Dell JSM-6060LV- User Manual

Dell JSM-6060LV

Dell JSM-6060LV– User Manual, read for free online in PDF format. We hope this helps you resolve any issues you may have. If you have further questions, please contact us through the contact form.

1 Page 1
2 Page 2
3 Page 3
4 Page 4
5 Page 5
6 Page 6
7 Page 7
8 Page 8
9 Page 9
10 Page 10
11 Page 11
12 Page 12
13 Page 13
14 Page 14
Page: / 14

Table of Contents:

  • Page 2 – No part of the sample should extend above the ruler end
  • Page 3 – The IR; EVAC
  • Page 4 – Recommended Procedure:
  • Page 5 – The slider might fully; The average brightness should; Auto Alignment (Auto; button in Alignment box
  • Page 9 – Knob Panel
  • Page 11 – and manual knob; BEIW; modes; Selecting the Operating Voltage; Click
Loading the manual

BASIC Step-by-Step Instructions for JEOL 6060LV

Version 4/13/11

General Operation


1.

Check System: Pumps running and EVAC light green (main panel near key switch)?
If not notify a staff member.

2.

If the SEM software is not running, start SEM Software (

SEM Main Menu

icon). It

takes several minutes to start, be patient. If you need to log into computer use the
account (USER:jeol;PASSWORD:jeol). Please do not log into other accounts. Use
portable media, ftp, or UIUC NetFiles to transfer your data.

3.

Prepare your samples before venting! Please keep the area clean, organized, and wear
GLOVES!! Anything that goes in the vacuum should be clean and handled only with
gloves or tweezers; keep the vacuum clean.

4.

Press the

Sample

button on toolbar.

5.

Press the

Vent

button; it takes a couple minutes to vent.

6.

When Vent light stops blinking (main panel near key switch), slide open stage
drawer.

"Loading the manual" means you need to wait until the file loads and becomes available for online reading. Some manuals are very large, and the time they take to appear depends on your internet speed.

Other Manuals for Dell JSM-6060LV

Summary

Page 2 - No part of the sample should extend above the ruler end

7. Slide sample holder onto stage dovetail. Note proper direction to slide on “dovetail” relative to step on stage platform. 8. Adjust sample height (Working Distance) using height gauge. 10 mm is the EDS working distance. Bottom end of ruler corresponds to the working distance (see photo). Adjust t...

Page 3 - The IR; EVAC

9. Specimen tilting is limited by working distance and specimen diameter. The IR chamber-scope must be used for sample tilting and Z adjustment to verify that safe clearance with detectors, etc. is maintained. Safe clearance is at least 5mm distance on LCD monitor. The following table shows the typi...

Page 4 - Recommended Procedure:

15. Set the Signal to SEI , if not already (You can change it to BEIW , if desired later). 16. Click the HT Ready icon to turn on the High Voltage, now should be Some intensity should be seen in the image window. If not press View and ACB . ACB takes several seconds to run. Something should be visib...