Operation Manual

Impor tant: Do not operate the

John Bunn DigiOx Finger Pulse

Oximeter without first reading

and understanding this manual!

Save this manual for future use.

Note: The most current version of this manual

can be found online at

www.g

rahamfield.com

DigiOx

Finger Pulse Oximeter

JB02008-INS-LAB-RevA09

CONTENTS

INTRODUCTION ............................................................ 3

INTENDED USE ....................................................... 3

CONTRAINDICATIONS ............................................ 3

IMPORTANT SAFETY GUIDELINES - PLEASE

READ BEFORE USE ............................................ 3

OVERVIEW .................................................................... 7
APPEARANCE .............................................................. 7
FEATURES .................................................................... 7
NAME AND MODEL ...................................................... 8
BATTERY INSTALLATION ............................................ 8
OPERATION INSTRUCTIONS ...................................... 8
ACCESSORIES (INCLUDED) ..................................... 11
REPAIR, MAINTENANCE, AND DISPOSAL ............... 11

MAINTENANCE ...................................................... 11

CLEANING AND DISINFECTION .......................... 12

DISPOSAL .............................................................. 13

TROUBLESHOOTING ................................................. 13
SYMBOL KEY .............................................................. 15
TECHNICAL SPECIFICATIONS .................................. 16
LIMITED WARRANTY ................................................. 18
MANUAL DE OPERACIÓN (ESPAÑOL) ..................... 21

INTRODUCTION

Dear User: Please read and understand all instructions

before using the John Bunn DigiOx Finger Pulse

Oximeter, JB02008.

Intended use

The John Bunn DigiOx Finger Pulse Oximeter is used

to measure oxygen saturation (SpO

2

) and pulse rate in

home or clinical settings. It is designed for adult and

pediatric patients.

Contraindications

WARNING: The oximeter is not intended for

continuous patient monitoring. DO NOT use the

oximeter while patient is undergoing MRI or CT

scanning.

Important Safety Guidelines - Please Read Before Use

The safety statements presented in this chapter refer

to the basic safety information that the operator of the

John Bunn DigiOx Finger Pulse Oximeter shall pay

attention to and abide by. There are additional safety

statements in other chapters or sections, which may

be the same as or similar to the following, or specific

to the operations.

WARNING: Indicates a potential hazard situation

or unsafe practice that, if not avoided, could result in

death or serious injury.

Caution: Indicates a potential hazard or unsafe

practice that, if not avoided, could result in minor

personal injury or product/property damage.

Note: Provides application recommendations or other

useful information to ensure that you get the most from

your product.

WARNINGS:

Important! Read and understand this manual before

using the oximeter. Do not use this device without

proper instruction from a Healthcare Professional.

If the oximeter is not properly operated, personal

injury and/or damage to the oximeter could result.
If components are damaged or missing, contact your

dealer immediately. DO NOT use substitute parts.
Inspect the device before use to ensure that there is

no visible damage that may affect patient or user’s

safety or measurement performance. Thereafter,

inspect the device at least once a week.
GF Health Products, Inc. assumes no responsibility

for any damage or injury caused by improper instal-

lation or use of this product.
DO NOT open the oximeter housing. It contains

no serviceable parts. Service must be performed

by trained, authorized personnel only. Otherwise,

device failure and health hazard may occur.

EXPLOSION HAZARD: Do not use the oximeter in

the presence of flammable anesthetics, explosive

substances, vapors or liquids.
The oximeter is designed for real-time and

rapid measurement. It is not suitable for long-time

continuous patient monitoring. DO NOT apply the

sensor to the same finger for more than two hours.

If any readings are found to be abnormal, please

change the position of the sensor.
The oximeter is intended only as an adjunct in

patient assessment. It must be used in conjunction

with clinical signs and symptoms. It is intended as a

measuring, not a treatment, device.

Magnetic and electrical fields are capable of interfer

-

ing with the proper performance of the oximeter. For

this reason, ensure that all external devices operated

in the vicinity of the oximeter comply with the relevant

EMC requirements. Mobile phones, X-ray equipment,

and/or MRI devices are a possible source of interfer-

ence, as they may emit high levels of electromagnetic

radiation. This device complies with IEC60601-1-2.
DO NOT clip this device on edema or tender tissue.
The infrared light emitted from the device (it is invis-

ible) is harmful to the eyes. DO NOT stare into the light.
Keep this device out of the reach of children.

Cautions

To ensure patient safety, use only parts and ac-

cessories specified in this manual.
Keep the oximeter away from dust, vibration,

corrosive substances, explosive materials, high

temperature and moisture.
If the oximeter gets wet, stop operation im-

mediately and do not resume operation until it is

dry. When it is carried from a cold environment

to a warm and/or humid environment, allow it to

warm to room temperature before using it.
DO NOT use sharp objects to press the front

panel button.
DO NOT sterilize this device. Do not subject this

device to irradiation, steam, ethylene oxide, high

temperature, or immersion in liquid. Refer to the

Repair and Maintenance/Cleaning and Disinfec-

tion section of this manual for instructions for

cleaning and disinfection.

OVERVIEW
Appearance

power/display

button

battery cover

display screen

rubber

cushion

Figure 1

Figure 2

The DigiOx Finger Pulse Oximeter is shown above.

Features

The John Bunn DigiOx Finger Pulse Oximeter

provides a simple way to spot-check patients by

combining the sensor and monitor into one integrated,

compact, easy-to-use device with single-button

operation. The oximeter can accurately measure

oxygen saturation and pulse rate. When a finger is

inserted into the sensor’s rubber cushion, and the

Power/Display Button is pressed to turn the oximeter

on, then the SpO

2

value and pulse rate value display

on the screen. The oximeter features a bright-color

LED screen with adjustable brightness and multi-

directional displays that enable the user to read results

from any angle. The oximeter features a low voltage

indicator, powers off automatically in eight seconds

when not in use, and provides approximately 30 hours

of use with the use of alkaline batteries.

Name and Model

Name: John Bunn DigiOx Finger Pulse Oximeter
Model: JB02008

BATTERY INSTALLATION

1. Refer to Figure 3; insert

two AAA batteries, oriented

as shown, into the battery

compartment.

2. Replace the cover.

Note: Ensure that the batteries

are correctly installed. Incorrect

installation will prevent the

device from operating.

OPERATION INSTRUCTIONS

Notes:

Clean the test finger with isopropyl alcohol before and

after each measurement. Ensure that the finger is dry

before being inserted into the oximeter.
Do not shake the finger; keep finger motionless during

oximeter use.

Avoid placing the device on a limb which is wrapped

with a cuff for blood pressure measurement or undergo-

ing venous infusion.
Ensure that no obstruction (contamination, scar tissue,

fingernail polish, acrylic fingernail, etc.) exists at the

site where the sensor is placed; such obstruction could

affect the signal received by the sensor and result in an

incorrect measurement.
Vigorous exercise and electrosurgical device interfer-

ence may affect the measuring accuracy.

1. Open the clip as shown in

Figure 4.

2. Insert finger, nail up, into

the clip's rubber cushion

(ensure that the finger is

correctly positioned), and then gently clip the finger.

3. Press the Power/Display Button to turn the oxim-

eter on.

4. The screen will then display the

data as shown in Figure 5. After

the values become stable, the

user can read the information

on the display screen.

5. When the screen displays as shown in Figure 5,

pressing the Power/Display Button will change the

screen's orientation:

a. Press the Power/Display Button

once more; the screen will

rotate 180° and display as

shown in Figure 6.

b. Press the Power/Display

Button once more; the screen

will rotate 90° and display as

shown in Figure 7.

c. Press the Power/Display Button

once more; the screen will

rotate 180° and display as

shown in Figure 8.

d. Press the Power/Display Button

once more; the screen will

rotate -90° and also show pulse

as a filled wave-form bar graph

along bottom of screen, as

shown in Figure 9.

e. Press the Power/Display Button

once more; the screen will now

show pulse as a hollow wave-

form bar graph along bottom of

screen, as shown in Figure 10.

f. Press the Power/Display Button once more; the

screen will return to the original screen as shown

in Figure 5 on previous page.

6. To change the oximeter’s display screen bright-

ness, press the Power/Display Button for more

than one second. There are 10 brightness levels,

and the level, from “BR 1” to “BR 10”, will show

on the screen during brightness adjustment. The

default is level four. Keep the Power/Display Button

depressed for as long as it takes to cycle to the

desired brightness level.

7. The oximeter will automatically turn off if no finger

is sensed for 8 seconds. If the finger is removed

and the Power/Display Button is pressed, the

oximeter will remain on and change screens, but

will display “FINGER OUT”.

ACCESSORIES (INCLUDED)

1 One lanyard
2 Two AAA alkaline batteries
3 One operation manual

REPAIR, MAINTENANCE, AND DISPOSAL

WARNING: DO NOT open the oximeter housing.

It contains no serviceable parts. Service must be

performed by trained, authorized personnel only. Oth-

erwise, device failure and health hazard may occur.

Maintenance

In order to ensure the oximeter's longevity and

continued dependable operation, please observe the

following maintenance recommendations.

Change the batteries when the low-voltage indicator displays.

Clean the surface of the device before using: Wipe the device

with a clean, soft cloth saturated with a solution of 70% isopropyl

alcohol, then let it air dry or wipe it dry. Repeat as necessary.

Take out the batteries if the oximeter will not be used for a

long period of time.

Environment (Note:

Humidity may shorten

the life of the device,

or even damage it)

Operation: 41°F - 104°F

(5°C - 40°C), 15% - 80% RH

Storage: -4°F - 158°F

(-20°C - 70°C), < 93% RH

Cleaning and Disinfection

1. Do not immerse the device in liquid. Clean the de-

vice only by using a surface-clean technique: Wipe

with a clean, soft cloth saturated with a solution of

70% isopropyl alcohol.

2. Wipe with a clean, soft cloth saturated with clean

water and air dry, or dry with a clean, soft cloth.

Caution: Do not sterilize this device. Do

not subject this device to irradiation, steam,

ethylene oxide, high temperature, or immersion

in liquid.

Caution: Do not use the sensor if it is dam-

aged.

Disposal

Do not treat this device or its batteries as household

waste. Separate collection is required.

TROUBLESHOOTING

Problem

Possible Reason

Solution

SpO

2

and

Pulse Rate

display

unstable or

hard to see

Finger

incorrectly

positioned in

device

Reposition finger

correctly and try again

Finger is

shaking or

patient is

moving

Ensure patient remains

immobile

Patient’s SpO

2

value is too low

to measure

Remeasure; if

unsuccessful, yet device

is functioning correctly,

contact a medical

professional immediately

Excessive

ambient light

Reduce ambient light

Troubleshooting continued

Problem

Possible Reason

Solution

The device

does not

turn on

Batteries are

low

Replace batteries

Batteries

incorrectly

installed

Reinstall batteries

correctly

The device has

malfunctioned

Contact your GF Health

Products, Inc. distributor

The device

turns off

suddenly

The device

powers off

automatically

8 seconds

after finger is

removed

Normal

Batteries are

almost drained

Replace batteries

Screen

displays

“Error3”,

“Error4”,

“Error6” or

“Error7”

Device

malfunction

Contact your GF Health

Products, Inc. distributor

SYMBOL KEY

Symbol

Description

Type BF

Warning: See operation manual

The pulse oxygen saturation

Pulse rate (beats per minute)

Low battery voltage

SN

Serial number

Not for continuous monitoring

TECHNICAL SPECIFICATIONS

Display screen

Organic LED Display

Power supply

requirement

Two 1.5V AAA alkaline batteries

SpO

2

Parameter

Specifications

Measuring range: 0%-100%

Accuracy: 70%-100%, ±3% < 70%,

no definition

Pulse Rate

Parameter

Specifications

Measuring range: 30-235 bpm

Accuracy: 30-99bpm, ±2 bpm; 100-

235 bpm, ±2%

Dimensions (L x

W x H)

2.28" x 1.26" x 1.34" (58 mm x 32 mm

x 34 mm)

Environment

Operation: 41°F - 104°F (5°C - 40°C),

15% - 80% RH

Storage: -4°F - 158°F (-20°C - 70°C),

< 93% RH

Net weight with

batteries

2 oz. (50g)

TECHNICAL SPECIFICATIONS continued

Classification

The type of protection against electric

shock: Internally powered equipment

The degree of protection against

electric shock: Type BF applied part

The degree of protection against

harmful ingress of liquids: Ordinary

equipment without protection against

ingress of water

Electro-Magnetic Compatibility: Group

I, Class B

LIMITED WARRANTY

GF Health Products, Inc. warrants the John Bunn

DigiOx Finger Pulse Oximeter, Model JB02008, for a

period of one year against manufacturer's defects. If

a product is deemed to be under warranty, GF Health

Products, Inc. shall provide, at its option, (1) replace-

ment of any defective part or product or (2) a credit of

the original selling price made to GF Health Products,

Inc.'s initial customer. The warranty does not include

any labor charges incurred in replacement part(s) in-

stallation or any associated freight or shipping charges

to GF Health Products, Inc.

Graham-Field

and

John Bunn

are registered

trademarks of

GF Health Products, Inc.

Packaging, warranties, products, and specifications

are subject to change without notice.
GF Health Products, Inc. is not responsible for

typographical errors.

USA Corporate Headquarters:

GF Health Products, Inc.
2935 Northeast Parkway

Atlanta, Georgia 30360

© 2009 GF Health Products, Inc.

Made in China

telephone: 800-347-5678, 770-447-1609

fax: 800-726-0601, 678-291-3232

www.grahamfield.com

Figure 3

batter y installation

Figure 4

operation

Figure 5

Figure 6

Figure 7

Figure 8

Figure 9

Figure 10