Page 2 - SET
2 CO200-EU-EN v1.2 08/13 Introduction Congratulations on your purchase of this Model CO200 Carbon Dioxide Meter. This meter measures CO 2 level, air temp., humidity, date and time. With visible and audible alarms, this is an ideal instrument for indoor air quality (IAQ) diagnosis. This meter is ship...
Page 3 - Operation; POWER ON; b At; TAKING MEASUREMENT
3 CO200-EU-EN v1.2 08/13 Operation POWER ON Plug in the adaptor and the meter turns on automatically with a short beep. I f t h e v ol ta g e i s t o o hi gh o r l o w, “ b At “ will display on the LCD and an LED will flash. The LCD will display current CO 2 , temp., humidity, date and time. The air...
Page 4 - CO
4 CO200-EU-EN v1.2 08/13 3 RELAY OUTPUT The meter is designed with a relay for connection to an external indicating or controlling device. When CO 2 readings go over the upper limit the relay closes. The relay will open again when CO 2 readings fall below the lower limit. The relay output port requi...
Page 5 - TEMPERATURE SCALE
5 CO200-EU-EN v1.2 08/13 P1. 3 CO 2 ALARM: BEEP ALARM P1. 3 is used for setting the CO 2 upper limit for the BEEPER ALARM level. The current set value will be blinking on the display. Press the ▲ or MIN/MAX ▼ button to increase or decrease the value. Each press adjusts in 100ppm increments. The alar...
Page 7 - Calibration; It is not suitable to use; Manual Calibration
7 CO200-EU-EN v1.2 08/13 Calibration CO 2 CALIBRATION The meter is calibrated to a standard 400ppm CO 2 concentration at the factory NOTE: When the accuracy becomes a concern or after a year of use, return the meter to Extech for a standard calibration. CAUTION: Do not calibrate the meter in an atmo...
Page 8 - CLEANING AND STORAGE
8 CO200-EU-EN v1.2 08/13 Specifications Function Range Resolution Accuracy CO 2 0 to 9999ppm 1ppm ±(5%rdg +50ppm) Temperature -10 to 60°C 14 to 140°F 0.1° ±0.6°C/0.9°F Humidity 0.1 to 99.9% 0.1% ±3%(10 to 90%) ±5%(< 10% or > 90%) Display LCD Sensor Type CO 2 : NDIR (non-dispersive infrared) te...
Page 9 - TROUBLESHOOTING
9 CO200-EU-EN v1.2 08/13 TROUBLESHOOTING Can’t power on: Check whether the adaptor is properly plugged in. Slow response: Check whether the air flow channels on the rear of the meter are blocked. "BAt” and green LED keep flashing: The adaptor output voltage is too high or too low. Please use t...
Page 10 - Copyright
10 CO200-EU-EN v1.2 08/13 CO 2 Levels and Guidelines Non-Enforced Reference levels: 250 - 350 ppm – background (normal) outdoor air level 350- 1,000 ppm - typical level found in occupied spaces with good air exchange. 1,000 – 2,000 ppm - level associated with complaints of drowsiness and poor ...