Page 2 - Milliamps; See Specifications for overload protection.; To return to autorange, press
70/73 English Instruction Sheet Page 2 Input Jacks 300 mA 10A COM FUSED V ! Milliamps Amps Volts, Ohms, Diode Test Common Terminal 600V 70 70 hr2f.eps See Specifications for overload protection. Pushbutton Use the pushbutton to either select a fixed range or put the meterin the Automatic Touch Hold ...
Page 3 - Warning; To avoid electric shock, do not use the Touch Hold; Bar Graph
70/73 English Instruction Sheet Page 3 Automatic Touch Hold Mode W Warning To avoid electric shock, do not use the Touch Hold to determine if a circuit with high voltage is dead.The Touch Hold mode will not capture unstable or noisy readings. The Touch Hold mode automatically captures and di...
Page 6 - continuously. The meter beeps twice if it is in the Touch Hold
70/73 English Instruction Sheet Page 6 Continuity Test ( R ) + _ + _ hr5f.eps If continuity exists (resistance < 210 Ω ), the beeper sounds continuously. The meter beeps twice if it is in the Touch Hold mode. Current ( ? A ) W Warning To avoid injury, do not attempt a currentmeasurement if the ...
Page 7 - the meter face down in the holster.; Maintenance; Caution
70/73 English Instruction Sheet Page 7 Holster The snap-on holster absorbs shocks and protects the meter andcomes with a Flex-Stand . To protect the front of the meter, put the meter face down in the holster. hr15f.eps Maintenance W Warning To avoid electric shock, remove the test leadsbefore open...
Page 8 - Internal Fuse Test; Battery and Fuse Replacement; Note
70/73 English Instruction Sheet Page 8 Internal Fuse Test 300 mA Ω 5-8 Ω <0.5 OK OK 10 A OK 70 OK hr10f.eps Battery and Fuse Replacement Note Before opening the case, make sure the test leads areremoved and the rotary switch is turned to OFF. F1 F2 70 hr11f.eps
Page 9 - Cleaning; Item; Specifications; Accuracy specifications are given as:; Display
70/73 English Instruction Sheet Page 9 Cleaning To clean the meter, use a damp cloth and mild detergent; do notuse abrasives or solvents on the meter. Service and Parts For service information in the U.S.A. and Canada, call 1-800-44-FLUKE (1-800-443-5853), in Europe, 31 40 267 8200, from othercountr...
Page 12 - Overload protection for all functions and ranges: 600 V.; Function
70/73 English Instruction Sheet Page 12 Overload protection for all functions and ranges: 600 V. Function Input Impedance (Nominal) L , m L , K >10 M Ω , <50 pF Common ModeRejection Ratio (1k Ω Unbalanced) Normal ModeRejection L , m L >120 dB at dc, 50 Hz,or 60 Hz >60 dB at 50 Hz or 60 H...