Page 3 - CONTENTS
© Adam Equipment Company 2007 1 CONTENTS 1.0 INTRODUCTION ...................................................................................................... 2 2.0 SPECIFICATIONS.................................................................................................... 3 2.1 KEYBOARD AND...
Page 5 - Model; KEYBOARD AND DISPLAY
© Adam Equipment Company 2007 3 2.0 SPECIFICATIONS Model BFW 75 BFW 150 BFW 300 Maximum Capacity 75 kg 150 kg 300 kg Readability 5 g 10 g 20 g Repeatability (Std. Dev.) 5 g 10 g 20 g Linearity (±) 10 g 20 g 40 g Tare Range Full range Units of Measure 4 (kg, g, lb, oz) Stabilisation Time 2-3 seconds ...
Page 6 - The scale is at zero.
© Adam Equipment Company 2007 4 The display has the following symbols: 0 The scale is at zero. A weight has been tared, the display is showing the net weight. The battery needs to be re-charged. While the battery is getting re-charged, this symbol will be flashing. O O O O The weight is stable. Th...
Page 7 - SETTING UP THE SCALE; Unpack the scale by removing from the box.
© Adam Equipment Company 2007 5 2.3 SETTING UP THE SCALE Indicator Cable with Connector Holder Cable Pillar • Unpack the scale by removing from the box. • Place the pillar in the upright position into the rear socket of the platform base and fasten the screw. • Pull out the cable slightly from the h...
Page 8 - SYS UP; Uolt xx
© Adam Equipment Company 2007 6 3.0 OPERATIONS 3.1 POWER • Attach the main power cord to the power supply. The indicator will charge the internal battery whenever it is connected to the 220 VAC power supply. • To turn on the scale, press the power switch on the right back of the indicator. • A messa...
Page 10 - USER PARAMETERS
© Adam Equipment Company 2007 8 3.7 ALARM 3.7.1 Over-load alarming • When the sample weight on the scale is over the maximum capacity the beep will sound and the display will show “EEE----“. • Remove the weight immediately to avoid any damage to the scale. 3.7.2 Low-load alarming • If the platform i...
Page 11 - Parameters
© Adam Equipment Company 2007 9 Parameters Description 1. BLT Enable / disable the backlight 2. CHK Enable / disable the check-weighing facility 3. FIL Enable / disable the ADC filter rate 4. SDT Sets the RS-232 output function. Options are 1, 2, 3 and 4. 5. PAR Sets the following parity options- 8 ...
Page 13 - The standard Interface parameters are:; Normal Output; Complete Data Output:; SDT; ” not “; Command Description; or
© Adam Equipment Company 2007 11 6.0 RS-232 INTERFACE The BFW series of scales come with a bi-directional RS-232 interface. Default settings are 9600 baud, No Parity, 8 data bits, 1 stop bit (9600,N,8,1). The standard Interface parameters are: C onnection details are: Normal Output U : (For SDT 1, 2...
Page 14 - Enable check weighing.; TECHNICAL PARAMETERS; ZEO
© Adam Equipment Company 2007 12 C1 Enable check weighing. 7.0 TECHNICAL PARAMETERS To enter this section, press and hold the [Mode] and the [Tare] key while you turn on the scale. The display will show the revision number and then the first parameter Zero Tracking as “ ZEO ” 7.1 ZERO TRACKING • Whe...
Page 15 - WEl
© Adam Equipment Company 2007 13 7.3 CALIBRATION Next is the Calibration section. “ WEl ” will be displayed. • When WEI is displayed, press the [Unit] key to scroll through the options for selecting the calibration mass. • To select the desired mass press the [Mode] key. • The raw ADC count will be ...
Page 16 - PROBLEMS
© Adam Equipment Company 2007 14 8.0 TROUBLE-SHOOTING GUIDE PROBLEMS POSSIBLE CAUSES Display is blank On/Off switch on rear panel is off Scale not turned on Battery not charged No turn on test Battery not charged Power supply not plugged in Power supply faulty Display turned off Error message displa...
Page 17 - SERVICE INFORMATION; Details of your company
© Adam Equipment Company 2007 15 9.0 SERVICE INFORMATION This manual covers the details of operation. If you have a problem with the scale that is not directly addressed by this manual then contact your supplier for assistance. In order to provide further assistance, the supplier will need the follo...
Page 18 - WARRANTY INFORMATION
© Adam Equipment Company 2007 16 WARRANTY INFORMATION Adam Equipment offers Limited Warranty (Parts and Labour) for the components failed due to defects in materials or workmanship. Warranty starts from the date of delivery. During the warranty period, should any repairs be necessary, the purchaser ...
Page 20 - ADAM EQUIPMENT; is an ISO 9001:2000 certified global organisation with more than; ADAM; © Copyright by Adam Equipment Co. Ltd. All rights reserved. No part of
© Adam Equipment Company 2007 ADAM EQUIPMENT is an ISO 9001:2000 certified global organisation with more than 30 years experience in the production and sale of electronic weighing equipment. Products are sold through a world wide distributor network supported from our company locations in the UK, US...