Page 3 - Product Information; Degree of Protection Against Electric Shock: No applied part.
Product Information The design of the Dome E4c digital display takes into account every known measure to ensure your personal safety. Improper use of the display can result in electric shock, fire, or damage to the display. Read all instructions before setting up the display. Classification: Shock P...
Page 4 - Symbol explanations; Do not use household or municipal waste; BAROMETRIC PRESSURE.; Maximum operating altitude, up
iv | Dome E4c Display Symbol explanations Intended use The Dome E4c display is an AMLCD unit designed for viewing medical X-ray images. This unit should not be used near patients and should be kept outside of 1.83 m perimeter and 2.29 m vertical. This device must not be used in primary image diagnos...
Page 5 - Safety precautions; Safety requirements for medical electrical; Safety tips; • Never open the display case, even when the power is off. Do not drop or
Product Information | v Safety precautions External equipment intended for connection to signal input, signal output, or other connectors, must comply with the relevant IEC standard (EN/IEC 60601-1 series for medical electrical equipment). In addition, all such combinations (systems) must comply wit...
Page 6 - Unpacking and handling tips
vi | Dome E4c Display Unpacking and handling tips The Dome E4c display is a precision instrument that requires proper care to maintain adherence to specification. Unpack the display carefully, then set up and handle the unit properly to avoid damage to the LCD panel. • Use both hands to grasp the di...
Page 7 - Cleaning the display unit; WARNING
Product Information | vii Cleaning the display unit Periodically wipe the display housing with a soft, dry cloth to prevent dust and other material from clogging the venting holes. The display screen can be damaged by excessive scrubbing and by incompatible cleaning solvents. Dampen a clean, soft li...
Page 9 - About the Display; System requirements; • PCI Express x16 lane slot
About the Display The Dome E4c display is a 30-inch AMLCD color panel designed for widescreen diagnostic imaging. The combination of high resolution and high contrast ratio allow the viewing of color and grayscale images simultaneously. Bezel-free framing of 4 megapixels of data is presented in a la...
Page 10 - Display components; • Green. The display is connected and running properly.; LED on display front
2 | Dome E4c Display Display components The Dome E4c display has a light-emitting diode, or LED, on the front panel to indicate display status. • Green. The display is connected and running properly. • Amber. The video signal is not being received There is no power switch on the display unit. The un...
Page 11 - Dome CXtra; Pin assignment of DIN connector
About the Display | 3 The ports for connecting the power cord and the video cable are on the back. The power input port is an 8-pin DIN connector that drives power to the display. The power input is 24V ±5% (115 W typical).See detail below for pin assignment. The dual-link DVI connector drives the d...
Page 12 - Display Installation; Installing the controller and display
4 | Dome E4c Display Display Installation Before you install the display controller, remove any existing graphics card and its driver from your computer. After installation, the display turns on automatically when you turn on your computer system. Installing the controller and display 1 Remove the c...
Page 14 - Installing the display driver
6 | Dome E4c Display Installing the display driver Before you install the driver, remove any previously installed display driver for the display controller from your system. 1 Start the system. Click Cancel on the Found New Hardware Wizard. Click No when the system prompts you restart the computer. ...
Page 15 - Setting the display mode; Removing the driver
Display Installation | 7 Setting the display mode 1 Right-click the desktop and select Properties > Settings. 2 Select the monitor that you want to change. Then click Advanced. 3 Click the Monitor tab. Make sure that the check box for Hide modes that this monitor cannot display is empty (unchecke...
Page 16 - Installing Dome CXtra; Dome CXtra User’s Guide; Conserving energy use
8 | Dome E4c Display Installing Dome CXtra Bundled with the display system, the Dome CXtra software enhances the functionality of the Dome E4c display with a range of value-added services, such as DICOM calibration, error reporting, and backlight saver. Network management of Dome displays running th...
Page 17 - Troubleshooting; Problem; Display does not light up.
Troubleshooting | 9 Troubleshooting Problem Possible Cause Solution Display does not light up. Power is not on. Make sure that the computer is turned on. Activate the computer from sleep mode. Power cord is not securely connected. Tighten power cord connection and turn on computer. No image appears ...
Page 18 - Specification; Category
10 | Dome E4c Display Specification In locations where 240V outlets are used, use the E4c display only on a center-tapped, 240V, single-phase supply. Category Characteristic Item Specification Screen Screen size diagonal 756.228 mm (30 in.) Resolution 2560 x 1600 pixels (landscape) Pixel pitch 0.250...
Page 19 - Power supply
Specification | 11 Power supply CAUTION: Use only the power adapter supplied with the E4c display unit (JEC Korea Corporation, Type JMW1150KA2423F51). Reliability Environment Category Characteristic Item Specification Power input requirements Voltage selection Auto-ranging Voltage 100 – 240V AC Curr...
Page 20 - Regulatory Compliance
Regulatory Compliance Canada, European Union, United States This display has been tested and found to comply with IEC/EN 60601-1 and IEC/EN 60601-1-2 standards, and is certified to meet medical standard C22.2 No. 601.1-M1990 (C US Mark). The medical display, in addition to meeting medical requiremen...
Page 21 - system
The applicable safety standards for an MDD Class I display are IEC/EN 60601-1:1990 along with Amendments 1 and 2. To help the medical device designer evaluate the suitability of these displays, Planar has also conducted EMC testing to IEC 60601-1-2 as it can be applied. The display with its power su...
Page 22 - EU Declaration of Conformity for Medical Application
EU Declaration of Conformity for Medical Application A Declaration of Conformity has been filed for this product. For additional copies of the Declaration of Conformity document, contact Planar Systems. The Dome E4c digital flat-panel display meets the essential health and safety requirements, is in...
Page 24 - Standard Warranty; Summary; Standard Warranty Return Procedure
Standard Warranty Summary • Standard one-year “repair-and-return” warranty • Two-year backlight warranty * • Typical repair turnaround time of 10 business days Standard Warranty Return Procedure As a Planar Standard Warranty customer, you must follow the procedure below if you have a non-functioning...