Page 2 - Contents
a e f c d b Contents Thank you for purchasing the OMRON R3 Wrist Blood Pressure Monitor. The OMRON R3 is a compact and easy to use blood pressure monitor, operating on the oscillometric principle. It measures your blood pressure and pulse rate simply and quickly. For comfortable controlled inflation...
Page 3 - Important Safety Information
1 EN Important Safety Information Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia or arteriosclerosis.Please read this section carefully before using the unit. Warning: • Indicates a potentially hazardous situation which, if not avoided, could result in death or serio...
Page 5 - Overview; Main unit
3 EN 1. Overview Main unit DisplayBattery compartment coverWrist cuffMEM (Memory) buttonO/I START ( ) button SET ( ) buttonDate/Time setting ( ) button Display Systolic blood pressureDiastolic blood pressureMemory symbolAverage value symbolIrregular heartbeat symbolPulse displayMovement error symbol...
Page 6 - Preparation; Installing/Replacing the Batteries; Remove the battery cover.
4 2. Preparation 2.1 Installing/Replacing the Batteries 1. Remove the battery cover. 1) Push the hook on the bottom of the battery cover. 2) Pull the cover off the main unit. 2. Insert two 1.5V “AAA” alkaline (LR03) batteries as indicated in the battery compartment and then replace the battery cover...
Page 7 - Setting the Date and Time; Press the O/I START button to store the setting.
2. Preparation 5 EN 2.2 Setting the Date and Time 1. Set the monitor to the correct date and time before taking a measurement for the first time.Press the button. 2. Press the O/I START button to store the setting. Notes: • To reset the date and time, hold down the button while the power is off. • I...
Page 8 - Using the Unit; Applying the Wrist Cuff; Place the wrist cuff over your wrist.
6 3. Using the Unit 3.1 Applying the Wrist Cuff Do not apply over clothing. 1. Place the wrist cuff over your wrist. 2. Wrap the wrist cuff around your wrist. Make sure that the wrist cuff does not cover the protruding part of the wrist bone (ulna) on the outside of the wrist. --- Your thumb should ...
Page 9 - How to Sit Correctly
3. Using the Unit 7 EN 3.2 How to Sit Correctly To take a measurement, you need to be relaxed and comfortably seated, under comfortable room temperature. No eating, smoking or exercising 30 minutes before taking a measurement.• Sit on a chair with your feet flat on the floor.• Sit upright with your ...
Page 10 - Taking a Reading; Undo the wrist cuff and remove the unit.
3. Using the Unit 8 3.3 Taking a Reading Notes: • To cancel a measurement, press the O/I START button at any time during measurement.• Remain still while taking a measurement. 1. Press the O/I START button. The wrist cuff will start to inflate automatically. 2. Undo the wrist cuff and remove the uni...
Page 12 - Using the Memory Function
3. Using the Unit 10 3.4 Using the Memory Function The monitor automatically stores the result up to 60 sets. It can also calculate an average reading based on the measurements from the last three readings taken within 10 minutes. If there are only two readings in memory for that period, the average...
Page 13 - To delete all the values stored in memory
3. Using the Unit 11 EN To delete all the values stored in memory When the memory symbol ( ) appears, first press the MEM button. Then while holding it down, press the O/I START button simultaneously for about 2-3 seconds. Note: You cannot partially delete the stored readings. Second First HEM-6200-...
Page 14 - Troubleshooting and Maintenance; The Icons and Error Messages
12 4. Troubleshooting and Maintenance 4.1 The Icons and Error Messages Error Display Cause Remedy Irregular or weak pulses are detected. Remove the wrist cuff. Wait 2-3 minutes and then take another measurement. Repeat the steps in section 3.3. If this error continues to appear, contact your doctor....
Page 15 - Troubleshooting and Maintenance
4. Troubleshooting and Maintenance 13 EN Note: The irregular heartbeat symbol ( ) may also be displayed with error messages. Cuff is over inflated. Carefully read and repeat the steps listed under section 3.3. Movement during measurement. An Er mark with a code/number indicates the device has a hard...
Page 16 - Troubleshooting
4. Troubleshooting and Maintenance 14 4.2 Troubleshooting Problem Cause Remedy The reading is extremely low (or high). The wrist cuff is not at heart level. Measure while in the correct posture. Refer to section 3.2. The cuff is wrapped snugly around the wrist. Wrap the cuff correctly. Refer to sect...
Page 18 - Maintenance
4. Troubleshooting and Maintenance 16 4.3 Maintenance To protect your unit from damage, please avoid the following:• Subjecting your unit to extreme temperatures, humidity, or direct sunlight.• Washing the cuff or exposing the cuff or unit to water. • Disassembling the unit.• Subjecting the unit to ...
Page 20 - Technical Data
18 5. Technical Data Product Description Wrist Blood Pressure Monitor Model OMRON R3 (HEM-6200-E/HEM-6200-E8) Display LCD Digital Display Measurement Method Oscillometric method Measurement Range Pressure: 0 mmHg to 299 mmHg Pulse: 40 to 180 beats/min. Accuracy Pressure: 3 mmHg Pulse: 5% of read...
Page 23 - Some Useful Information about Blood Pressure; What is Blood Pressure?; What is Irregular Heartbeat?
21 EN 6. Some Useful Information about Blood Pressure What is Blood Pressure? Blood pressure is a measure of the force of blood flowing against the walls of the arteries. Arterial blood pressure is constantly changing during the course of the heart’s cycle. The highest pressure in the cycle is calle...
Page 25 - Some Useful Information about Blood Pressure
6. Some Useful Information about Blood Pressure 23 EN Classification of Blood Pressure by the World Health Organization The World Health Organization (WHO) and the International Society of Hypertension (ISH) developed the Blood Pressure Classification shown in this figure.This classification is base...