Page 2 - Press button to switch between modes.; Battery Saver
Model 16 English Instruction Sheet Page 2 Symbols Press button. Press button to switch between modes. Double insulation. Manual ranging mode. ip14i.eps Automatic Range Selection The meter defaults to autoranging when you turn it on. The 4000 mVrange can be entered only with manual range selection. M...
Page 3 - Also refer to V; DC; Resistance and Continuity; eR; Turn off circuit power before testing. Also refer to V; Low Impedance; Short; Detects shorts and opens
Model 16 English Instruction Sheet Page 3 AC and DC Voltage Also refer to V • Check. Volts AC Volts DC Input Impedance ≅ 5 M Ω Input Impedance ≅ 10 M Ω 50 Hz to 400 Hz 16 MULTIMETER OFF V 600V COM CAT + AC / DC DC AC 4000 mV 4V 40V 400V 600V 16 MULTIMETER OFF V 600V COM CAT + AC / DC ip01i.eps Resis...
Page 4 - Warning; or; Meter puts 2 k; Disable and re-enable V•Check; For , and
Model 16 English Instruction Sheet Page 4 W V • Check If a dc or ac voltage greater than about 4.5V is present across the inputswhen the meter is set to R , G , or e , the meter switches automatically to dc or ac voltage mode (V • Check mode). W Warning Repetitive transients on a dc bus will cause V...
Page 5 - Thermocouple; Note correct connector polarity.
Model 16 English Instruction Sheet Page 5 Temperature 80PK-1 Type K Thermocouple Probe Vent or Pipe ° C ° F 16 MULTIMETER OFF 600V COM CAT + A TEMPERATURE V • Check ip19f.eps Note correct connector polarity. To meet stated accuracy, the 80AK temperature adapter must be at thesame temperature as the ...
Page 6 - Note correct probe polarity for polarized capacitors.
Model 16 English Instruction Sheet Page 6 Capacitance E Turn off circuit power; then disconnect and discharge the capacitor beforemeasuring capacitance. + If the capacitor requires more discharging, diSC is displayed while the capacitor discharges. 1 µ F 10 µ F 100 µ F 10,000 µ F 16 MULTIMETER OFF 6...
Page 7 - Flame; μA AC
Model 16 English Instruction Sheet Page 7 Microamps µ A Range 0 to 200 µ A 16 MULTIMETER OFF 600V COM CAT + A TEMPERATURE V • Check Flame sensor probe Control module µA AC µA DC ip20i.eps Example: Flame rectification circuit.
Page 8 - Good Diode; Single beep; Forward Bias; Bad Diode; Shorted
Model 16 English Instruction Sheet Page 8 Diode Test G Turn off circuit power before testing. For best results diodesshould be measured out of circuit.Also refer to V • Check. Good Diode 16 MULTIMETER OFF 600V COM CAT + A TEMPERATURE V • Check 16 MULTIMETER OFF 600V COM CAT + A TEMPERATURE V • Check...
Page 9 - (Records the lowest and highest measurements); are disabled. Put the meter
Model 16 English Instruction Sheet Page 9 MIN MAX µ F (Records the lowest and highest measurements) V • Check , autoranging, and Battery Saver are disabled. Put the meter in the proper range before entering MIN MAX. When the reading changes more than about 50 digits, the meter gives ashort beep. W...
Page 12 - Replacement Parts
Model 16 English Instruction Sheet Page 12 Maintenance Clean the case with a damp cloth and detergent. Do not use abrasives orsolvents. Battery ReplacementRemove the test leads before disassembling the case. + COM 600V + ip13i.eps Replacement Parts Fluke TL-75 (Double-insulated leads) PN 855705 80AK...
Page 13 - C to
Model 16 English Instruction Sheet Page 13 Specifications Accuracy is specified for a period of one year after calibration, at 18 ° C to 28 ° C (64 ° F to 82 ° F) with relative humidity to 90%. AC conversions are ac-coupled, average responding, and calibrated to the rms value of a sinewave input. Ac...
Page 14 - MIN MAX Recording Accuracy and Response Time
Model 16 English Instruction Sheet Page 14 Function Range Resolution Accuracy Burden Voltage (50 Hz to 400 Hz) 0 to 200 µ A 0.1 µ A ± (2% + 3 counts) <5 mV/ µ A 0 to 200 µ A 0.1 µ A ± (1% + 2 counts) <5 mV/ µ A Function Overload Protection 1 Input Impedance (Nominal) f 600V rms >5 M Ω <1...
Page 15 - MIN MAX Recording with Elapsed Time; Elapsed Time; minute; Operating Temperature:; C or
Model 16 English Instruction Sheet Page 15 MIN MAX Recording with Elapsed Time Elapsed Time Resolution Accuracy 0 to 100 hours (99:59) 1 minute 0.3% typical Maximum VoltageBetween any Terminaland Earth Ground: 600V rms Display: 3 3/4-digits, 4000 counts, updates 4/sec Operating Temperature: -10 ° C ...
Page 16 - LIMITED WARRANTY & LIMITATION OF LIABILITY
Model 16 English Instruction Sheet Page 16 LIMITED WARRANTY & LIMITATION OF LIABILITY Each Fluke product is warranted to be free from defects in material and workmanship undernormal use and service. The warranty period is three years and begins on the date of shipment.Parts, product repairs and ...