Welch Allyn 3 Propaq Encore Reference Guide Contents General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Safety Summary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ...
Propaq Encore Reference Guide 9 Welch Allyn General Information Safety Summary This Safety Summary should be read by all Propaq Encore users. Specific warnings and cautions will be found throughout the Propaq Encore documentation where they apply. Caution Caution: Federal (U.S.A.) law restricts this...
Safety Summary Welch Allyn 10 Propaq Encore Reference Guide Warning Make frequent electrical and visual checks on cables and electrode wires. Avoid electrosurgery burns at monitoring sites by ensuring proper connection of the electrosurgery return circuit so that the return paths cannot be made thro...
Safety Summary Propaq Encore Reference Guide 11 Welch Allyn The Propaq Encore should be serviced only by a Welch Allyn service technician while under warranty. The Propaq Encore Service Manual (P/N 810-0696-XX) is available from Welch Allyn to assist the biomedical engineer during post-warranty peri...
Propaq Encore Documentation Propaq Encore Reference Guide 13 Welch Allyn Propaq Encore Documentation The Documentation Set The Propaq Encore documentation set consists of documents for the clinician, the biomedical technician, and the department head or purchaser of accessories for the Propaq Encore...
Propaq Encore Reference Guide 15 Welch Allyn Getting Started Introducing the Propaq Encore Intended Use Before using the Propaq Encore on a patient, be sure you understand the Safety Summary at the front of this book. It provides important information about safely using the Propaq Encore. The Propaq...
Introducing the Propaq Encore Welch Allyn 16 Propaq Encore Reference Guide Expansion Module The Propaq Encore Expansion Module attaches to the monitor and houses additional capabilities. The Expansion Module can be fitted with the SpO 2 , CO 2 , and Printer options. Propaq Encore Pulse Oximetry Opti...
Introducing the Propaq Encore Propaq Encore Reference Guide 17 Welch Allyn Impedance Pneumography (RESP) Option The RESP option detects the rate or absence of respiratory effort, and is configured with the Pulse Oximetry option. Printer Option The Expansion Module with Printer (EMP) provides a light...
Using the Propaq Encore Welch Allyn 18 Propaq Encore Reference Guide Using the Propaq Encore System Controls (Right Side Panel) On/Off Switch This switch turns the monitor on and off. The switch is recessed to prevent accidentally turning off the monitor, which would result in losing patient data. I...
Using the Propaq Encore Propaq Encore Reference Guide 19 Welch Allyn Defib Sync Connector This connector allows connection with a LIFEPAK 5 or LIFEPAK 6s defibrillator for synchronized cardioversion. See page 131 for more information. Real-Time ECG Output Connector This connector provides a real-tim...
Using the Propaq Encore Welch Allyn 20 Propaq Encore Reference Guide Power-up Screen 1. If the Propaq Encore has been used for a previous patient, switch the monitor off, then on again. The monitor will turn on in the powerup patient mode with the associated settings. 2. Verify the monitor is in the...
Using the Propaq Encore Propaq Encore Reference Guide 21 Welch Allyn Powerup Equipment Alert: Program Fault, Settings Lost If a PROGRAM FAULT: SETTINGS LOST, TIME/DAY RESET equipment alert appears when you turn on the monitor, the monitor cannot recall the programmed custom settings and current time...
Using the Propaq Encore Welch Allyn 22 Propaq Encore Reference Guide Patient Connections The left side panels differ depending on the Propaq Encore model. All models have ECG, NIBP, and two temperature connectors. The Propaq Encore 204 left side panel includes one invasive-pressure connector, and th...
Using the Propaq Encore Propaq Encore Reference Guide 23 Welch Allyn Option Connectors Mainstream CO 2 Connector Sidestream CO 2 Connector NELLCORSENSORS / CAPTEURS ! SpO 2 ! CO 2 ! CO 2 Masimo SpO 2 Connector (motion tolerant) Cable: see Welch Allyn Products and Accessories Guide (810-0409-XX) Nell...
Using the Propaq Encore Welch Allyn 24 Propaq Encore Reference Guide Propaq Encore Display The display shows waveforms, vital sign numeric values, Propaq Encore status, and alarm information in different windows. Different vital sign numeric values (such as heart rate and blood pressures) have upper...
Using the Propaq Encore Propaq Encore Reference Guide 25 Welch Allyn You can select up to three waveforms to be shown on the Propaq. When only one waveform is selected, a trend window automatically appears below the waveform. While changing Propaq Encore settings, a status window may appear below th...
Using the Propaq Encore Welch Allyn 26 Propaq Encore Reference Guide The five buttons below the screen, and their associated labels located on the screen, provide access to the menus. Later in this manual, the notation A , B , C is used as a shorter way to say “Press Button A , then B , then C .” Pr...
Using the Propaq Encore Propaq Encore Reference Guide 27 Welch Allyn Key-press Route to Setup Menu 1 STAT SET SUSPEND ALL ALRM LIMITS PREV MENU SETUP NIBP ECG/RESP INV PRS SpO2/CO2 ALARMS MENU RESUME UP DOWN ON/OFF NEXT NEXT PAGE ALARM LIMITS MENU* TRENDS MENU PRINT NXT TRND OXYCRG SETUP MENU 1 MORE...
Using the Propaq Encore Welch Allyn 28 Propaq Encore Reference Guide Key-press Route to Setup Menu 2 ON/OFF NEXT INSERV PREV MENU MORE NIBP TEST IBP TEST SpO2 TST CO2 TEST NEXT CHANGE PR TREND MORE PREV MENU SETUP NIBP ECG/RESP INV PRS SpO2/CO2 MORE MORE NEXT CHANGE WAVE SEL PRINTER WAVE SELECT MENU...
Monitor Setup Propaq Encore Reference Guide 29 Welch Allyn Monitor Setup Setup Menu 1 is accessed by pressing the SETUP button on the Main Menu. STATSCALE Automatically readjusts all waveform scales. ALARMS Allows access to the Alarms menu. WAVE SEL Allows you to turn on and off desired waveforms or...
Monitor Setup Welch Allyn 30 Propaq Encore Reference Guide HR/PR TONE Sets the Heart Tone volume to HIGH, MEDIUM, LOW, or OFF. PATIENT MODE Pressing CHANGE in this selection displays the following Patient Mode window: If you press NEONATAL , PED , or ADULT , a confirmation window appears, requiring ...
Monitor Setup Propaq Encore Reference Guide 31 Welch Allyn Selecting Waveforms for Display To select waveforms for display, press SETUP , WAVE SEL . Use the NEXT and ON/OFF buttons to turn on the desired waveforms in the wave select window: Display Priorities You can turn on more than three waveform...
Monitor Setup Welch Allyn 32 Propaq Encore Reference Guide Time/Day Settings and Trends When you change the hour/minute/second setting for the monitor in the Time/Day window, the monitor deletes any patient trend data that is older than five hours for non-NIBP trends or older than eight hours for NI...
Monitor Setup Propaq Encore Reference Guide 33 Welch Allyn Setting the Current, Custom, and Powerup Modes The Propaq Encore has two sets of patient mode settings: • Factory patient modes. The powerup settings and alarm limits for these patient modes are preset and cannot be changed. They are listed ...
Monitor Setup Welch Allyn 34 Propaq Encore Reference Guide From the Main Menu, press SETUP , MORE, CHANGE, SETUP . The Mode Setup window appears: The asterisk (*) indicates which patient mode is currently selected for powerup. Whenever you change the patient mode, the alarm limit settings automatica...
Printer Functions Propaq Encore Reference Guide 35 Welch Allyn Printer Functions Press SETUP , MORE , PRINTER to display the printer menu and setup window. NEXT Selects the next setting in the status window. CHANGE Changes the currently selected display setting. PR TREND Prints all trends turned on ...
Printer Functions Welch Allyn 36 Propaq Encore Reference Guide PRINTER FAULT Messages These PRINTER FAULT messages can appear in an equipment alert window. The front panel of the printer lets you control the basic printer functions. If you press FREEZE prior to pressing SNAPSHOT , the printer prints...
Learning the Propaq Encore Propaq Encore Reference Guide 37 Welch Allyn Learning the Propaq Encore Using In-Service Mode You can practice using the Propaq Encore without a patient simulator by using the Propaq's in-service mode of operation. The in-service mode cannot be activated while you are moni...
Learning the Propaq Encore Welch Allyn 38 Propaq Encore Reference Guide NIBP For noninvasive pressure measurements, keep the Propaq Encore in manual NIBP operating mode and take pressure measurements by pressing the START button. You can also press the NIBP Menu's TURBOCUF button to consecutively ta...
Propaq Encore Reference Guide 39 Welch Allyn Patient Monitoring ECG/RESP Intended Use–Impedance Pneumography (RESP) The Respiration channel is intended to detect the rate or absence of respiratory effort, deriving the signal by measuring the AC impedance between the selected terminals of the ECG ele...
ECG/RESP Welch Allyn 40 Propaq Encore Reference Guide Intended Use–ECG The Propaq Encore is intended for ECG monitoring of either a five-lead or three-lead configuration, including the Marriott configuration 1 (MCL1 requires all three electrodes). The five lead configuration can derive one of seven ...
ECG/RESP Propaq Encore Reference Guide 41 Welch Allyn Even though the Propaq Encore contains fully isolated patient-connected circuitry, it has not been specially designed for direct cardiac application. The Propaq Encore can be used on patients with pacemakers. See page 47. ECG Connector and Applic...
ECG/RESP Welch Allyn 42 Propaq Encore Reference Guide Preparation Preparing for ECG monitoring with the Propaq Encore requires you to prepare the monitor, prepare the patient, set up the ECG channel, and then set the ECG alarms. Preparing the Monitor 1. Inspect the ECG cable for wear, breakage, or f...
ECG/RESP Propaq Encore Reference Guide 43 Welch Allyn 4. Apply the electrodes to the patient. 5. Support the ECG cable so it does not stress the electrode wires, ECG cable connectors, or electrodes. Some electrodes may be subject to large offset potentials due to polarization. This effect is most li...
ECG/RESP Welch Allyn 44 Propaq Encore Reference Guide 6. If an electrosurgical unit is going to be used, place the ECG cable and electrode wires as far as possible from the surgical site and from the electrosurgical return electrode and its cables. This will minimize interference. By now there shoul...
ECG/RESP Propaq Encore Reference Guide 45 Welch Allyn How ECG/RESP is Displayed Because of the critical nature of monitoring ECG, it is always displayed in the top part of the waveform display area. The ECG waveform is the only waveform that cannot be turned off using the Setup Wave Select Menu. Res...
ECG/RESP Welch Allyn 46 Propaq Encore Reference Guide When you press MORE in the first ECG/RESP Menu, a status window appears showing you the current ECG/RESP settings and additional selections. The Size function “increases” or “decreases” the ECG or RESP waveform size. Each time you press a SIZE bu...
ECG/RESP Propaq Encore Reference Guide 47 Welch Allyn Using the Propaq Encore With Pacemaker Patients Pacemaker Signals If the patient being monitored has a pacemaker, the Propaq Encore detects and can indicate the occurrence of pacemaker signals. With the Propaq, pacemaker signals are not counted a...
ECG/RESP Welch Allyn 48 Propaq Encore Reference Guide Noise on the signal Noise on the ECG signal may be detected as pacer signals, causing the pacer indicator to appear on the display. If you don't need to indicate pacemaker signals, you may want to turn off the pacemaker indicator for a better dis...
ECG/RESP Propaq Encore Reference Guide 49 Welch Allyn RESP Messages LEAD FAIL. One or more electrodes are making very poor or no contact. Check for proper connection; replace electrodes if needed. INAPPROPRIATE ECG CABLE. ECG cable appears not to contain 1K current limiting resistors. These resistor...
Invasive Pressure Welch Allyn 50 Propaq Encore Reference Guide Invasive Pressure This section applies only to Propaq Encore Models 204EL and 206EL. If you don't have one of these models, you can skip this section. Intended Use The Propaq Encore invasive pressure channel is intended for measuring art...
Invasive Pressure Propaq Encore Reference Guide 51 Welch Allyn Preparation Preparing for invasive pressure monitoring with the Propaq Encore requires you to prepare the transducer, zero the transducer, set up the pressure channel, and set the invasive pressure alarm limits. Preparing the Transducer ...
Invasive Pressure Welch Allyn 52 Propaq Encore Reference Guide 3. Wait for a tone to briefly sound and the word ZEROED to appear in the blood pressure numerics window. You will then see the pressure scale to the left of the waveform, and the pressure numerics appear. 4. Close the transducer's stopco...
Invasive Pressure Propaq Encore Reference Guide 53 Welch Allyn To remove the ZERO REJECTED message and to restore the invasive pressure numerics during an invasive pressure alarm, you must return to the invasive pressure menu and press CANCEL . This will restore the invasive pressure numerics. How I...
Invasive Pressure Welch Allyn 54 Propaq Encore Reference Guide Invasive Pressure Messages The following messages can appear in the numerics window. NOT ZEROED (or NO ZERO). No zero reference has been established. The monitor displays the pressure waveform, but to protect against erroneous readings, ...
Invasive Pressure Propaq Encore Reference Guide 55 Welch Allyn ZERO REJECTED (or REJECT). Unable to establish a zero reference value. The message remains until the CANCEL button is pressed. CANCELED. This message appears if CANCEL is pressed while the channel is zeroing. These equipment messages can...
NIBP Welch Allyn 56 Propaq Encore Reference Guide NIBP Intended Use The Propaq Encore noninvasive blood pressure channel (NIBP) indirectly measures arterial pressures using an inflatable cuff. If ECG is also monitored, the Propaq Encore synchronizes the NIBP measurement process to the occurrences of...
NIBP Propaq Encore Reference Guide 57 Welch Allyn NIBP measurements are affected by normal physiological pressure variations from reading to reading. Normal respiration may affect pressure by as much as 10 to 20 mmHg. Patient’s emotional state, body position, and cuff fit may also adversely affect N...
NIBP Welch Allyn 58 Propaq Encore Reference Guide If artifact is so severe while Smartcuf is enabled that it affects the accuracy of an NIBP measurement, that measurement is marked with a special symbol on the display and on printouts There may be some situations where it is desirable to disable Sma...
NIBP Propaq Encore Reference Guide 59 Welch Allyn For information about patient mode specifications, see “NIBP Specifications” on page 142. Preparation Setting up for noninvasive blood pressure monitoring requires three steps: place the cuff on the patient and connect the cuff to the monitor, set up...
NIBP Welch Allyn 60 Propaq Encore Reference Guide 3. Screw the hose connector onto the NIBP connector on the monitor's left side. 4. If motion artifact such as shivering, coughing, or vehicle motion interferes with NIBP readings, do the following: Position the patient’s limb away from the body so th...
NIBP Welch Allyn 62 Propaq Encore Reference Guide Set the NIBP Alarm Limits Set the alarm limits according to your hospital's standards. NIBP Display Default Settings To select which vital sign waveforms are displayed, press MAIN MENU, SETUP, WAVE SEL to display the Wave Select window. To enable the...
NIBP Propaq Encore Reference Guide 63 Welch Allyn AIR LEAK, CHECK HOSE (ERR# 1). The Propaq Encore could not properly inflate cuff. Check the hose and cuff for obvious leaks, such as O-rings in the hose connections. CUFF NOT DETECTED (ERR# 2). During cuff inflation the detected pressure did not suff...
NIBP Welch Allyn 64 Propaq Encore Reference Guide LOW BATTERY, NIBP DISABLED (ERR# 11). The battery lacks sufficient voltage to be able to operate the NIBP channel. Connect the Encore to the ac power adapter. SERVICE REQUIRED, NIBP DISABLED (ERR# 12). Have the monitor serviced. CUFF TOO LARGE FOR PA...
NIBP Propaq Encore Reference Guide 65 Welch Allyn NIBP IN PROGRESS Message The monitor displays this message when noise or artifact such as vehicle motion causes a delay while measuring NIBP. To remove the message, press any button below the screen. To cancel the NIBP measurement, press STOP . SpO 2...
Temperature Welch Allyn 66 Propaq Encore Reference Guide Temperature Intended Use Propaq Encore monitors provide two temperature channels (except for the HP-side panel option). When both channels are active, the difference temperature ( ∆ T) is also displayed. You can select °C or °F. Temperature Co...
Temperature Propaq Encore Reference Guide 67 Welch Allyn Temperature Messages The following messages can appear in an equipment alert window. A temperature caution message will also appear in the temperature numeric window when one of these messages appears (except PROBE NOT DETECTED ). PROBE NOT DE...
Pulse Oximetry (SpO2) Welch Allyn 68 Propaq Encore Reference Guide Pulse Oximetry (SpO 2 ) Each SpO 2 sensor is designed for application to a specific site on the patient within a certain size range. To obtain optimal performance, use an appropriate sensor and apply it as described in the sensor ’s ...
Pulse Oximetry (SpO2) Propaq Encore Reference Guide 69 Welch Allyn If NIBP will be monitored while using SpO 2 , place the NIBP cuff on a different limb than the SpO 2 sensor to help reduce unnecessary SpO 2 alarms. For optimal measurements, avoid placing the SpO 2 sensor on the same limb as an arte...
Pulse Oximetry (SpO2) Welch Allyn 70 Propaq Encore Reference Guide 4. Press SIZE to adjust the waveform size for best viewing (1x, 2x, 4x, or 8x). 5. Adjust the placement of the sensor until a good SpO 2 waveform is displayed. A waveform with artifact may cause erroneous oxygen saturation readings. ...
Pulse Oximetry (SpO2) Propaq Encore Reference Guide 73 Welch Allyn Perform SpO 2 “Spot-Check” Monitoring The SpO 2 Standby Mode allows you to remove the SpO 2 sensor from a patient without having to disable all alarms or disconnect the SpO 2 sensor cable from the Propaq monitor. You can therefore pe...
Pulse Oximetry (SpO2) Welch Allyn 74 Propaq Encore Reference Guide STANDBY is displayed in the numeric window when the SpO 2 sensor is disconnected from the patient, an alarm occurs, and you press the STANDBY button. STANDBY is also displayed if you first plug the SpO 2 sensor cable into the monitor...
Capnography (CO2) Welch Allyn 76 Propaq Encore Reference Guide Mainstream CO 2 Option The Mainstream CO 2 option measures the carbon dioxide content of a patient’s inhaled and exhaled breath. A mainstream sensor is attached to an airway adapter in series with a ventilator ’s patient breathing circui...
Capnography (CO2) Propaq Encore Reference Guide 77 Welch Allyn Displayed values of ETCO 2 and INCO 2 are the highest and lowest values (respectively) of CO 2 measured during the time interval set by the RESPONSE setting on the CO 2 Menu. You can set upper and lower alarm limits for ETCO 2 , and an u...
Capnography (CO2) Welch Allyn 78 Propaq Encore Reference Guide UNCAL indicates the monitor has detected a problem such as a lack of calibration, an obstruction, or a low battery. WARM UP indicates Mainstream CO 2 has been activated and is preparing for operation. This typically requires 30 seconds a...
Capnography (CO2) Propaq Encore Reference Guide 79 Welch Allyn Sidestream CO 2 Menu and Status Window When Sidestream CO 2 is active, press MORE from the first CO 2 menu to access the SCO 2 menu and the SCO 2 status window: SWEEP SPEED The CO 2 and RESP display sweep speed (set by the mm/s button) B...
Capnography (CO2) Propaq Encore Reference Guide 81 Welch Allyn 3. Plug in the CO 2 sensor cable to the Mainstream CO 2 connector on the Propaq left side panel. 4. Set up the CO 2 channel and alarms. Follow the steps described on page 81. How to Set Up the CO 2 Channel and Alarm Limits After you conn...
Capnography (CO2) Welch Allyn 82 Propaq Encore Reference Guide 3. Press mm/s to select either 3.13, 6.25 or 12.5 mm/sec. The default setting is 6.25 mm/ sec. This sweep speed also applies to RESP. 4. Press MORE to view the CO 2 status window. 5. If either O 2 or N 2 O is being administered to a pati...
Capnography (CO2) Propaq Encore Reference Guide 83 Welch Allyn Motion artifact or other factors can cause false RR/BR alarms. To help minimize false alarms, the Propaq delays or “holds off” triggering a RR/BR alarm for 5 seconds. After this holdoff period begins, if the Propaq detects that the patie...
Capnography (CO2) Welch Allyn 84 Propaq Encore Reference Guide 3. Connect to a non-intubated patient. OR 4. Connect to an intubated patient. Use only accessories provided or recommended in the Welch Allyn Products and Accessories booklet. Always refer to the manufacturer's Directions for Use. Connec...
Capnography (CO2) Propaq Encore Reference Guide 85 Welch Allyn Connect to a Non-Intubated Patient 1. Position the cannula on the patient according to the manufacturer ’s instructions. 2. Connect the sample line to the cannula and the watertrap. Make sure that the sample line is firmly connected. War...
Capnography (CO2) Welch Allyn 86 Propaq Encore Reference Guide Connect to an Intubated Patient 1. Connect the gas sampling elbow and elbow connector into the patient’s breathing circuit according to the manufacturer ’s instructions. 2. Connect the sample line to the elbow connector and the watertrap...
Capnography (CO2) Propaq Encore Reference Guide 89 Welch Allyn OCCLUSION - CHECK WATERTRAP/TUBING . Blockage has been detected on the Sidestream CO 2 input. Check the watertrap, sample line, and any connected tubing for occlusion. PUMP FAILURE, SERVICE REQUIRED . The pump is not able to maintain the...
Propaq Encore Reference Guide 91 Welch Allyn Alarms and Limits Alarms Window and Menu You can set the alarms in normal operating mode or in-service mode. You can also have the Propaq Encore automatically print the patient's vital sign information when an alarm occurs. Alarm Status Window This status...
Alarms Window and Menu Welch Allyn 92 Propaq Encore Reference Guide To adjust the alarm tone volume, press SETUP , then MORE . The status window appears. Use the NEXT button to select ALARM TONE, and the CHANGE button to change the setting. Warning At the highest volume alarm level, the sound pressu...
Types of Propaq Alarms Propaq Encore Reference Guide 93 Welch Allyn Types of Propaq Alarms Patient Alarms: Definitions and Indications The Propaq Encore can alert you to changing patient conditions through its programmable alarm functions. After you set alarm limits, any vital sign that violates any...
Types of Propaq Alarms Welch Allyn 94 Propaq Encore Reference Guide The alarm tones are summarized in the table: Life-Threatening Alarms Life-threatening alarms are automatically turned on when the channel is activated, and cannot be turned off by the user. A life-threatening alarm is the highest pr...
Types of Propaq Alarms Propaq Encore Reference Guide 95 Welch Allyn Responding to Patient Alarms Suspend an Alarm Tone You can temporarily turn off an alarm tone by pressing the SUSPEND/RESUME key to the left of the screen or the SUSPEND button in the Patient Alarm Menu. The tone is suspended for 90...
Types of Propaq Alarms Welch Allyn 96 Propaq Encore Reference Guide • Manually change the violated alarm limit value by selecting the violated limit value and adjusting it • Turn off all alarm limits by pressing MAIN MENU, SETUP , ALARMS , ALL ALRM Responding to Life-Threatening Alarms During a life...
Alarm Holdoffs Propaq Encore Reference Guide 97 Welch Allyn Alarm Holdoffs Propaq Audio and Visual Alarm Holdoffs Motion artifact or other factors can cause false vital sign alarms. To help minimize false alarms, the Propaq briefly delays or “holds off” triggering alarms for certain vital sign limit...
Setting Alarm Limits Welch Allyn 98 Propaq Encore Reference Guide Setting Alarm Limits Adjust Limits with STAT SET When you want to quickly set all alarm limits, the Propaq Encore can calculate new alarm limits using the patient’s current values. Press the STAT SET button in the Alarms Menu ( SETUP ...
Setting Alarm Limits Propaq Encore Reference Guide 99 Welch Allyn Turning On and Off All Limits If you want to turn all limits on or off without changing their values, press ALL ALRM in the Alarms Menu (a confirmation window appears requiring you to confirm your selection). The Alarms Status Window ...
Setting Alarm Limits Welch Allyn 100 Propaq Encore Reference Guide Customizing Propaq Encore Alarm Limits Although adjusting the Propaq's alarm limits can be automatic with STAT SET , you can also customize the limits you want to take effect every time you turn on the monitor. See “Setting the Curre...
Equipment Alerts Propaq Encore Reference Guide 101 Welch Allyn Equipment Alerts Definitions and Indications The Propaq Encore can alert you to changing equipment conditions, such as disconnected or faulty sensors, low battery voltage, lost programmed settings, and other conditions that can affect pa...
Propaq Encore Reference Guide 103 Welch Allyn Trends Every two minutes, the Propaq Encore collects the monitored vital sign numerics and stores them in its trend memory, which can save the last five hours of trend information. (See “NIBP Trends “on the next page for exceptions.) All this information...
How Trends are Accumulated Welch Allyn 104 Propaq Encore Reference Guide How Trends are Accumulated For each of the four continuous trends, numerics are sampled every two minutes to a maximum of 150 samples (up to five hours). When the maximum is reached, the trending continues but the older data is...
Displaying Trends Propaq Encore Reference Guide 105 Welch Allyn Selecting a Trend (NXT TRND) Select the trend you want displayed by pressing the NXT TRND button. The trend is identified by a label at the top of the table. P1 TREND P2 TREND NIBP TREND RESP TREND TEMP TREND Time Time Time Time Time HR...
Propaq Encore Reference Guide 107 Welch Allyn Printing Printing Patient Data Printing Waveforms If you have an expansion module set up with your monitor, you can print any waveform by pressing either the SNAPSHOT or START/STOP button on the expansion module when the waveform is displayed on the moni...
Printing Patient Data Propaq Encore Reference Guide 109 Welch Allyn Printing NIBP Measurements You can print the results of an NIBP measurement each time one occurs. This printout is called the NIBP Ticket. The NIBP Ticket must be turned on in the Printer Setup window. 1. To turn on NIBP TICKET, pre...
Printing Patient Data Welch Allyn 110 Propaq Encore Reference Guide Printing the Apnea Ticket The Apnea Ticket documents the length of each apnea episode. To set the Propaq to print an Apnea Ticket after an apnea event, follow these steps. 1. From the main menu, press the following buttons: SETUP , ...
Printing Patient Data Propaq Encore Reference Guide 111 Welch Allyn Printing When a Patient Alarm Occurs To set up the printer to print on a patient alarm, follow these steps. 1. From the Main Menu, press the following buttons: SETUP , MORE , PRINTER . 2. Press the NEXT button to select ALARM PRINT ...
Printing Patient Data Welch Allyn 112 Propaq Encore Reference Guide OxyCRG The OxyCRG is a graphical printout of two minutes of continuous HR/PR and SpO 2 numerics, and a condensed respiratory waveform. If any of the parameters have been completely inactive for the two minutes prior to the initiatio...
Printing Trends Propaq Encore Reference Guide 113 Welch Allyn Printing Trends Printed trends are useful for reviewing the patient's vital signs over the last several minutes to the last five hours. The Propaq enables you to print one trend or several trends with a press of a button, or automatically...
Printing Trends Welch Allyn 114 Propaq Encore Reference Guide Automatic Trend Prints To automatically print trends at 4-hour intervals, activate AUTO TREND and select the print times. Use the CHANGE button to set the trend print times according to the start time (clock hour) of each shift or to OFF....
Propaq Encore Reference Guide 115 Welch Allyn Acuity Central Monitoring System Intended Use The Acuity system is used as a central monitoring system for Propaq Encore monitors configured to interface with Acuity. The Acuity system may be used to monitor all patients. For neonatal patients, use all A...
Press NET OFF to Disconnect from Acuity Propaq Encore Reference Guide 117 Welch Allyn Key-press Route to Acuity Menu Press NET OFF to Disconnect from Acuity When you want to disconnect the monitor from Acuity, be sure to use the NET OFF button as described below. The NET OFF button is a safety featu...
Printing at Acuity Welch Allyn 118 Propaq Encore Reference Guide 2. Within 15 seconds, disconnect the Acuity network cable from either the Propaq Encore side panel or the bedside jack. If the patient will no longer be monitored with the Propaq Encore, turn off the monitor to erase trend information....
Propaq Encore Reference Guide 119 Welch Allyn Power Sources For in-hospital operation and recharging from ac mains, an ac power adapter plugs into the monitor. Use only a Welch Allyn ac power adapter to ensure protection against risk (leakage) current hazards. The Propaq Encore can also be powered a...
Power Adapter Intended Use Welch Allyn 120 Propaq Encore Reference Guide Verifying Proper Power Adapter Configuration Prior to using the power adapter, check it for proper voltage selection by looking in the small indicator window on the front end (by the power switch). If the number in the window d...
Power Adapter Intended Use Propaq Encore Reference Guide 121 Welch Allyn Procedure 1. Unplug the power adapter ’s removable cord from the ac mains outlet and the power adapter. 2. Turn the power adapter so you can see the window that indicates the voltage setting. 3. Using a small, flat-blade screwd...
Battery Care Propaq Encore Reference Guide 123 Welch Allyn Battery Care Recharging Time The battery charges to full capacity within eight hours (if the monitor remains off). Monitor Functions Resumed In most instances, most monitor functions are usable immediately after plugging in the ac power adap...
Battery Care Welch Allyn 124 Propaq Encore Reference Guide Monitor Functions Based on Battery Voltage As battery voltage drops during extended monitor battery operation, error messages are displayed and monitor functions are discontinued in order of priority. Checking Battery Voltage The Propaq’s ba...
Propaq Encore Reference Guide 125 Welch Allyn Care and Maintenance Inspect and Clean the Monitor and Accessories Before cleaning, thoroughly inspect the monitor and all accessories for any signs of damage, cracks, or improper mechanical function of keypads, switches, connectors, and printer paper do...
Maintenance Welch Allyn 126 Propaq Encore Reference Guide Maintenance Service Interval Recommendations At the intervals recommended below, verify the Propaq Encore for proper operation of all channels and internal circuitry. Such checks and verifications should only be carried out by a qualified bio...
Monitor Care Propaq Encore Reference Guide 127 Welch Allyn Product Recycling You can return a Propaq Encore to Welch Allyn for recycling when the monitor reaches the end of its life (7 to 10 years). You can also return an Encore’s internal battery to Welch Allyn for recycling when it reaches the end...
Printer Maintenance Welch Allyn 128 Propaq Encore Reference Guide Printer Paper Removal Printer Maintenance Loading Paper Paper is loaded through the bottom of the printer. 1. Lay the monitor on its back to gain access to the bottom of the printer. 2. Squeeze the locks on the paper door toward each ...
Customer Services Welch Allyn 130 Propaq Encore Reference Guide Customer Services Ordering and Customer Service For ordering information, for the location of your nearest Welch Allyn sales representative or service center, or for more information on other Welch Allyn products, contact: Welch Allyn85...
Propaq Encore Reference Guide 131 Welch Allyn Defibrillator Synchronization These instructions are not intended to replace existing hospital procedures relative to the provision of cardiac electrical therapy and operation of the specific models of defibrillators. Use all safety standards and clinica...
Removing the Defib Sync Interface from the LIFEPAK 5 Welch Allyn 132 Propaq Encore Reference Guide 2. Slide the Propaq Encore Defib Sync interface onto the left side of the LIFEPAK 5 until it snaps in place. 3. Connect the cable end to the Propaq’s DEFIB SYNCHRO connector on the right side panel. Re...
Synchronized Cardioversion Using the LIFEPAK 5 Propaq Encore Reference Guide 133 Welch Allyn Synchronized Cardioversion Using the LIFEPAK 5 1. Set up the LIFEPAK 5 Defibrillator and any other instrumentation according to institutional procedures and manufacturer ’s operating instructions. Use the in...
Synchronized Cardioversion Using the LIFEPAK 6s Welch Allyn 134 Propaq Encore Reference Guide 5. Follow hospital procedures and LIFEPAK 5 instructions for cardioversion. 6. If subsequent cardioversion must be performed, repeat steps 3 through 5. Synchronized Cardioversion Using the LIFEPAK 6s 1. Set...
Synchronized Cardioversion Using the LIFEPAK 6s Propaq Encore Reference Guide 135 Welch Allyn 2. Verify the integrity of the ECG patient electrodes and the fidelity of the ECG waveform for tall, distinct R-waves and minimal artifact. 3. Connect the Propaq Encore end of the Welch Allyn Defib Sync cab...
Defib Sync Message Welch Allyn 136 Propaq Encore Reference Guide 7. Follow hospital procedures and LIFEPAK 6s instructions for cardioversion. 8. If subsequent cardioversion must be performed, repeat steps 5 through 7. Defib Sync Message The Defib Sync message can appear in an equipment alert window....
Propaq Encore Reference Guide 137 Welch Allyn Specifications ECG The ECG channel meets all the requirements for Cardiac Monitors Heart Rate Meters and Alarms specified ANSI/AAMI EC13-1992, except for Standardizing Voltage (section 3.2.9.9). The channel also meets the American National Standard, Safe...
Impedance Pneumography (RESP) Welch Allyn 140 Propaq Encore Reference Guide Impedance Pneumography (RESP) Impedance Pneumography (RESP) Specifications Characteristic Specification Display Characteristics Sweep speed 3.13, 6.25, 12.5 mm/sec; user-selectable Amplitude range 1x, 2x, 4x, 8x, 16x Perform...
Invasive Pressure Propaq Encore Reference Guide 141 Welch Allyn Invasive Pressure Applies only to models 204 and 206. Invasive Pressure Specifications Characteristic Specification Transducer Type Strain-gauge resistive bridge, or HP quartz (with HP Option). a a.Transducers with 40 µV/V/mmHg sensitiv...
NIBP Welch Allyn 142 Propaq Encore Reference Guide NIBP NIBP Specifications Characteristic Specification Method Oscillometric Control Automatic and manual measurement control Auto Intervals 1, 2, 3, 5, 10, 15, 30, and 60 minutes Turbocuf Maximum measurements allowable in a 5-minute period Displayed ...
Temperature Propaq Encore Reference Guide 143 Welch Allyn Temperature Temperature Specifications Minimum Time between automatic measurements 30 seconds (Auto Mode)2 seconds (Turbo Mode) Electrosurgery InterferenceSuppression Included in all models: 202EL, 204EL, 206EL NIBP Performance Per EN 1060-1,...
Pulse Oximetry (SpO 2 ) Welch Allyn 144 Propaq Encore Reference Guide Pulse Oximetry (SpO 2 ) Masimo SpO 2 Specifications Characteristic Specification Saturation (% SpO 2 ) RangeResolutionAlarm Limits 1% to 100%1%52% to 100% (upper)50% to 98% (lower) Probe Accuracy (25° to 41° C) Adults, Pediatrics:...
Pulse Oximetry (SpO 2 ) Propaq Encore Reference Guide 145 Welch Allyn Nellcor SpO 2 Specifications Characteristic Specification a a.When performing SpO2 monitoring, the monitor SpO2 channel may not (in rare instances) initially detect a damaged SpO2 sensor or extension cable (as described in EN865:1...
Capnography (CO 2 ) Welch Allyn 146 Propaq Encore Reference Guide Capnography (CO 2 ) General CO 2 Specifications Characteristic Specification CO 2 Display Screen Display CO 2 waveform and ETCO 2 and INCO 2 (when in alarm) numerics Numeric Display Ranges ETCO 2 0-99 mmHg, 0-13.2 kPa, 0-23.1% INCO 2 ...
Alarms Welch Allyn 150 Propaq Encore Reference Guide Alarms Alarms Specifications Characteristic Specification Indicators Flashing red ALARM light indicates patient alarm; continuously on ALARM light indicates patient alarms are suspended. Continuously on ALARM(S) OFF light indicates one or more ala...
Trends Propaq Encore Reference Guide 151 Welch Allyn Trends Trends Specifications Display Display Specifications Characteristic Specification a a.Assumes SpO 2 and CO 2 functions are present. Model 202 Parameters NIBP, T1, T2, ∆ T, HR (heart rate/pulse rate), SpO 2, End-tidal CO 2, Inspired CO 2 , B...
Monitor (Environmental) Welch Allyn 152 Propaq Encore Reference Guide Monitor (Environmental) Monitor Environmental Specifications Characteristic Specification Operating Temperature 0° to 40° C Shipping and Storage Temperature -20° to 60° C Operating Altitude -2,000 to 15,000 ft (-610 to 4,572 m) Sh...
Monitor (Physical) Propaq Encore Reference Guide 153 Welch Allyn Monitor (Physical) Monitor Physical Specifications Characteristic Specification Protection Classifications, all Configurations a a.Per EN 60601-1 unless otherwise stated. Type of Protection against Electric Shock:—Power Adapter Power a...
Printer Welch Allyn 154 Propaq Encore Reference Guide Printer Printer Specifications Characteristic Specification Operation Operating Modes Continuous, Snapshot, Freeze Print, Auto Print, Auto Trend, Tabular Trend, Alarm Print, NIBP Ticket, Apnea Ticket, OxyCRG, OxyCRG on Alarm Auto Print Intervals ...
Power Propaq Encore Reference Guide 155 Welch Allyn Power Monitor Power Specifications Electromagnetic Compatibility (EMC) Per IEC 601-1-2, which is a collateral standard of IEC 601-1, for electromagnetic compatibility. Paper Short-term Storage Environment (up to 7 days) -20 to 40°C; 5% to 80% nonco...
Power Adapters Welch Allyn 156 Propaq Encore Reference Guide Power Adapters Power Adapter General Specifications Characteristic Specification Protection Classifications, all Adapters a a.Per EN 60601-1 unless otherwise stated. Type of Protection Against Electric Shock Class I, (Protectively Earthed)...
Power Adapters Propaq Encore Reference Guide 157 Welch Allyn Power Adapter Physical Specifications Characteristic Specification Universal Power Adapter, Part No. 503-0054-00, 503-0093-XX Length 5.0 in (12.7 cm) Width 3.6 in (9.1 cm) Height 3.1 in (7.9 cm) Weight 3.1 lb (1.4 kg) Rated Input 100V-120V...
Factory Default Settings Welch Allyn 158 Propaq Encore Reference Guide Factory Default Settings Factory Default Settings Characteristic Specification Date a MO/DA/YR, DA.MO.YR, or YR/MO/DA Decimal a . (Period) HR/PR Sweep 25 mm/sec RR/BR Sweep 6.25 mm/sec Alarm Tone MEDIUM HR/PR TONE LOW HR/PR SOURC...
In-Service Simulated Values Welch Allyn 160 Propaq Encore Reference Guide In-Service Simulated Values In-Service Values Channel Display Initial Value Alternate Value ECG Waveform Normal sinus rhythm, 1mV, Lead II Normal sinus rhythm ECG Heart Rate 80 beats per minute 125 beats per minute RESP Respir...
Propaq Encore Reference Guide 161 Welch Allyn Glossary ∆ T Difference temperature. The difference between T1 and T2. AAMI Association for the Advancement of Medical Instrumentation (United States of America). AC Power Adapter The device that plugs into the 12-28V dc receptacle on the Propaq Encore's...
Propaq Encore Reference Guide 163 Welch Allyn CO 2 A patient channel indicating the by-product of respiration, carbon dioxide, which is exhaled by the lungs. Configuration The patient channels included with each Propaq Encore model. A table in Chapter 1 lists the configuration of each Propaq Encore ...
Propaq Encore Reference Guide 167 Welch Allyn PA Pulmonary artery (label for an invasive blood pressure channel). Parameter See Vital Sign Parameter. Patient Alarm The condition that exists when a vital sign parameter numeric violates an alarm limit. Patient Channel ECG, P1, P2, T1, T2, SpO 2 , CO 2...
Welch Allyn 168 Propaq Encore Reference Guide Rescale Mode The method used in invasive pressure display to show each waveform against its own scale. The scale is automatically selected for best viewing of the entire waveform. Respiration The exchange of oxygen and carbon dioxide in the lungs and wit...
Welch Allyn 170 Propaq Encore Reference Guide Window An area on the display screen in which information is displayed. YSI An acronym for Yellow Springs Instrument Company. Zeroing The process by which an invasive pressure zero reference is obtained so that pressures can be related to atmospheric pre...
Welch Allyn 171 Propaq Encore Reference Guide Index A ac adapter connector 18Accessories Products & Accessories book 13 Acuity System 115 Acuity menu 117connecting 115Intended use 115Network Fault message 118printing at Acuity 118 Adult Mode 56Alarm Holdoffs 97 , 150 ALARM light 19Alarm Limits M...
Index Propaq Encore Reference Guide 172 Welch Allyn CO2 Menu 1 mm/sec button 78 , 82 RANGE button 78 , 81 CO2 Menu 2 FLOWRATE button 79GAS COMP button 78 , 79 , 82 RESPONSE button 78 , 79 , 82 SOURCE button 78 , 79 Connectors ac adapter 18defibrillator synchronization 131 , 135 ECG 41INV PRS 50NIBP ...
Index Propaq Encore Reference Guide 176 Welch Allyn PRINT button ( see also Trends Menu) 113 Printer Setup window 110RESP waveforms 108Setup Menu 2 109speed 107SpO2 waveforms 107sweep speeds with SNAPSHOT 108trends 113trends according to shift 114with an expansion module 107 Probes, temperature 66Pr...
Index Propaq Encore Reference Guide 178 Welch Allyn SpO2 70 , 71 Waveforms display priorities 31ECG 31scales 54 , 81 size 45 Windows Propaq screen 24 Z ZERO (P1/P2) buttons 54Zeroing messages 51transducers 51
Welch Allyn Manuals
-
Welch Allyn 202el
Manual
-
Welch Allyn 5200-103A
Manual
-
Welch Allyn 1500
Manual
-
Welch Allyn LXi
Manual
-
Welch Allyn 246
Manual
-
Welch Allyn Software version 8.2x
Manual
-
Welch Allyn CP 200
Manual
-
Welch Allyn 48780
Manual
-
Welch Allyn 692
Manual
-
Welch Allyn 48766
Manual
-
Welch Allyn 48728
Manual
-
Welch Allyn REF 420
Manual
-
Welch Allyn 6920
Manual
-
Welch Allyn Model 408
Manual
-
Welch Allyn V 1.3
Manual
-
Welch Allyn 408
Manual
-
Welch Allyn 8.1x
Manual
-
Welch Allyn 406
Manual
-
Welch Allyn 621NO
Manual
-
Welch Allyn 206el
Manual