Page 2 - This service manual covers the following APS Models:; IMPORTANT NOTICE
This service manual covers the following APS Models: 76” & 80” Full Electric with Pedal Lock Caster System & Grid Decks = APS98174; 76” & 80” Full Electric with Pedal Lock Caster System, Grid Decks, and Underbed Light = APS981741; 76” & 80” Full Electric with Pedal Lock Caster System...
Page 3 - Zenith aps seRies; TABLE OF CONTENTS
Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual Zenith aps seRies 3 TABLE OF CONTENTS Safety & Warnings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 ID/Warning Labels . . . . . . . . . . . . . . . . . . . . ....
Page 4 - IMPORTANT SAFETY AND WARNING INFORMATION
Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual 4 This product is a variable height, adjustable mattressplatform, which will provide comfort and conveniencefor residents/patients and caregivers in long termcare settings. The MA...
Page 5 - ETL; TYPICAL APS 9000 BED IDENTIFICATION LABELS; ETL
Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual 5 GF HEALTH PRODUCTS, INC. DUTY CYCLE - 2 min Continuous/18 min MODE OF OPERATION - Intermittent 600 lbsSAFE WORKING LOAD = IPX4 MODEL NO 806-0000-000000 SERIAL NO 000000000000 MF...
Page 6 - CAUTION; TYPICAL SAFETY/WARNING BED LABELS
Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual 6 WARNING! DO NOT LOWER BED WHEN OBJECTS AREBENEATH BED. FAILURE TO INSPECT UNDERBED CAN RESULT IN DAMAGE TO PROPERTYOR PERSONAL INJURY. ATTENTION: S’assurer de ne pas faire desce...
Page 7 - Hospital; ENTRAPMENT & COMPLIANCE INFORMATION
Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual 7 On March 10, 2006, the FDA (U.S. Food and Drug Adminis-tration) released long-awaited guidelines for reducing the riskof bed entrapment: “Hospital Bed System Dimensional and Ass...
Page 9 - UNPACKING YOUR BED
Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual 9 Zenith aps seRies UNPACKING YOUR BED • Make sure all parts/components are included. • Check all bed components for obvious damage. • Inspect the Power Supply Cord for any cuts a...
Page 10 - HEADBOARD; HEAD- AND FOOTBOARD ASSEMBLY
Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual 10 1. HEADBOARD INSTALLATION • The headboard comes with four pre-installed inserts - consider this the inside of the board. • Position 2 mounting tubes on the outside of the headb...
Page 11 - PLUGGING IN THE FOOTBOARD NURSE/STAFF CONTROL
$OLJQ $UURZV 3KLOOLSV6FUHZV DWWDFK6KURXG 6WDII&RQWURO $VVHPEO\IURP LQVLGHRIERDUG 6WDII&RQWURO $VVHPEO\ 6KURXG &DEOH&RYHU )RRW%RDUG ZLWKFXWRXW &DEOH7LH 6WDII&RQWURO $VVHPEO\ FDEOHWR PRXQWLQJWXEH )227 (1' /RRS0DOH6WDII&RQURO&DEOH DQGFDEOHWLHWKURXJKPRXQWLQJ WXEHKROHZLWK)...
Page 12 - STANDARD MATTRESS RETAINER INSTALLATION
Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual 12 STEP 1. HEAD RETAINER INSTALLATION 1. Position retainer with curved end toward headboard. 2. Squeeze ends of the retainer toward the inside of the bed and lower between the dec...
Page 13 - STANDARD WIREFORM WALLSAVER INSTALLATION; STEP 1. WALLSAVER ASSEMBLY; HEAD END; POSITIONS
Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual 13 Zenith aps seRies :LUHIRUP:DOOVDYHU %HQGIDFHVXSZDUG '(7$,/$ 6WDUWZLWKWDEKHOG KRUL]RQWDOO\ '(7$,/% (QGZLWK :DOOVDYHU UHVWLQJRQ IORRU STANDARD WIREFORM WALLSAVER INSTALLATION STE...
Page 14 - OPTIONAL WALLSAVER ADAPTOR BRACKET INSTALLATION TO; ASSEMBLY
Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual 14 Zenith aps seRies :LUHIRUP:DOOVDYHU %HQGIDFHVXSZDUG '(7$,/$ 6WDUWZLWK WDEKHOG KRUL]RQWDOO\ '(7$,/% (QGZLWK :DOOVDYHU UHVWLQJRQ IORRU OPTIONAL WALLSAVER ADAPTOR BRACKET INSTALLA...
Page 15 - BED OPERATIONS - APS PENDANT/HAND CONTROLLER
Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual 15 NOTE: THE ZENITH 9000 APS BED COMES STANDARD WITH A FOOTBOARD NURSE/STAFF CONTROL(PLEASE MAKE SURE TO ORDER WITH APPROPRIATE BOARD STYLE). THE BACKLIT PENDANTWITH UNDERBED LIGH...
Page 16 - BED OPERATIONS - NURSE/STAFF CONTROL PANEL
Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual 16 Press “Underbed Light” button (#14) to turn underbed light on or off.Press “CPR Heart” button (#7) to Level/Flatten the bed horizontally for CPR position. “Safe Mode” (#15) LED...
Page 17 - BED OPERATIONS - CHAIR POSITION; CPR POSITION
Zenith aps seRies BED OPERATIONS - CHAIR POSITION Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual 17 1. Make sure the Chair position on the staff control panel is not locked out ( orange lock icon ). To unlock, press the Key ic...
Page 18 - BED OPERATIONS - TRENDELENBERG/REVERSE TRENDELENBERG
75(1'(/(1%(5* )227 (1' 5(9(56(75(1'(/(1%(5* +($' (1' )227 (1' +($' (1' Zenith aps seRies BED OPERATIONS - TRENDELENBERG/REVERSE TRENDELENBERG Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual 18 1. Make sure the Hi/Lo function on...
Page 19 - APS 9000 SERIES ELECTRICAL COMPONENTS (LINAK)
Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual 19 ! DO NOT use unauthorized parts, accessories, or adaptors other than those specified/authorized by GF Health Products, Inc. DO NOT open assemblies such as the Actuators, Hand C...
Page 20 - APS9000 SERIES ELECTRICAL CABLING - STANDARD
Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual 20 T-Cable Hi/Lo (Ft) Foot Hi/Lo (Hd) Head Power Cable GroundWire T-Cable (HB)Hi/Lo (Ft) - Port 4Foot - Port 3Hi/Lo (Hd) - Port 2Head - Port 1 Power Cable &Battery Cable -Side...
Page 21 - APS9000 SERIES ELECTRICAL CABLING - WITH UNDERBED LIGHT
Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual 21 APS9000 SERIES ELECTRICAL CABLING - WITH UNDERBED LIGHT UBL-Cable Hi/Lo (Ft) Foot Hi/Lo (Hd) Head Power Cable GroundWire Light T-Cable - Port 2 in Underbed Light BoxHi/Lo (Ft) ...
Page 23 - OPTIONAL PIVOT ASSIST BAR INSTALLATION & OPERATION; INSTALLATION
+($''(&. 6/((3685)$&( +($' '(&. (1' +($' '(&. (1' +($' (1' 5HFRPPHQG 3LYRW7RZDUG +HDGERDUG %ODFN.QREWR /RFNRU5HOHDVH +($' (1' E E D D Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual 23 Zenith aps seRies OPTIONAL...
Page 24 - OPTIONAL TRAPEZE SUPPORT INSTALLATION
Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual 24 Zenith aps seRies 0DLQ)UDPH +($' (1' OPTIONAL TRAPEZE SUPPORT INSTALLATION (The Trapeze Support is attached to the head section of the bed.) Please Order Kit Number ZA79000 Tra...
Page 25 - OPTIONAL BATTERY BACKUP REPLACEMENT CONFIGURATIONS
Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual 25 Zenith aps seRies OPTIONAL BATTERY BACKUP REPLACEMENT CONFIGURATIONS Configuration 1: Replacing Single BatteryCable for a “Portable”Battery Backup Unit Configuration 2: Replaci...
Page 26 - TROUBLESHOOTING - NOTHING WORKS - NO POWER; STAFF CONTROL LIGHTS ARE BLINKING; ALWAYS UNPLUG THE POWER CORD BEFORE DOING
Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual 26 Zenith aps seRies TROUBLESHOOTING - NOTHING WORKS - NO POWER TROUBLESHOOTING STAFF CONTROL - BLINKING LIGHTS 1. NOTHING WORKS - NO POWER a. Check to make sure you have power co...
Page 27 - NOTHING ON STAFF CONTROL WORKS BUT PENDANT WORKS; HAND CONTROLLER FUNCTION DOESN’T
Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual 27 Zenith aps seRies TROUBLESHOOTING PENDANT & DOUBLE T-CABLE NOTHING ON STAFF CONTROL WORKS BUT PENDANT WORKS 1. STAFF CONTROL WORKS BUT PENDANT/ HAND CONTROLLER FUNCTION DOE...
Page 28 - TROUBLESHOOTING MOTOR CABLES & CONTROL BOX; NOT WORKING; QUICK REFERENCE
Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual 28 Zenith aps seRies TROUBLESHOOTING MOTOR CABLES & CONTROL BOX 1. HEAD, FOOT, OR HI/LO FUNCTION NOT WORKING a. Check plugs to make sure they are firmly seated in all ports on...
Page 29 - TROUBLESHOOTING/REPLACING YOUR APS STAFF CONTROL
Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual 29 Zenith aps seRies 6WDII&RQWURO %HYHO 6FUHZV 6WDII&RQWURO $VVHPEO\ )RRW%RDUG ZLWKFXWRXW 6KURXG6FUHZV 6WDII&RQWURO %HYHO6FUHZV ,03257$175XQD&DEOH 7LHWKURXJKWKHHQG...
Page 30 - WARRANTY; Years on Motors and Control Box; - 3 Years on Other Electronics; RECOMMENDED MAINTENANCE AND INSPECTION
Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual 30 } Maintenance Inspection of All Components (Receipt of shipment ) • Make sure all parts/components are included (Please see page 9 - “Unpacking Your Bed”). • Check all bed comp...
Page 31 - DECKING
Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual 31 Zenith aps seRies SERVICE/REPLACEMENT PART RECORD (Page 1) DECKING Part Number Description QTY Order Date Price 999-0844-920SP Head/Back Deck & Hardware Pack 1 Deck 999-084...
Page 32 - UNDERBED LIGHT
Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual 32 Zenith aps seRies SERVICE/REPLACEMENT PART RECORD (Page 2) DOUBLE T-CABLE Part Number Description QTY Order Date Price 999-0806-200SP Double “T” Control Cable 1 Cable Deck Hard...
Page 33 - WHEN ORDERING REPLACEMENT PARTS WITH CUSTOMER SERVICE,; SERVICE/REPLACEMENT PART DIAGRAM - ELECTRONICS
Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual 33 Zenith aps seRies WHEN ORDERING REPLACEMENT PARTS WITH CUSTOMER SERVICE, PLEASE HAVE YOUR BED’S SERIAL NUMBER AVAILABLE TO CONFIRM WHETHER THE PART IS COVERED UNDER WARRANTY. (...
Page 34 - SERVICE/REPLACEMENT PART DIAGRAM - DECKING
Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual 34 Zenith aps seRies WHEN ORDERING REPLACEMENT PARTS WITH CUSTOMER SERVICE, PLEASE HAVE YOUR BED’S SERIAL NUMBER AVAILABLE TO CONFIRM WHETHER THE PART IS COVERED UNDER WARRANTY. (...
Page 35 - MAIN FRAME CAPPING
Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual 35 Zenith aps seRies SERVICE/REPLACEMENT PART DIAGRAM - MAIN FRAME CAPPING CAPS OR PLUGS ARE NOT COVERED UNDER WARRANTY (PACKS OF 12 CAN BE ORDERED IN THE EVENT YOU DAMAGE OR LOSE...
Page 36 - CAPS OR PLUGS ARE NOT COVERED UNDER WARRANTY; CAPPING ON ARMS & LEGS
Basic American Medical Products, a division of GF Health Products, Inc. www.grahamfield.com 999-0831-190A Service Manual 36 Zenith aps seRies SERVICE/REPLACEMENT PART DIAGRAM - CAPPING ON HEAD, KNEE, AND FOOT DECKS CAPS OR PLUGS ARE NOT COVERED UNDER WARRANTY (PACKS OF 12 CAN BE ORDERED IN THE EVENT...