Page 4 - Avoid using the flowmeter in an; Converter LF602 dimensions and weights:
LF410/LF600 LF410/LF602 4 Cable connection ports: Cable glands — LF600 and LF602 without cFMus Approval: Provided as standard OD of cable φ 11~13mm Material Nylon 66 G (PF) 1/2 male screws. Note: When PROFIBUS option is specified , cable grands size is φ 6~8mm for signal cable, φ 11~13mm for power c...
Page 5 - ANSI class 150 and class 300 dimensions:; The dimension of L1 is changed when the material of
LF410/LF600 LF410/LF602 5 Installation Dimensions Unit : mm L1 Φ D1 (L 2) 64 143 (35) 225 131 BS16, DIN PN16 and JIS 10K dimensions: Meter size (mm) L1 (mm) L2 (mm) D1 (mm) Weight (kg) 15 70 252 49 approx. 5 Note: 1 inch = 25.4mm ANSI class 150 and class 300 dimensions: Meter size (inch) L1 (inch) L...
Page 8 - Figure 9. Separate type converter LF602; External Connections; Use the appropriate switch rating as shown below:; flowmeters Wiring Diagram
LF410/LF600 LF410/LF602 8 矢視A 信号ケーブルグランド 電源ケーブルグランド 封止栓 励磁ケーブルグランド 電流出力ケーブルグランド 接地端子(M4) 取付金具 4-φ11 取付金具 82 (36 ) 15 6 13 15 1 17 8 20 0 151 74±0.3 25 LCD 表示器 赤外線スイッチ Excitation cable ground 15 1 151 74±0.3 4 – Φ 11 Attachment LCD display 17 8 20 0 IR Switch Cable ground 15 6 (36) 25 82 13 I/O cable...
Page 9 - flowmeters wiring Diagram
LF410/LF600 LF410/LF602 9 Terminal board Connected detector Excitation cable IV wire 5.5 mm 2 or more Power switch (External double-pole power switch ) Thick walled steel conduit Signal cable Power cable (CVV) Digital input cable (CVV-S) Current output (4~20mAdc)or PROFIBUS Grounded with 100 Ω or le...
Page 10 - Wiring Precautions; Figure 12. Electrical Conductivity and
LF410/LF600 LF410/LF602 10 Wiring Precautions (1) Explosion proof type flowmeters are not provided cable glands. Refer to the part Cable connection port at detector and converter. (2) Connect the grounding wire (IV wire 5.5mm² or more) to a good earth ground (100 Ω or less ground resistance). Make t...
Page 11 - Combined type; Separate type; Grounding of the Combined type and Separate type
LF410/LF600 LF410/LF602 11 Piping Precautions (1) Design piping so that the flowmeter detector pipe is always filled with the fluid being measured, whether the fluid is flowing or not. (2)The detector has no adjustable piping mechanism. Install an adjustable short pipe where needed. L:Required strai...
Page 12 - Make sure the full-scale flow rate used for the; Figure 13. Flow Rate and Flow Velocity; Calibration Range; Table 1. Standard Flow Range; PED matrix in each flange connection.; Table 2 PED matrix in each flange connection
LF410/LF600 LF410/LF602 12 Meter Size To select the meter size: See Figure 13 and find meter sizes within the velocity of 0.1 to 10m/s for a specified full-scale (measuring range high limit) flow. Select one that has its full-scale velocity between 1 and 3m/s. Note: Make sure the full-scale flow rat...
Page 13 - About establishment environment; Ordering Information; Consult Toshiba representative before ordering
LF410/LF600 LF410/LF602 13 About establishment environment Do not store or install the flowmeter in: • Places where there is direct sunlight. • Places where excessive vibration or mechanical shock occurs. • Places where high temperature or high humidity conditions obtain. • Places where corrosive at...
Page 14 - Table 3. Specification Code (Wafer type detector LF410 Series)
LF410/LF600 LF410/LF602 14 Table 3. Specification Code (Wafer type detector LF410 Series) Detector category Model Specification Code Normal type Ex. type 1 2 3 4 5 6 7 8 9 10 11 12 13 14 Description Gr.-A Gr.-B Gr.-C Gr.-D L F 4 1 0 Normal specification type √ √ L F 4 1 4 Hazardous location certific...
Page 15 - Table 4. Specification Code for converters
LF410/LF600 LF410/LF602 15 Table 4. Specification Code for converters Model Specification Code 1 2 3 4 5 6 7 8 9 10 11 12 13 14 Contents L F 6 0 Electromagnetic flowmeter converter LF600 type LF602 type 0 Integral (Combined) type – 2 Separate (Remote) type – A F Purpose Standard cFMus class I, Divis...