Page 2 - Lumex; are registered trademarks, and; Patriot; CONTENTS
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 2 GF Health Products, Inc. is not responsible for typographical errors. Packaging, warranties, products and specifications are subject to change without notice. Graham-Field and Lumex are registered trademarks, and Patriot is a trademark, of GF Health Pro...
Page 3 - SAFETY
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 3 1 SAFETY The safety statements presented in this chapter refer to the basic safety information that the op-erator of the Patriot Homecare Bed shall pay attention to and abide by. There are additional safety statements in other chapters or sections, whic...
Page 4 - INTRODUCTION, PATRIOT HOMECARE BED USER MANUAL; THERMAL MOTOR PROTECTION
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 4 2 INTRODUCTION, PATRIOT HOMECARE BED USER MANUAL This manual contains assembly and maintenance instructions for your Model US0208 / US0208PL Semi-Electric and US0458 / US0458PL Full-Electric / Low Patriot bed. Read the entire manual care-fully before us...
Page 5 - ASSEMBLY INSTRUCTIONS; HEAD AND FOOT ASSEMBLY; tion’s rivet; on both sides; . Ensure both rivets are completely seated before continuing.; straighten; shown above. Position head wing as shown above.
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 5 3 ASSEMBLY INSTRUCTIONS HEAD AND FOOT ASSEMBLY foot section head section motor cam hook rivet hook rivet Assembly, head and foot sections Note: Grid deck not shown for clarity 1. Place the head and foot sections on their sides at approximately right ang...
Page 6 - MOTOR ASSEMBLY; WARNING: Assemble motor on bed as follows. While it is possible to; motor; then unplug the power cord; Install Motors; Install motor; foot; sembly into position on the cam.; Head
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 6 MOTOR ASSEMBLY WARNING: Assemble motor on bed as follows. While it is possible to install the motor backwards, avoid it! Incorrect installation could cause the bed to move in unexpected ways. Note that the motor ends are labeled and embossed with head a...
Page 7 - Position the caps as shown at right so that; WARNING: Ensure that the motor is assem-; Problem; Install motor slide caps
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 7 Install motor slide caps 7. Position the caps as shown at right so that the grooves in the slide cap are aligned with the grooves in the motor. 8. Apply straight downward pressure to each end of the cap while sliding into the closed position. 9. Repeat ...
Page 8 - Low bed height
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 8 FULL-ELECTRIC BED ONLY (US0458 / US0458PL): CASTER POSITION Any full-electric Patriot bed can be used as either a standard-height or low-height bed by moving the casters. fig 2: casters mounted at low height, bed in lowest position fig 3: casters mounte...
Page 9 - BED END ASSEMBLY; All bed models: Assemble bed head end; Insert one locking and one non-locking caster into bed end.; Locking casters must be installed; See previous page for full-electric caster position options.; bed head end; Tilt the bed end slightly back toward you as shown above.; Engage lower rivet; All models: Assemble bed foot end; After bed head end has been attached, perform steps 1-6 in; All models: Assemble bed head end; above to install the foot end.
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 9 BED END ASSEMBLY All bed models: Assemble bed head end 1. Insert one locking and one non-locking caster into bed end. Locking casters must be installed diagonally from each other. See previous page for full-electric caster position options. 2. Standing ...
Page 10 - All models: Install bed end locking hardware; After bed head end has been attached, install; Bed End Locking Hardware; to act as a latch; Lanyard
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 10 All models: Install bed end locking hardware 1. After bed head end has been attached, install Bed End Locking Hardware to act as a latch for each corner.Bed End Locking Hardware, part number 690-1000-100, includes four lanyards, each as-sembled from a ...
Page 12 - Route cord under cross-angle
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 12 Full-electric bed only (US0458 / US0458PL): Assemble hi/lo motor shaft to foot end gear box notches in hi/lo motor shaft coupler slide over gear box shaft pin gear box hi/lo motor shaft hi/lo motor Attach hi/lo motor shaft to foot end gear box FOOT END...
Page 14 - All models: Attach hi/lo rod to gearbox or motor shaft pins; end of hi/lo motor shaft
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 14 All models: Attach hi/lo rod to gearbox or motor shaft pins Both ends of the hi/lo rod assembly have a slot to mount on the foot end motor shaft pins and bed end gearbox shaft pins. Follow steps 2-3 to mount the hi/lo rod assembly on the gearboxes. foo...
Page 15 - BED OPERATION; Plug the power cord into any grounded outlet.; WARNING: Do not leave the bed unattended when plugged in.; PENDANT HAND CONTROL OPERATION
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 15 4 BED OPERATION WARNING: Ensure that bed is secure and assembled as instructed on previous pages before operating. Plug the power cord into any grounded outlet. WARNING: Do not leave the bed unattended when plugged in. Info: In the event of a power fai...
Page 16 - crank
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 16 CRANK OPERATION: SEMI-ELECTRIC BED ONLY (US0208 / US0208PL) Use the crank at the center foot of the semi-electric bed (shown at right) to change bed height (raise or lower entire deck). TRENDELENBURG POSITION, ALL MODELS(HEAD END AT LOWER ELEVATION THA...
Page 17 - * Specifications are subject to change without notice.; Bed length; Deck width; * Maximum patient weight
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 17 5 SPECIFICATIONS, ALL MODELS * * Specifications are subject to change without notice. Deck height High position Deck height: High Position US0208 / US0208PL: 24.0 inches US0458 / US0458PL with cast- ers in low position: 20.0 inches US0458 / US0458PL wi...
Page 18 - ACCESSORIES: BED RAILS; Please read the following bed rail safety information.; BED RAIL SAFETY; and; Assist Bar
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 18 6 ACCESSORIES: BED RAILS Patriot Homecare Bed decks are equipped with patent-pending slots specifically for bed rail instal-lation, providing a fixed place to mount GF bed rails. Two styles of Lumex bed rails are described in Section 6: 1) The Lumex Pa...
Page 20 - LUMEX UNIVERSAL BED RAILS; result in personal injury and damage to the bed rails.; HEADBOARD; FOOTBOAR; Full-Length Side Rail; FOOTBOAR; No Gap Side Rail
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 20 LUMEX UNIVERSAL BED RAILS If you install Lumex universal bed rails, follow the installation instructions that accompany them; the following instructions are provided to describe the differences between the installation of bed rail crossbraces on Patrio...
Page 21 - Low Universal Half Rail; Head end
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 21 GF6580A-1: Low Universal Half Rail GF6580A-1 Low Universal Half Rail Cable tie Spring Buttons with cable tie Secure crossbrace to head deck angle FOOTBOAR D Spring Buttons G3 A G2 HEADBOARD B Follow same procedure for mounting half rails on foot-end us...
Page 22 - SERVICE PARTS, ALL MODELS; BED COMPONENTS; HEAD; Part number
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 22 7 SERVICE PARTS, ALL MODELS Service parts for all bed models are shown below and on the following two pages. See pictures and tables for appropriate service parts. Info: To order service parts, contact our Customer Service department at 800-365-2338 (f...
Page 23 - BED END COMPONENTS; Item Part number; panels
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 23 BED END COMPONENTS 3 5 6 7 4 1 2 8 HEAD FOOT Item Part number Description Semi-electric Full-electric US0208 US0208PL US0458 US0458PL 1 690-7001-901 Head bed end, semi-electric, walnut panels X 690-7601-901 Head bed end, semi-electric, plastic panels X...
Page 24 - BED SERVICE PARTS; Motor and pendant hand control, semi-electric
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 24 BED SERVICE PARTS 1 HEAD FOOT 2 3 4 Item Part number Description Semi-electric, H7920SE Full-electric, H7920FE US0208 / US0208PL US0458 / US0458PL 1 690-2001-913 Motor and pendant hand control, semi-electric X 690-2001-411 Pendant hand control, semi-el...
Page 25 - MAINTENANCE AND SAFETY CHECKS; Check all controls to ensure that all functions work; Cleaning; periodically remove mattress and clean mattress deck.
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 25 8 MAINTENANCE AND SAFETY CHECKS Perform the following as needed or between patient placements, whichever happens sooner: Electronics and Motors Check all controls to ensure that all functions work properly. Foot controlHead controlHi/Lo (if applicable)...
Page 26 - WARRANTY
US0208-INS-LAB-RevD11 • Patriot Homecare Bed 26 9 WARRANTY GF Health Products, Inc. warrants the Patriot US0208 / US0208PL Semi-Electric Homecare Bed Patriot US0458 / US0458PL Full-Electric / Low Bed as follows: Two-year limited warranty for defects in workmanship and materials of mechanical compone...