Page 2 - Intended Use
2 Intended Use Medical Purpose This is a medical device that measures the brachial blood pressure in a non-invasive manner. Intended User Legally certi ed: such as doctor, nurse and medical expert. Patient Population It is used on adult patients only. Environment The instrument is used in physicians...
Page 3 - Exemptions
3 Exemptions OMRON will not bear any responsibilities on the following matters. When a problem or damage occurs caused by the maintenance and/or repair con-ducted by a person other than OMRON or the dealer speci ed by OMRON The problem or damage of OMRON product caused by the product of other manu-f...
Page 4 - Names and Functions of the Parts; Connector; Main unit
4 Names and Functions of the Parts HIDE P-SET 100 AUTO 140 180 220 260 280 MODE AVG. MANU. CHECK SINGLE ON/OFF A VG . /1s t / 2 nd /3rd D E F L AT I O N START STOP mmHg SYS mmHg DIA BPM PULSE AC DC8V 1. LCD display 2. HIDE (non-display) Button 3. DC jack 4. P-SET (pressure setting) Volume 5. MODE Se...
Page 5 - Battery level; Display; SYS; Pulse rate; Number of irregular pulse waves:; Charging
5 Battery level Displays for the usable level. Displays for the low level. Displays for the unusable level. Names and Functions of the Parts Display HIDE BPM mmHg PULSE SYS mmHg DIA AC Systolic blood pressure Diastolic blood pressure Contents when the AVG.Mode is selected AVG. : Mean value 1st : Fir...
Page 6 - Explanation of Functions; Target value setting:
6 Names and Functions of the Parts Explanation of Functions (1) Pressure setting function P-SET In ation can be set by AUTO (automatic setting) or the target value. AUTO (automatic setting): In the SINGLE (single measurement), AVG. (average), and MANU. (auscultation) Modes, the monitor estimates the...
Page 7 - Items to set; Procedure to change the set values
7 Names and Functions of the Parts (4) Setting mode change function You can set the number of times to measure, the waiting time until the start of measurement, and the interval between measurements. Items to set Set value F1 Number of measurements 2 times, 3 times F2 Waiting time until the start of...
Page 8 - Notes on Safety; Warning sign; Examples of signs; Warning
8 Notes on Safety The warning signs and the sample icons shown here are listed for you to use the product safely and correctly as well as to prevent the risk and the damage to you and others from happening.The icons and meanings are as follow. Warning sign Contents Warning Indicates matters in which...
Page 9 - Caution
9 Notes on Safety Caution When you are not going to use the unit for a long period of time, be sure to remove the AC adapter from the electric outlet. You may suffer electric shock, or electric leak or re may arise because of deteriorated insulation. Be sure to remove the AC adapter from the electri...
Page 10 - Requests from OMRON
10 Notes on Safety Caution Be sure to use the power supply of 230 VAC. It may cause re or electric shock. Do not share an electric outlet with other unit or electric appliance. It may cause re or electric leak. Do not pull the power cord when pulling out the AC adapter from the electric outlet. The ...
Page 11 - Maintenance
11 Notes on Safety Maintenance Be sure to inspect the unit on regular basis. Check accuracy on a 1 year interval.If the unit has not been used for a while, be sure to con rm that the unit operates normally and safely before use Durability The durability of this unit is ve years. (Arm cuff is a consu...
Page 12 - Components of the Product; Options
12 Components of the Product Main unit Accessories HIDE P-SE T 100 AUTO 140 180 220 260 280 MODE AVG . MA NU. CHE CK SIN GLE ON/OFF AV G ./ 1 st /2 nd / 3rd D E F L ATIO N START STOP Medium size arm cuff (with built-in air bag) Model: HEM-9CMCArt.no.: 4928640 - 0Applicable arm circumference: 22 to 3...
Page 13 - Securely connect the air tube.; How to Apply the Arm Cuff
13 Select the arm cuff according to the arm circumference of the patient. Arm circumference Name of the arm cuff 17 - 22 cm22 - 32 cm32 - 42 cm Small size adult arm cuff (option)Adult arm cuffLarge size adult arm cuff (option) Be sure to use the arm cuff suitable for the size of the patient’s arm.If...
Page 14 - Adjust the level of the arm cuff to the level of the hear t.; How to use the Power Source
14 How to Apply the Arm Cuff (cont.) RANGE ART. RANGE ART. INDEX 22~32cm (9"~13") REF HEM-CR19 Adjust the level of the arm cuff to the level of the hear t. Keep the level of the arm cuff at the same level as the heart during the measure-ment. How to use the Power Source (AC adapter exclusive...
Page 15 - How to Measure Blood Pressure; List of Measurement Modes; Refer to Page
15 How to Measure Blood Pressure Warning If any abnormal matter occurs during the mea-surement such as the in ation does not stop, remove the arm cuff or pull the air tube from the main unit. You may suffer peripheral neuropathy. Requests from OMRON Be sure that the patient should not touch the unit...
Page 17 - To stop the measurement during the measurement:; Push the STOP Button.; To start the stopped measurement again:; Push the START Button.
17 Wrap the arm cuff over the patient’s arm. 1. Set the MODE Selector to “AVG.”. 3. ON/OFF AVG./1st /2nd/3rd D EFL ATIO N START STOP HIDE P-SET 100 AUTO 140 180 220 260 280 MODE AVG. MANU. CHECK SINGLE Push the START Button to star t the measurement. Set the unit to the start time of the rst measure...
Page 19 - What you need to prepare; How to use the blood pressure monitor; Open the air release valve of rubber ball to de ate.; CHECK result
19 How to Measure Blood Pressure (in CHECK Mode) Display of pressure can be con rmed by the CHECK Mode (yearly procedure). What you need to prepare (1) Well-adjusted reference pressure monitor (including rubber ball), (2) T-shaped tube, (3) Two rubber tubes, and (4) something in the cylindrical shap...
Page 20 - Installation and Replacement of Battery Pack; Batter y life
20 Installation and Replacement of Battery Pack Remove the batter y cover on the back of the main unit. Installation: To install the battery pack, connect the battery pack connector to the connector in the battery cover. Replacement: Remove the battery pack from the connector and replace with a new ...
Page 21 - Then wipe the unit with a soft dr y cloth.; How to Clean the Unit after Use
21 Clean the blood pressure monitor,the cuff and the cables using a cloth dampened with water,diluted disinfectant alcohol or diluted mild detergent. To disinfect, wipe the unit with a soft cloth squeezed well after moistened with the following disinfectant. Benzalkonium chloride 0.01 - 0.2 w/v% Chl...
Page 22 - List of Error Codes; Error code
22 List of Error Codes Error code Cause How to correct Er1 In ation error When the pressure does not exceed 15 mmHg after operating a pump for 15 seconds from the start of in ationWhen the in ation does not reach the set pressure of the arm cuff within the speci ed time after starting the in ation •...
Page 23 - Troubleshooting
23 If a trouble occurs while using the unit, please check the following. Problem What to inspect How to correct The unit in ates to abnormally high (low) pressure. Is the arm cuff wrapped correctly? Wrap the arm cuff correctly, and measure again. Is the patient moving his/her arm or body during in a...
Page 24 - Speci cations
24 Speci cations Name : OMRON Digital Automatic Blood Pressure Monitor HEM-907 Model : HEM-907-E / HEM-907-E7 Display : Digital display Measurement : Oscillometric method Measurement Range : Pressure; 0 to 299 mmHg Pulse rate: 30 to 199 beats/min Accuracy : Pressure; Within +/-3 mmHg Pulse rate; Wit...
Page 25 - Made in Japan; EC
25 Manufacturer OMRON HEALTHCARE Co., Ltd. 53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAPAN EU-representative OMRON HEALTHCARE EUROPE B.V. Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDSwww.omron-healthcare.com Production facility OMRON HEALTHCARE Co., Ltd. Mie, JAPAN Subsidiary OMRON HEALTHCA...