Page 2 - Contents
760000-INS-LAB-RevA08 Contents INtRoductIoN ..................................................................................... 3 INteNded uSe of thIS devIce ........................................................ 3 ImpoRtANt SAfety pRecAutIoNS .................................................. 3...
Page 3 - IntroduCtIon; Intended use of thIs devICe; Important safety preCautIons; DANGER: Indicates an imminent hazard situation
760000-INS-LAB-RevA08 3 IntroduCtIon Use this manual for set-up and operation of the Alternating Pressure / Low Air Loss Mattress System. Read all instruc-tions before using the mattress system. Save this manual for future reference. Intended use of thIs devICe The intended use of the Alternating Pr...
Page 4 - danger; to reduce the risk of electrocution:; immediately; WarnIng
4 760000-INS-LAB-RevA08 danger to reduce the risk of electrocution: DANGER: Always unplug this product immediately after using.DANGER: Do not use this product while bathing.DANGER: Do not place or store this product where it can fall or be pulled into a tub or sink.DANGER: Do not place this product ...
Page 5 - only; CautIon; equIpment symbols
760000-INS-LAB-RevA08 5 WARNING: Never block the air openings of this product or place it on a soft surface, such as a bed or couch, where the openings may be blocked. Keep the air opening free of lint, hair, and other similar particles.WARNING: Never drop or insert any object into any opening or ho...
Page 6 - produCt desCrIptIon
6 760000-INS-LAB-RevA08 For U.S. and Canada only E228589 53DG Medical Equipment: Air Pumpwith respect to electrical shock, fire and mechanical hazards only in accordance with UL60601-1 and CAN/CSA C22.2 No. 601.1 produCt desCrIptIon pump and mattress system pump and mattress system CPR valves mattre...
Page 7 - There are four therapy mode settings:; Autofirm; (when the head is raised to
760000-INS-LAB-RevA08 7 pump: rear main power switch power cord receptacle fuse hanger filter pump: front panel cycle time adjust button; cycle time indicators at right of button decrease pressure button; pressure level indicators at right of button panel lock button; panel lock indicator at right o...
Page 8 - Front panel fourth row; Note: Main power switch on side of pump; Low pressure indicator:
8 760000-INS-LAB-RevA08 increase pressure under the buttocks). Press the therapy adjust button at the left end of the row to cycle between the settings. The selected therapy mode setting's indicator will illuminate. Front panel fourth row Panel lock button: The panel lock protects the panel set- tin...
Page 9 - contact your Graham-field equipment provider.; InstallatIon; unpaCkIng; Fold open the wire hangers on
760000-INS-LAB-RevA08 9 Ser vice indicator: Mechanical failure, service required; contact your Graham-field equipment provider. InstallatIon unpaCkIng 1. Before unpacking, check for obvious damage to the car- ton or its contents. If damage is evident, please notify the carrier and your Graham-Field ...
Page 10 - Plug the power cord into a properly grounded electrical; operatIon; Turn the main power
10 760000-INS-LAB-RevA08 s Caution: Place the device in a position where the care-giver and/or doctor can access it easily. 5. Connect the mattress-pump air hose quick connector; ensure that the connector has clicked securely into place before continuing. s Caution: Ensure that the air hoses are not...
Page 11 - pressure adjustment; weight distribution; quICk dIsConneCt
760000-INS-LAB-RevA08 11 Note: The mattress system will automatically enter Autofirm mode for the quickest inflation at set-up; initial inflation takes approximately twenty minutes. 3. The pump will begin delivering air into the mattress.4. When the selected pressure level is reached, the pump will ...
Page 12 - Disconnect the quick connector; maIntenanCe; CleanIng; Perform the following cleaning procedures before use.; Pump; : Wipe the pump with a clean, damp cloth and mild; Caution: Do not immerse or soak the pump.; : Wipe the mattress unit with a clean, damp cloth; Caution: Do not clean with phenolic products.
1 760000-INS-LAB-RevA08 transport Cap In case of power failure or trans-port: Disconnect the quick connec-tor and cap the mattress with the transport cap to slow deflation. Cpr funCtIon If CPR must be performed on the patient while the mattress is in use, to immediately deflate the mat-tress, either...
Page 13 - general maIntenanCe; Fuse replacement; Air filter replacement
760000-INS-LAB-RevA08 13 general maIntenanCe 1. Check main power cord for abrasion or excessive wear.2. Check mattress cover for signs of wear or damage. En- sure mattress cover and tubes are connected correctly. 3. Check airflow from the air hose connector. The airflow should alternate between each...
Page 14 - storage; Store mattress and pump in a dust-free environment; speCIfICatIons
14 760000-INS-LAB-RevA08 storage 1. Lay the mattress out flat and upside down.2. Roll from the head end toward the foot end.3. Stretch the foot-end strap around the rolled mattress to prevent unrolling. 4. Store mattress and pump in a dust-free environment with no exposure to direct sunlight. s Caut...
Page 15 - Model; Warranty
760000-INS-LAB-RevA08 15 Mattress Specification Model 8" mattress Dimensions (L x W x H) 78.7" x 35.4" x 8" (200 x 90 x 20.3 cm) Weight 16.5 lb (7.5 kg) Warranty GF Health Products, Inc. warrants the Lumex AltaDyne Plus 760000 Alternating Pressure / Low Air Loss Mattress Sys-tem for ...