Page 2 - Use and Operation; Warning

. 80PK-7: Industrial Surface Probe 80PK-8 Pipe Clamp Probe (For pipe diameter 1/4” to 1 7/8”) 80CK-M: Male Mini-Connector (for attachment of K-typethermocouple wire) Use and Operation Thermocouples rely upon the difference in temperaturebetween the measuring junction and a reference junction.Traditi...

Page 3 - 0TK Calibration Procedure; Location of Calibration Adjustments; Calibration Procedure 1.; Table 1. Required Equipment

80TK Calibration Procedure A calibration cycle of 1 year is recommended to maintain the80TK within specifications. The required equipment is listedbelow. Use Calibration Procedure 1 to calibrate the 80TK foruse with any K-type thermocouple. Use Calibration Procedure2 if the 80TK will be used exclusi...

Page 4 - LIMITED WARRANTY & LIMITATION OF LIABILITY

80TK-4012 80TK Schematic Diagram Replaceable Parts Ref Des PN Description 735845 CASE TOP 735852 CASE BOTTOM 735860 COVER SWITCH 735944 SHIELD, BOTTOM 735910 SHIELD, TOP 735928 SHIELD CONTACT 747501 SCREW,SCREW,PH,PO,STL,4-40,.250 AR1 802691 IC,OP AMP,TLC271A,PROGRAMMABLE,LOW POWER,SO8 BT1 696534 BA...