Page 2 - and; Lumex; are registered trademarks of GF Health Products, Inc.; CONTENTS
LF1050-INS-LAB-RevE13 2 GF Health Products, Inc. is not responsible for typographical errors. For the most updated and current information on packaging, warranties, products and specifications, including the most current version of these instructions, please visit our website at www.grahamfield.com....
Page 3 - INTRODUCTION; Intended use
LF1050-INS-LAB-RevE13 3 1 INTRODUCTION Congratulations on your purchase of the Lumex LF1050 Patient Lift. The following pages will provide you with important safety and operating instructions on the use of patient lifts, slings, and accessories, as well as maintenance and warranty information. Read ...
Page 4 - SAFETY PRECAUTIONS; CARE AND MAINTENANCE; Significance of safety statements; WARNINGS
LF1050-INS-LAB-RevE13 4 2 SAFETY PRECAUTIONS IMPORTANT: Before using patient lift, please read and adhere to the following safety precautions and warnings. Failure to do so could result in serious personal injur y or damage to your patient lift. Always consult your healthcare professional to determi...
Page 6 - FEATURES; LIFT COMPONENTS; Lowering Device; CONTROL BOX COMPONENTS; RESET Button
LF1050-INS-LAB-RevE13 6 3 FEATURES LIFT COMPONENTS a. Boom b. Mast c. Base d. Spreader bar e. Pendant f. Actuator g. Control box h. Caster brake i. Front caster j. Emergency Lowering Device CONTROL BOX COMPONENTS RESET Button Battery Testing Charger Plug-in Actuator Plug-in Pendant Plug-in
Page 7 - HANDLING PROCEDURES; SHIPPING AND TRANSPORTATION INSTRUCTIONS; Unpacking; from the box without assistance.; carrier or your Graham-Field authorized distributor.; Inspection
LF1050-INS-LAB-RevE13 7 4 HANDLING PROCEDURES SHIPPING AND TRANSPORTATION INSTRUCTIONS If the patient lift is to be reshipped by common carrier, it should be packed in the same carton in which it was received. Unpacking WARNING: The lift is heavy. To avoid personal injur y, DO NOT attempt to remove ...
Page 8 - ASSEMBLY INSTRUCTIONS; lift without assistance.
LF1050-INS-LAB-RevE13 8 5 ASSEMBLY INSTRUCTIONS WARNING: The lift is heavy. To avoid personal injur y, DO NOT attempt to assemble the lift without assistance. 1. Engage the lift brakes, as shown above.2. Remove the bolts from the bottom of the mast and base. 3. Pull the mast, boom, and spreader bar ...
Page 11 - CHARGING THE BATTERIES; three hours to be fully charged.; NOTICE: Do not charge the batteries for more than twelve hours.; Unplug the charger FIRST before using the lift.; Important Charging Information; CHECKING BATTERY CAPACITY
LF1050-INS-LAB-RevE13 11 CHARGING THE BATTERIES Charge the batteries after assembly.1. Ensure the battery box power switch is ON (i.e., the red button is up).2. Insert the charger plug into the battery box charging socket (the top slot).3. Insert the other charger plug into the power supply.4. All L...
Page 12 - OPERATING INSTRUCTIONS; the lift in the widest position.; Preparation before lifting; button to ensure the lift is operating correctly.; LIFTING THE PATIENT; WARNING: Use only Lumex slings with the Lumex LF1050 Patient Lift.
LF1050-INS-LAB-RevE13 12 6 OPERATING INSTRUCTIONS WARNING: During lifting and lowering, whenever possible, always keep the base legs of the lift in the widest position. WARNING: Use only Lumex slings specially designed for the Lumex LF1050 Patient Lift. WARNING: Do not put anything (e.g. cushion, pa...
Page 13 - Lifting the patient from a bed; center patient on sling.; should be in the maximum open position.; Lowering the patient onto a bed; feet on the base, straddling the column for extra stability.
LF1050-INS-LAB-RevE13 13 Lifting the patient from a bed 1. Position patient face-up in center of bed. Roll patient onto their side facing away from you.2. Fold sling lengthwise, outside out, and place it halfway under patient with lower edge just beneath knees and sling handle on the outside, facing...
Page 14 - Lowering the patient to the floor; Reverse the above procedures to lower.; transferring a patient to and from a wheelchair.; Then cross the loops, one through the other.; Removing the sling
LF1050-INS-LAB-RevE13 14 Lowering the patient to the floor 1. Place a pillow on the base to protect the patient’s head and neck (this is unnecessary when using a sling with a head support). 2. Reverse the above procedures to lower. Transferring the patient from a chair to a wheelchair or commode cha...
Page 16 - EMERGENCY RELEASE; Turn clockwise to lower
LF1050-INS-LAB-RevE13 16 EMERGENCY RELEASE In case of lift failure while using the Lumex LF1050 Patient Lift, follow the procedure below to safely lower the user. The Emergency Lowering Device is located at the top of the actuator shaft. It is intended for use if the actuator fails to operate while ...
Page 17 - CARE AND MAINTENANCE; BEFORE EACH USE
LF1050-INS-LAB-RevE13 17 7 CARE AND MAINTENANCE Proper care and maintenance are essential to keeping your Lumex LF1050 Patient Lift in a safe operating condition. In addition to inspecting the unit before each use, periodic maintenance checks should be done. We further recommend that this unit be se...
Page 19 - TROUBLESHOOTING
LF1050-INS-LAB-RevE13 19 8 TROUBLESHOOTING The following list of encountered problems and their solutions will assist you in determining what may be causing your Lumex lift not to function as designed. If you have a problem occurring which is not listed below, please contact your distributor or GF H...
Page 20 - SPECIFICATIONS
LF1050-INS-LAB-RevE13 20 9 SPECIFICATIONS Info: All dimensions are given in inches unless otherwise specified. Base width Open: 37.5"Closed: 26" Base height clearance 4.75" Base length 44" Lifting height range Info: As measured from sling hooks. Minimum: 27"Maximum: 69"Range:...
Page 21 - 0 LIMITED WARRANTY; GF Health Products Inc.
LF1050-INS-LAB-RevE13 21 10 LIMITED WARRANTY GF Health Products Inc. (“Graham-Field”) warrants this product to be free from manufacturing defects in material and workmanship for its period of warranty that is in accordance with industry standards. This warranty is extended only to the original purch...