Page 2 - Description
RHT10-EN v4.0 4/14 2 Introduction Congratulations on your purchase of the Relative Humidity and Temperature Datalogger. This datalogger measures and stores up to 16,000 relative humidity and 16,000 temperature readings over 0 to 100%RH and -40 to +70°C (-40 to +158°F) measurement ranges. The user ca...
Page 3 - Operation; INSTALLING THE GRAPHING SOFTWARE; USBXPressInstaller.exe; Follow the defaults to load this driver.
RHT10-EN v4.0 4/14 3 Operation INSTALLING THE GRAPHING SOFTWARE Install the supplied Windows PC Datalogger Software by placing the supplied program disk in the PC’s CD-ROM drive. If the installation program does not automatically open and provide on-screen prompts, open and browse the CD-ROM drive, ...
Page 4 - RUNNING THE GRAPHING SOFTWARE; Press “Connect” to open the main screen,
RHT10-EN v4.0 4/14 4 RUNNING THE GRAPHING SOFTWARE With the RHT10 inserted in the PC’s USB port, double-click the Datalogger Graph icon to launch the program. The opening screen will appear as shown below. Press “Connect” to open the main screen, Press “About” to view the version number, Press “View...
Page 5 - MENU BAR DESCRIPTION; The main menu for the software appears as follows:; LOGGER SETUP; Click on the computer icon on the menu bar (2; from left) or select LOGGER SET from the LINK pull-down menu.
RHT10-EN v4.0 4/14 5 MENU BAR DESCRIPTION The main menu for the software appears as follows: From left to right the MENU BAR icons represent DATA DOWNLOAD, LOGGER SETUP, FILE OPEN, FILE SAVE-AS, FILE PRINT, VIEW ALL, and ZOOM. In addition, the FILE, VIEW, LINK, and HELP pull-down menus are available...
Page 6 - DATA DOWNLOAD; Open the graphing software program if it is not still running.; on the left) or select Data Download from the LINK pull-down menu
RHT10-EN v4.0 4/14 6 Once the changes have been made, click on the SETUP button to save them. Press the DEFAULT button to set the RHT10 back to factory default conditions. Press the CANCEL button to abort the setup. Note: Any stored data will be permanently erased when the user clicks on Setup. To s...
Page 9 - FILE PRINT; optional
RHT10-EN v4.0 4/14 9 To customize the graph window select SHOW TRACES, BACKGROUND, GRID LINES, and MASK POINTS from the VIEW pull-down menu. These options provide the following features: SHOW TRACES: Allows the user to select which data to view as a trace (Temperature & Alarms). BACKGROUND: Sele...
Page 10 - No LEDs flashing; Record Alarm; Yellow double flash every 10 sec. ‡
RHT10-EN v4.0 4/14 10 RHT10 LED STATUS GUIDE LEDs Meaning Action Record Alarm No LEDs flashing -No logging in process Or -No battery found Or -Battery completely discharged Fit battery in place Replace the battery and download the data Record Alarm Green single flash every 10 sec. * -Logging, no ala...
Page 11 - Specifications; Selectable sampling interval: From 2 seconds up to 24 hours
RHT10-EN v4.0 4/14 11 Specifications Relative Humidity Overall Range 0 to 100% Accuracy (0 to 20 and 80 to 100%) ±5.0% Accuracy (20 to 40 and 60 to 80%) ±3.5% Accuracy (40 to 60%) ±3.0% Temperature Overall Range -40 to 158 o F (-40 to 70 o C) Accuracy (-40 to -10 and +40 to +70 o C) ±2 o C Accuracy ...
Page 12 - Battery Replacement
RHT10-EN v4.0 4/14 12 Battery Replacement It is highly recommended that the battery be replaced every 12 months, or prior to logging critical data. The RHT10 does not lose its stored readings when the battery weakens or when the battery is replaced; the data logging process will however be stopped a...
Page 14 - Garantía; Servicios; Líneas; Copyright
RHT10-EN v4.0 4/14 14 Garantía FLIR Systems, Inc. , garantiza este dispositivo marca Extech Instruments para estar libre de defectos en partes o mano de obra durante tres años a partir de la fecha de embarque (se aplica una garantía...