Page 2 - LIMITED WARRANTY AND LIMITATION OF LIABILITY
LIMITED WARRANTY AND LIMITATION OF LIABILITY Each Fluke product is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is two years and begins on the date of shipment. Parts, product repairs, and services are warranted for 90 days. This war...
Page 3 - Table of Contents
i Title Page How to Contact Fluke . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1Specification...
Page 4 - ii; Service Information
Page 5 - Introduction; XW; How to Contact Fluke; To contact Fluke, call one of the following telephone numbers:; Safety Information; identifies conditions and
1 Introduction XW Warning To avoid electric shock or personal injury, do not perform the calibration verification tests or calibration procedures described in this manual unless you are qualified to do so. The information provided in this manual is for the use of qualified personnel only. This servi...
Page 7 - Symbol
Electrical Tester Safety Information 3 Table 1 is a list of symbols used on the Product and in this document. Table 1. Symbols Symbol Description W WARNING. RISK OF DANGER. X WARNING. HAZARDOUS VOLTAGE. Risk of electric shock. Consult user documentation. FieldSense Measurement: Fluke voltage/curre...
Page 8 - Specifications
T6-1000/T6-600 Service Information 4 Specifications Battery Type and Life ............................................2 x AA (IEC LR6) 360 hr continuous, typical is 200 hr using the FieldSense function Fork Opening .........................................................17.8 mm Temperature Operatin...
Page 9 - Required Equipment; Annunciator and Hazardous Voltage Indicator
Electrical Tester Required Equipment 5 Required Equipment Table 2 is a list of equipment requirements for the performance tests of the Product . Table 2. Required Equipment Performance Tests Fluke recommends a 1-year cycle for the Performance Tests. If the Product fails any test, return the Product ...
Page 10 - Voltage Function; Perform these tests as follows:
T6-1000/T6-600 Service Information 6 Voltage Function Perform these tests as follows: 1. Set the calibrator NORMAL output to the voltage listed in Table 5, Step 1. Table 5. AC Voltage Tests 2. Set the Product to the voltage (Auto Select) measurement mode. 3. Apply the voltage to the Product. Verify ...
Page 11 - FieldSense Current Function; Note
Electrical Tester Performance Tests 7 FieldSense Current Function The tests in this section require the use of a 10/50-turn coil. See Figure 1. Note Make measurements on the 10-turn coil side of the 10/50-turn coil. The Product shows 10X more current than the calibrator supplies to the coil. Figure ...
Page 13 - FieldSense Voltage Function
Electrical Tester Performance Tests 9 FieldSense Voltage Function The tests in this section require the use of a PRV240FS Proving Unit (Fluke P/N 4910310) and a Fluke 8846A Multimeter. 1. Set the 8846A to the AC Voltage measurement mode (ACV button) and range manually to the 1000 V range. 2. Set the...
Page 14 - Low Battery Indicator
T6-1000/T6-600 Service Information 10 Low Battery Indicator To verify the correct operation of the low battery indicator: 1. Remove the Tester batteries. 2. Set the dc power supply to 3.0 V. Apply this voltage to the Tester battery terminals and observe the correct polarity. 3. Set the Tester to the...
Page 15 - Maintenance; How to Clean
Electrical Tester Maintenance 11 Maintenance If the Product is used appropriately it does not require special maintenance or repair. See www.fluke.com for locations and contact information of Fluke Service Centers worldwide. How to Clean W Caution For safe operation and maintenance, clean the case w...
Page 16 - Replacement Parts; Replacement parts and accessories are listed in Table 10; Table 10. Replacement Parts; Battery Door
T6-1000/T6-600 Service Information 12 Replacement Parts Replacement parts and accessories are listed in Table 10 . To order parts and accessories, see How to Contact Fluke . Table 10. Replacement Parts Ref. Description Fluke Part or Model Number Battery Door 4944370 Battery Door Screw 1618579 ...