Page 2 - CONTENTS
2 750000-INS-LAB-RevC11 CONTENTS INTRODUCTION ..................................................................................... 3 INTENDED USE OF THIS DEVICE ........................................................ 3 CONTRAINDICATION .................................................................
Page 3 - INTRODUCTION; INTENDED USE OF THIS DEVICE; Contraindication
750000-INS-LAB-RevC11 3 INTRODUCTION Use this manual for set-up and operation of the Lumex AltaDyne 750000 Alternating Pressure / Low Air Loss Mattress System. Read all instructions before using the mattress system. Save this manual for future reference. INTENDED USE OF THIS DEVICE The intended use ...
Page 4 - IMPORTANT SAFETY PRECAUTIONS; DANGER; To reduce the risk of electrocution:; immediately
4 750000-INS-LAB-RevC11 IMPORTANT SAFETY PRECAUTIONS The safety statements presented in this chapter refer to the basic safety information that should be observed by those using the Alternating Pressure / Low Air Loss Mattress System. There are additional safety statements in other chapters or secti...
Page 5 - WARNING; only
750000-INS-LAB-RevC11 5 WARNING To reduce the risk of burns, electrocution, fire, or personal injury: WARNING: Do not leave this product unattended when plugged in. WARNING: Always use close supervision when this product is used by, on, or near children or those who require close supervision. WARNIN...
Page 6 - NOTICE
6 750000-INS-LAB-RevC11 WARNING: Patient entrapment with bed side rails may cause injury or death. The bed frame and its components, including the mattress, bed side rails, head and foot board, bedding, and any accessories added to the bed, can all affect the risk of entrapment. Thorough patient ass...
Page 7 - EQUIPMENT SYMBOLS
750000-INS-LAB-RevC11 7 EQUIPMENT SYMBOLS Attention! Consult accompanying documents (this manual) “BF”symbol: Indicates that this product is in accordance with the degree of protection against electric shock for type BF equipmentFunctional earth (for UL only) Class II (for UL only) Disposal of Elect...
Page 8 - PRODUCT DESCRIPTION; PUMP AND MATTRESS SYSTEM; pump and mattress system
8 750000-INS-LAB-RevC11 PRODUCT DESCRIPTION PUMP AND MATTRESS SYSTEM pump and mattress system CPR valves mattress-pump connectors mattress foot end symbol pump Pump: front quick connector
Page 9 - Pressure adjust knob; : The pressure adjust knob controls; Normal pressure indicator:; The green; normal pressure LED; The yellow; low pressure LED
750000-INS-LAB-RevC11 9 Pump: rear Pump: top panel pressure adjustment knob normal pressure indicator low pressure indicator main power switch static function switch Pressure adjust knob : The pressure adjust knob controls the air pressure output. Turn it clockwise to increase output pressure (firme...
Page 10 - to; Main power switch; : The main power switch toggles
10 750000-INS-LAB-RevC11 Static function switch : Press the static function switch to suspend alternating function, if needed; the pressure inside the air tubes will then be adjusted to identical pressure. Press the static function switch again to change back to alternating mode. The static function...
Page 11 - INSTALLATION; UNPACKING; Before unpacking, check for obvious damage to the
750000-INS-LAB-RevC11 11 INSTALLATION UNPACKING 1. Before unpacking, check for obvious damage to the carton or its contents. If damage is evident, please notify the carrier and your Graham-Field equipment provider. 2. Remove all loose packing from the carton.3. Carefully remove all components from t...
Page 12 - PUMP AND MATTRESS INSTALLATION; Place the mattress on top; NOTICE: Place the device in a position where the
12 750000-INS-LAB-RevC11 PUMP AND MATTRESS INSTALLATION 1. Place the mattress on top of the bed frame; ensure that the foot end symbol faces up at foot end as shown at right. 2. Fold open the wire hangers on the back of the pump as shown at right. 3. Either hang the pump on the bed's foot end as sho...
Page 13 - Plug the power cord into a properly grounded electrical
750000-INS-LAB-RevC11 13 WARNING: Locate all cords so that they will not be stepped on, tripped over, or otherwise subjected to damage or stress. WARNING: Do not use a household extension cord if the electrical cord does not reach the power outlet. Use of an improper extension cord could result in f...
Page 14 - OPERATION; Turn the power switch; PRESSURE ADJUSTMENT; weight distribution
14 750000-INS-LAB-RevC11 OPERATION 1. Turn the power switch to the ON position. 2. Adjust the pressure adjust knob to the maximum position. Info: Set the pressure adjust knob to the maximum position for the quickest inflation at set-up; the mattress can then be adjusted to the desired pressure. 3. T...
Page 15 - Disconnect the quick connector
750000-INS-LAB-RevC11 15 under the patient's buttock area; your fingers should just fit. If the fit is too loose, decrease the setting. If your fingers do not fit well, increase the setting two notches firmer. Wait ten minutes and then check again. You should only have to adjust this setting one tim...
Page 16 - MAINTENANCE; CLEANING; Pump; GENERAL MAINTENANCE
16 750000-INS-LAB-RevC11 MAINTENANCE CLEANING Perform the following cleaning procedures before use. Pump : Wipe the pump with a clean, damp cloth and mild detergent, and keep it away from dust. Air dry. s NOTICE: Do not use phenolic products or corrosive or powdered cleansers to clean the pump. s NO...
Page 17 - Fuse replacement; Air filter replacement; STORAGE
750000-INS-LAB-RevC11 17 Fuse replacement tool needed: small screwdriver 1. If you suspect a blown fuse, disconnect the plug from the wall outlet immediately. 2. Use a small screwdriver to remove the cover of the fuse holder (fuse locations are shown in picture at right). 3. Insert a new fuse of the...
Page 18 - SPECIFICATIONS
18 750000-INS-LAB-RevC11 SPECIFICATIONS Pump Specification Power Supply 120V System: AC 120V 60 Hz Info: see rating label on product Fuse Rating T1A, 250V Cycle time 8 minutes Air input >4 LPM Dimensions (L x W x H) 9.8" x 5.3" x 3.7" (25 x 13.5 x 9.5 cm) Weight 4 lb (1.8 kg) Environm...
Page 19 - WARRANTY
750000-INS-LAB-RevC11 19 WARRANTY GF Health Products, Inc. warrants the Lumex AltaDyne 750000 Alternating Pressure / Low Air Loss Mattress System for a period of twelve months for defects in workmanship and materials. If a product is deemed to be under warranty, GF Health Products, Inc. shall provid...