  • Page 2 – INSTALLATION; Cutting the cable may void the warranty.
  • Page 6 – FIGURE11. Wiring sensors through a remote junction box; WIRING THROUGH A REMOTE JUNCTION BOX
  • Page 7 – INSERTION - RETRACTION ASSEMBLIES; the sensor an additional half turn once the gasket is seated.; with the following exception:; Insert the sensor and tube assembly into the retraction chamber.; INSTALLATION REQUIREMENTS
  • Page 8 – FIGURE 14. Manual Retraction Assembly; RETRACTION; Place the sensor; NOTE; WARNING; Retraction chamber contents may be under pressure.
  • Page 9 – Tighten the union nut.; Do not loosen cap screws or collar when pressurized.
  • Page 10 – REPLACING SEALS IN MANUAL AND MECHANICAL RETRACTION ASSEMBLIES.
  • Page 11 – FIGURE 16. Mechanical Retraction Assembly Replacement Parts
  • Page 12 – FIGURE 17. Manual Retraction Assembly Replacement Parts
  • Page 13 – CALIBRATION; If wall effects are present; , calibrating a toroidal sensor with a standard
  • Page 14 – CAUTION; TROUBLESHOOTING; PROBLEM
  • Page 16 – ON-LINE ORDERING NOW AVAILABLE ON OUR WEB SITE
Loading the manual

Model 228

Submersion/Insertion Toroidal Sensor

Instruction Sheet

PN 51A-228/rev.P
December 2010

SPECIFICATIONS - SENSORS

MODEL

Wetted Materials

Temperature

Pressure CRN pressure*

228-02 PEEK (glass-filled)

248°F (120°C)

295 psig (2135 kPa abs)

220 psig (1618 kPa abs)

228-03 PEEK (glass-filled)

392°F (200°C)

295 psig (2135 kPa abs)

220 psig (1618 kPa abs)

228-04 Tefzel (glass-filled)

248°F (120°C) 200 psig (1480 kPa abs)

150 psig (1135 kPa abs)

228-05 Tefzel (unfilled)

248°F (120°C) 200 psig (1480 kPa abs)

150 psig (1135 kPa abs)

SPECIFICATIONS - ADAPTERS

SPECIFICATIONS - RETRACTION ASSEMBLIES

Operating Conditions

Retraction/Insertion Conditions

PN

Wetted Materials

Temperature

Pressure

Temperature

Pressure

23311-00

392°F

295 psig

392°F

295 psig

(mechanical)

(200°C)

(2135 kPa abs)

(200°C)

(2135 kPa abs)

23311-01

316 SS, Teflon, EPDM

392°F

295 psig

266°F

35 psig

(manual)

(200°C)

(2135 kPa abs)

(130°C)

(343 kPa abs)

PN

Wetted Materials Temperature

Pressure CRN pressure*

23242-02

PEEK, Viton, 316 SS 392°F (200°C)

295 psig (2135 kPa abs) 220 psig (1618 kPa abs)

23242-03 PEEK, Viton, 316 SS 392°F (200°C)

295 psig (2135 kPa abs) 220 psig (1618 kPa abs)

2001990 CVPC, PEEK, Viton 100°F (38°C)

100 psig (791 kPa abs) NA

185°F (85°C)

45 psig (412 kPa abs)

For additional information, please visit our website

at

www.emersonprocess.com/raihome/liquid/

.

*

Maximum pressure for applications requiring CRN registration

*

Maximum pressure for applications requiring CRN registration

316 SS, Teflon, EPDM

316 SS, Teflon, EPDM

NA

CVPC, PEEK, Viton

2001990

WARNING

Before removing the sensor, be absolutely certain
that the process pressure is reduced to 0 psig and
the process temperature is lowered to a safe level!

CAUTION

The wetted sensor materials may not be compatible
with process com position and operating conditions.
Application compat ibility is entirely the responsi-
bility of the user.

CAUTION

SENSOR/PROCESS

APPLICATION COMPATIBILITY

"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.

Summary

Page 2 - INSTALLATION; Cutting the cable may void the warranty.

MODEL 228 INSTALLATION INSTALLATION Keep at least 1 in. (2.5 cm) between the sensor and pipe walls. If the clearance is too small, calibrate the sensorin place. Ensure that the sensor is completely submerged in liquid. Mounting the sensor in a vertical pipe run withthe flow from bottom to top is bes...

Page 6 - FIGURE11. Wiring sensors through a remote junction box; WIRING THROUGH A REMOTE JUNCTION BOX

MODEL 228 WIRING FIGURE11. Wiring sensors through a remote junction box Wire sensors point to point. For wiring at the analyzer end, refer to the appropriate analyzer wiring diagram. For interconnecting cable 23294-00, use the 2265-54 wiring diagram. For interconnecting cable 23294-04 and 23294-05, ...

Page 7 - INSERTION - RETRACTION ASSEMBLIES; the sensor an additional half turn once the gasket is seated.; with the following exception:; Insert the sensor and tube assembly into the retraction chamber.; INSTALLATION REQUIREMENTS

7 MODEL 228 USE OF INSERTION - RETRACTION ASSEMBLIES INSERTION - RETRACTION ASSEMBLIES MANUAL RETRACTION ASSEMBLY (PN 23311-01) INSTALLATION 1. Loosen the collet nut and retract the sensor tube into the retraction chamber. See Figure 14. 2. Loosen the union nut and separate the retraction chamber fr...

Other Emerson Models