Page 2 - COntents
John Bunn O Premier Oxygen Concentrator All specifications and product configurations are subject to change without notification. COntents 1 safety ................................................................................... 3 Important Information ...............................................
Page 3 - safety; Important Information; Before Installation
3 John Bunn O Premier Oxygen Concentrator 1 safety The John Bunn O2 Premier oxygen concentrator is intend-ed for use as an oxygen supplement device in a home or care facility. Use the concentrator only as prescribed by a physician. This user manual serves as a reference for your oxygen concentrator....
Page 4 - Placement; fire Warning & explosion
4 John Bunn O Premier Oxygen Concentrator s Damage: If the electrical power source becomes unstable, discontinue use and find an alternate source. s Damage: The oxygen concentrator cabinet contains no serviceable parts! The oxygen concentrator cabinet should ONLY be opened by an authorized equipment...
Page 5 - Maintenance
5 John Bunn O Premier Oxygen Concentrator WARNING: To reduce the risk of fire, special care must be taken when using oxygen concentrators. DO NOT SMOKE while using this device. Keep all matches, lighted cigarettes or other sources of ignition out of the room in which this concentrator is located. In...
Page 6 - additional safety Warnings
6 John Bunn O Premier Oxygen Concentrator additional safety Warnings WARNING: TO REDUCE THE RISK OF BURNS, ELECTROCUTION, FIRE OR INJURY TO PERSONS, avoid using while bathing. If continuous usage is required by the direction of a physician, the concentrator must be located in another room at least 9...
Page 8 - featuRes; summary; The John Bunn O; Characteristics
8 John Bunn O Premier Oxygen Concentrator featuRes summary The John Bunn O 2 Premier oxygen concentrator is intend- ed for use as an oxygen supplement device in a home or care facility. Use the concentrator only as prescribed by a physician.The concentrator separates oxygen from ambient air, providi...
Page 9 - HandlIng; Apparent Loss or Damage; unpacking; Note: Unless the oxygen concentrator is to be used
9 John Bunn O Premier Oxygen Concentrator 3 HandlIng gf Health Products, Inc. freight Policyfor your Protection, Read Carefully The carrier accepted this merchandise “in good condition” and is responsible for safe delivery.Before signing the freight bill, inspect the shipment care-fully for damage o...
Page 10 - Inspection
10 John Bunn O Premier Oxygen Concentrator Inspection Examine exterior of the oxygen concentrator for nicks, dents, scratches or other damages. Inspect all compo-nents. storage Store the repackaged oxygen concentrator in a dry area. DO NOT place anything on top of the repackaged concen-trator.
Page 11 - OPeRatIOn and InstallatIOn; Insert the humidifier bottle into the elastic belt on the
11 John Bunn O Premier Oxygen Concentrator 4 OPeRatIOn and InstallatIOn Preparing the Concentrator(refer to above picture) 1. Unscrew the humidifier bottle cover, if so equipped (hu- midifier bottle is optional). 2. Fill the humidifier bottle with purified water (or dis- tilled water) between the ma...
Page 12 - turning the Concentrator On; Press the power switch to start the concentrator. There; turning the Concentrator Off
1 John Bunn O Premier Oxygen Concentrator trator’s AC connector into a properly grounded power outlet. turning the Concentrator On 1. Press the power switch to start the concentrator. There will be a brief delay before the concentrator starts. The alarm system emits a tone of one second duration, th...
Page 13 - MaIntenanCe; Cleaning the Cabinet; Bottom filter; with clear water until all dirt is removed.; Wait for filter to dry.; When dry, install filter carefully as
13 John Bunn O Premier Oxygen Concentrator 5 MaIntenanCe Cleaning the Cabinet Clean the cabinet at least once a month.1. Turn off the power switch.2. Clean the cabinet with a mild household cleaner and a non-abrasive cloth or sponge. Do not spill liquid inside the cabinet. Cleaning or Replacing the ...
Page 14 - Side filters; Note: For earlier models only, you will need a small blade; Wait for filter to dry before installing
14 John Bunn O Premier Oxygen Concentrator Side filters Note: For earlier models only, you will need a small blade screwdriver to open the side filter access door. For later models, the filter grill simply pulls out. If this operation is too difficult for you to perform, please contact your equipmen...
Page 15 - Intake filter; Note: Please have your Graham-Field equipment provider
15 John Bunn O Premier Oxygen Concentrator • A wood stove is used• Kerosene or oil is used• Pets are present• Tobacco is used Intake filter Note: Please have your Graham-Field equipment provider replace the intake filter. This operation requires opening the intake filter access door. Replace, do not...
Page 16 - tROuBlesHOOtIng guIde
16 John Bunn O Premier Oxygen Concentrator 6 tROuBlesHOOtIng guIde syMPtOM CHeCK / aCtIOn Concentrator power is Check if concentrator is unplugged; plug in. on and alarm is on, but concentrator doesn’t Check if circuit breaker is tripped; fix circuit work breaker. If concentrator still doesn’t work,...
Page 17 - Note: If you experience a problem with your concentrator
17 John Bunn O Premier Oxygen Concentrator syMPtOM CHeCK / aCtIOn Concentrator doesn’t discontinue use; contact graham-field equipment work; alarm sounds; provider immediately. Oxygen concentration is too optional O sensor red low for safe use. warning light illuminates (JB0160-015B, JB0160-015, JB0...
Page 18 - BIll Of MateRIals; sensor already installed.; Replacement Parts and Optional accessories
18 John Bunn O Premier Oxygen Concentrator 7 BIll Of MateRIals Your concentrator includes the following components:• Oxygen concentrator (one each)• User manual (one each)• Humidifier bottle hose (one each)• Humidifier bottle adapter (one each)The concentrator comes with one bottom filter, two side ...
Page 19 - sPeCIfICatIOns; • Pressure Relief Mechanism Operational at:
19 John Bunn O Premier Oxygen Concentrator 8 sPeCIfICatIOns • Flow Range: 1 to 5L/min• Oxygen Concentration: 1-4 LPM 94% ± 2% 5 LPM 92% ± 2% • Maximum Outlet Pressure: 6.52 psi ± .065 psi (45kPa ± 4.5kPa) • Pressure Relief Mechanism Operational at: 36.26 psi ± 3.63 psi (250kPa ± 25kPa) • Altitude: U...
Page 20 - lIMIted WaRRanty
0 John Bunn O Premier Oxygen Concentrator 9 lIMIted WaRRanty This warranty is extended only to the original purchaser of our products. This warranty gives you specific legal rights and you may also have other legal rights which may vary from state to state. GF Health Products, Inc. warrants its John...
Page 21 - Note: The above warranty has been written to comply with
1 John Bunn O Premier Oxygen Concentrator as identification, within thirty (30) days of the return autho-rization date. Do not return products to GF Health Products, Inc. or its Authorized Equipment Providers, without prior consent. C.O.D. shipments will be refused. All return ship-ments must prepay...
Page 23 - nOtes
Page 24 - regIster; yOUr; • Receive updates with product information, mainte-; on the inside, and the address on the outside.
4 John Bunn O Premier Oxygen Concentrator regIster yOUr PRODUCT! The benefits of registering:• Safeguard your investment.• Ensure long-term maintenance and servicing of your purchase. • Receive updates with product information, mainte- nance tips, and industry news. • GF Health Products, Inc. can co...
Page 25 - QuestIOnnaIRe
QuestIOnnaIRe 1. Method of purchase: (check all that apply) � Medicare � Insurance � Medicaid � Other 2. This product was purchased for use by: (check one) � Self � Parent � Spouse � Other 3. This product was purchased for use at: � Home � Facility � Other 4. I purchased a Graham-Field product becau...