Instruction Manual

226

LIQ_MAN_ABR_226

April 2014

Submersion/Insertion Toroidal Sensor

For additional information, please visit our website at www.rosemountanalytical.com

CAUTION

SENSOR/PROCESS APPLICATION COMPATIBILITY

The wetted sensor materials may not be compatible with process composition and operating conditions.

Application compatibility is entirely the responsibility of the user.

CAUTION

Before removing the sensor, be absolutely certain that the process pressure is reduced to 0 psig and the process

temperature is lowered to a safe level!

Specifications

Wetted Material:

PEEK (glass-filled), EPDM gasket

Temperature:

248 °F (120 °C) maximum

Pressure:

295 psig (2135 kPa abs)

Installation

Keep at least 2.4 in. (60 mm) between the sensor and pipe walls. If the clearance is too small, calibrate the sensor in place.
Ensure that the sensor is completely submerged in the liquid. Mounting the sensor in a vertical pipe run with the flow
from bottom to top is best. If the sensor must be installed in a horizontal pipe run, mount the sensor in the 3 o’clock or 9
o’clock position.

Figure 1. Submersion Installation Showing Mounting Hardware

Screw the rear facing thread of the PEEK adapter into the pipe. Do not let the back end of the sensor get wet.

"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.

Summary

Page 2 - Instruction Manual; Figure 2. Insertion Installation Showing Mounting Hardware; Wiring; Figure 3. Wire Functions

2 226 Instruction Manual April 2014 LIQ_MAN_ABR_226 Figure 2. Insertion Installation Showing Mounting Hardware Wiring Keep sensor wiring away from ac conductors and high current demanding equipment. Do not cut cable. Cutting the cable may void the warranty. Note For additional wiring information on ...

Page 6 - Figure 10. Wiring sensors through a remote junction box

6 226 Instruction Manual April 2014 LIQ_MAN_ABR_226 Figure 10. Wiring sensors through a remote junction box Wire sensors point to point. For wiring at the analyzer end, refer to the appropriate analyzer wiring diagram. For interconnect-ing cable 23294-00, use the 226-54 wiring diagram. For interconn...

Page 7 - Calibration; metal pipe; Maintenance; CAUTION

7 Instruction Manual 226 LIQ_MAN_ABR_226 April 2014 Calibration The nominal cell constant of the 226 sensor is 1.2/cm. The error in the cell constant is about ±10%, so conductivity readings made using the nominal cell constant will have an error of at least ±10%. Wall effects, discussed below, will ...