Page 3 - LIMITED WARRANTY & LIMITATION OF LIABILITY
LIMITED WARRANTY & LIMITATION OF LIABILITY Each Fluke product is warranted to be free from defects in material and workmanship undernormal use and service. The warranty period is one year and begins on the date ofshipment. Parts, product repairs and services are warranted for 90 days. This warra...
Page 4 - Caution; Interference Information; Bescheinigung des Herstellers/Importeurs
P Caution This is an IEC safety Class 1 product. Before using, theground wire in the line cord or rear panel binding postmust be connect to an earth ground for safety. Interference Information This equipment generates and uses radio frequency energy and if not installed and used in strictaccordance ...
Page 5 - SAFETY TERMS IN THIS MANUAL; SYMBOLS MARKED ON EQUIPMENT:
SAFETY TERMS IN THIS MANUAL This instrument has been designed and tested in accordance with IEC publication1010-1, Safety Requirements for Electrical Measuring, Control and Laboratory Equipment. This Users Manual contains information, warnings and cautions. Useof this equipment in a manner not speci...
Page 6 - Use only a power cord that is in good condition.
DC POWER SOURCE The instrument may also be operated from a 9V to 16V dc power source wheneither the rear panel ground binding post or the power cord grounding conductoris connected properly. USE THE PROPER FUSE To avoid fire hazard, for fuse replacement use only a 1/4 ampere, 250V non-timedelay line...
Page 7 - Table of Contents; Chapter
i Table of Contents Chapter Title Page 1 Overview ........................................................................................... 1-1 1-1. Introduction .......................................................................................... 1-3 1-2. Instrument Features and Capabilities...
Page 8 - Reviewing and Setting the Subnet Mask and Default Gateway.. 2-38
2640A/2645A NetDAQ Users Manual ii 2 Preparing for Operation .................................................................. 2-1 2-1. Introduction .......................................................................................... 2-3 2-2. Instrument Preparation ..............................
Page 9 - Contents
Contents (continued) iii 2-44. Setting Host Computer Networking Parameters .............................. 2-51 2-45. Installing NetDAQ Logger ............................................................... 2-52 2-46. Installing NetDAQ Logger with Trumpet.................................... 2-53 2-47...
Page 10 - Add the New Folder to the Trend Link
2640A/2645A NetDAQ Users Manual iv 3-35. Default Configuration Settings ............................................................ 3-273-36. Using Configuration Lockout .............................................................. 3-283-37. Saving an Instrument’s Configuration as a Text File. ......
Page 12 - Appendices
2640A/2645A NetDAQ Users Manual vi 6-29. Totalizer Sensitivity Test ............................................................ 6-28 6-30. Master Alarm Output Test ............................................................... 6-29 6-31. Trigger Input Test ............................................
Page 13 - List of Tables; Table
vii List of Tables Table Title Page 1-1. Models, Options and Accessories ................................................................... 1-14 2-1. Front Panel Key Descriptions ......................................................................... 2-17 2-2. Annunciator Display Descriptions ......
Page 15 - List of Figures; Figure Title
ix List of Figures Figure Title Page 1-1. 2640A/2645A NetDAQ Networked Data Acquisition Units .......................... 1-3 1-2. 2640A/2645A Front Panel .............................................................................. 1-4 1-3. Typical Front Panel Display During Scanning and Monitoring...
Page 17 - Chapter 1; Overview
1-1 Chapter 1 Overview Contents Page 1-1. Introduction .......................................................................................... 1-3 1-2. Instrument Features and Capabilities ................................................... 1-4 1-3. Analog Channels ..................................
Page 19 - Introduction; Networked Data Acquisition Units are
Overview Introduction 1 1-3 Introduction 1-1. The 2640A and 2645A NetDAQ Networked Data Acquisition Units are 20-channel front ends that operate in conjunction with NetDAQ Logger forWindows (hereafter known as NetDAQ Logger) to form a data acquisitionsystem. The instruments measure dc volts, ac vo...
Page 20 - Instrument Features and Capabilities
2640A/2645A NetDAQ Users Manual 1-4 Instrument Features and Capabilities 1-2. The following describes the front and rear panels of the instrument and itscapabilities (Figures 1-2 to 1-4). NetDAQ NETWORKED DATA ACQUISITION UNIT COMM DIO MON ENTER CAL ENABLE REVIEW LAST MAX MIN REM AUTO SCAN MON SET M...
Page 23 - Analog Channels; The average of a group of channels,; Channel Numbering
Overview Instrument Features and Capabilities 1 1-7 Analog Channels 1-3. The analog channel (1 to 20) measurement connections are made via theUniversal Input Module. External signal conditioning for the analog inputs is notnecessary. The host computer configures all analog channels using NetDAQLogge...
Page 24 - Alarms
2640A/2645A NetDAQ Users Manual 1-8 Alarms 1-7. Two alarms, Alarm 1 and Alarm 2, can be applied to any configured channel. Analarm condition occurs when a measurement falls below a low alarm value orrises above a high alarm value. You can use alarms to trigger scanning (see“Alarm Triggering”) and to...
Page 25 - Trigger Input
Overview Instrument Features and Capabilities 1 1-9 overflows (reaching the maximum count), the display briefly shows OL(overload) and begins counting from zero again A totalizer input from contact closures increments on the “open” portion of theswitch sequence close-open. To prevent switch contact ...
Page 26 - Master Alarm
2640A/2645A NetDAQ Users Manual 1-10 Master Alarm 1-13. Master Alarm is an instrument output line that is logic low (nominal +0.8V dc)for as long as any channel is in alarm while scanning is active. The connectionuses the ALARM/TRIGGER I/O terminals MA and GND (Figure 1-4). This TTLoutput interfaces...
Page 27 - NetDAQ Logger Features and Capabilities; Some major features of NetDAQ Logger include:; Operating a NetDAQ Data Acquisition System
Overview NetDAQ Logger Features and Capabilities 1 1-11 You can combine Alarm Trigger with External Trigger and Interval Trigger. Forexample, set the Interval Trigger for 60 seconds (Interval 1) and the AlarmTrigger for 10 seconds (Interval 2). Scanning is at 60-second intervals exceptwhen a channel...
Page 28 - Isolated Networks
2640A/2645A NetDAQ Users Manual 1-12 operations refer to the instrument by BCN. NetDAQ Logger supports up to 20instruments for operation. You cannot operate an instrument from more than onehost computer at a time. Isolated Networks 1-19. An isolated network consists of only NetDAQ instruments and ho...
Page 29 - Host Computer Requirements; Scanning and Logging; Interval Trigger, where an elapsed interval timer triggers a scan.
Overview Host Computer Requirements 1 1-13 Designate one instrument in the group as the Master and the others as Slaves. TheMaster controls scanning operations, including the scan intervals and method ofscan triggering. You can create only one group instrument. Scanning and Logging 1-24. When a scan...
Page 30 - A Hard disk drive with 2 MB of free disk space.; Options and Accessories
2640A/2645A NetDAQ Users Manual 1-14 Host Computer Requirements 1-26. The host computer used for instrument operations must meet the followingminimum requirements: • IBM PC with an Intel 386 microprocessor or greater, running Windows 95,Windows NT, or Windows 3.1. • Color VGA Monitor. • A Hard disk ...
Page 31 - Instrument Connector Set
Overview Options and Accessories 1 1-15 Instrument Connector Set 1-28. The 2620A-100 is a complete set of input connectors: one Universal InputModule, one ALARM/TRIGGER I/O connector, and one DIGITAL I/Oconnector. A 2620A-100 Instrument Connector Set comes with each instrument.You can wire additiona...
Page 33 - Chapter 2; Preparing for Operation
2-1 Chapter 2 Preparing for Operation Contents Page 2-1. Introduction .......................................................................................... 2-3 2-2. Instrument Preparation ......................................................................... 2-3 2-3. Unpacking and Inspecting...
Page 37 - Instrument Preparation; Unpacking and Inspecting the Instrument
Preparing for Operation Instrument Preparation 2 2-5 Unpacking and Inspecting the Instrument 2-3. Verify the contents of the shipping package against the checklist in the package. Ifany items are missing or damaged, report the problem immediately to your Flukerepresentative. Carefully remove the ins...
Page 38 - AC Power; Warning
2640A/2645A NetDAQ Users Manual 2-6 WARNING: IF VOLTAGES GREATER THAN 50V (2640A ONLY) ARE TO BE MEASURED, A SEPARATE EARTH GROUND MUST BE ATTACHED TO THIS REAR PANEL GROUND CONNECTOR WHEN THE INSTRUMENT IS OPERATED FROM BATTERY POWER. Line Cord (AC Operation) External Battery (DC Operation) + – Gro...
Page 39 - DC Power; Loosen the wire clamp screw for the associated terminal.; Grounding and Common Mode Voltage; Universal Input Module Connections; Loosen the two large screws on top and open the module.
Preparing for Operation Instrument Preparation 2 2-7 DC Power 2-7. The instrument operates from any dc voltage between 9 and 16 volts. Powerconsumption is a nominal 6 watts. To connect the ALARM/TRIGGER I/Oconnector to the rear panel, complete the following procedure: 1. Remove the ALARM/TRIGGER I/O...
Page 43 - Shielded Wiring; Digital I/O Connections
Preparing for Operation Instrument Preparation 2 2-11 Shielded Wiring 2-10. Use shielded wires and sensors (such as thermocouples) in environments whereelectrical noise is present, and connect the wire shield to the chassis groundterminal. Also refer to Appendix B "Noise, Shielding and Crosstalk...
Page 44 - Totalizer
2640A/2645A NetDAQ Users Manual 2-12 2. Loosen the wire clamp screw for the associated terminal. 3. Feed the wire into the gap between the connector body and the wire clamp. 4. Tighten the wire clamp; do not overtighten and crush the wire. 5. Repeat steps 2 through 4 for each wire. 6. Insert the con...
Page 46 - Trigger Output
2640A/2645A NetDAQ Users Manual 2-14 + 9-16V DC PWR MA TO TI ALARM/TRIGGER I/O TR 3 2 1 0 – + +–MATOTI DC Positive InputDC Negative InputMaster Alarm OutputTrigger Out OutputTrigger In InputSignal Ground Instrument Connector +–23TR Instrument Function Connector Figure 2-6. ALARM/TRIGGER I/O Connecto...
Page 47 - External Trigger Wiring for a Group Instrument
Preparing for Operation Instrument Preparation 2 2-15 External Trigger Wiring for a Group Instrument 2-19. External Trigger Wiring for a group instrument refers to the triggeringconfiguration in which you connect the Master TO (Trigger Out) line to eachSlave TI (Trigger In) line and provide a common...
Page 49 - Front Panel Controls
Preparing for Operation Instrument Preparation 2 2-17 Front Panel Controls 2-21. Use the front panel controls (Figure 2-8) to enter configuration parameters, andchoose monitoring functions. Table 2-1 summarizes the front panel controlfunctions. ENTER Selection Set/Review COMMunication Parameters Dis...
Page 50 - Front Panel Indicators
2640A/2645A NetDAQ Users Manual 2-18 Front Panel Indicators 2-22. The front panel indicators (Figure 2-9) consist of two five-digit displays and a setof annunciators. Table 2-2 summarizes the front panel indicator functions. REVIEW LAST MAX MIN REM AUTO SCAN MON SET Mx+B FUNCALARM ° C ° F RO mV AC D...
Page 52 - Rear Panel Controls; Rear Panel Indicators
2640A/2645A NetDAQ Users Manual 2-20 Rear Panel Controls 2-23. The rear panel has a single control: the power switch (Figure 2-10). The powerswitch controls both ac and dc power inputs. Power Switch Applies AC and/or DC power to the instrument. Figure 2-10. Rear Panel Controls Rear Panel Indicators ...
Page 53 - Front Panel Operating Procedures; There are three power-on options as listed below:
Preparing for Operation Instrument Preparation 2 2-21 Front Panel Operating Procedures 2-25. Power-On Options 2-26. There are three power-on options as listed below: • Normal Power-On Turn power switch on. The instrument communicationparameters are the same as when the instrument was last turned off...
Page 54 - Displaying a Monitor Channel
2640A/2645A NetDAQ Users Manual 2-22 Displaying a Monitor Channel 2-27. Perform the procedure in Figure 2-12 to monitor an instrument analog channel (01to 20) or computed channel (21 to 30). See Figure 2-13 for examples. • Channel Display When you press the MON key, the first monitor channeldisplaye...
Page 56 - Displaying the Digital I/O Status
2640A/2645A NetDAQ Users Manual 2-24 Displaying the Digital I/O Status 2-28. Perform the procedure in Figure 2-14 to display an instrument Digital I/O linestatus. The instrument updates the DIO display once per second. (See Figure 2-15for examples.) • Number Of DIO Lines There are eight DIO lines: D...
Page 58 - Displaying the Totalizer Status
2640A/2645A NetDAQ Users Manual 2-26 Displaying the Totalizer Status 2-29. Perform the procedure in Figure 2-16 to display the instrument Totalizer status.The instrument updates the Totalizer display once per second. (See Figure 2-15for examples.) To clear the Totalizer count, cycle the instrument p...
Page 59 - Reviewing and Setting the Base Channel Number
Preparing for Operation Instrument Preparation 2 2-27 Reviewing and Setting the Base Channel Number 2-30. Perform the procedure in Figure 2-17 to review or set the Base Channel Number(BCN). The BCN identifies the instrument. The BCN is also the first two digits ofthe Global Channel Number (GCN), whi...
Page 61 - Reviewing and Setting the Line Frequency
Preparing for Operation Instrument Preparation 2 2-29 Reviewing and Setting the Line Frequency 2-31. Perform the procedure in Figure 2-19 to review or set the line frequency. Linefrequency selection allows the instrument to optimize internal circuitry formaximum precision. (See Figure 2-20 for examp...
Page 63 - Reviewing and Setting the Network Type
Preparing for Operation Instrument Preparation 2 2-31 Reviewing and Setting the Network Type 2-32. Perform the procedure in Figure 2-21 to review or set the network type to isolated.Perform the procedure in Figure 2-23 to review or set the network type to general.An isolated network consists of only...
Page 68 - Reviewing and Setting the General Network Socket Port
2640A/2645A NetDAQ Users Manual 2-36 Reviewing and Setting the General Network Socket Port 2-33. Perform the procedure in Figure 2-25 to review or set the general network SocketPort (1024 to 65535). The default is 04369. In order to communicate with eachother, a host computer and an instrument must ...
Page 69 - Reviewing and Setting the General Network IP Address
Preparing for Operation Instrument Preparation 2 2-37 Reviewing and Setting the General Network IP Address 2-34. Perform the procedure in Figure 2-26 to review or set the instrument’s generalnetwork Internet Protocol (IP) address. (See Figure 2-24 for examples.) • General Network IP Address Enter th...
Page 70 - Reviewing and Setting the Subnet Mask and Default Gateway; Note
2640A/2645A NetDAQ Users Manual 2-38 Reviewing and Setting the Subnet Mask and Default Gateway 2-35. If communication between the host computer and the NetDAQ instrument passesthrough a router or gateway, you must set the subnet mask and default gatewayaddress on both the host computer and the instr...
Page 73 - Viewing the Instrument Ethernet Address
Preparing for Operation Instrument Preparation 2 2-41 Viewing the Instrument Ethernet Address 2-36. Perform the procedure in Figure 2-28 to view the Instrument Ethernet address.(See Figure 2-29 for examples.) The network administrator must know theinstrument Ethernet address when the instrument oper...
Page 75 - Host Computer and Network Preparation; Installing Host Computer Ethernet Adapter; To install an Ethernet adapter, use the following procedure:
Preparing for Operation Host Computer and Network Preparation 2 2-43 Host Computer and Network Preparation 2-37. This section contains information for preparing your host computer and setting upnetwork communication, as summarized in Figure 2-30. Installing Host Computer Ethernet Adapter 2-38. Skip ...
Page 77 - Instrument and Host Computer Interconnection; Host Computer/Instrument Direct Connection
Preparing for Operation Host Computer and Network Preparation 2 2-45 Instrument and Host Computer Interconnection 2-39. You may interconnect NetDAQ instruments and host computer(s) with either10Base2 (coaxial) or 10BaseT (twisted pair) wiring. If your site is already wired,you will probably use the ...
Page 80 - Interconnection Using 10Base2 (Coaxial) Wiring
2640A/2645A NetDAQ Users Manual 2-48 Interconnection Using 10Base2 (Coaxial) Wiring 2-41. P Caution Connect the ground lug on the 50-Ohm terminator to theinstrument ground terminal only when there is no othershield ground connected for the network. 10Base2 interconnection uses 50-Ohm coaxial cables ...
Page 82 - Installing Host Computer Networking Software; Install a driver for the adapter
2640A/2645A NetDAQ Users Manual 2-50 Installing Host Computer Networking Software 2-43. To establish Ethernet communication in your host computer, you must do thefollowing: • Install a driver for the adapter • Install a TCP/IP protocol stack and Windows Socket (Winsock) software • Set host computer ...
Page 83 - Setting Host Computer Networking Parameters; Click OK to exit Network Setup.
Preparing for Operation Host Computer and Network Preparation 2 2-51 Setting Host Computer Networking Parameters 2-44. This section discusses how to set your host computer networking parameters afteryou install your adapter and networking software. I f you plan to install NetDAQLogger for general ne...
Page 84 - Installing NetDAQ Logger; To install NetDAQ Logger, complete the following procedure:
2640A/2645A NetDAQ Users Manual 2-52 Installing NetDAQ Logger 2-45. The NetDAQ Logger setup program automatically determines whether to installthe 32-bit (for Windows 95 and Windows NT) or 16-bit version of the software.The setup program will ask you to select isolated or general network operation.S...
Page 85 - Installing NetDAQ Logger with Trumpet; Select the Ethernet adapter you are using.
Preparing for Operation Host Computer and Network Preparation 2 2-53 6. Check the setup. If the setup is correct, click Next to begin file transfer. If you want to change the setup, click Back to go to the appropriate screen andmake the changes. Click Next and check the setup again. If it is correct...
Page 86 - Changing from an Isolated Network to a General Network; You may want to copy or backup your setup and data files first.
2640A/2645A NetDAQ Users Manual 2-54 6. Tell the setup program whether to modify your autoexec.bat file for the packetdriver. Click Next. 7. Select the program folder. The default name of the program folder or programgroup is Fluke NetDAQ Logger. You can select or type a different name.Click Next. 8...
Page 87 - Installing Trend Link for Fluke (Optional); Enter the directory location where you installed Trend Link.; Testing and Troubleshooting; Testing the Installation
Preparing for Operation Testing and Troubleshooting 2 2-55 “Installing Host Computer Networking Software” and “Setting HostComputer Networking Parameters” to configure the host computer for generalnetwork operation. Installing Trend Link for Fluke (Optional) 2-48. The Trend Link setup program automa...
Page 90 - Troubleshooting Network Problems
2640A/2645A NetDAQ Users Manual 2-58 Troubleshooting Network Problems 2-51. Review the troubleshooting information below to help locate any networkproblems. Table 2-4 summarizes network messages reported by NetDAQ Loggerfor Windows. Table 2-5 is a summary of how to use the Ethernet LED indicatorson ...
Page 95 - Chapter 3; Configuring NetDAQ Logger for
3-1 Chapter 3 Configuring NetDAQ Logger for Windows Contents Page 3-1. Introduction .......................................................................................... 3-3 3-2. Starting NetDAQ Logger ................................................................. 3-3 3-3. The Main Window .....
Page 97 - Configuring NetDAQ Logger for Windows; Starting NetDAQ Logger
Configuring NetDAQ Logger for Windows Introduction 3 3-3 Introduction 3-1. NetDAQ Logger provides an easy method for building configuration databases,transferring the configuration to the instrument, collecting data from theinstrument, and managing the collected data. This chapter provides instructi...
Page 98 - Configuring Network Communications; maintained by NetDAQ Logger. This file is common to
2640A/2645A NetDAQ Users Manual 3-4 ds002s.bmp Figure 3-1. NetDAQ Logger Main Window Configuring Network Communications 3-5. Before you can configure the instruments on your network and create setup filesfor these instrument configurations, you must configure your networkcommunications. This involve...
Page 99 - The Communications Configuration Dialog Box; Adding an Instrument to the Network
Configuring NetDAQ Logger for Windows Configuring Network Communications 3 3-5 The Communications Configuration Dialog Box 3-6. Use the Communications Configuration dialog box (Figure 3-2) to record yournetwork configuration. To open the Communications Configuration dialog box, selectSetup | Communi...
Page 100 - Click Close to return to the Main Window.
2640A/2645A NetDAQ Users Manual 3-6 ds007s.bmp 3. Click OK to return to the Communications Configuration File dialog box,which now lists the new instrument. For a general network, the IP addressappears after the instrument model number. 4. If you are on a general network, you may need to change the ...
Page 101 - Configuring the Current Setup; Deleting an Instrument from the Network; Highlight the instrument you want to remove.; Verifying Network Communications
Configuring NetDAQ Logger for Windows Configuring the Current Setup 3 3-7 Deleting an Instrument from the Network 3-8. Complete the following procedure to remove an instrument from the Instrumentson Network List, which is common to all setup files. Any instruments you deletein this procedure will no...
Page 102 - Creating an Instrument Icon; To create an instrument icon, use the following procedure:
2640A/2645A NetDAQ Users Manual 3-8 Configuring the Current Setup 3-10. NetDAQ Logger keeps track of all instrument configuration information (e.g.,reading rate and measurement functions) for all the instruments represented on theicon bar. This collection of information is called the current setup. ...
Page 103 - Deleting an Instrument Icon; Wired Group
Configuring NetDAQ Logger for Windows Configuring the Current Setup 3 3-9 ds004s.bmp Deleting an Instrument Icon 3-12. To delete an instrument icon, select Setup | Delete Instrument Icon. A messageappears warning you that the instrument’s configuration will be lost. If youchoose to delete the icon, ...
Page 104 - Unwired Group
2640A/2645A NetDAQ Users Manual 3-10 When a grouped instrument is externally wired, you can select anycombination of trigger types for the Master instrument: Interval Trigger,External Trigger, or Alarm Trigger. NetDAQ Logger configures the Slavesfor External Trigger so they will respond to the trigg...
Page 105 - Setup Files; Saving the Current Setup in a File
Configuring NetDAQ Logger for Windows Setup Files 3 3-11 Setup Files 3-14. NetDAQ Logger allows you to save the current setup in a file. You can open thesetup file later to reestablish the saved setup. Using setup files, you can create anynumber of setups in advance and use them when applicable. You...
Page 106 - Opening a Setup File
2640A/2645A NetDAQ Users Manual 3-12 ds039s.bmp Opening a Setup File 3-16. If NetDAQ Logger is already active, you can load a saved setup into the currentsetup by opening a setup file. Before you open a setup file, you must stop loggingon all instruments. Use any of the following methods to open a s...
Page 107 - Starting NetDAQ Logger with a Setup File
Configuring NetDAQ Logger for Windows Setup Files 3 3-13 ds001s.bmp Starting NetDAQ Logger with a Setup File 3-17. On Windows 95, there are several ways to start NetDAQ Logger and open aspecific setup file at the same time: • In the Windows Explorer or My Computer, you can double-click on the nameof...
Page 108 - Starting Logging Automatically; netdaq
2640A/2645A NetDAQ Users Manual 3-14 3. On the Shortcut tab, add the setup file name to the end of the commandline in the Target box (see “NetDAQ Logger Command Line” for thecommand lines). • You can use Start | Settings | Taskbar to modify the command line of NetDAQLogger to open a setup file whene...
Page 109 - To activate NetDAQ Logger with the default setup:; Configuring an Instrument; Dimmed Configuration Commands
Configuring NetDAQ Logger for Windows Configuring an Instrument 3 3-15 To activate NetDAQ Logger with the default setup: “pathname\netdaq32” To activate NetDAQ Logger with a specific setup file (for example, mysetup.stp): “pathname\netdaq32” “mysetup.stp” To activate NetDAQ Logger with a specific se...
Page 110 - Interval Only
2640A/2645A NetDAQ Users Manual 3-16 ds017s.bmp Figure 3-3. Instrument Configuration Dialog Box 1. Select Celsius or Fahrenheit to use as temperature units for thermocouple andRTD temperature measurements. 2. Select a monitor channel. This channel will automatically be displayed on theinstrument’s f...
Page 111 - External Only
Configuring NetDAQ Logger for Windows Configuring an Instrument 3 3-17 • External Only A scan is triggered immediately after the external trigger input is pulledlow (active). While the external trigger line remains low, each successivescan is triggered after Interval 2 elapses since the start of the...
Page 112 - Configuring Channels; The Channels Configuration Dialog Box
2640A/2645A NetDAQ Users Manual 3-18 Configuring Channels 3-23. These procedures provide instructions for configuring an instrument’s analogchannels and computed channels. These instructions include selecting themeasurement function, enabling Mx+B scaling, configuring alarms, and enteringchannel lab...
Page 113 - Configuring Analog Channel Functions; Click Function to open the Function Selection dialog box.
Configuring NetDAQ Logger for Windows Configuring Channels 3 3-19 Configuring Analog Channel Functions 3-25. Analog channel functions include volts dc, volts ac, ohms, temperature(thermocouple and RTD), frequency, and dc current. To configure analog channelfunctions, complete the following procedure...
Page 114 - Configuring Computed Channel Functions; Click Function to open the Computed Function dialog box.; Defining a Computed Channel Equation
2640A/2645A NetDAQ Users Manual 3-20 Configuring Computed Channel Functions 3-26. Computed channel functions include average, difference, difference from average,and equation. To select computed channel functions, complete the followingprocedure: 1. Select a computed channel in the Channels Configur...
Page 115 - Equation Syntax
Configuring NetDAQ Logger for Windows Configuring Channels 3 3-21 3. If the equation contains an error, an error message appears. Click OK in theerror message box. If possible, NetDAQ Logger highlights the location of theerror in the equation. Change the equation as necessary and click OK to returnt...
Page 116 - Allowed; Configuring Mx+B Scaling
2640A/2645A NetDAQ Users Manual 3-22 Allowed Not Allowed (-2*c21/c2)/(2*((c1+1)-(c21/c2)*(c1-1))) (-2*(abs(c1/(c2-c3)))) See Appendix E for syntax definition and more detailed information. Configuring Mx+B Scaling 3-29. Mx+B scaling multiplies a measurement by a multiplier (M) and then applies anoff...
Page 117 - Configuring Alarms; To configure alarms, complete the following procedure:
Configuring NetDAQ Logger for Windows Configuring Channels 3 3-23 ds024s.bmp Configuring Alarms 3-30. Each channel has two alarms, each of which may be set to HI, LO, or OFF. Analarm condition occurs when a measurement falls below a low alarm value (LOalarm) or rises above a high alarm value (HI ala...
Page 118 - Assigning Channel Labels; Configuring Mx+B Scaling From a File; Select the desired instrument icon on the Icon Bar.
2640A/2645A NetDAQ Users Manual 3-24 3. Select a digital output if you would like to associate an alarm with a digitalI/O line. The instrument will set the digital I/O line to a TTL logic low for aslong as the alarm conditions is met. 4. Check Use Channel as Alarm Trigger if you would like an alarm ...
Page 119 - The values in the Mx+B file must have the following format:; Entering an Instrument’s Description; Highlight the desired instrument icon on the Icon Bar.
Configuring NetDAQ Logger for Windows Entering an Instrument’s Description 3 3-25 The values in the Mx+B file must have the following format: <channel number> <M value> <B value> <cr> <lf> Each parameter is separated by spaces or tabs and each definition ends with acarr...
Page 120 - Copying a Channels Configuration; Highlight the icon of the source instrument on the Icon Bar.
2640A/2645A NetDAQ Users Manual 3-26 ds034s.bmp Copying a Channels Configuration 3-34. Complete the following procedure to copy a complete instrument channelsconfiguration from one instrument to another instrument. This includes channelFunctions, Mx+B Scaling, Alarms, and Channel Labels. For example...
Page 121 - Default Configuration Settings
Configuring NetDAQ Logger for Windows Default Configuration Settings 3 3-27 Default Configuration Settings 3-35. When you create a new instrument icon, NetDAQ Logger initializes the settingsfor that instrument to the default parameters. (See Table 3-1.) Table 3-1. NetDAQ Logger Default Instrument Co...
Page 122 - Using Configuration Lockout
2640A/2645A NetDAQ Users Manual 3-28 Using Configuration Lockout 3-36. Use The Configuration Lockout feature to prevent accidental instrumentconfiguration changes. The following commands in the Setup menu will bedisabled: Create Instrument Icon, Delete Instrument Icon, Group Instruments,Communicatio...
Page 123 - Configuring the netdaq.ini File; You can modify the
Configuring NetDAQ Logger for Windows Configuring the netdaq.ini File 3 3-29 The configuration file is now saved as an ASCII text file. You can open the textfile for viewing or printing from an application such as Notepad. Refer to Figure3-5 for a typical text file. INSTRUMENT CONFIGURATION - 02 Des...
Page 124 - The following describes the parameters in the
2640A/2645A NetDAQ Users Manual 3-30 The following describes the parameters in the netdaq.ini file. If you alter any of the settings, save the old file under a different name in case you want to revertback to the default settings. • MemoryWrapAround=True This is the instruction for the instrument to...
Page 125 - Chapter 4; Operating NetDAQ Logger for
4-1 Chapter 4 Operating NetDAQ Logger for Windows Contents Page 4-1. Introduction .......................................................................................... 4-3 4-2. Starting and Stopping Logging............................................................. 4-3 4-3. Starting or Stoppi...
Page 127 - Operating NetDAQ Logger for Windows
Operating NetDAQ Logger for Windows Introduction 4 4-3 Introduction 4-1. This chapter provides instructions for operating your system using NetDAQLogger after you have set up communications and configured the instruments.This includes starting and stopping logging, viewing and storing your data, and...
Page 128 - Select the instrument’s icon.; Starting or Stopping all Instruments at Once; Select the icon for the Master instrument.
2640A/2645A NetDAQ Users Manual 4-4 To stop logging for an individual instrument, complete the following procedure: 1. Select the instrument’s icon. 2. Select Logging | Stop Instrument or click the Stop Instrument button on thetoolbar. If the menu command and button are dimmed, the instrument is not...
Page 129 - Starting and Stopping Logging; Clearing an Instrument’s Totalizer Value; Highlight the desired instrument icon.
Operating NetDAQ Logger for Windows Starting and Stopping Logging 4 4-5 Clearing an Instrument’s Totalizer Value 4-5. If you check Logging | Clear Totalizer on Start, NetDAQ Logger will clear thetotalizer count in each instrument when you start logging for that instrument.Otherwise, the totalizer va...
Page 130 - Click OK to return to the Main Window.
2640A/2645A NetDAQ Users Manual 4-6 ds029s.bmp 3. Check Stop Scanning after Interval. Enter the hours/minutes/seconds for theduration. The maximum entry is 999 hours, 99 minutes, 99 seconds(999:99:99). 4. Click OK to return to the Main Window. Real-Time Displays 4-8. The Real-Time displays let you m...
Page 131 - The Logging Status Window
Operating NetDAQ Logger for Windows Real-Time Displays 4 4-7 • Trend Link provides a wide range of features for trending real time data orpreviously collected data. See Chapter 5 of this manual for more informationon Trend Link. • Dynamic Data Exchange (DDE) links measurement data to other applicati...
Page 132 - The Readings Table Window
2640A/2645A NetDAQ Users Manual 4-8 Instrument listed by BCN Number of scans in the instrument scan queue Number of scans retrieved and recorded. Listing of channels where alarms haveoccurred since the last status update. AlarmBell is displayedwhen at least one channel is in alarm. ds073c.eps The Re...
Page 133 - The Spy Window; Complete the following procedure to use the Spy Window.
Operating NetDAQ Logger for Windows Real-Time Displays 4 4-9 ds300s.bmp 2. Select the icon of the instrument for which you would like to view incomingdata. If this instrument is not currently logging, select Logging | StartInstrument. The Spy Window 4-11. The Spy Window displays up-to-date measureme...
Page 135 - Quick Plot; A summary of Quick Plot attributes is shown below.
Operating NetDAQ Logger for Windows Real-Time Displays 4 4-11 Quick Plot 4-12. Quick Plot provides a quick way to get a simple graphical view of your loggingdata. Quick Plot samples the incoming scan data at a specified interval anddisplays a graphical trend of the data. For a more detailed view of ...
Page 137 - Alarm1 (Alarm 1 status, decimal equivalent of the 30 Alarm 1 states); Data Files; A summary of each data file type is as follows.
Operating NetDAQ Logger for Windows Data Files 4 4-13 Items: ‘Cnn’ (nn is the channel number 01 to 30) DIO (Digital I/O status, decimal equivalent of the 8 I/O lines) Totalizer (Totalizer count, 0 to 4294967295) Alarm1 (Alarm 1 status, decimal equivalent of the 30 Alarm 1 states) Alarm2 (Alarm 2 sta...
Page 138 - Trend Link; Configuring a Data File; Click Data File to open the Data File Configuration dialog box.
2640A/2645A NetDAQ Users Manual 4-14 • Trend Link The Trend Link data file format requires using the real-time trendingcapability offered by the Trend Link trending package. You can record datainto Fast Binary and then convert it into the Trend Link format. When using the Trend Link format, you may ...
Page 139 - Click Auto Rollover to open the Auto Rollover dialog box.
Operating NetDAQ Logger for Windows Data Files 4 4-15 3. Select the desired File Format. For the Fast Binary or ASCII (CSV) format,select Append to File or Overwrite File. This file mode determines whathappens when logging starts if a data file of the same name exists. Enter textcharacters (79 maxim...
Page 140 - Click OK to return to the Data File Configuration dialog box.
2640A/2645A NetDAQ Users Manual 4-16 6. Select the desired Auto Rollover type and its associated value. The data filemay be automatically rolled over when it reaches a specified size, when atime interval elapses from the last rollover, or at a particular time each day. To safeguard your data when re...
Page 141 - Entire logging session.
Operating NetDAQ Logger for Windows Data Files 4 4-17 ds301.bmp 12. Enter a batch name and the desired batch definition and click OK to return tothe Data File Configuration dialog box. Use the batch definition to tell NetDAQ Logger when to mark the start andstop points of batches: • Entire logging s...
Page 142 - Converting Data Files; Binary to ASCII; Viewing and Printing a Data File; Minimize NetDAQ Logger and open the Notepad application.
2640A/2645A NetDAQ Users Manual 4-18 Converting Data Files 4-16. You can convert data files to other types for compatibility with your datamanagement applications. NetDAQ Logger performs the following conversions: • Binary to ASCII • ASCII to Text (timestamp conversion) • Binary to Trend Link Comple...
Page 143 - Optimizing Performance
Operating NetDAQ Logger for Windows Optimizing Performance 4 4-19 The example below is a typical data file in text format. Off-scalemeasurements are recorded as +1.0E+9 (+overload) and -1.0E+9 (-overload).Open thermocouple measurements are recorded as +9.0E+9. Data File Name Number of Columns(An asy...
Page 144 - Optimizing Performance for Speed; Increasing Scanning Rate
2640A/2645A NetDAQ Users Manual 4-20 Optimizing Performance 4-18. You can set up instrument operations to optimize either your system’s datacollection rate or measurement precision, depending on your requirements. Theprocedures in this section describe the options for optimizing performance. Optimiz...
Page 145 - Increasing Network Speed; Optimizing Performance for Precision; Using Online Help
Operating NetDAQ Logger for Windows Using Online Help 4 4-21 In addition, try to avoid running other tasks on the computer when you are writingto the data file. The following suggestions are: • Avoid using the Quick Plot, Spy, Readings Table, or Logging Statuswindows. • Avoid using the Trend Link wi...
Page 147 - Chapter 5; Using Trend Link for Fluke
5-1 Chapter 5 Using Trend Link for Fluke Contents Page 5-1. Introduction .......................................................................................... 5-3 5-2. Displaying a Trend Link Chart During Logging.............................. 5-3 5-3. Playing Back a Trend Link File in Trend Lin...
Page 149 - Displaying a Trend Link Chart During Logging
Using Trend Link for Fluke Introduction 5 5-3 Introduction 5-1. Trend Link for Fluke (Trend Link) graphs real time and historical data in theMicrosoft Windows environment. This chapter supports Trend Link operationwith a series of procedures to use in conjunction with NetDAQ Logger. Refer tothe Tren...
Page 150 - Close Trend Link by selecting File | Exit.; Playing Back a Trend Link File in Trend Link; Select File | Open to display the Chart Open dialog box.
2640A/2645A NetDAQ Users Manual 5-4 Note If this is the first use of the Trend Link data file, allow a few secondsfor Trend Link to create the folder and files used for the chart. 4. To change the look of the chart, see "Getting the Right Look for your TrendLink Chart" later in this chapter....
Page 151 - Select the chart file to play back. For example, select; Playing Back a Fast Binary File in Trend Link
Using Trend Link for Fluke Introduction 5 5-5 ds123s.bmp 3. Select the chart file to play back. For example, select c:\Program Files\Fluke\netdaq\_data.set\data.cht 4. Click OK. 5. Observe that the selected chart opens. Use the scroll bar at the bottom of thechart to display the measurement area of ...
Page 152 - Playing Back an ASCII (CSV) File in Trend Link; Create a New Trend Link Data Set Directory; Data Set name = Data Set path
2640A/2645A NetDAQ Users Manual 5-6 Playing Back an ASCII (CSV) File in Trend Link 5-5. To play back a NetDAQ Logger ASCII (CSV) data file in Trend Link, refer to“Importing Trend Link Data Files” in this chapter. Specify Fluke Format ( *.csv ) for the Import File Type. Be sure you do NOT import the ...
Page 153 - Import the NetDAQ ASCII (CSV) File into Trend Link; Import File Type Select Fluke Format (; Click OK when you have made all selections.; Display the Trend Link Chart for the Imported File; . Repeat this step for each curve that you
Using Trend Link for Fluke Introduction 5 5-7 Import the NetDAQ ASCII (CSV) File into Trend Link 5-9. Open Trend Link from the Programs list. Select Utilities | Import CSV Data todisplay the Import *.CSV File dialog box. Make the following selections: Target Data Set Double-click the Data Set you en...
Page 154 - Title the Trend Link Chart
2640A/2645A NetDAQ Users Manual 5-8 ds126s.bmp Examine the chart and perform any standard Trend Link procedures to view, print,and so forth. (See "Getting the Right Look for Your Trend Link Chart.") Title the Trend Link Chart 5-11. To open the Chart Title dialog box, select Preferences | Cha...
Page 155 - Save the Trend Link Chart; Exporting Trend Link Data Files; and Spread Sheet Column Oriented (
Using Trend Link for Fluke Introduction 5 5-9 ds139s.bmp Save the Trend Link Chart 5-12. To save the Trend Link chart, select File | Save As. Enter the path to your DataSet directory for the *.cht chart. For example, enter c:\Program Files\Fluke\netdaq\_data.set\data.cht. Click OK. ds150s.bmp Export...
Page 156 - Select File | Open and select the desired Trend Link chart. Click OK.
2640A/2645A NetDAQ Users Manual 5-10 Link format. If this is not the case, see "Playing Back a Fast Binary (BIN) File inTrend Link" or "Playing Back an ASCII (CSV) File in Trend Link" for conversionprocedures. 1. Open Trend Link from Start | Programs. 2. Select File | Open and select...
Page 157 - Deleting Old Trend Link Files; file
Using Trend Link for Fluke Introduction 5 5-11 ds151s.bmp Deleting Old Trend Link Files 5-14. NetDAQ Logger creates a *.set file and a directory for every new Trend Link datafile. The directory name is identical to the Trend Link data file name, except foran underscore "_" character prefix. ...
Page 158 - Repeat Steps 2 and 3 for each Trend Link Data Set being deleted.; Getting the Right Look for Your Trend Link Chart; Using the Trend Link Control Bar
2640A/2645A NetDAQ Users Manual 5-12 3. Repeat Steps 2 and 3 for each Trend Link Data Set being deleted. Getting the Right Look for Your Trend Link Chart 5-15. When you display a Trend Link chart, you can modify the chart appearance to suityour application. Use the Control Bar or various menu select...
Page 159 - line in the Data Set box to display the
Using Trend Link for Fluke Getting the Right Look for Your Trend Link Chart 5 5-13 Control Bar Historical Mode: Click to freeze the chart during the Real TimeMode, allowing you to use the scroll bar at the bottom of the chart toback up to any portion on the recorded chart. If you are collectingmeasu...
Page 161 - Using the Trend Link Menus
Using Trend Link for Fluke Getting the Right Look for Your Trend Link Chart 5 5-15 Using the Trend Link Menus 5-17. The Trend Link menus (below) allow you to modify the chart display appearanceand functionality. This section does not discuss menu functions repeated on theControl Bar. (See "Using...
Page 162 - and add the line
2640A/2645A NetDAQ Users Manual 5-16 ds132s.bmp To change the default value so that all created charts will have a selected presetvalue in the Time Interval Window, use the Notepad accessory to modify c:\Program Files\Fluke\netdaq\tldef.cht . Find [Section: Curve Window], and add the line Curve Wind...
Page 163 - Configuring the Curve Status Display
Using Trend Link for Fluke Getting the Right Look for Your Trend Link Chart 5 5-17 ds133s.bmp Configuring the Curve Status Display 5-20. Complete the following procedure to select the elements of the Curve Statusdisplay. The Curve Status display is the window at the bottom of the chartcontaining the...
Page 164 - Curve Preferences
2640A/2645A NetDAQ Users Manual 5-18 ds134s.bmp 2. Select the desired features of the Curve Status display by checking orunchecking each of the Sections Displayed listings. Click OK. Curve Preferences 5-21. Complete the following procedure to select the curve preferences. You cancustomize various at...
Page 167 - Color Click here to select the color of the curve trace.; Background Preferences
Using Trend Link for Fluke Getting the Right Look for Your Trend Link Chart 5 5-21 the left side of the curve window will include a display of scale values.Otherwise, a band without numbers will appear. Curve Type Select the type of curve: normal, highlight, shaded, or SPC(Statistical Processing Con...
Page 168 - The dialog box entries are as follows:; Bkgrnd Color: Choose the background color.
2640A/2645A NetDAQ Users Manual 5-22 ds137s.bmp 2. The dialog box entries are as follows: • Bkgrnd Color: Choose the background color. • Grid Color: Choose the grid color. • Display Grid: This is the same as selecting Grid Lines On/Off on theControl Bar. • Calculate Interval: Select this option to e...
Page 169 - Real Time Frequency Update; use this procedure to enter any custom title.
Using Trend Link for Fluke Getting the Right Look for Your Trend Link Chart 5 5-23 • Constant Interval: Select this option to enter the exact time interval thata vertical grid line should represent. • Pixel Width: Select this option to enter the distance between vertical gridlines in pixels. • Use s...
Page 170 - station one; Using the Note System; When the message “Add note permanently to curve?” appears, click Yes.; Printing a Chart; Bring into view the portion of the chart you wish to print.
2640A/2645A NetDAQ Users Manual 5-24 1. Select Preferences | Chart Title to open the Chart Title dialog box. Enter thechart title name in the text box, for example, station one . Click OK. 2. Observe that the entered text appears at the top of the chart in the Title Bar. Using the Note System 5-25. ...
Page 172 - Click OK to print the chart.
2640A/2645A NetDAQ Users Manual 5-26 ds141f.bmp Click OK to print the chart.
Page 173 - Chapter 6; Maintenance
6-1 Chapter 6 Maintenance Contents Page 6-1. Introduction .......................................................................................... 6-3 6-2. Self-Test Diagnostics and Error Codes ................................................ 6-3 6-3. Cleaning.........................................
Page 176 - Cleaning; Fuse Replacement; Reinstall the cables removed in Step 1.
2640A/2645A NetDAQ Users Manual 6-4 Cleaning 6-3. Warning Keep the instrument dry to avoid electrical shock topersonnel or damage to the instrument. To preventdamage, never apply solvents to the instrument housing. Wipe the instrument with a cloth lightly dampened with water or mild detergent.Do not...
Page 177 - Performance Test
Maintenance Performance Test 6 6-5 1 Remove all Cables 2 Remove BottomScrews (4 places) 3 Remove Rear Bezel Screws (2 places) 4 Remove Rear Bezel and Case for Fuse Access Fuse (15/100A, 250V Time Delay) Bottom Top Figure 6-1. Replacing the Fuse
Page 181 - Initializing the Performance Test Setup; Apply Host Computer Power Apply power to the host computer.
Maintenance Performance Test 6 6-9 Initializing the Performance Test Setup 6-7. Complete the following procedure to initialize the performance test setup. It isassumed you have configured the host computer and instrument as described in"Configuring the Performance Test Setup" (above). Testin...
Page 183 - Accuracy Performance Tests
Maintenance Performance Test 6 6-11 Accuracy Performance Tests 6-8. This accuracy performance test assumes you have completed "Initializing thePerformance Test Setup" above. Do not begin this test until the instrument hasbeen temperature stabilized for a minimum of 30 minutes. Do not use the...
Page 185 - Volts AC Accuracy Test
Maintenance Performance Test 6 6-13 3. Verify Accuracy Configure the 5700A for the output values below andverify the Spy window measurement is between the minimum and maximumvalues. Change the channel 1 range as required (see Step 1). Volts DC Range 5700A Output Minimum Reading Maximum Reading 90 mV...
Page 186 - Frequency Accuracy Test
2640A/2645A NetDAQ Users Manual 6-14 3. Verify Accuracy Configure the 5700A for the output values below andverify the Spy window measurement is between the minimum and maximumvalues. Change the channel 1 range as required (see Step 1). Volts AC Range 5700A Output Minimum Reading Maximum Reading 300 ...
Page 187 - Analog Channel Integrity Test; Repeat Test Repeat steps 4 to 5 for channels 29 and 30.
Maintenance Performance Test 6 6-15 Analog Channel Integrity Test 6-13. Complete the following procedure to test the integrity of each analog channel (2to 20) to verify each analog channel is capable of making measurements. 1. Configure Channel for Ohms In NetDAQ Logger, configure channels 2(then 3,...
Page 188 - Thermocouple Temperature Accuracy Test; This test checks the Open Thermocouple response.
2640A/2645A NetDAQ Users Manual 6-16 Thermocouple Temperature Accuracy Test 6-15. Ensure that the Accuracy Tests (above) have been completed before performingthis test. 1. Connect a Thermocouple Remove the Universal Input Module from theinstrument and connect the supplied type T thermocouple to the ...
Page 198 - Digital Input/Output Tests; Digital I/O Output Test; This test checks the Digital I/O lines when used as outputs.
2640A/2645A NetDAQ Users Manual 6-26 4. Verify Accuracy Insert the RTD and a mercury thermometer in a room-temperature bath. Allow 20 minutes for thermal stabilization. The valuedisplayed on the mercury thermometer should equal the value in the SpyWindow +0.25 ° C (2640A) or +0.80 ° C (2645A) plus s...
Page 199 - Digital Input Test; This test checks the Digital I/O lines when used as inputs.
Maintenance Performance Test 6 6-27 8. Short Channel Inputs Remove the Universal Input Module from theinstrument and apply a short to channels 1 through 8. Reinstall the UniversalInput Module. 9. Start Instrument Scanning Select Logging | Start Instrument to start theinstrument scanning. The instrum...
Page 200 - Totalizer Tests; Totalizer Count Test; This test checks the ability of the Totalizer feature to count.; Increase Totalizer Count Connect the; Totalizer Sensitivity Test
2640A/2645A NetDAQ Users Manual 6-28 3. Verify Digital I/O Input for all Set Lines In sequence, individually groundeach DIO line to the GND line using the DIO wires connected in Step 1. Notethe change in the DIO status reported in the Spy window as follows: None grounded Reported DIO Status = 255DIO...
Page 201 - Connect Test Leads At the DIGITAL I/O connector, connect the; Master Alarm Output Test
Maintenance Performance Test 6 6-29 2. Connect Test Leads At the DIGITAL I/O connector, connect the Σ (Totalizer) test lead and GND test lead to a signal generator’s outputterminals. Adjust the signal generator for an output of 1.5V rms sine wave at10 Hz. 3. Open Spy Window Select the Spy command fr...
Page 202 - Trigger Input Test
2640A/2645A NetDAQ Users Manual 6-30 Trigger Input Test 6-31. This test checks the ability of the Trigger Input line to trigger measurementscanning. 1. Configure Trigger Input In NetDAQ Logger, configure the scan parametersfor External Trigger with an Interval 2 of 1 second. Be sure Interval Trigger...
Page 203 - Calibration
Maintenance Calibration 6 6-31 4. Verify Trigger Output is Enabled In NetDAQ Logger, select Setup |Instrument Config. In the Instrument Configuration dialog box, verify theTrigger Out box is checked. Click OK to return to the Main Window. (See“The Instrument Configuration Dialog Box" in Chapter ...
Page 204 - Adding Calibration to the Utilities Menu
2640A/2645A NetDAQ Users Manual 6-32 Note Do not press CAL ENABLE unless you have a copy of the ServiceManual and intend to calibrate the instrument. If you have activatedcalibration and wish to exit, press CAL ENABLE until the CALannunciator is removed from the display (or turn the power off). Addi...
Page 205 - Instrument Calibration Dialog Box; Volts dc Click to perform volts dc calibration.; Calibration Steps Dialog Box
Maintenance Calibration 6 6-33 Instrument Calibration Dialog Box 6-35. The Instrument Calibration dialog box (below) sets up the RS-232 port forcalibrating the instrument and provides calibration function selection. See theService Manual (PN 942615) for the calibration procedure. ds306s.bmp RS-232 P...
Page 206 - Variations in the Display; After prolonged periods of displaying the same information.; Service
2640A/2645A NetDAQ Users Manual 6-34 ds308s.bmp Apply Displays the suggested value to apply to the instrument from the calibrator. Actual Enter the actual value applied to the instrument from the calibrator. Use avalue as close as possible to the suggested value for best calibration results. Perform...
Page 207 - For written assistance or information, contact:
Maintenance Service 6 6-35 Centers is provided in Appendix K of this manual. Be sure to pack the instrumentsecurely; use the original container if available. Include a description of theproblem. Fluke assumes no responsibility for damage in transit. For application or operation assistance or informa...
Page 208 - Replacement Parts
2640A/2645A NetDAQ Users Manual 6-36 Fluke offers a 1-year warranty for the 264XA-801 (Ethernet plug-in card) and264XA-802 (Ethernet Parallel-to-LAN adapter), and 264XA-803 (PCMCIAAdapter) options. Defective units may be returned to either Fluke or to the optionmanufacturer for repair or replacement...
Page 209 - Appendix
Appendices Appendix Title Page A Specifications ....................................................................................... A-1 B Noise, Shielding, and Crosstalk Considerations .................................. B-1 C True-RMS Measurements ...................................................
Page 211 - Appendix A; Specifications
A-1 Appendix A Specifications Introduction A-1. Specifications are divided into three sections. The first section contains thecombined specifications that apply equally to both the 2640A and 2645Ainstruments. The second section contains specifications that apply only to the2640A instrument. The thir...
Page 215 - Trigger In
Specifications 2640A/2645A Combined Specifications A A-5 Table A-3. 2640A/2645A DIGITAL I/O Specification (cont) Specification Characteristic Output Voltage - TTL Logical Zero 0.8V maximum for an I out of -1.0 mA (1 LSTTL load) Output Voltage - TTL Logical One 3.8V minimum for an I out of 0.05 mA (1...
Page 216 - Trigger Out
2640A/2645A NetDAQ Users Manual A-6 Trigger Out A-8. Table A-5 provides a summary of the Trigger Out specifications. The Trigger Outoutput is located on the ALARM/TRIGGER I/O connector, terminals TO andGND. Table A-5. 2640A/2645A Trigger Out (TO) Specification Specification Characteristic TTL Logica...
Page 218 - 640A DC Voltage Measurement Specifications
2640A/2645A NetDAQ Users Manual A-8 2640A DC Voltage Measurement Specifications A-13. Tables A-9 to A-11 provide 2640A specifications for the dc voltage measurementfunction. Table A-9. 2640A DC Voltage Measurement General Specifications Specification Characteristic Input Impedance 100 M Ω in paralle...
Page 219 - 640A AC Voltage Measurement Specifications
Specifications 2640A Specifications A A-9 Table A-11. 2640A DC Voltage Accuracy Specifications Accuracy, 3 σ ± (% input + V) 18 ° C to 28 ° C -10 ° C to 60 ° C Range 90 Day 1 Year 1 Year Slow Fast Slow Fast Slow Fast 90 mV .01%+7 µV .01%+17 µV .013%+8 µV .013%+18 µV .042%+18.2µV .042%+44.2µV 300 mV ...
Page 222 - 640A 4-Wire Resistance Measurement Specifications
2640A/2645A NetDAQ Users Manual A-12 2640A 4-Wire Resistance Measurement Specifications A-15. Tables A-15 to A-17 provide 2640A specifications for the 4-wire resistancemeasurement function. The 4-wire measurements use 2 input channels a decadeapart, e.g., channels 4 and 14. Table A-15. 2640A 4-Wire ...
Page 223 - 640A 2-Wire Resistance Measurement Specifications
Specifications 2640A Specifications A A-13 2640A 2-Wire Resistance Measurement Specifications A-16. The 2640A specifications for the 2-wire resistance measurement function is basedon the 4-wire resistance measurement specification (above) except you add anominal 5-Ohm (10-Ohm maximum) positive offse...
Page 226 - 640A Frequency Measurement Specifications
2640A/2645A NetDAQ Users Manual A-16 2640A Frequency Measurement Specifications A-20. Tables A-22 to A-23 provide 2640A specifications for the frequency measurementfunction. Table A-22. 2640A Frequency Accuracy Specifications Frequency Measurement Accuracy, 1 Year, -10 º C to 60 º C Range Resolution...
Page 227 - 645A DC Voltage Measurement Specifications
Specifications 2645A Specifications A A-17 2645A Specifications A-21. This section includes specifications specific to the 2644A instrument bymeasurement function. 2645A DC Voltage Measurement Specifications A-22. Tables A-24 to A-26 provide 2645A specifications for the dc voltage measurementfunctio...
Page 229 - 645A AC Voltage Measurement Specifications
Specifications 2645A Specifications A A-19 2645A AC Voltage Measurement Specifications A-23. Tables A-27 to A-29 provide 2645A specifications for the ac voltage measurementfunction. Table A-27. 2645A AC Voltage General Specifications Specification Characteristic Input Impedance 1 M Ω in parallel wit...
Page 231 - 645A 4-Wire Resistance Measurement Specifications
Specifications 2645A Specifications A A-21 2645A 4-Wire Resistance Measurement Specifications A-24. Tables A-30 to A-32 provide 2645A specifications for the 4-wire resistancemeasurement function. The 4-wire measurements use 2 input channels a decadeapart, e.g., channels 4 and 14. Table A-30. 2645A 4...
Page 232 - 645A 2-Wire Resistance Measurement Specifications
2640A/2645A NetDAQ Users Manual A-22 2645A 2-Wire Resistance Measurement Specifications A-25. The 2645A specifications for the 2-wire resistance measurement function is basedon the 4-wire resistance measurement specification (above) except you add a 700to 1000 Ohm positive offset. This value varies ...
Page 235 - 645A Frequency Measurement Specifications
Specifications 2645A Specifications A A-25 2645A Frequency Measurement Specifications A-28. Tables A-37 to A-38 provide 2645A specifications for the frequency measurementfunction. Table A-37. 2645A Frequency Accuracy Specifications Frequency Measurement Accuracy, 1 Year, -10 º C to 60 º C Range Reso...
Page 237 - Appendix B; Considerations
B-1 Appendix B Noise, Shielding, and Crosstalk Considerations Introduction B-1. This appendix supplies information on how you can setup your instruments tominimize measurement errors. The topics include static versus dynamicmeasurement, noise and shielding, and how to minimize crosstalk effects Stat...
Page 238 - Route signal wiring away from power mains wiring.; Crosstalk Considerations; can be greatly affected by
2640A/2645A NetDAQ Users Manual B-2 reading rate and scan interval that is an exact multiple of the power line periodand cause line frequency noise to be aliased to a very low frequency, where itappears as drift, or instability, in the readings. For example, if the instrument is setto Reading Rate =...
Page 239 - Noise, Shielding, and Crosstalk Considerations
Noise, Shielding, and Crosstalk Considerations Crosstalk Considerations B B-3 the routing of your input wiring: keep it away from or shielded from all acpower mains wiring and other sources of noise. 3. Channel Spacing When possible, leave as many unused channels as possiblebetween your ac voltage c...
Page 241 - Appendix C; True-RMS Measurements
C-1 Appendix C True-RMS Measurements Introduction C-1. The instrument measures the true rms value of ac voltages. In physical terms, therms (root-mean-square) value of an waveform is the equivalent dc value thatcauses the same amount of heat to be dissipated in a resistor. True rmsmeasurement greatl...
Page 245 - Appendix D; RTD Linearization; may be specified via NetDAQ
D-1 Appendix D RTD Linearization Introduction D-1. This appendix discusses Fixed-385 and Custom-385 RTD measurements withNetDAQ. Fixed-385 RTD D-2. The Fixed-385 RTD linearization uses segmented polynomials to approximate theCallendar Van-Dusen equation for a 385 RTD. R 0 may be specified via NetDAQ...
Page 247 - Using the Custom-385 RTD with Other Platinum RTDs
RTD Linearization Custom-385 RTD D D-3 Using the Custom-385 RTD with Other Platinum RTDs D-4. Although the Custom-385 RTD linearization uses a β and δ from a 385 RTD, other RTD types can be used. The Custom-385 linearization will compensate forthe change in α , but errors will be introduced due to t...
Page 251 - Appendix E; Computed Channel Equations; Computed channel equations must observe the following syntax:
E-1 Appendix E Computed Channel Equations Introduction E-1. An equation is converted into a stored binary format which is sent to theinstrument where the calculations are performed during the processing of eachscan. Constants are passed to the instrument as single precision (4 byte) floatingpoint nu...
Page 252 - enclose groups of elements to clarify the scope of other symbols
2640A/2645A NetDAQ Users Manual E-2 ″ ″ enclose literal characters or strings which must appear as shown (except forcase) () enclose groups of elements to clarify the scope of other symbols <equation> = <add-term> { (“+” | “-”) <add-term> } <add-term> = <mul-term { (“*” | ...
Page 253 - Appendix F; Data File Format
F-1 Appendix F Data File Format Introduction F-1. The ASCII (CSV) data file is maintained in comma-separated ASCII format. TheASCII format is compatible with data analysis and management tools runningsimultaneously in other computers on a network. A conversion utility, Binary toASCII Conversion (Uti...
Page 255 - Appendix G
G-1 Appendix G Dynamic Data Exchange (DDE) Introduction G-1. Dynamic Data Exchange (DDE) provides NetDAQ Logger data to other Windowsapplications in real time. For example, with a DDE connection, data collected byNetDAQ Logger can be displayed in a spreadsheet as it is being collected from theinstru...
Page 256 - DDE Example Using Excel
2640A/2645A NetDAQ Users Manual G-2 DDE reference includes the Service Name, Topics, and Items: (cont) Service Name Topics Items NETDAQ for 16-bit NETDAQ32 for 32-bit Alarm1(The status of Alarm 1 for all30 channels is summarized asa 30-bit binary numberexpressed in decimal.) Alarm2(The status of Ala...
Page 258 - Continue by entering the following:
2640A/2645A NetDAQ Users Manual G-4 5. Continue by entering the following: Cell Formula A4 =NETDAQ32|bcn25!DIO A5 =NETDAQ32|bcn25!Totalizer A6 =NETDAQ32|bcn25!Alarm1 A7 =NETDAQ32|bcn25!Alarm2 A8 =NETDAQ32|bcn25!ElDateTime 6. Your entries will appear on the spreadsheet as shown below for the conditio...
Page 259 - A note about the Alarm 1 and Alarm 2 values reported in Excel:
Dynamic Data Exchange (DDE) DDE Example Using Excel G G-5 A note about the Alarm 1 and Alarm 2 values reported in Excel: The alarm value is the decimal equivalent of a 30-bit binary numberrepresents the alarm conditions for each of the 30 instrument channels (20analog and 10 computed channels). In t...
Page 261 - Appendix H; Ethernet Cabling
H-1 Appendix H Ethernet Cabling Introduction H-1. This appendix details the unshielded twisted-pair (UTP) cables used for 10BaseTEthernet interconnection. UTP cable configurations are summarized in Figure H-1. UTP Cables H-2. There are two types of UTP cables: patch cord cables and direct connection...
Page 263 - Appendix I; Network Considerations; Assign a destination address.; NetDAQ and Network Communication Techniques
I-1 Appendix I Network Considerations Introduction I-1. This is a tutorial on commonly used Network communication technology and howto get started using the NetDAQ instruments. It’s not necessary to know the detailsof Network communication methods in order to use NetDAQ effectively.However, there ar...
Page 264 - NetDAQ and Your Network Administrator
2640A/2645A NetDAQ Users Manual I-2 low system data throughput because of the processing rate of host computers anddisk systems, and instruments that are physically close to the host computer. Now, the increased processing capability of affordable host computers, increasedchannel measurement rates, ...
Page 265 - Will all the routers route IP packets?; Network Primer
Network Considerations Network Primer I I-3 2. Does the network contain more than one subnet? If so, what’s the IP address ofmy default gateway or router device? 3. Will all the routers route IP packets? 4. What are my domain and host names (optional)? The minimum information you need to get is the ...
Page 266 - Physical Layer Wiring Schemes Supported by NetDAQ
2640A/2645A NetDAQ Users Manual I-4 the computers and other devices attached to the network wiring. The Ethernetstandard consists of a set of low-level addressing and data transmission protocolsthat run on several different wiring schemes. Computers on a network are usually referred to as a “host.” ...
Page 267 - Basic Network Packet and Frame Contents
Network Considerations Network Primer I I-5 • A Repeater is a device used to physically isolate and extend the length of anetwork segment, but it does not divide the network into subnets. • A hub is a device that allows communication among multiple 10BaseTnetwork hosts and allows these hosts to comm...
Page 268 - IP Addresses and Segmented Networks
2640A/2645A NetDAQ Users Manual I-6 strips off the Frame information around the packet. So, unless you’re doing low-level analysis of network traffic, you never need to deal with Frame information. For any protocol, the Packet consists of a destination address, a source address,some protocol type in...
Page 270 - Host PC
2640A/2645A NetDAQ Users Manual I-8 ID portion to be the higher order 24 bits of the IP address (a class C address).Using the Subnet Mask, the TCP/IP software is able to detect if a destinationaddress is in the local subnet or if the packet must be sent to another subnetthrough a Router or Gateway. ...
Page 271 - Ethernet Adapter Hardware for the PC
Network Considerations Network Primer I I-9 software sends out a broadcast frame that every TCP/IP host and router receives.This frame contains a request for the host with the unknown IP address to return aframe to the ARP requester. This response frame contains the Ethernet address ofthe sourcing h...
Page 273 - Link Support; TCP/IP Protocol Stack
Network Considerations Network Primer I I-11 While a protocol section might look like the following: [NETMANAGE]DRIVERNAME=NETMNG$BINDINGS=EtherLinkII These two protocol.ini file sections describe a hardware driver characteristic for a 3COM EtherLink II hardware card and specifies that the NetManage...
Page 275 - The; WINSOCK DLL and Application Programming Interface; . This DLL contains a set of functions that support; Notes to Network Administrators
Network Considerations Notes to Network Administrators I I-13 matching network ID. This process is generally referred to as routing the packet,and devices that do this are also referred to as Routers or Gateways. In many modern networks, the use of subnet masking and the default gateway isnot necess...
Page 276 - that supports version 1.1 of the WINSOCK standard. A
2640A/2645A NetDAQ Users Manual I-14 ARP protocol or have static routing tables, you don’t have to use these parameters.In the Proxy ARP protocol, the routing path between any two hosts isautomatically established by the routers during the ARP process. The NetDAQ PC software communicates with the Ne...
Page 277 - Running Two Network Cards in a PC; file for the NDIS driver, or in the driver section
Network Considerations Running Two Network Cards in a PC I I-15 If your network contains routers, bridges, gateways, or other devices that dividethe network into subnets, and you intend to have the PC on a different subnet fromthe NetDAQ instrument(s), you need to make sure that the devices can rout...
Page 278 - Troubleshooting Information
2640A/2645A NetDAQ Users Manual I-16 Network drivers typically use PC soft interrupts to communicate with the protocolsoftware. If you’re using two network cards, and as a result, two drivers, you mustselect two different soft interrupt levels for the drivers. For example, Packetdrivers (used with T...
Page 279 - file (value is seconds) to increase the timeout up to 1000
Network Considerations Troubleshooting Information I I-17 NetDAQ instrument and the PC default gateway addresses to make sure theyare set correctly. Check all network wiring. Make sure that 10Base2 terminators are placed atthe two open ends of the coax. Make sure that there is no more than onetermin...
Page 280 - Interaction with advanced display adapter driver packages.
2640A/2645A NetDAQ Users Manual I-18 Move your instrument to a less loaded subnet of your network or isolate theNetDAQ instruments on their own network. If possible, increase the scan interval. Use the Fast Binary file format rather than CSV or Trend Link formats. Make sure that there are few or no ...
Page 281 - unidir; Handling Token Ring Networks
Network Considerations Handling Token Ring Networks I I-19 7. Failure to communicate when using the Parallel to LAN Adapter (PLA) In some computers the speed of the parallel port hardware is insufficient tosupport the normal bi-directional communication used by the Silicom PLA. Inthese rare cases, y...
Page 282 - Once hardware setup is complete, continue the installation.; Newt Networking Software
2640A/2645A NetDAQ Users Manual I-20 If you install Newt networking software on a token ring network on Windows 3.1,the Newt installation procedure must be modified slightly. When prompted for theinterface type in the “Add Interface” dialog box, use the list pull-down control todisplay the other typ...
Page 283 - associated with the
Network Considerations Newt Networking Software I I-21 Ethernet Components and Wiring The NetDAQ instruments and host computershould be connected to the network as shown in Chapter 2 of this manual. Instrument Preparation Each NetDAQ instrument should be connected to an acoutlet or dc source with a ...
Page 284 - Link Driver EXP16ODI
2640A/2645A NetDAQ Users Manual I-22 Link Driver EXP16ODI Port 300Frame Ethernet_802.3Frame Ethernet_IIProtocol IPX 0 Ethernet_802.3 Newt TCP/IP vs. Microsoft LANManager I-19. If you have the Microsoft LANManager software already installed on your PC,you do not need to install a new driver. However,...
Page 285 - hardware driver name
Network Considerations Newt Networking Software I I-23 After configuring Vines for NDIS driver support, install Newt. Select "None"when asked for a hardware Vendor. Once Newt is installed, edit the protocol.ini file in the Vines directory to add the following section: [NETMANAGE] DRIVERNAME=...
Page 286 - Yes; Newt TCP/IP vs. Windows for Workgroups and Novell NetWare
2640A/2645A NetDAQ Users Manual I-24 • autoexec.bat Netmanage directory added to the path. • system.ini netmanage.dos added to the transport= line under the [Network Drivers] section. LoadRMDrivers= will be changed to Yes . • protocol.ini [NETMANAGE] section added that should have a BINDINGS= line t...
Page 287 - Accept the default directory
Network Considerations Newt Networking Software I I-25 d. Accept the default directory c:\netmanag . Click Continue. e. Acknowledge the Setup is Complete! message by clicking OK. 3. Select Interface | Add. Accept the Ethernet defaults. Click OK. If you are already using Novell NetWare Client softwar...
Page 290 - File
2640A/2645A NetDAQ Users Manual I-28 Write down the Physical (Ethernet) Address from the Custom window inside theback cover of this manual. Select File | Exit to close Newt.
Page 293 - Glossary; library, which itself gives users
Network Considerations Glossary I I-31 Glossary I-24. 10Base2 Network wiring scheme using 50 ohm coaxial cable (10 MHz,Baseband, 200 meter max. length). 10BaseT Also called UTP. Network wiring scheme using unshielded, twisted paircable (10 MHz, Baseband, Twisted Pair). API Application Programming In...
Page 294 - Host Any computer system attached to a network.
2640A/2645A NetDAQ Users Manual I-32 DNS Domain Name Service. A database and associated software packages thatprovides the means of establishing and accessing the relationship between thenames and IP addresses of network hosts. Domain A group of computers on the network may be assigned a domain name...
Page 295 - IPX A protocol used by Novell NetWare LANs.
Network Considerations Glossary I I-33 IP A unit of information passed across an Internet. The IP portion of the TCP/IPprotocol is responsible for transmitting blocks of data from the source to thedestination using the IP addresses specified. It also provides for fragmentation andreassembly of data ...
Page 296 - A file with a standard format used to hold configuration
2640A/2645A NetDAQ Users Manual I-34 Packet Driver A network device driver specification defined by 3COM. Packetdrivers support operation of only a single protocol stack over the driver andhardware interface. PC Personal Computer. A much misused term usually used to refer to a computerbuilt to compl...
Page 297 - Basic Data Transfer; Token Bus One type of physical network standardized by IEEE-802.4.
Network Considerations Glossary I I-35 change to TCP/IP is contemplated, an RFC is issued over the Internet andeveryone is invited to comment on the change before its implementation. Router A router provides a communication path between two or more sections ofa network. It directs packets between se...
Page 299 - Appendix J; Error Messages & Exception
J-1 Appendix J Error Messages & Exception Conditions Introduction J-1. This appendix describes error messages and how NetDAQ Logger handlesexception conditions during operation. Communications Connection J-2. If the communications connection to an instrument is broken during logging, amodeless w...
Page 300 - Error and Status Messages
2640A/2645A NetDAQ Users Manual J-2 invalid values into the fields of a dialog box. If a critical instrument error occurs(RAM constants failure or A/D not responding), a modeless window appears witha warning message. For a description of each error message, refer to Error andStatus Messages in the o...
Page 301 - Error Messages & Exception Conditions; W Measurements not allowed on Channels 11-20!
Error Messages & Exception Conditions Error and Status Messages J J-3 [File Name] is not a valid ASCII file! You are trying to use an incompatible ASCII file for this conversion. Make sure you are selecting a NetDAQ ASCII (CSV) file. Other CSV formatssuch as Trend Link cannot be used with the Ne...
Page 302 - All associated configurations will be lost. Continue?
2640A/2645A NetDAQ Users Manual J-4 A/D Reference Balance Values Invalid! Continue? Your instrument has detected a fault during internal performance monitoring. This message does not appear under normal operating conditions. Note themessage and click Yes to continue operations or No to terminate ope...
Page 303 - All Slave instruments will become Async. Continue?; Your instrument may have experienced a failure.; Are you sure you want to Delete All?
Error Messages & Exception Conditions Error and Status Messages J J-5 All Slave instruments will become Async. Continue? You are setting a Master instrument back to an Asynchronous instrument. A group instrument consists of a Master and one or more Slaves. If you redefinethe Master as an Asynchr...
Page 304 - Trend Link could not start.
2640A/2645A NetDAQ Users Manual J-6 Associated configuration will change to reflect new instrument model.Continue? You are modifying an instrument model number in the CommunicationsConfiguration File, which will also modify the instrument model number for theassociated icon on the Icon Bar. Associat...
Page 305 - Calibration Constants Corrupted! Continue?; Your
Error Messages & Exception Conditions Error and Status Messages J J-7 Boot ROM Failure! Continue? Your instrument has detected a fault during internal performance monitoring. This message does not appear under normal operating conditions. Note themessage and click Yes to continue operations or N...
Page 306 - Cannot change Configuration while logging!; You tried to copy channels to an instrument that was logging.; Cannot create data file!; Enter a valid file name and directory path for the output file.
2640A/2645A NetDAQ Users Manual J-8 A Data Set is opened for each instrument in a group instrument and each data setis created by appending 1 of 25 letters in the alphabet. If more than 25 letters arerequired, then this message appears. To delete old Trend Link files, see "DeletingOld Trend Link...
Page 307 - Cannot open data file!
Error Messages & Exception Conditions Error and Status Messages J J-9 Cannot write to Trend Link Data Set! Check amount of free diskspace available. You are trying to write to a Trend Link data file on a disk drive that has little orno storage capacity. To use this disk drive, delete unneeded fi...
Page 308 - Communication Interruption!
2640A/2645A NetDAQ Users Manual J-10 changes in the Auto Rollover dialog box and then cancels the Data FileConfiguration dialog box, the message will appear. Channel [GCN] must be defined for calculation of computed channel[GCN] You tried to reference a channel set to OFF in a computed channel defin...
Page 309 - Configuration will be lost! Delete anyway?; Your network cannot support the connection with the instrument.; Connection Successful!
Error Messages & Exception Conditions Error and Status Messages J J-11 Computed channel [Channel #]: Instrument equation buffer size limitswill be exceeded! The instrument’s buffer space for storing equation tokens (1000 bytes) has beenexceeded. Delete or modify the equations until they occupy n...
Page 310 - Conversion is complete!; You are changing a Slave instrument into a Master instrument.
2640A/2645A NetDAQ Users Manual J-12 Connection Successful, but instrument has older software that willnot support computed channel equations or custom RTDs! You have successfully established communications with an instrument and cancontinue to operate, but you will not be able to use computed chann...
Page 312 - Error registering window class!; The Trend Link application has experienced an error.; Ethernet Parameters Corrupted! Continue?; Your equation contains too many sets of nested parentheses.
2640A/2645A NetDAQ Users Manual J-14 Error registering window class! Windows was unable to open NetDAQ Logger due to a conflict with otherversions of Windows, or Windows is operating with marginal memory. Error returned from Trend Link function [Error] The Trend Link application has experienced an e...
Page 313 - Existing data file is not compatible for appending!; You entered an invalid File Size into the Auto Rollover dialog box.; Front Panel Display Not Responding! Continue?
Error Messages & Exception Conditions Error and Status Messages J J-15 Existing data file is not compatible for appending! You tried to log data to a data file in the Append mode but have added orsubtracted instrument channels from the configuration since the previous loggingsession. Change the ...
Page 314 - Instrument BCN does not match! Reconfigure Communications!
2640A/2645A NetDAQ Users Manual J-16 Instrument [BCN] already has this filename! Your have configured more than one instrument with the same data file name. Check your data file names and make sure each instrument has a unique data filename, or uncheck Enable Data File Recording in the Data File Con...
Page 315 - Instrument model does not match! Reconfigure Communications!; If the network connection stopped working, check the following:; Instrument Scan Queue Overflow! Newest scans discarded.; file” in
Error Messages & Exception Conditions Error and Status Messages J J-17 Instrument model does not match! Reconfigure Communications! The instrument model does not match the entry in the CommunicationsConfiguration File. Modify the instrument model number in the Communications Configuration File(s...
Page 316 - Instrument Scan Queue Overflow! Oldest scans overwritten!
2640A/2645A NetDAQ Users Manual J-18 Instrument Scan Queue Overflow! Oldest scans overwritten! NetDAQ Logger is unable to retrieve scans from the instruments quickly enoughand the instrument scan queue has overflowed. In this condition, the instrument isoverwriting the oldest scans and saving the ne...
Page 317 - Invalid Channel Number!
Error Messages & Exception Conditions Error and Status Messages J J-19 Invalid Channel Number! Your equation contains a channel number that is not expressed with the correctsyntax. When used within an equation, channel numbers must be preceded by the letter‘c’ (or ‘C’). The channel number itself...
Page 318 - Invalid Minimum Value for Scaled Range!; You have entered a Plot Maximum value that is excessively large.; Invalid Plot Minimum Value!; You have entered a Plot Minimum value that is excessively large.; IP Address segments must range from 0 to 255!; You tried to enter an IP address segment that is over 255.
2640A/2645A NetDAQ Users Manual J-20 Invalid Minimum Value for Scaled Range! You have entered an invalid Minimum Value for the Mx+B Input Rangecalculation. Enter a Minimum Value that is less than +9.9999E+9. Invalid Minimum Value for Input Range! You have entered an invalid Minimum Value for the Mx+...
Page 319 - The current Windows memory level cannot support your operations.; Maximum number of Instruments reached!; You tried to add more than 20 instruments to the Icon Bar.; Min. Free Disk Space must be between 0 and 100000!; Amend the equation to refer to another channel.; Missing left parenthesis!
Error Messages & Exception Conditions Error and Status Messages J J-21 Low on memory! The current Windows memory level cannot support your operations. Close all applications and restart Windows. Maximum number of Instruments reached! You tried to add more than 20 instruments to the Icon Bar. The...
Page 320 - Missing right parenthesis!; You selected a Monitor Channel that has the function OFF.; NetDAQ Logger software is outdated for this instrument!
2640A/2645A NetDAQ Users Manual J-22 Missing right parenthesis! One of the sets of parentheses in your equation is not complete, leaving a leftparenthesis without a matching right one. Check each set of parentheses in your equation to locate the one that is missingthe right parenthesis. Must Select ...
Page 321 - No instruments are available for association!; Windows timer resources have been exceeded.; Not responding to command!
Error Messages & Exception Conditions Error and Status Messages J J-23 No instruments are available for association! You tried to create an instrument icon from the Available Instruments list whenthere were no instruments listed. Instruments on the network are either shown as icons on the Icon B...
Page 322 - Out of Memory! Close another application and try again!; You have run out of Windows memory.; Plot Maximum must be larger than Plot Minimum!
2640A/2645A NetDAQ Users Manual J-24 Out of Memory! Close another application and try again! You have run out of Windows memory. Close other applications to free Windows memory and continue operations. PC time earlier than existing Trend Link data! Verify PC time, wait [nn]seconds, or create a new D...
Page 323 - Primary Interval must range from 0 to 24 hours (86400 seconds)!; You have an error in your Communications Configuration File.
Error Messages & Exception Conditions Error and Status Messages J J-25 Primary Interval must range from 0 to 24 hours (86400 seconds)! You tried to enter a Primary Interval greater than 86400 or tried to enterextraneous characters. Enter numbers only between 00000.000 and 86400.000 for the Prima...
Page 324 - Sampling interval must range from 1 to 3600 seconds!
2640A/2645A NetDAQ Users Manual J-26 occurs, you must reconfigure communications to add the network parameters(select the Communications Config command from the Setup menu). ROM Failure! Continue? Your instrument has detected a fault during internal performance monitoring. This message does not appe...
Page 325 - Select at least one trigger type!; Your scan parameters do not include a type of trigger.; Setup File Is Invalid!; Your setup file is corrupted.; Shunt Resistance value must range from 10 to 250 Ohms!; Your host computer is not able to connect to the network.
Error Messages & Exception Conditions Error and Status Messages J J-27 Select at least one trigger type! Your scan parameters do not include a type of trigger. Configure your scan parameters to include at least one of the triggers: IntervalTrigger, External Trigger or Alarm Trigger. Selected Tre...
Page 326 - There are no Trend Link data files to be initialized!
2640A/2645A NetDAQ Users Manual J-28 Socket Port Number must range from 1024 to 65535! You tried to enter a Socket Port number less than 1024, greater than 65535 orincluded extraneous characters. Enter a Socket Port number between 1024 and 65535. The default is 04369. There are no Trend Link data fi...
Page 327 - Token String is too Large!; . This is necessary because the data set directory is the same name; Trigger Interval must range from 0 to 24 hours (86400 seconds)!
Error Messages & Exception Conditions Error and Status Messages J J-29 Token String is too Large! This equation exceeds the limit of 1000 bytes after it’s been converted into binaryformat. Once equations are converted into binary format (so they can be read by theinstrument) they cannot exceed 1...
Page 329 - Unable to load Trend Link DLL!
Error Messages & Exception Conditions Error and Status Messages J J-31 Unable to load Trend Link DLL! NetDAQ Logger is unable to load the Trend Link c:\Program Files\Fluke\tl\tbase.dll into memory. Check that you have sufficient memory space to load tbase.dll , that the tbase.dll has not been co...
Page 330 - Unable to Open Setup File!; Your NetDAQ Logger installation is missing the required; Unable to stop scanning!
2640A/2645A NetDAQ Users Manual J-32 Unable to Open Setup File! You have specified a setup file name or directory that does not exist. Check your files and directories for the correct name and path. If you are usingthis setup file on the NetDAQ Logger command line, change the command line toa valid ...
Page 331 - Unable to Write to Setup File!; You have experienced an unknown communications error.
Error Messages & Exception Conditions Error and Status Messages J J-33 in the Auto Rollover dialog box. You can change the minimum amount of freedisk space that can exist while logging to a Trend Link file by changing thesettings in the Low Disk Space for Trend Link section of the Auto Rollover ...
Page 332 - Window creation failed!
2640A/2645A NetDAQ Users Manual J-34 Window creation failed! Windows was unable to open NetDAQ Logger due to insufficient resources suchas insufficient memory in Windows. Close all applications and restart Windows to clear the resource problem. Winsock.dll not found! The winsock.dll file cannot be f...
Page 333 - Appendix K; Fluke Service Centers
K-1 Appendix K Fluke Service Centers USA CaliforniaFLWFluke Calibration CenterC/o FLW Service Corporation3505 Cadillac Ave., Bldg ECosta Mesa, CA 92626TEL: (714) 751-7512FAX: (714) 755-7332 DallasFluke Service Center2104 Hutton DriveSuite 112Carrollton, TX 75006TEL: (214) 406-1000FAX: (214) 247-5642...
Page 339 - Index
1 Index —1— 10Base2 (Coax) wiring, interconnec- tion, 2-48 10BaseT (twisted-pair) Ethernet wiring , interconnection, 2-48 —2— 2640A 2-Wire resistance measurement specifications, A-13 2640A 4-Wire resistance measurement specifications, A-12 2640A ac voltage measurement specifications, A-9 2640A dc vo...
Page 347 - Instrument Parameter Record (Isolated Network)
Instrument Parameter Record (Isolated Network) Instrument BCN: Note: Instrument BCN: Note: Instrument BCN: Note: Instrument BCN: Note: Instrument BCN: Note: Instrument BCN: Note: Instrument BCN: Note: Instrument BCN: Note: Instrument BCN: Note: Instrument BCN: Note: Instrument BCN: Note: Instrument ...
Page 348 - Instrument Parameter Record (General Network)
Instrument Parameter Record (General Network) Instrument BCN: IP Address: . . . Subnet Mask: . . . Default Gateway: . . . Ethernet Address: : : : : : Note: Instrument BCN: IP Address: . . . Subnet Mask: . . . Default Gateway: . . . Ethernet Address: : : : : : Note: Instrument BCN: IP Address: . . . ...