Page 6 - SPECIFICATIONS; GBC Models
P a g e 4 © Adam Equipment Company 2011 SPECIFICATIONS GBC Models Model # GBC 16 GBC 32 GBC 60 GBC 35a GBC 70a GBC 130a Maximum Capacity 16kg 32kg 60kg 16kg/35lb 32kg/70lb 60kg/130lb Readability 0.5g 1g 2g 0.5g/0.001lb 1g/0.002lb 2g/0.005lb Repeatability (Std Dev) 0.5g 1g 2g 0.5g/0.001lb 1g/0.002lb ...
Page 7 - GFC Models
P a g e 5 © Adam Equipment Company 2011 GFC Models Model # GFC 75 GFC150 GFC 300 GFC 165a GFC 330a GFC 660a Maximum Capacity 75kg 150kg 300kg 75kg / 165lb 150kg / 330lb 300kg / 660lb Readability 5g 10g 20g 5g / 0.01lb 10g / 0.02lb 20g / 0.05lb Repeatability (Std Dev) 5g 10g 20g 5g / 0.01lb 10g / 0.0...
Page 8 - UNPACKING
P a g e 6 © Adam Equipment Company 2011 3.0 INSTALLATION 3.1 UNPACKING The user’s application and the technical specifications of the platform or load cell will determine the necessary configuration. 3.2 LOCATING • The scales should not be placed in a location that will reduce the accuracy. • Avoid ...
Page 9 - SETTING
P a g e 7 © Adam Equipment Company 2011 3.3 SETTING UP THE SCALES • The pillar is attached to the base using a bracket that must be attached to the base frame first using the 4 bolts supplied. The pillar is secured to the bracket using 2 sets of screws. The cable from the base to the indicator modul...
Page 10 - DESCRIPTIONS; Keys
P a g e 8 © Adam Equipment Company 2011 4.0 KEY DESCRIPTIONS Keys Functions [0-9, .] Numeric entry keys, used to manually enter a value for tare weights, unit weight, and sample size. [CE] Used to clear the unit weight or an erroneous entry. [M+] Add the current count to the accumulator. Up to 99 va...
Page 11 - The scales have three sections. These are; “Weight”, “Unit Weight” and “Count”; WEIGHT
P a g e 9 © Adam Equipment Company 2011 5.0 DISPLAYS The scales have three sections. These are “Weight”, “Unit Weight” and “Count” . The LCD design is: WEIGHT UNIT WEIGHT COUNT 5.1 WEIGHT DISPLAY It has 6-digit display to indicate the weight on the scale. Net Weight Display, "Net" Stability ...
Page 12 - “Mem”
P a g e 10 © Adam Equipment Company 2011 5.2 UNIT WEIGHT DISPLAY This display will show the unit weight of a sample. This value is either input by the user or computed by the scale. The unit of measurement is grams if grams if kilograms is selected for the weighing unit, or pounds if pounds is selec...
Page 13 - “Unit; ZEROING; “Weight”; TARING
P a g e 11 © Adam Equipment Company 2011 6.0 OPERATION NOTE: GBC/GFC-a Scales only SETTING THE WEIGHING UNIT, lb or kg The scale will turn on displaying the last weighing unit selected, either kilograms or pounds. To change the weighing unit press the [U. Wt./Units] key when the “Unit Weight” displa...
Page 14 - Net”; “Count”
P a g e 12 © Adam Equipment Company 2011 • The user can enter a preset tare value if the display is at zero or gross weight. Enter the value for preset tare using the keypad, then press the [Tare] key to set the preset zero value. • If the display was at zero the weight will show a negative value eq...
Page 15 - B. Entering a known Unit Weight
P a g e 13 © Adam Equipment Company 2011 B. Entering a known Unit Weight • If the unit weight is already known then it is possible to enter that value using the keypad. • Enter the value of the unit weight in grams, using the numeric keys followed by pressing the [U. Wt./Units] key. The "Unit We...
Page 16 - “diSPEr”
P a g e 14 © Adam Equipment Company 2011 6.3.4 Manually Accumulated Totals • The values (weight and count) shown on the display can be added to the values in the memory by pressing the [M+] key. The "Weight" display will show the total weight, the "Count" display will show the total ...
Page 17 - PARAMETERS; “F1” “CAL”
P a g e 15 © Adam Equipment Company 2011 7.0 USER PARAMETERS The parameters are set to customise the scale to suit the weighing applications. You need to enter a secure menu by entering a password when requested. • Press [Tare] once, during the initial counting of the display after the power is turn...
Page 18 - SETTING OF PIN; ENABLE WEIGHING UNITS
P a g e 16 © Adam Equipment Company 2011 SETTING OF PIN • Display will show “F2” “Pin” . This parameter allows setting of a new password number. The default password is “0000” . • Press [Tare] . The “Weight” display will show "Pin 1" . • Enter the new password number. The “Unit Weight” displ...
Page 20 - INTERFACE
P a g e 18 © Adam Equipment Company 2011 9.0 RS-232 INTERFACE The GBC/GFC Series are supplied with a RS-232 bi-directional interface. The scale when connected to a printer or computer through the RS-232 interface, outputs the weight, unit weight and count. Specifications: Connector: The scale can be...
Page 21 - TOTAL
P a g e 19 © Adam Equipment Company 2011 Data Format-Normal Output: Data Format- Memory Recall Print: <lf><cr> <lf><cr> DATE 12/09/2006 TIME 14:56 <lf><cr> GROSS Wt 1.234 Kg Net Wt. if net weight is displayed Unit Wt. 123 g Kg or g for metric weights and lb for po...
Page 23 - INPUT
P a g e 21 © Adam Equipment Company 2011 9.1 INPUT COMMANDS FORMAT The scale can be controlled with the following commands. The commands must be sent in upper case letters, i.e. “T” not “t”. Press the Enter key of the PC after each command. T<cr><lf> Tares the scale to display the net we...
Page 24 - SETUP; Weight
P a g e 22 © Adam Equipment Company 2011 9.2 RS-232 SETUP The RS-232 interface uses parameters set by the user for language, baud rate and date format. Press and hold the [Print] key for 4 seconds to access the parameters. Press [U. Wt./Units] to scroll through the options and [Tare] to confirm the ...
Page 25 - ACCUMULATION FUNCTIONS
P a g e 23 © Adam Equipment Company 2011 The scale will perform the following functions depending on the Accumulation and Print settings: ACCUMULATION FUNCTIONS PRINT FUNCTIONS AC Auto AC mAn AC oFF Print Auto Accumulate and print automatically Print automatically ; Accumulate and print when [M+] is...
Page 26 - REAL; Setting up the clock
P a g e 24 © Adam Equipment Company 2011 9.3 REAL TIME CLOCK SETUP The Real Time Clock (RTC) is used only for the RS-232 output. The Date and Time can be set as required. The scale will keep the clock running even when the power is off. Setting up the clock • Press and hold the [CE] key for 4 second...
Page 27 - AUTO; AND; “LO bAT”; “Unit Weight”
P a g e 25 © Adam Equipment Company 2011 9.4 AUTO SLEEP FUNCTION This function may be enabled or disabled by the user. If enabled, when the scale is not used for some time (as pre-set by the user under this function) it automatically switches off. To set this parameter- • Press and hold the [Zero] k...
Page 28 - FOR
P a g e 26 © Adam Equipment Company 2011 10.2 BACKLIGHT FOR LCD • The backlight of the LCD can be set to be- “1” : ON at all the time, “2” : ON only when a weight is placed on the scale or “3” : Turned off. • To set the backlight press and hold [Pst] key for 4 seconds. • The weight display will show...
Page 29 - CODES; ERROR CODE; Err 1
P a g e 27 © Adam Equipment Company 2011 11.0 ERROR CODES During the initial power-on testing or during operation, the scale may show an error message. The meaning of the error messages is described below. If an error message is shown, repeat the step that caused the message, turning the balance on,...
Page 30 - TO
P a g e 28 © Adam Equipment Company 2011 12.0 SERVICE PARAMETERS 12.1 USING “0000” TO ENTER THE CALIBRATION PARAMETER • Press the [Tare] key during the display counting when turned on. The display will ask for a code number, “Pin “ on the Weight Display. • Enter the number 0000 when “Pin “ is displa...
Page 32 - PARTS; Equipment. A partial list of such items is as follows-; Main Power cord
P a g e 30 © Adam Equipment Company 2011 13.0 REPLACEMENT PARTS AND ACCESSORIES If you need to order any spare parts and accessories, contact your supplier or Adam Equipment. A partial list of such items is as follows- • Main Power cord • Power supply (GFB/GFC-a) • Replacement Battery • Stainless st...
Page 33 - INFORMATION; following information which should be kept ready:
P a g e 31 © Adam Equipment Company 2011 14.0 SERVICE INFORMATION This manual covers the details of operation. If you have a problem with the scale that is not directly addressed by this manual then contact your supplier for assistance. In order to provide further assistance, the supplier will need ...
Page 35 - Manufacturer’s Declaration of Conformity
© Adam Equipment Company 2011 Manufacturer’s Declaration of Conformity This product has been manufactured in accordance with the harmonised European standards, following the provisions of the below stated directives: Electro Magnetic Compatibility Directive 2004/108/EC Low Voltage Directive 2006/95/...