Page 2 - LIMITED WARRANTY AND LIMITATION OF LIABILITY; AS FITNESS FOR A PARTICULAR PURPOSE, ARE EXPRESSED OR IMPLIED.
LIMITED WARRANTY AND LIMITATION OF LIABILITY This Fluke product will be free from defects in material and workmanship for one year from the date of purchase. This warranty does not cover fuses, disposable batteries, or damage from accident, neglect, misuse, alteration, contamination, or abnormal con...
Page 3 - Table of Contents; Title
i Table of Contents Title Page Introduction ................................................................................................................. 1 How to Contact Fluke ................................................................................................. 1 Safety Information ...
Page 4 - Users Manual; ii
Page 5 - Introduction; Warning; How to Contact Fluke
1 Introduction The Fluke 59 MAX and 59 MAX + Infrared Thermometers (the Product) can determine the surface temperature by measuring the amount of infrared energy radiated by the target’s surface. Warning Read all safety information before you use the Product. How to Contact Fluke To contact Fluke,...
Page 6 - Safety Information
59 MAX/59 MAX + Users Manual 2 Or, visit Fluke's website at www.fluke.com . To register your product, visit http://register.fluke.com . To see, print, or download the latest manual supplement, visit http://us.fluke.com/usen/support/manuals . Safety Information A Warning identifies conditions and pro...
Page 9 - Maintenance
Infrared Thermometer Maintenance 5 Maintenance Warning To prevent possible electrical shock, fire, or personal injury, have an approved technician repair the Product. Caution To avoid damage to the Product, do not leave the thermometer on or near objects of high temperature. How to Change the B...
Page 10 - Specifications
59 MAX/59 MAX + Users Manual 6 Specifications 59 MAX 59 MAX + Temperature Range -30 ° C to 350 ° C (-22 ° F to 662 ° F) -30 ° C to 500 ° C (-22 ° F to 932 ° F) Accuracy (Calibration geometry with ambient temperature 23 ° C ± 2 ° C) ≥ 0 ° C: ± 2.0 ° C or ± 2.0 % of reading, whichever is greater( ≥ 32...
Page 12 - Standards and Agency Approval
59 MAX/59 MAX + Users Manual 8 Standards and Agency Approval Ingress Protection Rating ........................ IP40 per IEC 60529 Vibration and Shock ................................ IEC 68-2-6 2.5 g, 10 to 200 Hz, IEC 68-2-27, 50 g, 11 ms Compliance ............................................. EN...
Page 13 - Nominal Surface Emissivity
Infrared Thermometer Nominal Surface Emissivity 9 Nominal Surface Emissivity Material Value Material Value Default**** 0.95 Glass (plate) 0.85 Aluminum* 0.30 Iron* 0.70 Asbestos 0.95 Lead* 0.50 Asphalt 0.95 Oil 0.94 Brass* 0.50 Paint 0.93 Ceramic 0.95 Plastic** 0.95 Concrete 0.95 Rubber 0.95 Copper*...
Page 14 - The Product
59 MAX/59 MAX + Users Manual 10 The Product 1 2 hdi01.eps
Page 16 - OFF
59 MAX/59 MAX + Users Manual 12 7 10 min OFF hdi16.eps