Emerson LIQ_MAN_ABR_226 - Manuals

Emerson LIQ_MAN_ABR_226 – Manual in PDF format online.

Manuals:

1 Page 1
2 Page 2
3 Page 3
4 Page 4
5 Page 5
6 Page 6
7 Page 7
8 Page 8
9 Page 9
10 Page 10
11 Page 11
12 Page 12

Summary

Page 2 - Instruction Manual; Figure 2. Insertion Installation Showing Mounting Hardware; Wiring; Figure 3. Wire Functions

2 226 Instruction Manual April 2014 LIQ_MAN_ABR_226 Figure 2. Insertion Installation Showing Mounting Hardware Wiring Keep sensor wiring away from ac conductors and high current demanding equipment. Do not cut cable. Cutting the cable may void the warranty. Note For additional wiring information on ...

Page 6 - Figure 10. Wiring sensors through a remote junction box

6 226 Instruction Manual April 2014 LIQ_MAN_ABR_226 Figure 10. Wiring sensors through a remote junction box Wire sensors point to point. For wiring at the analyzer end, refer to the appropriate analyzer wiring diagram. For interconnect-ing cable 23294-00, use the 226-54 wiring diagram. For interconn...

Page 7 - Calibration; metal pipe; Maintenance; CAUTION

7 Instruction Manual 226 LIQ_MAN_ABR_226 April 2014 Calibration The nominal cell constant of the 226 sensor is 1.2/cm. The error in the cell constant is about ±10%, so conductivity readings made using the nominal cell constant will have an error of at least ±10%. Wall effects, discussed below, will ...

Emerson Manuals