Page 4 - Check following components!
!" Vérifier les composants suivants !Prüfen Sie folgende Teile des Lieferumfangs!Controllare i componenti indicati di seguito!¡Compruebe los siguientes componentes! Controleer de volgende onderdelen!' ()* +/0! English Français Deutsch Italiano Español Nederlands 1( +2 Türkçe Automatic Blood Pres...
Page 5 - Contents
Contents Thank you for purchasing the OMRON M6 AC Automatic Blood Pressure Monitor. The OMRON M6 AC is a compact, fully automatic blood pressure monitor, operating on the oscillometric principle. It measures your blood pressure and pulse rate simply and quickly. For comfortable controlled inflation ...
Page 6 - Important Safety Information
1 EN Important Safety Information Warning: Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury. (General Usage) Consult your physician before using the device in pregnancy including pre-eclampsia, or if diagnosed with arrhythmia or arteriosclero...
Page 9 - Know Your Device; Monitor; Arm cuff; Display
4 1. Know Your Device Monitor DisplayCuff wrap guide lampMemory buttonWeekly average buttonSTART/STOP buttonUp/Down buttonsUSER ID selection switchDate/Time setting buttonBlood pressure colour indicatorAir jackBattery compartmentAC adapter jack Arm cuff Arm cuff (Arm circumference 22-42cm) Air plugA...
Page 10 - Display Symbols
5 1. Know Your Device EN Display Symbols Irregular Heartbeat Symbol ( ) When the monitor detects an irregular rhythm two or more times during the measurement, the irregular heartbeat symbol ( ) will appear on the display with the measurement values. An irregular heartbeat rhythm is defined as a rhyt...
Page 12 - Preparation; Battery Installation; Remove the battery cover.
7 EN 2. Preparation 2.1 Battery Installation 1. Remove the battery cover. 2. Insert 4 “AA” batteries as indicated in the battery compartment. 3. Replace the battery cover. Notes: • When the low battery symbol ( ) appears on the display, turn the monitor off, then replace all batteries at the same ti...
Page 13 - Using the AC Adapter
8 2. Preparation 2.2 Using the AC Adapter 1. Insert the AC adapter plug into the AC adapter jack on the rear side of the monitor. 2. Plug the AC adapter into an electrical outlet. To disconnect the AC adapter, unplug the AC adapter from the electrical outlet first and then remove the AC adapter plug...
Page 14 - Setting the Date and Time
9 2. Preparation EN 2.3 Setting the Date and Time Set the monitor to the correct date and time before taking a measurement for the first time. 1. Press the Date/Time setting ( ) button. The year flashes on the Date/Time display. 2. Push or button to change the year. Push button to confirm the year a...
Page 15 - Using the Device; Applying the Arm Cuff; Insert; Secure closed with the fabric fastener.; Open both the front and rear covers to read the following:
10 3. Using the Device 3.1 Applying the Arm Cuff Remove tight-fitting clothing or tight rolled up sleeve from your left upper arm.Do not place the arm cuff over thick clothes. 1. Insert the air plug into the air jack securely. 2. Wrap the arm cuff firmly in place around your left upper arm. The bott...
Page 16 - How to Sit Correctly
11 3. Using the Device EN 3.2 How to Sit Correctly To take a measurement, you need to be relaxed and comfortably seated, under comfortable room temperature. Avoid bathing, drinking alcohol or caffeine, smoking, exercising or eating 30 minutes before taking a measurement. • Sit on a chair with your f...
Page 17 - Taking a Measurement
12 3. Using the Device 3.3 Taking a Measurement Notes: • To cancel a measurement, press the START/STOP button to release the air in the arm cuff.• Remain still while taking a measurement. The monitor is designed to take measurements and store the measurement values in the memory for 2 people using U...
Page 18 - Press the START/STOP button to turn the monitor off.; Using the Guest Mode
13 3. Using the Device EN 3. Remove the arm cuff. 4. Press the START/STOP button to turn the monitor off. The monitor automatically stores the measurement result in its memory.It will automatically turn off after 2 minutes. Note: Wait 2-3 minutes before taking another measurement. Waiting between me...
Page 19 - Using the Memory Function; To View the Measurement Values Stored in Memory; Press the
14 3. Using the Device 3.4 Using the Memory Function The monitor automatically stores the results up to 100 sets for each user (1 and 2). It can also calculate an average value based on the last 3 measurement values taken within 10 minutes. Notes: • If there are only 2 measurement values in the memo...
Page 20 - To View the Average Value; Press and hold the
15 3. Using the Device EN To View the Average Value 1. Select your USER ID (1 or 2). 2. Press and hold the button for more than 3 seconds. Notes: • If the previous measurement was taken without setting the date and time, the average value is not calculated. • If there are no measurement values store...
Page 21 - To View the Weekly Average Value
16 3. Using the Device To View the Weekly Average Value The monitor calculates and displays weekly averages for the measurements taken in the morning and evening within 8 weeks for each user. Note: The week begins Sunday at 2:00. 1. Select your USER ID (1 or 2). 2. Press the button. The morning aver...
Page 23 - To Delete All the Values Stored in Memory; Press the Memory button, while the memory symbol (
18 3. Using the Device To Delete All the Values Stored in Memory The values stored in the memory are deleted by USER ID. 1. Select your USER ID (1 or 2). 2. Press the Memory button, while the memory symbol ( ) appears. 3. While holding the button down, press the START/STOP button for more than 3 sec...
Page 24 - Error Messages and Troubleshooting; Error Messages
19 EN 4. Error Messages and Troubleshooting 4.1 Error Messages Display Cause Solution Irregular heartbeats are detected. Remove the arm cuff. Wait 2-3 minutes and then take another measurement. Repeat the steps in section 3.3. If this error continues to appear, contact your physician. Movement durin...
Page 25 - Error Messages and Troubleshooting
20 4. Error Messages and Troubleshooting Air plug is disconnected. Insert the air plug securely.Refer to section 3.1. Arm cuff is applied too loosely. Apply the arm cuff tighter.Refer to section 3.1. Air is leaking from the arm cuff. Replace the arm cuff with a new one.Refer to section 5.3. Movement...
Page 26 - Troubleshooting
21 4. Error Messages and Troubleshooting EN 4.2 Troubleshooting Problem Cause Solution The measurement result is extremely high (or low). Arm cuff is applied too loosely. Apply the arm cuff tighter. Refer to section 3.1. Movement or talking during measurement. Remain still and do not talk during mea...
Page 28 - Maintenance and Storage; Maintenance
23 EN 5. Maintenance and Storage 5.1 Maintenance To protect your device from damage, please observe the following: • Store the device and the components in a clean, safe location.• Do not use any abrasive or volatile cleaners. • Do not wash the device and any components or immerse them in water.• Do...
Page 29 - Maintenance and Storage; Storage; Unplug the air plug from the air jack.
24 5. Maintenance and Storage 5.2 Storage Keep the device in its storage case when not in use. 1. Unplug the air plug from the air jack. 2. Gently fold the air tube into the arm cuff. Note: Do not bend or crease the air tube excessively. 3. Place the monitor and the arm cuff in the storage case. Do ...
Page 30 - Optional Medical Accessories
25 5. Maintenance and Storage EN 5.3 Optional Medical Accessories (within the scope of EC Medical Device Directive 93/42/EEC) Arm cuff Arm circumference 22-42 cm AC adapter Easy Cuff L-9911729-4 (Model: HEM-RML31) Adapter S-9515336-9 Adapter UK-9983666-5
Page 31 - Specifications
26 6. Specifications Product description Automatic Blood Pressure Monitor Model OMRON M6 AC (HEM-7322-E) Display LCD digital display Measurement method Oscillometric method Measurement range Pressure: 0 to 299 mmHgPulse: 40 to 180 beats / min. Accuracy Pressure: ± 3 mmHg Pulse: ± 5% of display readi...
Page 34 - Warranty
29 EN 7. Warranty Thank you for buying an OMRON product. This product is constructed of high quality materials and great care has been taken in its manufacturing. It is designed to give you every satisfaction, provided that it is properly operated and maintained as described in the instruction manua...
Page 35 - Some Useful Information about Blood Pressure; What is Blood Pressure?; Why is it a Good Thing to measure Blood Pressure at Home?
30 8. Some Useful Information about Blood Pressure What is Blood Pressure? Blood pressure is a measure of the force of blood flowing against the walls of the arteries. Arterial blood pressure is constantly changing during the course of the heart’s cycle. The highest pressure in the cycle is called t...