Page 3 - Before You Begin; This manual describes how to install a Micro Motion; Customer service; Refer to the transmitter manual for troubleshooting help.
Installation Manual 1 Before You Begin Before You Begin This manual describes how to install a Micro Motion ® ELITE ® sensor. The following information is provided in this manual: Customer service For technical assistance, phone the Micro Motion Customer Service department: • In the U.S.A., phone 80...
Page 4 - Specifications
2 Micro Motion ® ELITE ® Sensors Before You Begin Specifications Full product specifications can be found in the product data sheet, which is available from the Micro Motion web site at www.micromotion.com . Definitions The term MVD ™ transmitter refers to the following transmitter models: • Models ...
Page 5 - Figure 1
Installation Manual 3 Before You Begin Figure 1 ELITE sensor with Model 2400S transmitter or enhanced core processor Flow direction arrow Purge fittings or rupture disks (optional) Calibration tag Thru holes for alternative mounting Process connection CMF010 Approval tag Model 2400S transmitter or e...
Page 6 - Figure 2
4 Micro Motion ® ELITE ® Sensors Before You Begin Figure 2 ELITE sensor with core processor Flow direction arrow Purge fittings or rupture disks (optional) Calibration tag Core processor housing Thru holes for alternative mounting Process connection CMF010 CMF025, CMF050, or CMF100 Purge connections...
Page 7 - Figure 3
Installation Manual 5 Before You Begin Figure 3 ELITE sensor with junction box Flow direction arrow Purge connections or rupture disks (optional) Calibration tag Junction box Approval tag Thru holes for alternative mounting Process connection Junction box Approval tag CMF025, CMF050, or CMF100 Flow ...
Page 8 - Figure 4; Determining a Location; Full flow tubes
6 Micro Motion ® ELITE ® Sensors Determining a Location Figure 4 High-temperature or extreme high-temperature ELITE sensor Step 1 Determining a Location Choose a location for the sensor based on the requirements described in this section. The following general guidelines can help you select an appro...
Page 9 - Figure 5; except
Installation Manual 7 Determining a Location Figure 5 ELITE sensor ambient and process temperature limits (all models except high-temperature and extreme high-temperature models) Ambien t temper ature of core processor or tr ansmitte r in °F (°C) Maximum process temperature in °F (°C) Mount transmit...
Page 10 - Figure 6; . IECEx and NEPSI approvals also use the ATEX “T” rating. UL and CSA; Table 1
8 Micro Motion ® ELITE ® Sensors Determining a Location Figure 6 High-temperature and extreme high-temperature ELITE sensor ambient and process temperature limits Hazardous area approvals may impose additional limits on ambient and process temperature. For the ATEX “T” rating, refer to the ATEX docu...
Page 11 - Maximum wiring distances; integrally mounted on the sensor, so Table 2 does not apply.; Pipe run
Installation Manual 9 Determining a Location Maximum wiring distances If the transmitter is mounted remotely from the sensor, the maximum distance between the sensor and transmitter depends on cable type. See Table 2. For high-temperature and extreme high-temperature ELITE sensors, note the followin...
Page 12 - Orienting the Sensor; Flow direction arrow
10 Micro Motion ® ELITE ® Sensors Orienting the Sensor Step 2 Orienting the Sensor The sensor will function properly in any orientation if the sensor tubes remain filled with process fluid. Micro Motion recommends orienting ELITE sensors as shown in Table 3. Flow direction arrow The sensor has a flo...
Page 13 - Mounting the Sensor; Figure 7
Installation Manual 11 Mounting the Sensor Step 3 Mounting the Sensor Use your common practices to minimize torque and bending load on process connections. Figure 7 illustrates how to mount the sensor. To reduce the risk of condensation or excessive moisture, the conduit opening should not point upw...
Page 14 - Installing extended mount electronics; it aside. You will use it later to secure the extender to the sensor.; Figure 9
12 Micro Motion ® ELITE ® Sensors Mounting the Sensor Installing extended mount electronics If you ordered a sensor with an extended mount for the electronics, you will need to install the extender. To install the extended mount electronics: 1. There is a metal clamping ring around the base of the f...
Page 15 - Mounting the temperature extender onto the sensor; flexible conduit. This conduit is required for agency approval.
Installation Manual 13 Mounting the Sensor Figure 10 Mounting the temperature extender onto the sensor Mounting the electronics of high-temperature and extreme high-temperature sensors High-temperature and extreme high-temperature ELITE sensors come with a 32 ″ (812 mm) pre-installed flexible condui...
Page 16 - Installing wafer-style sensors; • 4 flange bolts
14 Micro Motion ® ELITE ® Sensors Mounting the Sensor Figure 11 Electronics mounting for high-temperature and extreme high-temperature ELITE sensors Installing wafer-style sensors A wafer-style sensor, which has no flanges or fittings, lets you “clamp” the sensor between process connections in the p...
Page 17 - Tightening the sensor alignment rings
Installation Manual 15 Mounting the Sensor Figure 12 Wafer-style assembly Figure 13 Tightening the sensor alignment rings Process connection Sensor wafer Flange bolt Alignment ring Gasket (user-supplied) Flange nut Rotate the ring... ...to push the bolts outward Sensor wafer Sensor alignment ring Fl...
Page 18 - Wiring; Hazardous area installations; The sensor has one of the following configurations:; Core processor to a 4-wire remote transmitter; Remove the cover from the core processor housing.
16 Micro Motion ® ELITE ® Sensors Wiring Step 4 Wiring Hazardous area installations If you are installing the sensor in a hazardous location, verify that the hazardous classification information printed on the sensor tag matches the environment in which the sensor will be installed. Improperly seale...
Page 19 - Micro Motion cable gland and heat shrink; and drain wires exposed, and separate the wires.; Wrapping the shield drain wires; cover the drain wires.
Installation Manual 17 Wiring Figure 14 Micro Motion cable gland and heat shrink 4. For connection at the core processor housing, prepare shielded cable as follows (for armored cable, omit steps d, e, f, and g): a. Strip 4 1/2 inches (114 mm) of cable jacket. b. Remove the clear wrap that is inside ...
Page 20 - Applying the heat shrink
18 Micro Motion ® ELITE ® Sensors Wiring Figure 16 Applying the heat shrink g. Position gland clamping insert so the interior end is flush with the heat shrink. h. Fold the cloth shield or braid and drain wires over the clamping insert and approximately 1/8 inch (3 mm) past the O-ring. Figure 17 Fol...
Page 21 - Connecting the wires at the core processor; Reinstall and tighten the core processor housing cover.
Installation Manual 19 Wiring Figure 19 Connecting the wires at the core processor 7. Reinstall and tighten the core processor housing cover. 8. Additional wiring instructions for the transmitter can be found in the transmitter manual. Note: Never ground the 4-wire cable shield and shield drain wire...
Page 22 - Grounding; • Connect ground leads directly to earth, or follow plant standards.; Pressure Ratings; Table 4 shows the pressure ratings for ELITE sensors.; Table 4
20 Micro Motion ® ELITE ® Sensors Grounding Step 5 Grounding The sensor can be grounded via the piping if the joints in the pipeline are ground-bonded. If the sensor is not grounded via the piping, connect a ground wire to the internal or external grounding screw, which is located on the transmitter...
Page 23 - Purge Fittings; Removing a purge plug; Leave the outlet connection open.
Installation Manual 21 Purge Fittings Purge Fittings If the sensor has purge fittings, they should remain sealed at all times. After a purge plug has been removed, the sensor case should be purged with argon or nitrogen and resealed. Purging the case protects internal components. The sensor is purge...
Page 24 - Rupture Disks; Case purging procedure; Table 5
22 Micro Motion ® ELITE ® Sensors Rupture Disks Rupture Disks If the sensor has rupture disks, they are installed in the sensor purge fitting openings. The rupture disks should remain installed at all times. If you remove a rupture disk from the sensor case, it will be necessary to re-purge the case...
Page 25 - Return Policy; New and unused equipment
Installation Manual 23 Return Policy Return Policy Micro Motion procedures must be followed when returning equipment. These procedures ensure legal compliance with government transportation agencies and help provide a safe working environment for Micro Motion employees. Failure to follow Micro Motio...
Page 26 - Micro Motion Inc. USA; Micro Motion Europe
Micro Motion Inc. USA Worldwide Headquarters 7070 Winchester CircleBoulder, Colorado 80301T +1 303-527-5200+1 800-522-6277 F +1 303-530-8459 Micro Motion Europe Emerson Process ManagementNeonstraat 16718 WX EdeThe NetherlandsT +31 (0) 318 495 555 F +31 (0) 318 495 556 Micro Motion Japan Emerson Proc...