Fluke 16 - Manuals

Fluke 16 – Manual in PDF format online.

Manuals:

1 Page 1
2 Page 2
3 Page 3
4 Page 4
5 Page 5
6 Page 6
7 Page 7
8 Page 8
9 Page 9
10 Page 10
11 Page 11
12 Page 12

Summary

Page 2 - Multimeter with Temperature; Function

16 Multimeter with Temperature 2 Accuracy specifications are given as follows: ±([% of reading] + [number of least significant digits]). Table 1. Specifications (cont) Function Range Resolution Accuracy f (50 to 400 Hz) 4000 mV 1 4.000 V 40.00 V 300.0 V 400.0 V 600.0 V 1 mV 0.001 V 0.01 V 0.1 V 0.1 ...

Page 3 - Specifications

Specifications 3 Table 1. Specifications (cont) Function Overload Protection 1 Input Impedance (Nominal) Common ModeRejection Ratio (1 k Ω Unbalanced) 2 Normal Mode Rejection 2 f 600 V dc >5 M Ω <100 pF 2 Automatic Selectionand LoZ = >2 k Ω <200 pF(ac coupled) 3 >60 dB at dc 50 Hz or6...

Page 4 - Disassembly and Reassembly; Caution; Cleaning; Warning; Recommended Test Equipment

16 Multimeter with Temperature 4 Disassembly and Reassembly Referring to Figure 1, disassemble the Meter as follows. APhillips-head screwdriver and small flat-blade screwdriverare required. 1. Remove the test leads and set the rotary knob to OFF. 2. Remove the Phillips-head screws from the case bott...

Fluke Manuals