  • Page 2 – DO NOT FORCE THIS LEVER,; TURNING THE BEAM ON –; ENTER; Microscope will automatically go to 5kV.
  • Page 3 – the; DIRECT; READ THE MANUAUL INSTRUCTIONS; CHANGING ACCELERATING VOLTAGE; BEAM ALIGN
  • Page 4 – Hit ENTER to move $ sign against; STIG ALIGN Y; Press; ENTER again; CHANGING THE DETECTOR; PF2 COLUMN COND; H. CHANGING THE WORKING DISTANCE; The working distance is changed by using the
  • Page 5 – To fine focus only press the; AUTO; button on the focus panel. This will fine focus the image.; DO NOT
  • Page 6 – FINISHING A SESSION; SC AIRLOCK VALVE; from; CLOSED; to
Loading the manual

HITACHI 4500-Il INSTRUCTIONS

The following instructions provide a guide to performing simple tasks on the microscope. Initial sections
(A to D) describe how to insert a sample and obtain an image. Subsequent sections provide information
on more COni~!CX operations such as changing the accelerating voltage or the detecton Sections K and
L provide infonnation on what to do at the end of a session.

Compared with the JEOL 840 SEM there are more variable parameters available to the operator.
Optimum imaging conditions will vary from user to user depending on their samples. It is
advisable that you spend some time attempting to determine which conditions are best for your
sample.


When using this microscope please note the following rules:

RULE 1: DON'T TOUCH A CONTROL IF YOU ARE NOT SURE OF THE

OUTCOME

OF

THAT

ACTION.

RULE 2: NEVER, EVER FORCE ANYTHING BEYOND FINGER STRENGTH

RULE 3: IF IN DOUBT; ASK FOR HELP


A. START-UP

1. Turn on the visual display by turning toggle switch

DISPLAY POWER

to 1. (These

toggle switches must be pulled out slightly before switching - DO NOT FORCE THEM)

2. Turn the toggle switch SC

AIRLOCK VALVE

from

CLOSED

to

AUTO

(OPEN)- a

few seconds later the green light on the HV off/ready display will come on. If it doesn't please
contact EM staff for help.


B. INSERT SAMPLE

1. Put specimen stub onto specimen holder> inc exact mechanism will vary depending upon
the specimen type. EM Unit staff will advise the best method for your sample when you art
being trained. The specimen holder and stub MUST be set so that so that thtt top of the
specimen is level with or below tIze gauge.
2. Ensure toggle switch on SC

AIRLOCK VALVE DISPLAY

is set to S.E.C.

3. Press

AIR

on

Sc AIRLOCK VALVE DISPLAY -

this will bleed air into the specimen

exchange chamber. It will take 2~3O seconds for the chamber to come to air.

4. Ensure stage settings

(X,Y,Z,R

and T) have been return to their default positions

(displayed on the specimen chamber).
5. Open specimen exchange chamber door Do not force.
6. Push specimen exchange rod forward slightly so it protrudes through the specimen exchange
chamber door and screw specimen holder on to screw thread of the specimen exchange rod.

PLEASE MAKE SURE YOU DO NOT GET

A

CROSS THREAD.

Pull specimen

exchange rod back into specimen exchange chamber door.

7. Close airlock door

"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.

Summary

Page 2 - DO NOT FORCE THIS LEVER,; TURNING THE BEAM ON –; ENTER; Microscope will automatically go to 5kV.

8. Press EVAC on SC AIRLOCK VALVE DISPLAY - this will pump the specimen exchange chamber. 9. Observe SEC vacuum display as status lights show vacuum goes from AIR to LOW to HIGH. 10. When SEC vacuum is FilCH (green light is on), open airlock door, by turning airlock level anticlockwise about one hal...

Page 3 - the; DIRECT; READ THE MANUAUL INSTRUCTIONS; CHANGING ACCELERATING VOLTAGE; BEAM ALIGN

D. TAKING A PHOTOGRAPH 1. Obtain the image you desire. 2. Press ABC button on the image adjust section to automatically alter brightness and contrast to the correct levels for the camera. 3. Press DIRECT switch on the photographic control section. 4. Wait for the line scan to perform on cycle and ex...

Page 4 - Hit ENTER to move $ sign against; STIG ALIGN Y; Press; ENTER again; CHANGING THE DETECTOR; PF2 COLUMN COND; H. CHANGING THE WORKING DISTANCE; The working distance is changed by using the

10. Hit ENTER to move $ sign against STIG ALIGN Y Press ENTER again 11. The image will oscillate on the screen, minimise oscillation with ALIGNMENT X and Y knobs. 12. Press PF16 EXIT to finish. G. CHANGING THE DETECTOR The 4500 has two secondary electron detectors, an upper detector which sits above...

Other Hitachi Models