Page 2 - DO NOT FORCE THIS LEVER,; TURNING THE BEAM ON –; ENTER; Microscope will automatically go to 5kV.

8. Press EVAC on SC AIRLOCK VALVE DISPLAY - this will pump the specimen exchange chamber. 9. Observe SEC vacuum display as status lights show vacuum goes from AIR to LOW to HIGH. 10. When SEC vacuum is FilCH (green light is on), open airlock door, by turning airlock level anticlockwise about one hal...

Page 3 - the; DIRECT; READ THE MANUAUL INSTRUCTIONS; CHANGING ACCELERATING VOLTAGE; BEAM ALIGN

D. TAKING A PHOTOGRAPH 1. Obtain the image you desire. 2. Press ABC button on the image adjust section to automatically alter brightness and contrast to the correct levels for the camera. 3. Press DIRECT switch on the photographic control section. 4. Wait for the line scan to perform on cycle and ex...

Page 4 - Hit ENTER to move $ sign against; STIG ALIGN Y; Press; ENTER again; CHANGING THE DETECTOR; PF2 COLUMN COND; H. CHANGING THE WORKING DISTANCE; The working distance is changed by using the

10. Hit ENTER to move $ sign against STIG ALIGN Y Press ENTER again 11. The image will oscillate on the screen, minimise oscillation with ALIGNMENT X and Y knobs. 12. Press PF16 EXIT to finish. G. CHANGING THE DETECTOR The 4500 has two secondary electron detectors, an upper detector which sits above...