP 8517B S-Parameter Test Set Manual 3-4 Figure 3-3 RF and IF Switching with Two Test Sets Table 3-3 RF and IF Switch Settings in Figure 3-3, Above New ADDRESS of Test Set 1, 2 1. Not all system connections are illustrated. 2. In dual source configurations, the second source may be multiplexed in a ...
Page 43 - Multiple Test-Set Connections; Selecting a Test Set
HP 8517B S-Parameter Test Set Manual 3-5 Multiple Test-Set Connections For dual test-set configurations, set each rear panel address switch on eachtest set to the address shown in Figure 3-3. Refer to Figure 3-4 on page 3-7for the configuration of two or more test sets. • Use the supplied test set i...
Page 44 - Selecting a Test Set When Multiple Units are Configured
HP 8517B S-Parameter Test Set Manual 3-6 Table 3-4 Selecting a Test Set When Multiple Units are Configured Test Set IF Switching A test set is selected via the built-in capability of the analyzer to generate an addressed command to each test set. Each time you change the {HP-IB ADDRESS} {TEST SET} f...
Page 45 - RF and IF Switching with Four Test Sets
HP 8517B S-Parameter Test Set Manual 3-7 Figure 3-4 RF and IF Switching with Four Test Sets Table 3-5 Coaxial Switch Settings for Figure 3-4, Above New ADDRESS of Test Set 1, 2, 3 1. Not all system connections are shown. 2. In dual source configurations, the second source can be multiplexed in a sim...
Page 46 - Measurement Calibration; sets the address of test set
HP 8517B S-Parameter Test Set Manual 3-8 Measurement Calibration After selecting the active test set, complete the system calibration procedureas usual. When you select a different test set, make sure that you recall thecal set that applies to that test set. NOTE Since the cal-set limited instrument...
Page 47 - Measuring High-Power Devices with Option 004
HP 8517B S-Parameter Test Set Manual 3-9 Measuring High-Power Devices with Option 004 The HP 8517B, Option 004 test set enables devices with output power up to30 dBm (1 watt) to be measured at Port 2, without adding externalattenuators. Reverse dynamic range (S 12 ) degrades by approximately 3 dB. W...
Page 51 - Making Operational Checks
HP 8517B S-Parameter Test Set Manual 3-13 Making Operational Checks To check the operation of multiple test-set configurations, do the following: 1. Connect a device with a known response to test set number 1. Press the following HP 8510C front-panel keys: INSTRUMENT STATE [LOCAL] AUXILIARY MENUS [S...
Page 52 - Performance Verification
HP 8517B S-Parameter Test Set Manual 3-14 Performance Verification Standard system performance verification procedures verify the operation ofan Option 001 test set used as test set number 1. To verify the performance of a different test set in the configuration, enter itsHP-IB address (as explained...
Page 53 - Using Anti-Rotation Clamps; Attach the first clamp; connector to be used.; Wiggle the clamp to ease it over the connector.; Do not twist the cable as you attach it to the test port.
HP 8517B S-Parameter Test Set Manual 3-15 Using Anti-Rotation Clamps During the performance verification procedure, use anti-rotation clamps tosecure RF connections at the test ports of each test set. While installed, eachclamp stabilizes the connector on RF cable to the large nut on the test set’sR...
Page 54 - Using a Torque Wrench Correctly; Positioning the thumb-screw
HP 8517B S-Parameter Test Set Manual 3-16 CAUTION Important! The test set RF connector becomes loosened easily. Hold the RFcable securely throughout the remainder of this procedure. Do not allow thecable to rotate. Figure 3-5 Using a Torque Wrench Correctly Positioning the thumb-screw See Figure 3-6...
Page 55 - Visually Aligning Clamp and Nut Flats; Positioning the connector; Mating the Clamp and Nut Flats
HP 8517B S-Parameter Test Set Manual 3-17 NOTE The flats may actually be in any orientation, with respect to the front panel. Figure 3-6 Visually Aligning Clamp and Nut Flats Positioning the connector See Figure 3-7. Maneuver the clamp over the RF connector and onto the testport connector. 1. Hold t...
Page 56 - Aligning the thumb-screw; Push the clamp toward the test set front panel.; Attaching the second clamp
3-18 HP 8517B S-Parameter Test Set Manual Operating the HP 8517B Test Set Using Anti-Rotation Clamps Aligning the thumb-screw See Figure 3-8. Ensure that the thumb-screw is aligned with the counter-sinkhole on the clamp’s body. 1. Push the clamp toward the test set front panel. 2. Finger-tighten the...
Page 58 - Mechanical Specifications; Front Panel Test Ports
HP 8517B S-Parameter Test Set Manual 4-2 Specifications 4 Specifications Specifications describe the warranted performance of the instrument. Theelectrical specifications of the test set when used with an HP 8510 networkanalyzer, are defined in the HP 8510C On-Site Service Manual. Mechanical Specifi...
Page 59 - Rear Panel RF Connectors
4-3 HP 8517B S-Parameter Test Set Manual Specifications Rear Panel RF Connectors Connector type precision 2.4 mm female Recommended connection torque 90 N-cm (8 in-lb), maximum for precision-2.4 mm hardware Damage input level 15 dBm 1 Nominal connector-nut size 8 mm User 1 Power Levels for Reference...
Page 61 - Troubleshooting the Test Set
HP 8517B S-Parameter Test Set Manual 5-1 5 Troubleshooting the Test Set The troubleshooting strategy for the HP 8517B S-Parameter test set is asystematic sequence of procedures. The information may be used aftersystem-level troubleshooting has identified the test set as the probleminstrument. Use th...
Page 62 - Theory of Operation; The RF Source Power
5-2 HP 8517B S-Parameter Test Set Manual Troubleshooting the Test Set Theory of Operation Theory of Operation RF paths for Port 1 and Port 2 are similar. This section discusses in detail thesource power signal, as it is routed via the RF IN connector, through the testset, and to Port 1 for S 11 and ...
Page 63 - The Test Signal Path; The S
HP 8517B S-Parameter Test Set Manual 5-3 Troubleshooting the Test Set Theory of Operation The Reference SignalPath • The reference signal is routed from the switch/splitter through the20-7 dB tapered pad (AT1) to attenuate any VTO feed-through signals. • The signal is then applied to A10, the a1 sam...
Page 64 - The LO Signal Control; The HP 8510C Network Analyzer; Electrostatic Discharge Sensitive (ESD) Assemblies
5-4 HP 8517B S-Parameter Test Set Manual Troubleshooting the Test Set Theory of Operation • If the test set is equipped with Option 001 (multiple test set capability),its begin state is inactive. That is, rather than using the test set’s internalIF, a switch is set in the test set to accept an IF fr...
Page 65 - Troubleshooting Sequence; Figure 5-2 Troubleshooting Flowchart
HP 8517B S-Parameter Test Set Manual 5-5 Troubleshooting the Test Set Troubleshooting Sequence Troubleshooting Sequence Use this troubleshooting flowchart, along with the procedures in thefollowing section, to help isolate faulty assemblies. Figure 5-2 Troubleshooting Flowchart
Page 66 - Equipment Required, But Not Supplied
5-6 HP 8517B S-Parameter Test Set Manual Troubleshooting the Test Set Troubleshooting Sequence Equipment NeededBut Not Supplied Gather the equipment listed in Table 5-1 to use for troubleshooting the testset (these items are not supplied with the instrument at shipment). Table 5-1 Equipment Required...
Page 67 - Troubleshooting Procedures; A15 Primary Regulator; Power Supply Voltages to A15
HP 8517B S-Parameter Test Set Manual 5-7 Troubleshooting the Test Set Troubleshooting Procedures Troubleshooting Procedures The troubleshooting procedures in this chapter are listed below: • “A15 Primary Regulator Board Assembly” on page 5-7 • “A27 Post-Regulator Board Assembly” on page 5-7 • “Self-...
Page 68 - Figure 5-3 Power Supply Fuses and Test Points; appears on the system display.
5-8 HP 8517B S-Parameter Test Set Manual Troubleshooting the Test Set Troubleshooting Procedures Figure 5-3 Power Supply Fuses and Test Points HP-IB Address Switch The HP-IB address switch is located on the rear-panel of the instrument. It iseasy to access, but unnecessary to change unless the error...
Page 69 - Figure 5-4 Instrument HP-IB Switch Setting; Fuse Location; Figure 5-5 Fuse and Voltage Cam Location
HP 8517B S-Parameter Test Set Manual 5-9 Troubleshooting the Test Set Troubleshooting Procedures Figure 5-4 Instrument HP-IB Switch Setting Fuse Location The fuse is located inside the line module at the rear panel of the test set. Thevalue rating and part number of the fuse is listed in Chapter 7, ...
Page 70 - Figure 5-6 The ACTIVE LED Location; If the Self-Test Fails to Run Properly; All LEDs flash briefly, then turn off
5-10 HP 8517B S-Parameter Test Set Manual Troubleshooting the Test Set Troubleshooting Procedures Figure 5-6 The ACTIVE LED Location If the Self-Test Fails to Run Properly If the portion of memory holding the self-test program is faulty, the self-testwill not run properly. If any of the following co...
Page 71 - Figure 5-7 HP 8518B Test Set, A4 Board Assembly LEDs
HP 8517B S-Parameter Test Set Manual 5-11 Troubleshooting the Test Set Troubleshooting Procedures Figure 5-7 HP 8518B Test Set, A4 Board Assembly LEDs Table 5-4 Instrument Self-Test Indicator State Definitions Self-Test Indication LSN TLK SRQ REM Time (Following Power-On) PWON ON ON ON ON 0 to 0.5 s...
Page 72 - Amplifier Test; Figure 5-8 Service Adapter Connections; Using the Service Adapter; Instrument Self-Test Indicator State Definitions
5-12 HP 8517B S-Parameter Test Set Manual Troubleshooting the Test Set Troubleshooting Procedures HP 85102 IF Amplifier Test The service adapter can substitute as a test set by returning the 20 MHz IFsignal from the HP 85102 to the amplifier on the HP 85102. Doing thisdetermines whether a fault is i...
Page 73 - Service Adapter Conclusions; Results Observations
HP 8517B S-Parameter Test Set Manual 5-13 Troubleshooting the Test Set Troubleshooting Procedures Service Adapter Conclusions Table 5-5 Results Observations If all channels look good The HP 85102 is working properly. The problem is most likely in the source or the test set. Refer to Procedure 4, “Un...
Page 74 - Unratioed Power Test; Figure 5-9 Diagram of IF Signal Path; Testing Signal Paths 1-4; Set the source for step mode. Press the network analyzer keys:
5-14 HP 8517B S-Parameter Test Set Manual Troubleshooting the Test Set Troubleshooting Procedures Unratioed Power Test We will examine six RF signal paths in the test set by observing theirassociated IF signals. Figure 5-10 through Figure 5-15 show the test settraces (and simplified block diagrams) ...
Page 76 - Testing Paths 5 and 6
5-16 HP 8517B S-Parameter Test Set Manual Troubleshooting the Test Set Troubleshooting Procedures Testing Paths 5 and 6 1. Connect a cable between Port 1 and Port 2. 2. To observe the b1 power-level trace through path 5, press: PARAMETER [MENU] {USER 4 b1} then {REDEFINE PARAMETER} {DRIVE} DRIVE: {P...
Page 83 - Finding the Faulty; Sample Diagnostics; Test Results for Path 1 through Path 6
HP 8517B S-Parameter Test Set Manual 5-23 Troubleshooting the Test Set Troubleshooting Procedures Finding the Faulty Assembly Compare the results recorded in Table 5-6 with the matrix in Table 5-7.There are 32 possible combinations, twelve of which point to probablefailures. These are shown in Table...
Page 86 - ESD-Safe Workstation Configuration; Performance Tests
6-2 HP 8517B S-Parameter Test Set Manual HP 8517B Test Set Replacement Procedures Figure 6-1 ESD-Safe Workstation Configuration Performance Tests Performance test information and procedures are located in Chapter 8 of theHP 8510C On-Site Service Manual. Note that the HP 8510C performance test softwa...
Page 87 - Equipment Needed But Not Supplied; Replacement-Procedure Tools Needed
HP 8517B S-Parameter Test Set Manual 6-3 HP 8517B Test Set Replacement Procedures Equipment Needed But Not Supplied Table 6-1 lists the equipment needed to replace the assemblies documentedin the following procedures. See Figure 6-2 to locate the test set assemblies. Table 6-1 Replacement-Procedure ...
Page 88 - HP 8517B Major Assembly Location Diagram
6-4 HP 8517B S-Parameter Test Set Manual HP 8517B Test Set Replacement Procedures Figure 6-2 HP 8517B Major Assembly Location Diagram
Page 89 - Assembly Replacement Procedures; Initial Precautions; A18 Switch/Splitter Assembly
HP 8517B S-Parameter Test Set Manual 6-5 HP 8517B Test Set Replacement Procedures Assembly Replacement Procedures Assembly Replacement Procedures Initial Precautions Before beginning any of the procedures in this chapter, do the following: 1. Turn off the HP 8517B test set. 2. Disconnect the test se...
Page 90 - A15 Regulator Board Assembly
6-6 HP 8517B S-Parameter Test Set Manual HP 8517B Test Set Replacement Procedures Assembly Replacement Procedures 6. Unplug the four harnessed (multi-colored) wire and socket assemblies. 7. Lift the frequency converter out of the test set. 8. Remove the frequency converter bracket by removing the tw...
Page 91 - mm Test Port Connectors; Pull out the center conductor assembly and replace it with a new; Diagram of 2.4 mm Test Port Connector
HP 8517B S-Parameter Test Set Manual 6-7 HP 8517B Test Set Replacement Procedures Assembly Replacement Procedures Test Port ConnectorReplacement 2.4 mm Test Port Connectors 1. Incline the rear of the test set approximately four inches and support it with a stable, solid object (such as a thick book)...
Page 92 - B1 Fan Assembly; that hold the transformer onto the rear panel.; bench near the test set.; T1 Power Transformer; Turn off the test set and disconnect the power cord from the mains.
6-8 HP 8517B S-Parameter Test Set Manual HP 8517B Test Set Replacement Procedures Assembly Replacement Procedures Fan AssemblyReplacement B1 Fan Assembly Turn off the test set and disconnect the power cord from the ac-power mains. WARNING Electrocution can result if ac power remains connected to the...
Page 93 - connected between the coupler A7 and the bulkhead connector J2.
HP 8517B S-Parameter Test Set Manual 6-9 HP 8517B Test Set Replacement Procedures Assembly Replacement Procedures WARNING Electrocution can result if ac power remains connected to the test set during this procedure. Disconnect the ac power cord before starting. 1. See Figure 6-2 for component locati...
Page 94 - Pry the O-ring out of the clamp with fine tweezers or a similar tool.; A1 Front Panel and Interface Board Assembly; of the assembly toward you.
6-10 HP 8517B S-Parameter Test Set Manual HP 8517B Test Set Replacement Procedures Assembly Replacement Procedures O-Ring Replacement Anti-Rotation Clamp O-Ring Replacement 1. Pry the O-ring out of the clamp with fine tweezers or a similar tool. 2. To insert the new O-ring (HP part number 0900-0007)...
Page 95 - Bias Tees Replacement; A7 and A9 Bias Tees; A6 and A8 Test Port Couplers
HP 8517B S-Parameter Test Set Manual 6-11 HP 8517B Test Set Replacement Procedures Assembly Replacement Procedures Bias Tees Replacement A7 and A9 Bias Tees 1. See Figure 6-2 for component locations. 2. Remove the two semi-rigid cables attached to the bias tee to be replaced. 3. De-solder the wires ...
Page 96 - A22 Input Amplifier Assembly; See the figure in Chapter 7, Table 7-17 for the location of the A22; A23 to A26 Buffer Amplifiers; Remove the two semi-rigid cables from the buffer amplifier to be
6-12 HP 8517B S-Parameter Test Set Manual HP 8517B Test Set Replacement Procedures Assembly Replacement Procedures Input AmplifierReplacement A22 Input Amplifier Assembly NOTE This procedure is for Option 007, Option 002 “plus” 007, and Option 004“plus” 007, only. 1. See the figure in Chapter 7, Tab...
Page 97 - A27 Post Regulator Board Assembly
HP 8517B S-Parameter Test Set Manual 6-13 HP 8517B Test Set Replacement Procedures Assembly Replacement Procedures Post Regulator BoardReplacement A27 Post Regulator Board Assembly NOTE This procedure is for Option 007, Option 002 “plus” 007, and Option 004“plus” 007, only. 1. See the figure in Chap...
Page 99 - Introduction; Replaceable Parts List; Part description, using the abbreviations found in Table 7-3.
HP 8517B S-Parameter Test Set Manual 7-1 7 HP 8517B Replaceable Parts Introduction This section provides information about ordering replacement parts. Partsthat can be replaced include major assemblies and chassis hardware, but notindividual pieces of major assemblies. Table 7-1 lists the major asse...
Page 100 - Ordering Information; Reference Designation Abbreviations
7-2 HP 8517B S-Parameter Test Set Manual HP 8517B Replaceable Parts Ordering Information To order parts listed in the replaceable parts table, record the followinginformation, then address the order to the nearest Hewlett-Packard Sales andService Office: the Hewlett-Packard part numberthe quantity r...
Page 101 - Reference Designation Abbreviations (Continued)
HP 8517B S-Parameter Test Set Manual 7-3 HP 8517B Replaceable Parts DC Directional Coupler MP Miscellaneous Mechanical Part VR Breakdown Diode (Zener), Voltage Regulator DL Delay Line P Electrical Connector (Movable Portion), Plug W Cable, Wire, Jumper DS Annunciator, Lamp, Light Emitting Diode (LED...
Page 102 - Standard Abbreviations
7-4 HP 8517B S-Parameter Test Set Manual HP 8517B Replaceable Parts Table 7-3 Standard Abbreviations A A Across Flats, Acrylic, Air (Dry Method), Ampere ADJ Adjust, Adjustment ANSI American National Standards Institute (formerly USASI-ASA) ASSY Assembly AWG American Wire Gage B BCD Binary Coded Deci...
Page 108 - Multipliers, Abbreviation and Description
7-10 HP 8517B S-Parameter Test Set Manual HP 8517B Replaceable Parts VAR Variable VDC Volts—Direct Current W W Watt, Wattage, White, Wide, Width W/SW With, Switch WW Wire Wound X X By (as used with dimensions), Reactance Y YIG Yttrium-Iron-Garnet Z ZNR Zener Diode Table 7-3 Standard Abbreviations (C...
Page 109 - Figure 7-1 Power Cable and Plug Part Numbers
HP 8517B S-Parameter Test Set Manual 7-11 HP 8517B Replaceable Parts Figure 7-1 Power Cable and Plug Part Numbers
Page 110 - Instrument Fuses
7-12 HP 8517B S-Parameter Test Set Manual HP 8517B Replaceable Parts Table 7-5 Instrument Fuses Fuse Name Location Amps for 100 to 120 Vac (part number) Amps for 220 to 250 Vac (part number) ac Line F1 Rear Panel 2.0 2110-0002 1.0 2110-0001 +5 Vdc F1 Regulator Board 1.5 2110-0043 1.5 2110-0043 - 5 V...
Page 111 - Figure 7-2 The Low Cost Rebuilt-Exchange Procedure
HP 8517B S-Parameter Test Set Manual 7-13 HP 8517B Replaceable Parts Figure 7-2 The Low Cost Rebuilt-Exchange Procedure Use this fast, efficient economicalmethod to keep your Hewlett-Packard instrument in service. *Hewlett-Packard pays postage on boxes mailed in the U.S.A.
Page 112 - Major Assembly Reference Designator and Parts Information
7-14 HP 8517B S-Parameter Test Set Manual HP 8517B Replaceable Parts Table 7-6 Major Assembly Reference Designator and Parts Information Ref Des HP Part Number QTY Description A1 08513-60005 1 Front Panel Interface Board Assembly A6, A8 5086-75185086-6518 2 50 GHz Coupler, (NEW)50 GHz Coupler, (R-E)...
Page 114 - Semi-Rigid Cable Assemblies in the Standard HP 8517B Test Set
7-16 HP 8517B S-Parameter Test Set Manual HP 8517B Replaceable Parts Table 7-7 Semi-Rigid Cable Assemblies in the Standard HP 8517B Test Set Ref Des HP Part Number Quantity Description W10 08517-20039 1 Cable Assembly AT1 to A10J2 W31 08517-20042 1 Cable Assembly A8J3 to A11J2 W32 08517-20041 1 Cabl...
Page 115 - Flexible Cable Assemblies in the Standard HP 8517B Test Set
HP 8517B S-Parameter Test Set Manual 7-17 HP 8517B Replaceable Parts Table 7-8 Flexible Cable Assemblies in the Standard HP 8517B Test Set Ref Des HP Part Number Quantity Description W1 08513-60121 1 Cable Assembly A12J3 to A1J11 W2 08513-60122 1 Cable Assembly A13J3 to A4J11 W3 08513-60123 1 Cable ...
Page 117 - Miscellaneous Parts, Front Panel HP 8517B Test Set
HP 8517B S-Parameter Test Set Manual 7-19 HP 8517B Replaceable Parts Table 7-9 Miscellaneous Parts, Front Panel HP 8517B Test Set Ref Des HP Part Number Quantity Description DS1,DS2 1990-0858 2 LED Lamp 25 MA, Maximum E33,E34 2110-0797 2 Fuse Holder F1,F2 2110-0012 2 Fuse 0.5 A 250 V J13,J14 1 – 1 e...
Page 118 - Test Port Connector Assembly Components
7-20 HP 8517B S-Parameter Test Set Manual HP 8517B Replaceable Parts Table 7-10 Test Port Connector Assembly Components Item HP Part Number Quantity Description 1 5062-7274 1 Test Port Connector 2 5001-39305001-39045001-39055001-3906 .0002 IN SHIM.0005 IN SHIM.001 IN SHIM.002 IN SHIM 3 5062-7276 1 T...
Page 119 - Miscellaneous Parts, Rear Panel HP 8517B Test Set
HP 8517B S-Parameter Test Set Manual 7-21 HP 8517B Replaceable Parts Table 7-11 Miscellaneous Parts, Rear Panel HP 8517B Test Set Ref Des HP Part Number Quantity Description E10 5062-7243 1 2.4 mm RF Connector Assembly FL1 9135-0217 1 Line Module-Filtered J10 1251-2197 1 Connector R&P 24F (Part ...
Page 121 - Figure 7-4 Detailed Views of Hardware Stack-Ups
HP 8517B S-Parameter Test Set Manual 7-23 HP 8517B Replaceable Parts Figure 7-4 Detailed Views of Hardware Stack-Ups
Page 122 - Miscellaneous Parts, Top Internal View
7-24 HP 8517B S-Parameter Test Set Manual HP 8517B Replaceable Parts Table 7-12 Miscellaneous Parts, Top Internal View Item HP Part Number Quantity Description 1 0360-0031 1 Terminal Crimp 2 2360-0115 5 Machine Screw 3 08512-20005 1 RFI Gasket 4 08513-00001 1 Top Deck 5 08513-00002 1 Transformer Bra...
Page 130 - Parts Unique to HP 8517B Option 007 Test Set (Continued)
7-32 HP 8517B S-Parameter Test Set Manual HP 8517B Replaceable Parts Ref Des HP Part Number Quantity Description Cable Assemblies Unique to Option 007 W45 08517-20023 1 Front Panel J7 to A22 Input Amplifier W46 08517-20024 1 A22 Input Amplifier to A18J5 W47 08517-2003 1 A6 Coupler to A23 W48 08517-2...
Page 131 - Parts Unique to Option 002 “plus” Option 007
HP 8517B S-Parameter Test Set Manual 7-33 HP 8517B Replaceable Parts Table 7-18 Parts Unique to Option 002 “plus” Option 007 There are no unique parts to Option 002 “plus” Option 007. All other parts are included in Option 002 or Option 007.
Page 132 - Parts Unique to Option 004 “plus” Option 007
7-34 HP 8517B S-Parameter Test Set Manual HP 8517B Replaceable Parts Table 7-19 Parts Unique to Option 004 “plus” Option 007 Ref Des HP Part Number Quantity Description W56 08517-20049 1 A17J1 to A25 All other parts are included in Option 004 or Option 007.
Page 133 - Instrument Chassis Replaceable Parts
HP 8517B S-Parameter Test Set Manual 7-35 HP 8517B Replaceable Parts Table 7-20 Instrument Chassis Replaceable Parts Item Number HP Part Number Quantity Description 1 08513-00040 1 Top Cover 2 08513-00041 1 Side Cover, Perforated 3 5021-8403 1 Front Frame 4 5062-3747 1 Bottom Cover 5 5062-3799 1 Han...
Page 134 - Instrument Chassis Replaceable Parts (Continued)
7-36 HP 8517B S-Parameter Test Set Manual HP 8517B Replaceable Parts 7 5041-8821 4 Bottom Foot 8 5062-3757 1 Side Cover Item Number HP Part Number Quantity Description 9 5021-5837 1 18-inch Corner Strut 10 5021-5804 1 Rear Frame 11 5041-8802 1 Trim Strip 12 5041-8821 2 Back Panel Foot 13 5021-8538 1...
Page 135 - Index
Index HP 8517B S-Parameter Test Set Manual Index-1 A abbreviations used in parts descriptions, 7-4accessories descriptions, 2-3active LED location, 3-1adapter part numbers, 1-11addressing a test set, 3-6altering signal path conditions, 3-10attaching anti-rotation clamps, 3-15 B bias supply connector...