Step 30. Select the Deviation Control Type Submenu

Press

d

. If flashing

_DEV

Deviation is displayed press

a

,

otherwise press

b

until flashing

_DEV

is shown. Now

press

d

to store and go to next menu item.

Step 31. Select the Latched Type Submenu

Press

d

. If flashing

UNLT

Unlatched is displayed press

a

,

otherwise press

b

until

UNLT

is displayed.

Press

d

to store and advance to next menu item.

Step 32. Select the Normally Open Type of Contact
Closure Submenu

Press

d

. If flashing

N.o.

Normally Open is displayed,

press

a

, otherwise press

b

until

N.o.

is displayed. Press

d

to store and advance to next menu item.

Step 33. Select the Above Type of Active Submenu

Press

d

. If flashing

ABoV

Above is displayed, press

a

,

otherwise press

b

until

ABoV

is displayed. Press

d

to store

and advance to next menu item.

Step 34. Enable Alarm 1 at Power On (

A.P.oN

)

Press

d

. If flashing

ENBL

is displayed, press

a

, otherwise

press

b

until

ENBL

is displayed. Press

d

to store and

advance to next menu item.

Step 35. Enter Alarm 1 High Submenu

Press

a

twice to skip

ALR.L

Alarm 1 Low value.

ALR.L

is

for below &

ALR.H

for above.

Step 36. Set the Alarm 1 High value (

ALR.H

)

Press

d

. Press

b

or

c

until value to set the display to

002.0

. Press

d

to save.

Step 37. Enter the Alarm 2 Menu

The display will show

ALR2

the top menu for Alarm 2.

Repeat steps from 28 to 36 to set for Alarm 2 the same
conditions as for Alarm 1.

Step 38. Configuration of Display Color Selection

Press

a

until the

COLR

Display Color Selection Menu

appears on the Display. Configure

COLR

as

N.CLR

/

GRN

(green),

1.CLR

/

RED

(red),

2.CLR

/

AMBR

(amber). Please

refer to the operator’s manual if needed.

Step 39. Run a Test

Press

a

until reset the controller and return to

RUN

Mode

to display

075.0

(Ambient Temperature). Now you are

ready to observe temperature as it rises 10°F higher than
displayed. Touch the tip of the Thermocouple to raise the
temperature above the Alarm 2 High value

082.0

, and AL2

will turn on, and Display Color will change from Green to
Amber. Continue touching the tip to raise the temperature
above the Alarm 1 High value

087.0

and Display Color will

change from Amber to Red.

Step 11. Enter to the Thermocouple Type Input Submenu

Press

d

to display flashing, previously selected Thermocouple

type.

Step 12. Scroll through available selection of TC types

Press

b

to sequence thru flashing Thermocouple types,

(select k -for type "K" CHROMEGA

®

/ALOMEGA

®

)

J K T E N DIN J R S B C - TC types

J k t E N dN J

R S b C - Display

Step 13. Store TC type

After you have selected the Thermocouple type press

d

to

store your selection, the instrument automatically advances
to the next menu item.

Step 14. Enter to Reading Configuration Menu

The display shows

RDG

Reading Configuration, which is the

top menu for 4 submenus: Decimal Point, Degree Units,
Filter Constant and Input/Reading Submenus.

Step 15. Enter to Decimal Point Submenu

Press

d

to show

DEC

Decimal Point.

Step 16. Display the Decimal Point position

Press

d

again to display the flashing Decimal Point position.

Step 17. Select the Decimal Point position

Press

b

to select

FFF.F

Decimal Point position.

Step 18. Store selected Decimal Point position

By pressing

d

momentarily the Decimal Point position will

be stored and the instrument will go to the next menu item.

Step 19. Enter to Temperature Unit Submenu

Display shows

TEMP

Temperature Unit.

Step 20. Display available Temperature Units

Press

d

to display the flashing Degree

°F

or

°C

.

Step 21. Scroll through Temperature Units selection

Press

b

to select

°F

Degree.

Step 22. Store the Temperature Unit

Press

d

to display momentarily that the Degree Unit has

been stored and the instrument will go automatically to the
next menu item.

Step 23. Enter the Filter Constant Submenu

Display shows

FLTR

Filter Constant Submenu.

Step 24. Display the Filter Constant Value Submenu

Press

d

to display the flashing, previously selected Filter

Constant.

Step 25. Scroll through available Filter Constants

Press

b

to sequence thru Filter Constants

0001

,

0002

,

0004

,

0008

,

0016

,

0032

,

0064

and

0128

.

Step 26. Store the Filter Constant

Press

d

momentarily to store

0004

Filter Constant and the

instrument will automatically go to the next menu item.

Step 27. Enter Alarm 1 Menu

The display will show

ALR1

the top menu for Alarm 1. In the

following steps we are going to enable Alarm 1, Deviation,
Unlatch, Normally Open, Active Above, Enable at power on
and +2°F High Alarm i.e. Process Value > Setpoint 1 Value
+2°F will activate Alarm 1.

Step 28. Enter Alarm 1 Enable/Disable Submenu

Press

d

to display flashing

DSBL

/

ENBL

.

Step 29. Enable Alarm 1 Submenu

If flashing

ENBL

is displayed, press

a

, if

DSBL

is displayed,

press

b

until

ENBL

is displayed, then press

d

to store and

go to the next menu item.

USA

MADE IN

Servicing North America:

USA:

DD

One Omega Drive, P.O. Box 4047

ISO 9001 Certified

D

Stamford CT 06907-0047

DD

TEL: (203) 359-1660DD

FAX: (203) 359-7700

DD

e-mail: [email protected]

Canada::

D

976 Bergar

DD

Laval (Quebec) H7L 5A1

DD

TEL: (514) 856-6928DD

FAX: (514) 856-6886

DD

e-mail: [email protected]

For immediate technical or application assistance:

USA and Canada:

D Sales Service: 1-800-826-6342 / 1-800-TC-OMEGA

®

DD

Customer Service: 1-800-622-2378 / 1-800-622-BEST

®

DD

Engineering Service: 1-800-872-9436 / 1-800-USA-WHEN

®

Mexico and

D

TEL: (001) 203-359-7803D

FAX: (001) 203-359-7807

Latin American:

D

e-mail: [email protected]

DD

Servicing Europe:

Benelux::

TEL: +31 20 3472121D

FAX: +31 20 6434643

D

Toll Free in Benelux: 0800 0993344D

D

e-mail: [email protected]

Czech Republic:

D

Frystatska 184, 733 01 Karviná

DD

TEL: +420 59 6311899DD

FAX: +420 59 6311114

DD

e-mail: [email protected]

OMEGAnet

®

On-Line Service

www.omega.com

Internet e-mail

[email protected]

®

®

France:

D

TEL: +33 1 61 37 29 00D

FAX: +33 1 30 57 54 27

D

Toll Free in France: 0800 466 342

D

e-mail: [email protected]

Germany/Austria:

D Daimlerstrasse 26, D-75392 Deckenpfronn, Germany

DD

TEL: +49 7056 9398-0DD

FAX: +49 7056 9398-29

DD

Toll Free in Germany: 0800 639 7678

DD

e-mail: [email protected]

United Kingdom:

D One Omega Drive

ISO 9001 Certified

D

River Bend Technology Centre

DD

Northbank, Irlam Manchester M44 5BD United Kingdom

DD

TEL: +44 161 777 6611D FAX: +44 161 777 6622

DD

Toll Free in England: 0800 488 488

DD

e-mail: [email protected]

MQS3446/0307

WARRANTY/ DISCLAIMER

OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of

one (1) year

from the date of purchase. In addition to OMEGA’s standard warranty period, OMEGA Engineering will

extend the warranty period for

four (4) additional years

if the warranty card enclosed with each instrument is

returned to OMEGA.

If the unit malfunctions, it must be returned to the factory for evaluation. OMEGA’s Customer Service Department will
issue an Authorized Return (AR) number immediately upon phone or written request. Upon examination by OMEGA,
if the unit is found to be defective, it will be repaired or replaced at no charge. OMEGA’s WARRANTY does not apply
to defects resulting from any action of the purchaser, including but not limited to mishandling, improper interfacing,
operation outside of design limits, improper repair, or unauthorized modification. This WARRANTY is VOID if the unit
shows evidence of having been tampered with or shows evidence of having been damaged as a result of excessive
corrosion; or current, heat, moisture or vibration; improper specification; misapplication; misuse or other operating
conditions outside of OMEGA’s control. Components which wear are not warranted, including but not limited to
contact points, fuses, and triacs.

OMEGA is pleased to offer suggestions on the use of its various products. However, OMEGA neither
assumes responsibility for any omissions or errors nor assumes liability for any damages that result from
the use of its products in accordance with information provided by OMEGA, either verbal or written. OMEGA
warrants only that the parts manufactured by it will be as specified and free of defects. OMEGA MAKES NO
OTHER WARRANTIES OR REPRESENTATIONS OF ANY KIND WHATSOEVER, EXPRESS OR IMPLIED,
EXCEPT THAT OF TITLE, AND ALL IMPLIED WARRANTIES INCLUDING ANY WARRANTY OF
MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE HEREBY DISCLAIMED. LIMITATION
OF LIABILITY: The remedies of purchaser set forth herein are exclusive, and the total liability of OMEGA with
respect to this order, whether based on contract, warranty, negligence, indemnification, strict liability or
otherwise, shall not exceed the purchase price of the component upon which liability is based. In no event
shall OMEGA be liable for consequential, incidental or special damages.

CONDITIONS: Equipment sold by OMEGA is not intended to be used, nor shall it be used: (1) as a “Basic
Component” under 10 CFR 21 (NRC), used in or with any nuclear installation or activity; or (2) in medical applications
or used on humans. Should any Product(s) be used in or with any nuclear installation or activity, medical application,
used on humans, or misused in any way, OMEGA assumes no responsibility as set forth in our basic
WARRANTY/DISCLAIMER language, and, additionally, purchaser will indemnify OMEGA and hold OMEGA harmless
from any liability or damage whatsoever arising out of the use of the Product(s) in such a manner.

RETURN REQUESTS/INQUIRIES

Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE
RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBTAIN AN AUTHORIZED RETURN (AR)
NUMBER FROM OMEGA’S CUSTOMER SERVICE DEPARTMENT (IN ORDER TO AVOID PROCESSING
DELAYS). The assigned AR number should then be marked on the outside of the return package and on any
correspondence.

The purchaser is responsible for shipping charges, freight, insurance and proper packaging to prevent breakage in
transit.

FOR WARRANTY RETURNS, please have the
following information available BEFORE
contacting OMEGA:
1.

Purchase Order number under which the product
was PURCHASED,

2.

Model and serial number of the product under
warranty, and

3.

Repair instructions and/or specific problems
relative to the product.

FOR NON-WARRANTY REPAIRS, consult OMEGA for
current repair charges. Have the following information
available BEFORE contacting OMEGA:
1.

Purchase Order number to cover the COST of the
repair,

2.

Model and serial number of product, and

3.

Repair instructions and/or specific problems
relative to the product.

OMEGA’s policy is to make running changes, not model changes, whenever an improvement is possible. This affords
our customers the latest in technology and engineering.

OMEGA is a registered trademark of OMEGA ENGINEERING, INC.

© Copyright 2007 OMEGA ENGINEERING, INC. All rights reserved. This document may not be copied, photocopied,
reproduced, translated, or reduced to any electronic medium or machine-readable form, in whole or in part, without
the prior written consent of OMEGA ENGINEERING, INC.

PATENT AND TRADEMARK NOTICE: This product is covered by one or more of the following patents: U.S. Pat.
No. Des. 336,895; 5,274,577; 6,243,021 / CANADA 2052599; 2052600 / ITALY 1249456; 1250938 / GERMANY DE
41 34398 C2 / SPAIN 2039150; 2048066 / UK Patent No. GB2 249 837; GB2 248 954 / FRANCE BREVET NO. 91
12756. The “Meter Bezel Design” is a trademark of Newport Electronics, Inc. USED UNDER LICENSE. Other U.S.
and International Patents pending or applied for.

This device is marked with the international caution symbol. It is important to read the
Setup Guide before installing or commissioning this device, as the guide contains important
information relating to safety and EMC.

WARNING:

These products are not designed for use in, and should not be used for, patient-

connected applications.

It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply.
OMEGA is constantly pursuing certification of its products to the European New Approach Directives.
OMEGA will add the CE mark to every appropriate device upon certification.
The information contained in this document is believed to be correct, but OMEGA Engineering, Inc.
accepts no liability for any errors it contains, and reserves the right to alter specifications without notice.

TRADEMARK NOTICE:

®

,

omega.com

®

,

,

and

®

are Trademarks of

OMEGA ENGINEERING, INC.

®

SPECIFICATION

Accuracy:

+0.5°C temp;
0.03% rdg. process typical

Resolution:

1°/0.1°; 10 µV process

Temperature Stability:

0.04°C/°C RTD;
0.05°C/°C TC @ 25°C (77°F);
50 ppm/°C process

Display:

4-digits, 9-segments LED,
21 mm (0.83") with red, green and
amber programmable colors

Input Types:

Thermocouple, RTD, Analog Voltage
and Current

TC: (ITS 90)

J, K, T, E, R, S, B, C, N, L

RTD: (ITS 68)

100/500/1000 ohm Pt sensor
2-, 3-, or 4-wire; 0.00385 or 0.00392
curve

Voltage:

0 to 100 mV, 0 to 1 V, 0 to 10 Vdc

Current:

0 to 20 mA (4 to 20 mA)

Output 1

:

Relay 250 Vac @ 3 A Resistive Load,
SSR, Pulse

Output 2

:

Relay 250 Vac @ 3 A Resistive Load,
SSR, Pulse

Only with -AL Limit Alarm option

Options: Communication

RS-232 / RS-485

or

10BaseT

or

Excitation: 24 Vdc @ 25 mA

EXC. not available for Low Power Option

Line Voltage/Power:

90 - 240 Vac ±10%, 50 - 400 Hz*,
or 110-375 Vdc,

4W

for Single Display;

5W

for Dual Display

* No CE compliance above 60 Hz

Low Voltage Power Option:

12 - 36 Vdc,

3 W**

for Single Display;

20 - 36 Vdc,

4 W**

for Dual Display

** Units can be powered safely with 24 Vac

but No Certification for CE/UL are claimed.

Dimensions:

48 H x 96 W x 127 D mm
(1.89 x 3.78 x 5")

Weight:

295 g (0.65 lb)

Approvals:

UL, C-UL, CE per EN 61010-1:2001

DPi8 Temperature & Process Monitor

CNi8-AL, CNi8D-AL

Temperature & Process Limit Alarm